HT9020 AC+DC TRMS Power/Harmonic Clamp up to 1000A
1. ELECTRICAL SPECIFICATIONS
Accuracy is calculated as [% rdg + (number of dgt) x resolution]. It is referred to 23°C ± 5°C, <80%RH
DC VOLTAGE
Range | Resolution | Accuracy | Input impedance | Overload protection |
---|---|---|---|---|
0.1-999.9V | 0.1V | ±(1.0%rdg + 3dgt) | 1MΩ | 1000VDC/ACrms |
AC (AC+DC) TRMS VOLTAGE
Range | Resolution | Accuracy | Input impedance | Overload protection |
---|---|---|---|---|
0.5-999.9V | 0.1V | ±(1.0%rdg + 3dgt) | 1MΩ | 1000VDC/ACrms |
Max crest factor: 1.41, Fundamental: 50/60Hz ± 15%, Frequency bandwidth: 42.5Hz - 1725Hz
AC/DC VOLTAGE – MAX/MIN/CREST
Range | Resolution | Accuracy | Response time | Overload protection |
---|---|---|---|---|
0.5-999.9V | 0.1V | ±(3.5%rdg + 5dgt) | 1s | 1000VDC/ACrms |
Input impedance: 1MΩ, Max crest factor: 1.41, Fundamental: 50/60Hz ± 15%, Frequency bandwidth: 42.5Hz - 1725Hz
DC CURRENT
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
0.1-999.9A | 0.1A | ±(2.0%rdg + 5dgt) | 1000ADC/ACrms |
AC (AC+DC) TRMS CURRENT
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
0.5-999.9A | 0.1A | ±(1.0%rdg + 5dgt) | 1000ADC/ACrms |
Max crest factor: 1.41, Fundamental: 50/60Hz ± 15%, Frequency bandwidth: 42.5Hz - 1725Hz
AC/DC CURRENT – MAX/MIN/CREST
Range | Resolution | Accuracy | Response time | Overload protection |
---|---|---|---|---|
0.5-999.9A | 0.1A | ±(3.5%rdg + 5dgt) | 1s | 1000ADC/ACrms |
Max crest factor: 1.41, Fundamental: 50/60Hz ± 15%, Frequency bandwidth: 42.5Hz - 1725Hz
DYNAMIC INRUSH CURRENT DC, AC+DC TRMS
Range | Resolution | PEAK Accuracy | Max RMS Accuracy | Overload protection |
---|---|---|---|---|
1.0-99.9A | 0.1A | ±(2.0%rdg + 5dgt) | ±(2.0%rdg + 5dgt) | 1000ADC/ACrms |
10-999A | 1A |
Crest factor: 3, Sample frequency: 4kHz, Response time: Peak: 1ms, Max RMS: calculated on: 16.7, 20, 50, 100, 150, 200ms
Accuracy declared for frequency: DC, 42... 69Hz
RESISTANCE AND CONTINUITY TEST
Range | Resolution | Accuracy | Buzzer | Overload protection |
---|---|---|---|---|
0.0Ω - 199.9Ω | 0.1Ω | ±(1.0%rdg + 5dgt) | 12 < 150Ω | 1000VDC/ACrms 1000ADC/ACrms |
200Ω - 1999Ω | 1Ω | |||
2.00kΩ-19.99kΩ | 0.01kΩ | |||
20.0kΩ - 29.9kΩ | 0.1kΩ |
FREQUENCY WITH TEST LEADS AND WITH JAWS
Range | Resolution | Accuracy | Overload protection |
---|---|---|---|
42.5 - 69.0Hz | 0.1Hz | ±(1.0%rdg + 5dgt) | 1000VDC/ACrms 1000ADC/ACrms |
Voltage range for frequency measurement: 0.5 ÷ 1000V / Current range for frequency measurement with jaws : 0.5 ÷ 1000A
2. GENERAL SPECIFICATIONS
Mechanical characteristics:
- Dimensions (L x W x H): 252 x 88 x 44mm
- Weight (including battery): 420g
- Max conductor size: 45mm
Power supply:
- Battery type: 2 batteries 1.5V type AAA IEC LR03
- Battery life: approx. 150 hours of continuous use in power/energy measures
- Auto Power Off: approx. 5 minutes of idleness
Display:
- Characteristics: graphic dot matrix, 128x128pxl with backlight
- Sample rate: 128 samples/period (@ 50Hz)
- Display update rate: 1 times/sec
- Conversion mode: TRMS
Climatic conditions:
