Inogen One G5 / Rove 6 Column Change Guide
Fast and easy maintenance - the Inogen benefit.
Contact Information
Inogen, Inc.
859 Ward Drive Suite 200.
Goleta, CA 93111
1-877-4-INOGEN (Toll Free)
1-877-446-6436 (Toll Free)
1-805-562-0515 (International)
Email: info@inogen.net
Website: www.inogen.com
Document ID: 96-09240-00-01 D
Instructions for Changing Columns
To change the columns on your Inogen One® G5/Rove 6, follow these steps:
- Turn off the Inogen One® G5/Rove 6 concentrator by pressing the power button to shut down the device.
- Remove the Inogen One® G5/Rove 6 concentrator from the carrying case (if attached).
- Remove the battery from the Inogen One® G5/Rove 6 concentrator.
- Place the Inogen One® G5/Rove 6 concentrator on its side so that the underside is visible. The metal column assembly can be seen on one side of the device.
- Unlock the column assembly by pushing the latch button away from the columns.
- While holding the button open, slide the column assembly out of the device by pulling on the column pull handle.
- Remove the columns completely from the Inogen One® G5/Rove 6. Both columns are removed as one piece.
- If tape is applied to the column assembly, remove the tape.
- Remove dust caps from the new column assembly. Ensure there is no dust or debris where the dust caps were located.
- Insert the column assembly into the Inogen One® G5/Rove 6 concentrator. Do not leave the column ends exposed; the column assembly should be inserted into the Inogen One® G5/Rove 6 as soon as the dust caps have been removed.
- Push the column assembly into the device such that the columns are fully seated into the Inogen One® G5/Rove 6 concentrator. The spring-loaded latch button should fully return to the closed position.
- Connect the AC power supply cord to the Inogen One® G5/Rove 6 and plug the power supply AC cord into an electrical outlet. Do not power on the Inogen One® G5/Rove 6 concentrator.
NOTE: Column change instructions are only to be used when maintenance is required and are not intended for practice purpose.
Image of device may vary.
Post-Procedure: Sieve Reset
After column replacement, a sieve reset is required.
Method 1 (Recommended):
- Press the alert/bell button [?] once. The screen will display "sieve reset success."
- Press the power button [▶️] to turn on the Inogen One® G5/Rove 6, and use normally.
Method 2 (Alternative):
- Press and hold the plus (+) and minus (-) buttons for 5 seconds. The screen will display a "sieve reset" message. Release buttons once the message is displayed on screen.
Continues on back [→]