ICON PROCESS CONTROLS IS-750D Series ProCon Free Chlorine Display Controller
Safety Information
- DO NOT exceed maximum temperature or pressure specifications
- ALWAYS wear safety goggles or face-shield during installation and/or service
- DO NOT alter product construction
Warning | Caution | Danger
Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, or failure, injury, or death.
Note | Technical Notes
Highlights additional information or detailed procedure.
Intended Use
When receiving the instrument, please open the package carefully, check whether the instrument and accessories are damaged by transportation and whether the accessories are complete. If any abnormalities are found, please contact our after-sales service department or regional customer service center, and keep the package for return processing. The technical data listed in the current data sheet are engaging and must be complied with. If the data sheet is not available, please order or download it from our homepage (www.iconprocon.com).
Personnel for Installation, Commissioning, and Operation
This controller is an analytical measurement and control instrument with highly precision. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument. Ensure that the power cable is physically separated from the power supply when connection or repair. Once the safety problem occurs, make sure that the power to the controller is off and disconnected.
Technical Specifications
Measurement Range | 0.005~20.00mg/L | 0.005~20.00ppm |
Measurement Unit | mg/L | PPM |
Resolution | 0.001mg/L | 0.001ppm |
Basic Error | ±1% F.S |
Temperature | 14 ~ 302ºF | -10 ~ 150.0oC (Depends on the Sensor) |
Temperature Resolution | 0.1°C |
Temperature Basic Error | ±0.3°C |
Current Output | 2 groups: 4-20mA |
Communication Output | RS485 MODBUS RTU |
Other Functions | Data Record & Curve Display |
Three Relay Control Contacts | 5A 250VAC, 5A 30VDC |
Power Supply | 9~36VDC (Standard) | 85~265VAC* | Power Consumption ≤ 3W |
Working Conditions | No strong magnetic field interference around except the geomagnetic field |
Working Temperature | 14 ~ 140oF | -10~60°C |
Relative Humidity | ≤90% |
Waterproof Rating | IP65 |
Weight | 0.8kg |
Dimensions | 144 x 114 x 118mm |
Installation Opening Size | 138 x 138mm |
Installation Methods | Panel | Wall Mount | Pipeline |
Dimensions
Cabinet Installation
Wall Mount Installation
- Install a mounting bracket for the instrument
- Wall screw fixation
Top view of mounting bracket. Pay attention to installation direction
Wiring
Terminal | Description |
P+, P- | VDC Power Supply |
TEMP1 | Temperature Connections |
CE,RE,WE | Sensor Input Connections |
I1, G, I2 | 4-20mA Output 1 & 2 |
Terminal | Description |
V+, V-, A1, B1 | Digital Input Channel 1 |
V+, V-, A2, B2 | Digital Input Channel 2 |
A3, B3 | RS485 Communication Output |
RLY3,RLY2,RLY1 | 3 Relay Terminals |
The connection between the instrument and the sensor: the power supply, output signal, relay alarm contact and the connection between the sensor and the controller are all inside the controller, and the wiring is as shown above. The length of the cable lead fixed by the electrode is usually 5-10 Meters, insert the line with corresponding label or color wire on the sensor into the corresponding terminal inside the instrument and tighten it.
Keypad Description
Keypad Operation
- Short Press: Release the key immediately after pressing (Default to short presses if not specified above).
- Long Press: Press the button for 3 seconds and then release it.
Display Descriptions
- All electrical connections should be checked before use. After the power is switched on, the meter will display as follows.
