NEPTUNE Multifunction Professional Safety Multimeter
HT Instruments
Product Overview
The HT Instruments NEPTUNE is an orange and black handheld digital multimeter. Its front features a large LCD screen displaying measurements, a rotary selector dial for choosing functions (e.g., V, Ω, LoZ, OFF), and several push buttons for modes like MODE, ZERO, and HOLD. Input terminals are located at the bottom, labeled COM, VΩ/+, and Ω.
This document details the electrical specifications, general specifications, and reference guidelines for the NEPTUNE multimeter.
1. Electrical Specifications
Accuracy is calculated as [%reading + (num. dgt* resolution)] at 23°C ±5°, <80%RH.
DC Voltage (Autorange)
Range [V] | Resolution [V] | Accuracy | Input Impedance | Overload Protection |
---|---|---|---|---|
0.0 ÷ 999.9 | 0.1 | ±(0.5% reading + 2 digits) | 5MΩ | 1000VDC/ACrms |
AC, AC+DC, LoZ TRMS Voltage (Autorange)
Range [V] | Resolution [V] | Frequency Range | Accuracy | Overload Protection |
---|---|---|---|---|
0.5 ÷ 999.9 | 0.1 | 32Hz ÷ 1kHz | ±(0.5% reading + 2 digits) | 1000VDC/ACrms |
Input impedance VAC function: 5MΩ, Input impedance LoZ function: 3.5kΩ. Auto detection DC mode, Max crest factor: 1.5.
Voltage/Current Frequency (Autorange)
Range [Hz] | Resolution [Hz] | Accuracy |
---|---|---|
32.00 ÷ 99.99 | 0.01 | ±(0.1% reading + 1 digit) |
100.0 ÷ 999.9 | 0.1 | ±(0.1% reading + 1 digit) |
Voltage range: 0.5V ÷ 999V, Current range: 0.5A ÷ 3000A (Flex clamp F3000U), 1mV ÷ 1000mV (STD Clamp).
DC, AC, AC+DC Current (Standard Rigid Clamp + Flex Clamp FS=1V) – (Autorange)
Range [mV] | Resolution [mV] | Accuracy (*) |
---|---|---|
1 ÷ 1000 | 1 | ±(0.5% reading + 2 digits) |
(*) For frequency >100Hz the accuracy is: ±(1.5% reading + 5 digits). Max crest factor: 3, Frequency bandwidth: 1kHz. Current zeroed for value <1%FS [A] (1V Flex clamp), Current zeroed for value <1%FS [A] (STD clamp).
AC TRMS Current (Flexible Clamp F3000U) – (Autorange)
Range [mV] | Resolution [mV] | Accuracy (*) |
---|---|---|
1 ÷ 3000 | 1 | ±(0.5% reading + 2 digits) |
(*) For frequency >100Hz the accuracy is: ±(1.5% reading + 5 digits). Max crest factor: 3, Frequency bandwidth: 1kHz. Current zeroed for value <1%FS [A].
Inrush Current – DC, AC, AC+DC TRMS (Standard Rigid Clamp + Flex Clamp FS=1V)
Range [mV] | Resolution [mV] | Accuracy (*) |
---|---|---|
1 ÷ 1000 | 1 | ±(2% reading + 2 digits) |
(*) Accuracy declared for frequency: DC, 42.5 ÷ 69Hz ; Max crest factor: 3; Sample frequency: 4kHz. Detection threshold: 1%FS [A] fixed. Response time: 1ms (Peak), 16.7ms, 20ms, 50ms, 100ms, 150ms, 175ms, 200ms (max RMS).
Inrush Current – AC TRMS (Flexible Clamp F3000U)
Range [mV] | Resolution [mV] | Accuracy (*) |
---|---|---|
1 ÷ 3000 | 1 | ±(2% reading + 2 digits) |
(*) Accuracy declared for frequency: DC, 42.5 ÷ 69Hz ; Max crest factor: 3; Sample frequency: 4kHz. Detection threshold: 1%FS [A] fixed. Response time: 1ms (Peak), 16.7ms, 20ms, 50ms, 100ms, 150ms, 175ms, 200ms (max RMS).
Resistance and Continuity Test (Autorange)
Range [Ω] | Resolution [Ω] | Accuracy | Buzzer |
---|---|---|---|
0.0 ÷ 199.9 | 0.1 | ±(1.0% reading + 5 digits) | <30Ω |
200 ÷ 1999 | 1 | ±(1.0% reading + 5 digits) |
2. Harmonic Voltage and Current (Autorange)
Harmonic Order | Fundamental Frequency | Resolution | Accuracy (*) (not zeroed values) |
---|---|---|---|
DC | 42.5Hz ÷ 69Hz | 0.1V / 0.1A / 0.1% | ±(5.0% reading + 20 digits) |
1 ÷ 25 | 0.1V / 0.1A / 0.1% | ±(5.0% reading + 10 digits) | |
THD% | 0.1% | ±(10.0% reading + 10 digits) |
Accuracy of harmonics amplitudes expressed in % is evaluated considering the accuracy of parameters ratio.
