RICE LAKE SCT-4X-MODBUS TCP IP Weight Transmitter User Guide

Electrical Schematic

Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals
Warranty information is available at www.ricelake.com/warranties

Key Functions

Indicator Light Descriptions

Configuration Menu

  1. Reboot the weight transmitter
  2.  Press the key when display shows the 888888 message

Function Mode

. Maximum Scale Capacity, Increment and Decimal Point Settings

Zeroing Mechanical Tare (pre-tare zeroing)

Inputs Settings

RS485 Port

Load Cell Exclusion (for dependent channel systems)

If a load cell is broken, it’s possible to temporarily exclude the channel where it is connected and continue to weigh, pending replacement.
! WARNING: this operation reduces the accuracy of the weighing system. We recommend use for liquid weighing or in applications
where the load is evenly distributed.

Example:

The unbalance condition is signalled via Modbus / Fieldbus or a digital output (refer to the complete manual to set the output).
This function is only available if EXCL.CH = NoNE.
Use this function only in systems where the load is evenly distributed

Programming Errors

MESSAGE DESCRIPTION SOLUTION
preC. Calibration error First calibrate the zero point (zero), then proceed with sample weight acquisition (span) (section 9).
Err.pn  

Calibration error

Check the connection of the load cell. Verify the load cell signal is stable, valid and greater than the previously acquired point.
Er 11 Calibration error Increase the calibration weight.
Er 12 Calibration error Check the signal from the load cell increases when weight is incremented on the scale.
Er 37 Calibration error Repeat calibration and verify capacity and division have been correctly set.
Er 39 Instrument not configured Transmitter needs to be configurated.
C.er.36 Calibration error Verify the signal from the load cell is not negative.
C.er.37 Calibration error Verify the signal from the load cell is not negative.
 

Err.mo

 

Weight unstable

Check in adC.uv parameter that the signal is stable.

If the connection of the cells is with 4 wires, check that the sense jumpers are inserted.

adC.err A/D converter error Converter failure. Reboot the instrument.
Cel.err Global load cell error Signal anomaly: check the load cells connection.
Er.Cel.1

Er.Cel.4

 

Load cell error

 

Signal anomaly: check the indicated load cell connection.

Modbus TCP

MODBUS TCP REGISTERS CH / ind.ch (1 SCALE)

Data Register DESCRIPTION
 

Gross weight

30001  

Bytes 1, 2, 3 and 4 contain the Gross Weight value.

30003
 

Net weight

30005  

Bytes 5, 6, 7 and 8 contain the Net Weight value.

30007
 

 

 

 

 

 

 

Input status register

 

 

 

 

 

 

 

30009

Bit 15(msb) Bit 14

Bit 13

Bit 12

Bit 11

Bit 10

Bit 9

Bit 8 (lsb)

Active channel.                                                       Bit 15 Bit 14     Active Channel

Active channel.

No function.                                                               0          0             Channel 1

No function.                                                               0          1            Channel 2

No function.

No function.                                                               1         0             Channel 3

Status of input n. 2.                                                    1          1            Channel 4

Status of input n. 1.

Bit 7(msb) Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0 (lsb)

1 = Scale unloaded (gross weight = 0). Tare PT (1 = PT tare is active).

Tare (1 = Tare is active).

Overload condition (0 = No; 1 = Overload). Underload condition (0 = No; 1 = Underload). Weight Stability (0 = Unstable; 1 = Stable).

Gross Weight Polarity (0 = “+”; 1 = “-”). Net Weight Polarity (0 = “+”; 1 = “-”).

 

 

 

Command status register

 

 

 

 

30011

Last received command.
Bit 7(msb) Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0 (lsb)

Last command result. Last command result. Last command result. Last command result.

Counting of processed commands. Counting of processed commands. Counting of processed commands. Counting of processed commands.

 

 

Output status register

 

 

 

30013

No Function.
Bit 7(msb)

Bit 2

Bit 1

Bit 0(lsb)

No function.

No function.

Digital output 2 status (0 = OFF; 1 = ON). Digital output 1 status (0 = OFF; 1 = ON).

