HIOKI RM3544 Resistance Meter Upgrade Procedures

HIOKI RM3544 Resistance Meter Upgrade Procedures

Before starting the version upgrade, read these instructions carefully.

Operating Precautions

  • Only one instrument can be connected to the computer at a time. Do not connect multiple resistance testers to the computer at the same time.
  • During the update, do not unplug the cable, or exit the updater software.

Introduction

The upgrade is performed using a computer via the USB interface. The procedure involves the following steps:

  • General preparations
  • Connecting the instrument
  • Executing the upgrade procedure
  • Checking the version

1. General Preparations

Before starting the upgrade process, ensure the following required items are available:

  • Unit to be upgraded: RM3544
  • USB Cable: A-miniB type. (Note: This type of USB cable is not included in the accessories of Model RM3544 but is included in those of Model RM3548 and others.)
  • A computer equipped with a USB interface (OS: Windows 10, 32bit/64bit)
  • Upgrade files set

2. Connecting the Instrument

In a power-off state, turn on the power while pressing the [COMP] and [AUTO] keys. Connect the maintenance terminal (MAINTENANCE ONLY) of the instrument and the computer's USB terminal using the USB cable. Note that the maintenance terminal is different from normal USB terminals.

Only one instrument can be connected to the computer at a time. Do not connect multiple resistance testers to the computer simultaneously. Disconnect all USB-connected Hioki products.

3. Executing the Upgrade Procedure

  1. Execute the "RM3544Vup.bat" file.
  2. While the program is being transferred, the progress is indicated by a percentage. When the transfer is completed, the indication will be 100% and the process terminates. During the update, do not unplug the cable, remove the batteries, or exit the updater software.
  3. Press the 'Enter' key on the computer's keyboard to terminate the version upgrade program.

Upgrade Process Output Example:

The upgrade process output typically shows information such as:

  • File size and checksums for boot software and main software.
  • Signature verification (e.g., HIOKI, OK).
  • Model information (e.g., Boot3548, RM3548).
  • Version information (e.g., V1.03).
  • Connection status (e.g., Connect 00M54, Reconnected...).
  • Upgrade progress (e.g., Now upgrading...100 %).
  • Status message (e.g., Successful).
  • Instruction to exit (e.g., Exit by pressing the return key.).

4. Checking the Version

The version number is displayed on the power-on screen. Verify that it is correct.

This completes the upgrade procedure.

The five files used during the upgrade are no longer needed and can be deleted from the computer. Repeated calibration or inspection due to the upgrade is not required.

If the Update Fails

If the update fails, review the "Operating Precautions" section. Repeat the process starting from "2. Connecting the Instrument."

If the instrument display shows "Err:090," connect it to the computer with the USB cable as is. The updater software will display "RM3544? [Y/N]." Enter "Y."

If the instrument does not turn on, it may need to be repaired.

Models: RM3544, RM3544 Resistance Meter, Resistance Meter, Meter

PDF preview unavailable. Download the PDF instead.

RM3544Vup en Microsoft Word 2013 Microsoft Word 2013

Related Documents

Preview HIOKI BT3554 & BT3554-01 Battery Tester Calibration Manual
Calibration manual for the HIOKI BT3554 and BT3554-01 Battery Testers, detailing status, recommended equipment, calibration items, and connection diagrams.
Preview HIOKI BT5525 Sample Application Manual: Installation, Operation, and Features
Detailed guide for the HIOKI BT5525 Sample Application, covering system requirements, installation, connecting the BT5525 instrument, Insulation Resistance Test, BDD Analysis, data saving options, and troubleshooting.
Preview Hioki CM4373-50 Digital Clamp Meter - Technical Specifications and Features
Detailed information on the Hioki CM4373-50 Digital Clamp Meter, including its features, specifications, and applications for electrical professionals in industrial and laboratory settings.
Preview HIOKI UK Declaration of Conformity for RM3545 Resistance Meter
Official UK Declaration of Conformity for the HIOKI RM3545 Resistance Meter, confirming compliance with relevant UK safety and EMC legislation.
Preview HIOKI EU Declaration of Conformity - RM3545 Resistance Meter
Official EU Declaration of Conformity for the HIOKI RM3545 Resistance Meter, confirming compliance with relevant EU directives and standards.
Preview Manual de Instruções HIOKI BT3554-50/51/52 Testador de Bateria
Guia completo para o Testador de Bateria HIOKI BT3554-50, BT3554-51, BT3554-52. Aprenda sobre medição de resistência, tensão, temperatura, gerenciamento de dados e recursos avançados para avaliação de baterias.
Preview HIOKI CM3286-50 AC 钳形功率计 用户手册
HIOKI CM3286-50 AC 钳形功率计用户手册,提供详细的操作指南、安全信息、测量功能(电流、电压、功率、频率、相位检测)及规格。支持可选的 Z3210 无线适配器和 GENNECT Cross 应用。
Preview HIOKI Battery Tester BT3554-50: Streamlined UPS and Lead-Acid Battery Diagnostics
The HIOKI Battery Tester BT3554-50 offers efficient and accurate diagnostics for UPS and lead-acid batteries using impedance methods. Features include audio guidance, streamlined data management via the GENNECT app, and precise measurement of internal resistance and voltage.