(*) Harmonic voltages are zeroed in the followed conditions:
- 1° harmonic: value <0.5V
- DC, 2° to 25° harmonics: harmonic value <0.5% fundamental value or value <0.5V
(*) Harmonic currents are zeroed in the followed conditions:
- 1° harmonic: value <1%FS[A]
- DC, 2° to 25° harmonics: harmonic value <0.5% fundamental value or value <1%FS[A]
3. Insulation Resistance (MΩ)
Test Voltage [V] | Range [MΩ] | Resolution [MΩ] | Accuracy |
---|---|---|---|
50 | 0.01 ÷ 9.99 | 0.01 | ±(5.0% reading + 2 digits) |
10.0 ÷ 99.9 | 0.1 | ±(2.0% reading + 2 digits) | |
100 | 0.01 ÷ 9.99 | 0.01 | ±(2.0% reading + 2 digits) |
10.0 ÷ 99.9 | 0.1 | ±(5.0% reading + 2 digits) | |
100.0 ÷ 199.9 | 0.1 | ±(2.0% reading + 2 digits) | |
250 | 0.01 ÷ 9.99 | 0.01 | ±(5.0% reading + 2 digits) |
10.0 ÷ 99.9 | 0.1 | ±(2.0% reading + 2 digits) | |
100 ÷ 499 | 1 | ±(5.0% reading + 2 digits) | |
500 | 0.01 ÷ 9.99 | 0.01 | ±(2.0% reading + 2 digits) |
10.0 ÷ 199.9 | 0.1 | ±(5.0% reading + 2 digits) | |
200 ÷ 499 | 1 | ±(2.0% reading + 2 digits) | |
1000 | 0.01 ÷ 9.99 | 0.01 | ±(5.0% reading + 2 digits) |
10.0 ÷ 199.9 | 0.1 | ±(2.0% reading + 2 digits) | |
200 ÷ 999 | 1 | ±(5.0% reading + 2 digits) | |
1000 ÷ 1999 | 1 | ±(5.0% reading + 2 digits) |
Open voltage: nominal test voltage -0% +10%
Nominal test current: >1mA at 1kΩ x Vnom (50V, 100V, 250V, 1000V), >2.2mA at 230kΩ @ 500V
Short circuit current: <6.0mA for each test voltage
Safety protection: error message for input voltage> 10V
4. Continuity of Earth Conductors (LoΩ)
Range [Ω] | Resolution [Ω] | Accuracy |
---|---|---|
0.00 ÷ 9.99 | 0.01 | ±(2.0% reading + 2 digits) |
10.0 ÷ 199.9 | 0.1 | ±(2.0% reading + 2 digits) |
Test current: >200mA DC for R≤5Ω; DC Current resolution : 1mA, Accuracy: ±(5.0% reading + 5 digits)
Open voltage: 4V ≤ V₀ ≤ 12V
Safety protection: error message for input voltage> 10V
5. Phase Sequence Rotation with 1-Wire Method
Voltage Range [V] | Frequency Range |
---|---|
100.0 ÷ 999.9 | 42.5 ÷ 69Hz |
(*) Measurement is only carried out by direct contact with metal live parts (not on insulation sheath).
6. General Specifications
Display
- 4 LCD, (max 9999 counts), sign, decimal point and bargraph
- Automatic polarity indication
- Backlight
- Refresh frequency: 2/s
- Conversion: TRMS
Features
- Data HOLD
- MAX/MIN
- PEAK (Voltage and Current), response time = 1ms
- Autorange
- Automatic detection of AC/DC signals
- Auto Power OFF after 15 minutes of idleness
Power Supply
- 4x1.5V alkaline batteries type AAA IEC LR03 ?
- Battery life: V, A, Ω, approx 132h (backlight OFF); V, A, Ω, approx 68h (backlight ON); MΩ (@500V) approx 400 test (backlight OFF); LoΩ approx 2000 test (backlight OFF)
Mechanical Specifications
- Dimensions (L x W x H): 175 x 85 x 55mm
- Weight (included batteries): 420g
- Mechanical protection: IP40
Environmental Conditions
- Reference temperature: 23°C ± 5°C
- Working temperature: 5°C ÷ 40°C
- Working humidity: <80%RH
- Storage temperature: -20°C ÷ 60°C
- Storage humidity: <80%RH
- Max height of use: 2000m
7. Reference Guidelines
- Safety: IEC/EN61010-1, IEC/EN61010-2-030, IEC/EN61010-2-033
- EMC: IEC/EN61326-1
- Test MΩ: IEC/EN61557-2
- Test LoΩ: IEC/EN61557-4
- Phase sequence rotation: IEC/EN 61557-7
- Insulation: double insulation
- Pollution degree: 2
- Category of measure: CAT IV 600V, CAT III 1000V to ground and between inputs
8. Compliance
This product conforms to the prescriptions of the European directive on low voltage 2014/35/EU and to EMC directive 2014/30/EU.
This product conforms to the prescriptions of the European directive 2011/65/EU (RoHS) and the European directive 2012/19/EU (WEEE).