The following is the menu structure of this instrument
Setting | Sensor | Type | FCL |
Unit | mg/L | ||
Temperature | Temperature Sensor | NTC2.252 KΩ | |
NTC10 KΩ | |||
PT100 | |||
PT1000 (Default) | |||
Temperature Offset | 0.0000 | ||
Temperature Input | Automatic (Default) | ||
Manual | |||
Temperature Unit | oC | ||
oF |
Calibration | Standard Solution Calibration | Point 1 | 0.01 (Default, can be modified) |
Point 2 | 1 (Default, can be modified) | ||
Point 3 | 10 (Default, can be modified) | ||
Point 4 | 100 (Default, can be modified) | ||
Point 5 | 1000 (Default, can be modified) | ||
Calibration Adjustment | Voltage 1 | ||
Voltage 2 | |||
Voltage 3 | |||
Voltage 4 | |||
Voltage 5 | |||
Field Calibration | Field Calibration | 00.001 (Default) | |
Offset Adjustment | 00.000 | ||
Slope Adjustment | 01.000 |
Set Point | Relay 1 | Status | ON |
OFF | |||
High/Low Set Point | High Alarm | ||
Low Alarm | |||
Clean | |||
Limit Value | 020.00 mg/L | ||
Hysteresis | 00.00 mg/L | ||
Relay 2 | Status | ON | |
OFF | |||
High/Low Set Point | High Alarm | ||
Low Alarm | |||
Clean | |||
Limit Value | 020.00 mg/L | ||
Hysteresis | 00.00 mg/L | ||
Relay 3 | Status | ON | |
OFF | |||
High/Low Set Point | High Alarm | ||
Low Alarm | |||
Clean | |||
Limit Value | 020.00 mg/L | ||
Hysteresis | 00.00 mg/L |
Output | Current 1 | Channel | Main |
Temperature | |||
Output Option | 4-20mA | ||
0-20mA | |||
20-4mA | |||
Upper Limit | |||
Lower Limit | |||
Current 2 | Channel | Main | |
Temperature | |||
Output Option | 4-20mA | ||
0-20mA | |||
20-4mA | |||
Upper Limit | |||
Lower Limit |
Output | RS485 | Baud Rate | 4800BPS |
9600BPS | |||
19200BPS | |||
Parity Check | None Parity | ||
Odd Parity | |||
Even Parity | |||
Stop Bit | 1 Bit | ||
2 Bit | |||
Network Node |
Data Log | Graphic Trend (Trend Chart) | Record Interval | Display according to interval settings 480 points/screen |
1h | |||
12h | |||
24h | |||
Record Query | Query by number of data | ||
Record Interval | 7.5s | ||
90s | |||
180s | |||
Data Output | 101600 Point | ||
Upload Data |
System | Language | English | |
Date/Time | Year-Month-Day | ||
Hour-Minute-Second | |||
Display | Display Speed | Low | |
Standard | |||
Medium | |||
High | |||
LED | Energy Saving | ||
Long Bright | |||
Information | Software Version | 22-8.0 | |
Password Settings | 0000 | ||
Serial number |
System | Factory Default | No | |
Yes | |||
Terminal Current Tuning | Current 1 | 4mA | The positive and negative ends of the ammeter are connected to the current 1 or current 2 output terminals of the instrument respectively, press [![]() |
|
Current 1 | 20mA | |||
Current 2 | 4mA | |||
Current 2 | 20mA | |||
Relay Test | Relay 1-Testing | Select three groups of relays and hear the sound of two switches ,the relay is normal. | |
Relay 2-Testing | |||
Relay 3-Testing |
Calibration
Press [MENU] to enter the setting mode and select the calibration
Calibration | Standard Solution Calibration | Point 1 | Enter given standard liquid value(Example:0.01) |
Point 2 | Enter given standard liquid value(Example:1.0) | ||
Point 3 | Enter given standard liquid value(Example:10.0) | ||
Point 4 | Enter given standard liquid value(Example:100.0) | ||
Point 5 | Enter given standard liquid value(Example:100.0) | ||
Field Calibration | |||
Offset Adjustment | |||
Slope Adjustment |
Standard Solution Calibration
- This function is used to calibrate the five calibration points of the sensor. It has been calibrated before delivery and users can use it directly. If calibration is required, prepare 5 suitable standard liquids with known value, press [MENU] to enter the setting mode and select the calibration point. Modify or enter the corresponding calibration value.
- After setting the calibration value, press [ENT] key to confirm and enter the calibration screen.
