ACCUJECT 3.0-1P Injector Set

Handling Instructions

Brand: ZEISS

Product Overview

The ACCUJECT 3.0-1P injector set is designed for the safe and efficient implantation of intraocular lenses (IOLs). This document provides handling instructions for the ACCUJECT 3.0-1P injector set, specifically detailing its compatibility and usage with ZEISS CT SPHERIS 204 and ZEISS CT ASPHINA 404 high diopter range IOLs.

Image Description: The ACCUJECT 3.0-1P injector set is a clear plastic, syringe-like medical device with a plunger and a specialized cartridge tip. It is designed for precise delivery of intraocular lenses during ophthalmic surgery.

Compatibility

The ACCUJECT 3.0-1P injector set is compatible with:

Handling Instructions

Follow these steps carefully for the proper use of the ACCUJECT 3.0-1P injector set:

  1. Step 1: Inject OVD

    Inject the Ophthalmic Viscosurgical Device (OVD) into the cartridge tip and along the internal channels of the injector.

    [Diagram Description: A close-up shows a syringe injecting a clear substance into the injector cartridge tip and channels.]

  2. Step 2: Place IOL

    Place the IOL in the center of the open cartridge. Use sterile forceps to gently push down on the edges of the IOL's optic and haptics, ensuring they are positioned correctly under the ledges of the cartridge.

    [Diagram Description: A close-up shows an IOL being placed in an open cartridge, with forceps used to position the optic and haptics under cartridge ledges.]

  3. Step 3: Close Cartridge

    Maintain the IOL in its position using folding forceps while closing the cartridge. Ensure that the haptics do not slip out from the cartridge grooves between the wings.

    [Diagram Description: A close-up shows the IOL held in place by forceps while the injector cartridge is being closed, with attention to haptic placement.]

  4. Step 4: Advance Plunger

    Advance the plunger until it begins to move the IOL forward. Confirm that the plunger is moving freely within the injector.

    [Diagram Description: A close-up shows the injector plunger being advanced, initiating forward movement of the IOL within the cartridge.]

  5. Step 5: Control Release

    To control the release of the IOL from the tip, slow down the movement of the plunger when the optic portion of the lens reaches the beveled end of the tip.

    [Diagram Description: A close-up shows the plunger movement being slowed as the IOL optic approaches the beveled injector tip, controlling the release.]

Manufacturer Information

Manufacturer: Carl Zeiss Meditec AG

Address: Goeschwitzer Str. 51-52, 07745 Jena, Germany

Website: www.meditec.zeiss.com/iol

Contact: www.meditec.zeiss.com/contacts

Regulatory Information: CE 0297

Product Codes: CT ASPHINA 404, CT SPHERIS 204

Document Reference: EN_32_022_0010I

The contents of this document may differ from the current status of approval of the product in your country. Please contact your regional representative for more information. Subject to change in design and scope of delivery and as a result of ongoing technical development. CT ASPHINA and CT SPHERIS are registered trademarks of Carl Zeiss Meditec. ACCUJECT is a trademark of Medicel (CE 0482). Printed on elemental chlorine-free bleached paper. ©2014 by Carl Zeiss Meditec AG. All rights reserved.

Models: CT SPHERIS 204 ACCUJECT 3.0-1P Injector Set, CT SPHERIS 204, ACCUJECT 3.0-1P Injector Set, 3.0-1P Injector Set, Injector Set, Set

File Info : application/pdf, 2 Pages, 98.80KB

PDF preview unavailable. Download the PDF instead.

ACCUJECT-3-0-1P-Injektor-Set-EN-32-022-0010I

References

Adobe PDF Library 15.0

Related Documents

Preview VISCOJECT-BIO 2.2 Injector Set Handling Instructions for 4-Point Haptic IOLs | ZEISS
Detailed handling instructions for the ZEISS VISCOJECT-BIO 2.2 Injector Set, designed for 4-point haptic IOLs. Learn step-by-step usage for ophthalmic procedures.
Preview ZEISS SKYJET Injector Set: Handling Instructions for CT ASPHINA 603P and CT SPHERIS 203P
Detailed handling instructions for the ZEISS SKYJET Injector Set, designed for preloaded CT ASPHINA 603P and CT SPHERIS 203P intraocular lenses. Learn step-by-step preparation and usage for ophthalmic procedures.
Preview ZEISS CT LUCIA 202 IOL Injector Handling Instructions
Detailed handling instructions for the ZEISS CT LUCIA 202 intraocular lens (IOL) injector system, covering preparation, loading, and implantation steps for ophthalmic surgery.
Preview ZEISS CT LUCIA Family IOLs, OVD, and Injectors: Technical Specifications
Detailed technical specifications for the ZEISS CT LUCIA 602 and 621P intraocular lenses (IOLs), OPHTHALIN® OVD, and associated injectors, including product features, materials, and performance data. Information on the ZEISS Digital Ordering Platform is also provided.
Preview ZEISS CT LUCIA 202 Handling Instruction - Medical Device Guide
Comprehensive handling instructions for the ZEISS CT LUCIA 202 IOL injector system, detailing preparation steps, IOL insertion, and implantation procedures for ophthalmic surgery.
Preview BLUEMIXS 180 Injector Handling Instructions from ZEISS
Detailed handling instructions for the ZEISS BLUEMIXS 180 injector, guiding medical professionals through the safe and effective use of the device for intraocular lens implantation.
Preview ZEISS CT LUCIA 602 & Z Cartridge IOL Delivery System Handling Instructions
Comprehensive handling and preparation guide for ZEISS CT LUCIA 602 and Z Cartridge IOL Delivery System, detailing step-by-step procedures for ophthalmic surgical use.
Preview ZEISS CT LUCIA 621P and 621PY Technical Specifications
Detailed technical specifications for ZEISS CT LUCIA 621P and CT LUCIA 621PY intraocular lenses, including optic design, material, dimensions, diopter range, and implantation details.