FlowTrax FT-2 Aliran, Pengujian Volume

FlowTrax -FT-2 -Flow-Volume-Testing-PRODACT-IMG

Penting nggunakake banyu suling utawa deionisasi ONLYFlowTrax- FT-2 -Flow-Volume-Pengujian-FIG-1

8 A. Pencet FUNCTION banjur tombol lembut CLEAN GLASS (Gbr. 2). B. Tarik alon-alon
tab ireng ing ndhuwur kaca Tube (Fig. 2) kanggo mbusak saka unit. C. Nggunakake solusi reresik & sikat kasedhiya, sregep scrub nang jero kabeh dawa saben tabung minimal 30 kaping.

D. Bilas Tabung Kaca nggunakake Botol Flush sing diisi cairan sing disetujoni. E. Garing njaba Tabung Kaca. Instal maneh menyang FT-2. Tekan EXIT soft key.FlowTrax- FT-2 -Flow-Volume-Pengujian-FIG-2

  • Bukak Occlusion Knob counterclockwise. Pencet tombol FUNCTION, tinimbang pencet tombol alus FLUSH. Jalanake pompa IV ing 900 ml / jam, volume 20 ml kanggo flush FT-2.
  • Program ulang pompa IV dadi 600 ml / jam, volume 50 ml. Miwiti pompa. Pencet tombol alus EXIT sawise 5 ml. FT-2 bakal ngasilake gelembung udara, pesen Perdhana Dibutuhake ilang lan nilai aliran ditampilake nalika rampung. Stop pompa IV.
  • Pencet tombol RESET FLOW kanggo mbusak asil sadurunge. Unit siap kanggo tes.
  1. Pencet UNITS kanggo milih unit tekanan sing dikarepake.
  2. Pencet RESET FLOW nganti nol nilai tekanan MAX.
  3. Program pompa IV menyang tingkat sing dikarepake lan miwiti pompa IV.
  4. Occlusion Elektronik: Pencet tombol Fungsi, banjur tombol Occlude 40 PSI, tekanan puncak kanthi grafik ditampilake. Pencet EXIT kanggo bali menyang layar utama.
  5. Kanggo test occlusion manual: Puter occlusion knob searah jarum jam nganti IV pump weker utawa occlusion knob flush karo panel ngarep. (Aja kenceng.)
    Cathetan: Zero Pressure biasane mung dibutuhake ing wiwitan saben dina kanthi FlowTrax tanpa cairan. Waca Op. Manual bagean Occlusion Testing kanggo informasi luwih lengkap.

TEKANAN METER TESTING

  1. Pencet FUNCTION, banjur tombol CLEAN GLASS kanggo nguras banyu.
  2. Puter occlusion knob searah jarum jam driji nyenyet. (Aja kenceng).
  3. Pencet UNITS kanggo milih unit tekanan sing dikarepake.
  4. Pencet tombol ZERO, banjur RESET tombol FLOW menyang tekanan MAX nol. Sambungake piranti ing test kanggo Input Luer. Wektu nyata lan tekanan Puncak bakal diukur.FlowTrax- FT-2 -Flow-Volume-Pengujian-FIG-3