- If the controller has been calibrated, the screen shows the calibration status. Press , press the [
] key to switch the calibration state of the calibration point. If recalibration is required in this state, press [ENT] key to enter re-calibration.Standard solution calibration has 5 calibration points. You can choose any one point to calibrate (at least choose one point for calibration)
- In the standard solution calibration mode, press the [
] key to switch the calibration points, press the [ENT] to start calibrating.
- If the monitor prompts you to enter the calibration safety password, press [
] or [
] key to set the calibration safety password, then press [ENT] to confirm the calibration safety password
Point 1 Calibration
After entering the calibration mode, the controller displays as shown in the above figure. The main value of the controller displays the point 1 known standard solution value. Place the electrode into the standard solution of the corresponding value, and the corresponding voltage mV value and calibration state will be displayed on the left side of the screen. After completion of calibration, (Done) will be displayed on the right side of the screen. If the next point needs to be calibrated, press [ ] to switch the calibration point. If only one point calibration is needed, after the calibration is completed, press [MENU] to exit. During the calibration process, if the calibration is failed, the screen will show (Error).
Field Calibration
Select field calibration methods: [Field calibration], [Offset adjustment], [Linear adjustment].Field Calibration
When the data from the portable instrument is taken into this setting, the instrument will automatically correct the data.
Calibration Results
- Confirm: When the “ENT” icon is green, press [ENT] to confirm
- Cancel : Press the [
] key to shift the green icon to ESC, and press [ENT] key to confirm
Offset Adjustment
Compare the data from portable instrument with the data measured by controller. If there are errors, the error data can be modified by this function.
Linear adjustment
Linear values after “field calibration” will be saved in this setting and the factory data is 1.00.
Graphic Trend (Trend Chart)
Data Log | Graphic Trend (Trend Chart) | Record Interval | Display according to interval settings 480 points/screen |
1h | |||
12h | |||
24h | |||
Record Query | Query by number of data | ||
Record Interval | 7.5s | ||
90s | |||
180s | |||
Data Output | 101600 Point | ||
Upload Data |
Press the [TREND] button in the measurement mode to view the trend chart of the saved data directly. Pressing the [ESC] button returns to the measurement screen. There are 480 sets of data record per page.
In the current screen, press the [ENT] key to move the data display line to the left and [
/TREND] key to move towards right. Press [INFO] to change the chart display range.
- Record Interval : Users can select the recording interval here. Once selected, the controller will save the data at the specified interval.
- Record Query : Enter the number of records to query here, then press [ENT] to retrieve the historical data.
MODBUS RTU
The hardware version number of this document is V2.0; the software version number is V5.9 and above. This document describes the MODBUS RTU interface in details and the target object is a software programmer.
MODBUS Command Structure
- Data format description in this document;
- Binary display, suffix B, for example: 10001B – decimal display, without any prefix or suffix, for example: 256 Hexadecimal display, prefix 0x, for example: 0x2A
- ASCII character or ASCII string display, for example: “YL0114010022”
Command Structure
The MODBUS application protocol defines the Simple Protocol Data Unit (PDU), which is independent of the underlying communication layer.
MODBUS protocol mapping on a specific bus or network introduces additional fields of protocol data units. The client that initiates the MODBUS exchange creates the MODBUS PDU, and then adds the domain to establish the correct communication PDU.
On the MODBUS serial line, the address domain contains only the slave instrument address. Tips: The device address range is 1…247
Set the device address of the slave in the address field of the request frame sent by the host. When the slave instrument responds, it places its instrument address in the address area of the response frame so that the master station knows which slave is responding. Function codes indicate the type of operation performed by the server. CRC domain is the result of the “ redundancy check” calculation, which is executed according to the information content.
MODBUS RTU Transmission Mode
When the instrument uses RTU (Remote Terminal Unit) mode for MODBUS serial communication, each 8-bit byte of information contains two 4-bit hexadecimal characters. The main advantages of this mode are greater character density and better data throughput than the ASCII mode with the same baud rate. Each message must be transmitted as a continuous string.
