SKYJET Injector Set
Handling instructions for preloaded CT ASPHINA 603P and CT SPHERIS 203P
Preparation and Use
Step 1: Check that the plunger is fully withdrawn and held by the safety clips.
Step 2: Insert the cassette into the injector following the arrows. Push all the way in until you hear a "click".
Step 3: Hold the breakable safety tab between thumb and index finger. Snap it in the clockwise direction. The tab breaks and releases the lens.
Step 4: Introduce viscoelastic into the hole on the back of the injector. Fill the first 3 mm of the injector nozzle up to the line.
Step 5: Gently arm the plunger to release it from its safety clips and slowly push to drive the lens into the injector nozzle.
Step 6: With the bevelled orifice facing downwards, insert the tip of the injector into the incision and inject the lens into the capsular bag.
Product Information
CE 0459
Product: SKYJET
Compatible IOLs: CT ASPHINA 603P, CT SPHERIS 203P