- Reference temperature: 23°C ± 5°C
- Operating temperature: 0°C ÷ 40°C
- Operating humidity: <80%RH
- Storage temperature: -10°C ÷ 60°C
- Storage humidity: <70%RH
Reference guidelines:
- Comply with: IEC/EN 61010-1, IEC/EN61010-2-032
- EMC: IEC/EN61326-1
- Safety of test leads: IEC/EN61010-031
- Insulation: double insulation
- Pollution degree: 2
- For inside use, max height: 2000m
- Installation category: CAT IV 600V to ground, max 1000V between inputs
This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU (LVD) and of EMC Directive 2014/30/EU
This instrument satisfies the requirements of 2011/65/EU (RoHS) directive and 2012/19/EU (WEEE) directive
Phase Sequence and Phase Conformity
Voltage range | Frequency range | Overload protection |
---|---|---|
100 - 1000V | 42.5 ÷ 69Hz | 1000VDC/ACrms |
Input impedance: 1MΩ
DC POWER
Range [kW] | Resolution [kW] | Accuracy |
---|---|---|
0.00 - 99.99 | 0.01 | ±(3.0%rdg + 3dgt) |
100.0 - 999.9 | 0.1 |
Input impedance: 1MΩ, Accuracy defined for voltage > 10V, current ≥ 2A
ACTIVE POWER, APPARENT POWER AC (AC+DC TRMS)
Range [kW, kVA] | Resolution [kW, kVAR, kVA] | Accuracy |
---|---|---|
0.00 - 99.99 | 0.01 | ±(2.0%rdg + 3dgt) |
100.0 - 999.9 | 0.1 |
Input impedance: 1MΩ, Accuracy defined for: sinusoidal waveform, 42.5..69Hz, Voltage ≥ 10V, Current ≥ 2A, Pf ≥ 0.5
REACTIVE POWER AC (AC+DC TRMS)
Range [kVAR] | Resolution [kW, kVAR, kVA] | Accuracy |
---|---|---|
0.00 - 99.99 | 0.01 | ±(2.0%rdg + 3dgt) |
100.0 - 999.9 | 0.1 |
Input impedance: 1MΩ, Accuracy defined for: sinusoidal waveform, 42.5..69Hz, Voltage ≥ 10V, Current ≥ 2A, Pf ≤ 0.9
ACTIVE ENERGY AC (AC+DC TRMS)
Range [kWh] | Resolution [kWh] | Accuracy |
---|---|---|
0.00 - 99.99 | 0.01 | ±(2.0%rdg + 3dgt) |
100.0 - 999.9 | 0.1 |
Input impedance: 1MΩ, Accuracy defined for: sinusoidal waveform, 42.5..69Hz, Voltage ≥ 10V, Current ≥ 2A, Pf ≥ 0.5
REACTIVE ENERGY AC (AC+DC TRMS)
Range [kVARh] | Resolution [kVARh] | Accuracy |
---|---|---|
0.00 - 99.99 | 0.01 | ±(2.0%rdg + 3dgt) |
100.0 - 999.9 | 0.1 |
Input impedance: 1MΩ, Accuracy defined for: sinusoidal waveform,42.5..69Hz, Voltage ≥ 10V, Current ≥ 2A, Pf ≤ 0.9
POWER FACTOR
Range | Resolution | Accuracy |
---|---|---|
0.20 ÷ 1.00 | 0.01 | ±(2.0%rdg+2dgt) |
Input impedance: 1MΩ, Accuracy defined for: sinusoidal waveform, 42.5..69Hz, Voltage ≥ 10V, Current ≥ 2A
VOLTAGE AND CURRENT HARMONICS
Harmonic order | Fundam. Freq. [Hz] | Resolution [V], [A] | Accuracy (values not zeroed) |
---|---|---|---|
0 | ±(5.0%rdg+20dgt) | ||
1÷25 | 42.5 ÷ 69.0 | 0.1V/0.1A | ±(5.0%rdg+10dgt) |
THD% | 0.1% | ±(10.0%rdg+10dgt) |
The accuracy of harmonics amplitude expressed in % is evaluated considering the accuracy of the parameters ratio
(*) Voltage harmonics are zeroed in the below conditions: • 1st harmonic: if value < 0.5V • DC, 2nd to 25th harmonics: if harmonic value <0.5% of fundamental value or if value < 0.5V Current harmonics are zeroed in the below conditions: • 1st harmonic: if value < 0.5A • DC, 2nd to 25th harmonics: if harmonic value <0.5% of fundamental value or if value < 0.5A