The format of each byte in RTU mode (11 bits):
- Coding system: 8-bit binary
- Each 8-bit byte in a message contains two 4-bit hexadecimal characters (0-9, A-F)
- Bits in each byte: 1 starting bit
- 8 data bits, the first minimum valid bits without parity check bits
- 2 stop bits
- Baud rate: 9600 BPS
How characters are transmitted serially:
Each character or byte is sent in this order (from left to right) the least significant bit (LSB)… Maximum Significant Bit (MSB)
Check Domain Structure: Cyclic Redundancy Check (CRC16)
Structure description:
Slave Instrument | Function Code | Data | CRC |
Address | 1 byte | 0…252 byte | 2 byte |
CRC Low byte | CRC High byte |
Fig.4 : RTU Information Structure
The maximum frame size of MODBUS is 256 bytes
MODBUS RTU Information Frame
In RTU mode, message frames are distinguished by idle intervals of at least 3.5 character times, which are called t3.5 in subsequent sections.
The entire message frame must be sent in a continuous character stream. When the pause time interval between two characters exceeds 1.5 characters, the information frame is considered incomplete and the receiver does not receive the information frame.
The RTU mode contains an error-detection domain based on a cyclic redundancy check (CRC) algorithm that performs on all message contents. The CRC domain checks the contents of the entire message and performs this check regardless of whether the message has a random parity check. The CRC domain contains a 16-bit value consisting of two 8-bit bytes. CRC16 check is adopted. Low bytes precede, high bytes precede.
Implementation of MODBUS RTU in Instrument
According to the official MODBUS definition, the command starts with a 3.5 character interval triggering command, and the end of the command is also represented by a 3.5 character interval. The device address and MODBUS function code have 8 bits. The data string contains n*8 bits, and the data string contains the starting address of the register and the number of read/write registers. CRC check is 16 bits.
Value |
Start |
Device Address |
Function |
Data |
Summary Check |
End |
|
No Signal bytes during 3.5 Characters |
1-247 1 |
Function Codes Confirming to MODBUS
Specification |
Data Confirming to MODBUS Specification |
CRCL |
CRCL |
No Signal bytes during 3.5 characters |
|
Byte | 3.5 | 1 | N | 1 | 1 | 3.5 |
Fig.7 : MODBUS definition of Data Transmission
Instrument MODBUS RTU Function Code
The instrument only uses two MODBUS function codes:
- 0x03: Read-and-hold register
- 0x10: Write multiple registers
- MODBUS Function Code 0x03: Read-and-hold Register
This function code is used to read the continuous block content of the holding register of the remote device. Request the PDU to specify the start register address and the number of registers. Address registers from zero. Therefore, the addressing register 1-16 is 0-15. The register data in the response information is packaged in two bytes per register. For each register, the first byte contains high bits and the second byte contains low bits.
Request:
Function Code | 1 byte | 0x03 |
Start Address | 2 byte | 0x0000 0xfffff |
Read Register Number | 2 byte | 1…125 |
Fig.8 : Read and hold register request frame
Response:
Function Code | 1 byte | 0x03 |
Number of bytes | 2 bytes | Nx2 |
Register Values | Nx2 byte | 1…125 |
N = Register Number
Figure 9 : Read and hold register response frame
The following illustrates the request frame and response frame with the read and hold register 108-110 as an example. (The contents of register 108 are read-only, with two byte values of 0X022B, and the contents of register 109-110 are 0X0000 and 0X0064)
Request Frame | Response Frame | ||
Number Systems | (Hexadecimal) | Number Systems | (Hexadecimal) |
Function Code | 0x03 | Function Code | 0x03 |
Start Address (High byte) | 0x00 | Byte Count | 0x06 |
Start Address (Low byte) | 0x6B | Register Value (High Bytes) (108) | 0x02 |
Number of Read Registers (High Bytes) |
0x00 |
Register Value (Low Bytes) (108) |
0x2B |
Number of Read Registers (Low Bytes) |
0x03 |
Register Value (High Bytes) (109) |
0x00 |
Register Value (Low Bytes) (109) | 0x00 | ||
Register Value (High Bytes) (110) | 0x00 | ||
Register Value (Low Bytes) (110) | 0x64 |
Figure 10 : Examples of read and hold register request and response frames
MODBUS Function Code 0x10 : Write Multiple Registers
This function code is used to write continuous registers to remote devices (1… 123 registers) block that specifies the value of the registers written in the request data frame. Data is packaged in two bytes per register. Response frame return function code, start address and number of registers written.