Selected page 30015 Shows the value of the selected page (3001).
µV Channel 1 30017 Bytes 16 and 17 contain the µV value of the channel 1.
µV Channel 2 30019 Bytes 18 and 19 contain the µV value of the channel 2.
µV Channel 3 30021 Bytes 20 and 21 contain the µV value of the channel 3.
µV Channel 4 30023 Bytes 22 and 23 contain the µV value of the channel 4.

MODBUS TCP REGISTERS ch (4 SCALES)

Data Register DESCRIPTION
 

 

 

 

 

 

 

Status register scale 1

 

 

 

 

 

 

 

30001

Bit 15(msb) Bit 14

Bit 13

Bit 12

Bit 11

Bit 10

Bit 9

Bit 8 (lsb)

Not used. Not used. Not used.

Scale active (0 = “no”; 1 = “yes”). Decimals (00 = 0; 01 = 1; 10 = 2; 11 = 3)

Unit of Measure (00 = “g”; 01 = “kg”; 10 = “t”; 11 = “lb”).

Bit 7(msb) Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0 (lsb)

Tare PT (1 = PT tare is active). Tare (1 = Tare is active).

Net Weight Polarity (0 = “+”; 1 = “-”).

1 = Scale unloaded (gross weight = 0). Overload condition (0 = No; 1 = overload). Underload condition (0 = No; 1 = underload). Stability (0 = “unstable”; 1 = “stable”).

Gross Weight Polarity (0 = “+”; 1= “-”).

Status register scale 2 30002 Same as Status register scale 1.
Status register scale 3 30003 Same as Status register scale 1.
Status register scale 4 30004 Same as Status register scale 1.
 

 

 

Input status register

 

 

 

 

30005

Bit 15(msb) Bit 14

Bit 13

Bit 12

Bit 11

Bit 10

Bit 9

Bit 8 (lsb)

Active channel. Active channel. No function.

No function. No function. No function.

Status of input n. 2. Status of input n. 1.

     
     
     
     
     
No function.
Command status register 30006 Same as Command status register page 10.
Output status register 30007 Same as Output status register page 10.
Selected page 30008 Shows the value of the selected page (2000).
 

Gross weight scale 1

30009  

Bytes 16, 17, 18 and 19 contain the Gross Weight of scale 1.

30010
 

Gross weight scale 2

30011  

Bytes 20, 21, 22 and 23 contain the Gross Weight of scale 2.

30012
 

Gross weight scale 3

30013  

Bytes 24, 25, 26 and 27 contain the Gross Weight of scale 3.

30014
 

Gross weight scale 4

30015  

Bytes 28, 29, 30 and 31 contain the Gross Weight of scale 4.

30016

DEVICENET REGISTERS FOR COMMAND SENDING

Data Register DESCRIPTION
 

 

 

 

 

 

 

Command

 

 

 

 

 

 

 

30001

Main available commands:

 

Value                 Command 00 Hex     No command 01 Hex          Scale zeroing 02 Hex                 Tare

03 Hex      Preset Tare

0A Hex     Setpoint 1 setting 0B Hex                 Setpoint 2 setting

19 Hex       Digital output setting

22 Hex     Reboot the weight transmitter

 

 

Parameter 1

 

30002

 

First parameter of the command.

Parameter is always expressed in absolute mode (no decimals, no sign).

 

30003

 

 

Parameter 2

 

30004

 

Second parameter of the command.

Parameter is always expressed in absolute mode (no decimals, no sign).

 

30005

  30006  

Used in advanced configuration, refer to the complete Fieldbus manual for further information.

30016

EXAMPLE 1

For zeroing the weight on the scale:
Set the command in byte 2

Register Value
 

30001

 

01 Hex

EXAMPLE 2

For setting a preset tare of 1000 lb:

  1. Set the tare value in parameter 1 (byte 3, 4, 5, 6)
  2. Set the command in byte 2
Register Value
 

30001

 

03 Hex

 

30002

 

00 Hex

 

30003

 

03E8 Hex

 

 

Read More About This Manual & Download PDF:

Documents / Resources

RICE LAKE SCT-4X-MODBUS TCP IP Weight Transmitter [pdf] User Guide
SCT-4X-MODBUS TCP IP Weight Transmitter, SCT-4X-MODBUS TCP, IP Weight Transmitter, Weight Transmitter, Transmitter