8-Second Dishwasher Safe Thermometer

Model #MT18L1-PRO

Distributed By: Microlife USA, Inc.

1617 Gulf to Bay Blvd., 2nd Floor, Clearwater, FL 33755

www.microlifeusa.com

Product Overview

This predictive clinical digital electronic thermometer is designed for the intermittent measurement and monitoring of human adult and pediatric body temperatures. It can be used orally, rectally, or under the arm (axillary).

The thermometer can be used with or without a tip cover. If using a cover, slide it onto the sensor tip until fully engaged and peel off the paper backing. Covers are for single use only. If the thermometer has been stored in a cold environment, allow it to warm up to room temperature for approximately 15 minutes before use. Avoid storing it at temperatures above 140°F (60°C).

Intended Use

Not Intended For Use With:

General Warnings

How to Take a Temperature

Press the power button to turn the unit on. The display should read "188.8" to confirm normal function. An "Err" message indicates the unit is not working properly. The last recorded measurement will show for 2 seconds, followed by a large "L" and a flashing "F" or "C", indicating that the thermometer is ready for use.

Oral Use:

Underarm Use:

Figure A Description: Illustrates the thermometer sensor tip placed correctly under the tongue for oral temperature measurement.

Figure B Description: Illustrates the thermometer sensor tip placed in the armpit with the arm lowered against the body for underarm temperature measurement.

Reading the Temperature

Changing from °F to °C

With the thermometer turned off, press and hold the power button for 5 seconds. Release and press again to toggle between °F or °C. When you have made your selection, wait 3 seconds for °F or °C to start blinking, then press the button again to confirm your selection. The thermometer will be set to your chosen unit the next time it is turned on.

How to Clean

Clean the thermometer before and after every use. Choose one of the following cleaning options:

How to Change Battery

Replace the battery when the [low battery symbol] appears on the display. Insert a coin into the groove on the battery lid and rotate counterclockwise to open. Replace the old battery with a new 3V, CR1632 battery, ensuring the Plus (+) sign faces upward. Replace the lid and rotate the coin clockwise to close.

Limited Warranty

Microlife USA Inc. warrants this thermometer for 5 years against manufacturer defects for the original purchaser only, from the date of purchase. The 5-year warranty applies to the thermometer itself; batteries and the protective case are not covered. There are no user-serviceable parts inside the device.

The warranty does not apply to consequential and incidental damages, or damage caused by batteries, improper handling, accidents, professional use, failure to follow operating instructions, or alterations made to the device or accessory by third parties.

Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.

Microlife USA Inc. will investigate your concern. A product determined to be out of specification under this warranty will be replaced and shipped to you at no cost. A product or accessory determined to be within specification will be returned to you with a report of findings, at no cost.

For any warranty concerns, please call customer service at 1-800-568-4147. Microlife USA Inc. requests that you contact them before sending any product back to better identify and more quickly process your concern.

This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.

Technical Specifications

Measurement range: 90.0°F (32.0°C) to 111.0°F (43.9°C)
Measurement Accuracy: ±0.2°F (±0.1°C) between 93.2°F (34°C) to 107.6°F (42°C)
Self test: Automatic internal check of sensor circuit. "Err" is displayed if there is an error.
Display: Liquid crystal display (LCD) with four digits
Auto Off: After 3 minutes of inactivity
Smallest unit of display: 0.1°F (0.1°C)
Signaling tone: Signals that the thermometer is ready to use and that the temperature increase is less than 0.2°F (0.1°C)/8 seconds.
Memory: Stores the last measured value
Operating Temperature: 50°F (10°C) to 104°F (40°C)
Storage temperature: -13°F (-25°C) to 140°F (60°C)
Humidity: 15% to 95% (non-condensing)
Battery: 3V, CR1632
Expected Product life: 5 years
Service: When the device is not in use, the battery/access may be replaced by the user. No other parts/components are accessible.
Disposal: Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
Reference to Standards: This medical device is compliant with medical device and thermometer standards IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E-1112, EN 12470-3, ISO 10993-10:2010, ISO 10993-5:2009, and ISO10993-12:2012.
Device Standard: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 ASTM E-1112, EN 12470-3, ISO 10993-10:2010, ISO 10993-5:2009, and ISO10993-12:2012
Ingress Protection (IP67): Totally protected against dust ingress, protected against the effect of immersion in water between 15cm and 1m.
Electromagnetic Compatibility: Device fulfills the stipulations of the standard IEC 60601-1-2.
Accuracy: Clinical accuracy characteristics and procedures are available from the manufacturer on request.

Test Protocol (for laboratory use only)

Contact Information

Questions? We’re here to help. Call toll-free 1-800-568-4147.

Models: MT18L1-PRO, 8-Second Dishwasher Safe Thermometer, Clinical Digital Electronic Thermometer

File Info : application/pdf, 2 Pages, 416.39KB

PDF preview unavailable. Download the PDF instead.

MT18L1-Instruction-Manual-041620-W QuarkXPress(R) 15.21

Related Documents

Preview Microlife MT 710 Digital Thermometer User Manual and Specifications
Comprehensive guide to the Microlife MT 710 digital thermometer, covering features, important safety instructions, technical specifications, usage methods, cleaning, and battery replacement.
Preview Microlife MT 800 Digital Thermometer Instructions for Use
Comprehensive guide for the Microlife MT 800 digital thermometer, covering important safety instructions, operating procedures, measurement methods, cleaning, disinfection, battery replacement, technical specifications, and warranty information.
Preview Microlife MT 800 Digital Thermometer User Manual and Specifications
Detailed instructions and specifications for the Microlife MT 800 digital thermometer, covering operation, safety, cleaning, battery replacement, and warranty information.
Preview Microlife MT 600 Digital Thermometer: Instructions and Specifications
Comprehensive guide to the Microlife MT 600 digital thermometer, covering features, safety instructions, usage methods (oral, rectal, axillary), cleaning, battery replacement, technical specifications, and warranty information.
Preview Microlife MT 19F1 Digital Thermometer User Manual
Comprehensive instructions for using the Microlife MT 19F1 digital thermometer, including operation, safety precautions, measurement methods, cleaning, and technical specifications. Supports oral, rectal, and axillary temperature readings.
Preview Microlife MT 800 Digital Thermometer User Manual and Instructions
Comprehensive guide for the Microlife MT 800 digital thermometer, covering safety instructions, measurement methods (oral, rectal, axillary), cleaning, battery replacement, and technical specifications.
Preview Microlife MT 800 Digital Thermometer: User Manual and Specifications
Comprehensive user manual and technical specifications for the Microlife MT 800 digital thermometer. Learn how to use, clean, and maintain your device safely and effectively for accurate temperature readings.
Preview Microlife MT 800 Digital Thermometer - User Manual
Comprehensive instructions for using the Microlife MT 800 digital thermometer, including safety guidelines, measurement methods (oral, rectal, axillary), cleaning, battery replacement, technical specifications, and warranty information.