Request:
Function Code | 1 byte | 0x10 |
Start Address | 2 byte | 2 byte |
Number of input registers | 2 byte | 2 byte |
Number of bytes | 1 byte | 1 byte |
Register Values | N x 2 bytes | N x 2 bytes |
Response:
Function Code | 1 byte | 0x10 |
Start Address | 2 byte | 0x0000 0xffff |
Register Number | 2 byte | 1…123(0x7B) |
N = Register Number
Figure 12 : Write Multiple Register Response Frames
The request frame and response frame are illustrated below in two registers that write the values 0x000A and 0x0102 to the start address of 2.
Request Frame | (Hexadecimal) | Response Frame | (Hexadecimal) |
Number Systems | 0x10 | Number Systems | 0x10 |
Function Code | 0x00 | Function Code | 0x00 |
Start Address (High byte) | 0x01 | Start Address (High byte) | 0x01 |
Start Address (Low byte) | 0x00 | Start Address (Low byte) | 0x00 |
Input Register Number (High bytes) | 0x02 | Input Register Number (High bytes) | 0x02 |
Input Register Number (Low bytes) | 0x04 | Input Register Number (Low bytes) | |
Number of bytes | 0x00 | ||
Register Value (High byte) | 0x0A | ||
Register Value (Low byte) | 0x01 | ||
Register Value (High byte) | 0x02 | ||
Register Value (Low byte) |
Figure 13 : Examples of writing multiple register request and response frames
Data Format in Controller
Floating Point
Definition: Floating point, conforming to IEEE 754 (single precision)
Description | Symbol | Index | Mantissa | SUM |
Bit | 31 | 30…23 | 22…0 | 22…0 |
Index Deviation | 127 |
Figure 14 : Floating Point Single Precision Definition (4 bytes, 2 MODBUS Registers)
Example: Compile decimal 17.625 to binary
Step 1:
- Converting 17.625 in decimal form to a floating-point number in binary form, first finding the binary representation of the integer part
- 17decimal= 16 + 1 = 1×24 + 0×23 + 0×22 + 0×21 + 1×20
- The binary representation of integer part 17 is 10001B
- Then the binary representation of decimal part is obtained 0.625= 0.5 + 0.125 = 1×2−1 + 0×2−2 + 1×2−3
- The binary representation of decimal part 0.625 is 0.101B.
- So the binary floating point number of 17.625 in decimal form is 10001.101B
Step 2:
Shift to find the exponent.
Move 10001.101B to the left until there is only one decimal point, resulting in 1.0001101B, and 10001.101B = 1.0001101 B× 24 . So the exponential part is 4, plus 127, it becomes 131, and its binary representation is 10000011B. Step 3:
Calculate the tail number
After removing 1 before the decimal point of 1.0001101B, the final number is 0001101B (because before the decimal point must be 1, so IEEE stipulates that only the decimal point behind can be recorded). For the important explanation of 23-bit mantissa, the first (i.e. hidden bit) is not compiled. Hidden bits are bits on the left side of the separator, which are usually set to 1
and suppressed.
Step 4:
Symbol bit definition
The sign bit of positive number is 0, and the sign bit of negative number is 1, so the sign bit of 17.625 is 0.
Step 5:
Convert to floating point number
bit symbol + 8 bit index + 23-bit mantissa
0 10000011 00011010000000000000000B (the hexadecimal system is shown as 0 x418d0000 )
Reference code:
- If the compiler used by the user has a library function that implements this function, the library function can be called directly, for example, using C language, then you can directly call the C library function memcpy to obtain an integer representation of the floating-point storage format in memory.
For example: float floatdata; // converted floating point number void* outdata; memcpy (outdata, & floatdata, 4);
Suppose floatdata = 17.625
If it is a small-end storage mode, after executing the above statement, the data stored in the address unit outdata is 0x00. Outdata + 1 stores data as 0x00
address unit (outdata + 2) stores data as 0x8D
address unit (outdata + 3) stores data as 0x41
If it is large-end storage mode, after executing the above statement, the data stored in outdata of address unit is 0x41 address unit (outdata + 1) stores data as 0x8D address unit (outdata + 2) stores data as 0x00 address unit (outdata + 3) stores data as 0x00 - If the compiler used by the user does not implement the library function of this function, the following functions can be used to achieve this function:
- void memcpy(void *dest,void *src,int n)
{ - char *pd = (char *)dest; char *ps = (char *)src; for(int i=0;i<n;i++) *pd++ = *ps++;
} - And then make a call to the above memcpy(outdata,&floatdata,4);
Example: Compile binary floating-point number 0100 0010 0111 1011 0110 0110 0110 10B to decimal number
- void memcpy(void *dest,void *src,int n)
Step 1: Divide the binary floating-point number 0100 0010 0111 1011 0110 0110 0110B into symbol bit, exponential bit and mantis-sa bit.
- 0 10000100 11110110110011001100110B
- 1-bit sign + 8-bit index + 23-bit tail sign bit S: 0 denotes positive number
- Index position E: 10000100B =1×27+0×26+0×25+0×24 + 0 × 23+1×22+0×21+0×20
- =128+0+0+0+0+4+0+0=132
- Mantissa bits M: 11110110110011001100110B =8087142
Step 2: Calculate the decimal number
- D = (−1)×(1.0 + M/223)×2E−127
- = (−1)0×(1.0 + 8087142/223)×2132−127
- = 1×1.964062452316284×32
- = 62.85
Reference Code:
- float floatTOdecimal(long int byte0, long int byte1, long int byte2, long int byte3){
long int realbyte0,realbyte1,realbyte2,realbyte3; char S; long int E,M; - float D; realbyte0 = byte3; realbyte1 = byte2; realbyte2 = byte1; realbyte3 = byte0; if((realbyte0&0x80)==0)
{ - S = 0;//positive number
} - else
{ - S = 1;//negative number
} - E = ((realbyte0<<1)|(realbyte1&0x80)>>7)-127;
- M = ((realbyte1&0x7f) << 16) | (realbyte2<< 8)| realbyte3;
- D = pow(-1,S)*(1.0 + M/pow(2,23))* pow(2,E);
- return D;
}
Function description: parameters byte0, byte1, byte2, byte3 represent 4 bytes of binary floating point number. The decimal number converted from the return value.
For example, the user sends the command to get the temperature value and dissolved oxygen value to the probe. The4 bytes representing the temperature value in the received response frame are 0x00, 0x00, 0x8d and 0x41. Then the user can get the decimal number of the corresponding temperature value through the following call statement.
- That is temperature = 17.625.
- Float temperature = floatTOdecimal( 0x00, 0x00, 0x8d, 0x41)
Read Instruction Mode
The communication protocol adopts MODBUS (RTU) protocol. The content and address of the communication can be changed according to the needs of customers. The default configuration is network address 01, baud rate 9600, even check, one stop bit, users can set their own changes;
Function code 0x04: This function enables the host to obtain real-time measurements from slaves, which are specified as single-precision floating-point type (i.e. occupying two consecutive register addresses), and to mark the corresponding parameters with different register addresses. Communication address is as follows:
- 0000-0001: Main Measured Value | 0002-0003: Temperature Value | 0004-0005: Main Voltage Value |
- 0006-0007: Temperature & Voltage Value
Communication examples:
Examples of function code 04 instructions:
Communication address = 1, Main Value = 20.0, Temperature = 10.0, Main voltage = 100.0, Temperature voltage = 200.0 Host Send: 01 04 00 00 08 F1 CC | Slave Response: 01 04 10 00 41 A0 00 41 20 00 42 C8 00 43 48 81 E8
Note:
- [01] Represents the instrument communication address;
- [04] Represents function code 04;
- [10] represents 10H (16) byte data;
- [00 00 00 41 A0] = 20.0; / Main Measured value
- [00 00 4120]= 10.0; // Temperature Value
- [00 00 42 C8] = 100.0; / / Main measured voltage value
- [00 00 43 48] = 200.0; / / Temperature & Voltage Values
- [81 E8] represents CRC16 check code;
Maintenance
According to the requirements of use, the installation position and working condition of the controller are relatively complex. In order to ensure that the controller is working normally, maintenance personnel should carry out regular maintenance on the controller. Please pay attention to the following matters during maintenance:
- Check the working environment of the controller. If the temperature exceeds the rated range of the controller, please take appropriate measures; otherwise, the controller may be damaged or its service life may be reduced;
- When cleaning the plastic shell of the controller, please use a soft cloth and a soft cleaner to clean the shell.
- Check whether the wiring on the terminal of the controller is firm. Pay attention to disconnect the AC or DC power before removing the wiring cover.
Warranty, Returns and Limitations
Warranty
Icon Process Controls Ltd warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Icon Process Controls Ltd for a period of one year from the date of sale of such products. Icon Process Controls Ltd obligation under this warranty is solely and exclusively limited to the repair or replacement, at Icon Process Controls Ltd option, of the products or components, which Icon Process Controls Ltd examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Icon Process Controls Ltd must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the one year from the date of replacement.
Returns
Products cannot be returned to Icon Process Controls Ltd without prior authorization. To return a product that is thought to be defective, go to www.iconprocon.com, and submit a customer return (MRA) request form and follow the instructions therein. All warranty and non-warranty product returns to Icon Process Controls Ltd must be shipped prepaid and insured. Icon Process Controls Ltd will not be responsible for any products lost or damaged in shipment.
Limitations
This warranty does not apply to products which:
- are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures outlined above;
- have been subjected to electrical, mechanical or chemical damage due to improper, accidental or negligent use;
- have been modified or altered;
- anyone other than service personnel authorized by Icon Process Controls Ltd have attempted to repair;
- have been involved in accidents or natural disasters; or
- are damaged during return shipment to Icon Process Controls Ltd
Icon Process Controls Ltd reserves the right to unilaterally waive this warranty and dispose of any product returned to Icon Process Controls Ltd where:
- there is evidence of a potentially hazardous material present with the product;
- or the product has remained unclaimed at Icon Process Controls Ltd for more than 30 days after Icon Process Controls Ltd has dutifully requested disposition.
This warranty contains the sole express warranty made by Icon Process Controls Ltd in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL Icon Process Controls Ltd BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF Icon Process Controls Ltd. This warranty will be interpreted pursuant to the laws of the province of Ontario, Canada.
If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty.
For additional product documentation and technical support visit: www.iconprocon.com
- e-mail: sales@iconprocon.com or support@iconprocon.com
- Ph: 905.469.9283
Corrosion-Free Instrumentation Equipment
25-0732 © Icon Process Controls Ltd.
FAQ
- Q: How do I ensure safety during installation?
A: Always wear safety goggles or face-shield and follow maximum temperature and pressure specifications. - Q: Who should handle the installation and operation of the controller?
A: Only skilled, trained, or authorized personnel should handle installation, setup, and operation.
Documents / Resources
![]() |
ICON PROCESS CONTROLS IS-750D Series ProCon Free Chlorine Display Controller [pdf] User Manual IS-750D, IS-750D Series ProCon Free Chlorine Display Controller, IS-750D Series, ProCon Free Chlorine Display Controller, Chlorine Display Controller, Display Controller, Controller |