Daviteq MBRTU-PHFLAT Datar pH Sensor Modbus Output
Pambuka
MBRTU-PHFLAT minangka sensor pH in-line (pengukuran terus-terusan) kanthi tujuan umum sing nawakake jangkoan aplikasi sensor datar sing terjangkau kanggo akeh aplikasi banyu kayata banyu ngombe, banyu industri, akuakultur, instalasi tank utawa aplikasi sing gegandhengan. Sensor lumahing warata asring diarani minangka "self-cleaning" nalika dipasang ing aliran proses amarga aliran banyu cenderung "nyukur" biofouling lan lebu liyane saka permukaan sensor. Output yaiku Modbus RTU supaya gampang digabung karo PLC, controller, SCADA, BMS utawa gateway IoT.
- Elektroda pH sing kuat kanggo pangukuran terus-terusan;
- Elektroda warata self-cleaning;
- Output ModbusRTU standar;
- Plug & Puter.
Aplikasi khas: Banyu ngombé, Banyu Limbah, Banyu Industri, Akuakultur,…
- PROSES PH FLAT SENSOR DENGAN KOMPENSASI SUHU MBRTU-PHFLAT
Spesifikasi
SPESIFIKASI PH SENSOR |
|
Teknologi Sensing |
Kaca, gabungke elektroda karo sensor suhu Pt100 |
Range pengukuran |
pH 0.. 14 |
Resolusi | pH 0.1 |
Akurasi | +/- 0.1 |
Suhu kerja |
0 .. 100 oC (kompensasi) |
Tekanan kerja | 0.. 100 psig |
Sambungan proses |
3/4″ NPT loro ujung |
Bagean teles |
PVC |
Kabel Sensor Kab |
6m karo konektor BNC |
Rating |
IP68 |
Ukuran Sensor |
D27 x 172 (mm) |
Sensor bobot net |
<200 gram |
SPESIFIKASI PH TRANSMITTER |
|
Input |
pH lan Pt100 |
Output |
RS485, Modbus RTU protokol, max 19200 baud |
sumber daya |
9..36VDC, rata-rata. <200mA |
Pemasangan |
DIN Rel |
Suhu kerja |
-40 .. 85 oC |
Kelembapan kerja |
0 .. 95% RH, non-kondensasi |
omah |
Plastik Rekayasa |
Ingress Protection |
IP20 |
ukuran |
93 x 40 (mm) |
bobot net |
<200 gram |
ukuran
Wiring
- Mangga kabel minangka kapacak ing ngisor iki:
Memmap ndhaptar
Kode Fungsi: 3 (Waca); 16 (Tulis)
Cathetan 1: Sadurunge kalibrasi sensor, atur calibEnb = 1, sawise kalibrasi sensor, atur calibEnb = 0
Cathetan 2:
Langkah 1: Kalibrasi nganggo larutan standar pH = 7.01, lan ngenteni udakara 3 menit nganti maca sensor stabil, lan tulis 7 kanggo umpan balik.
Langkah 2: Kalibrasi nganggo larutan standar pH = 4.01 utawa 10.01, ngenteni udakara 3 menit nganti wacan stabil, lan tulisake nilai pH pHFeedback saka solusi kasebut.
Cathetan 3: Yen ora ana sensor suhu, setel Manual_Temp_Enb = 1, banjur Suhu (prm 2) = Manual_Temp_Input
Peralatan lan Solusi sing dibutuhake
- pH meter (kompensator suhu dianjurake kanggo nambah akurasi pangukuran pHm)
- pH Buffer 4.01 utawa 10.01
- pH Buffer 7.01
- Solusi ngisi referensi (ndeleng dhaptar kanggo jinis elektroda pH tartamtu)
- Botol wisuh diisi banyu suling utawa de-ionisasi
- Laboratorium Magnetik stirrer lan Magnetik nglakoake bar
- Tisu lab
- Beakers resik
Preparation saka Elektroda kanggo nggunakake dhisikan
- Copot botol protèktif utawa tutup saka elektroda lan sak tenane mbilas elektroda nganggo banyu suling. Ngilangke kanthi teliti nganggo lap lab sing resik.
- Sajrone pangiriman, gelembung udara bisa uga wis migrasi menyang bohlam sensor elektroda. Tahan elektroda nganti cahya lan priksa bohlam sensing kanggo gelembung udara. Yen hawa katon, goyangake elektroda mudhun kanthi ati-ati (kaya termometer) kanggo ngilangi gelembung udara saka bohlam sensor ing pucuk elektroda.
- Kanggo model sing bisa diisi ulang, bukak port ngisi kanggo mbukak bolongan ngisi ruang referensi elektroda (kanggo elektroda sing disegel, diisi gel, ora nggatekake operasi iki).
- Isi ruangan referensi kanthi solusi ngisi referensi pH sing cocog. Elektroda sing wis diisi karo solusi pangisi sing salah ora dijamin ing garansi.
Pamilihan Solusi Ngisi Referensi
- Kanggo elektroda kombinasi pH karo Calomel lan Double Junction Ag/AgCI setengah sel referensi, gunakake solusi ngisi referensi 4 M KCI.
- Kanggo elektroda kombinasi pH kanthi sel setengah referensi Ag/AgCI tunggal, gunakake 4 M KCI jenuh karo solusi pangisi referensi AgCI
Kalibrasi
- Sadurunge digunakake, instrumen kasebut kudu dikalibrasi karo solusi standar;
- Kapisan yaiku kalibrasi kanthi larutan standar pH = 7.01. Tuku solusi standar ing kene:
- https://www.hannavietnam.com/detail-product/chuan-ph-701-500ml-19
- https://www.hannainstruments.co.uk/ph-7-01-buffer-solution.html
- Langkah sabanjure yaiku kalibrasi nganggo larutan standar pH = 4.01 utawa 10.01. Tuku solusi standar ing kene:
- https://www.hannavietnam.com/detail-product/chuan-ph-1001-500ml-20
- https://www.hannainstruments.co.uk/ph-10-01-buffer-solution.html
Cathetan suhu sekitar nalika kalibrasi kanggo input nilai pH standar miturut suhu (dituduhake ing awak botol standar)
Gunakake alat master modbus kanggo kalibrasi sensor. Utawa gunakake piranti lunak Modbus Daviteq, bebarengan karo kabel konfigurasi…
Cara nggunakake Modbus Configuration Tool (Klik kene)
Langkah 1: Ngimpor Cithakan file menyang Modbus Configuration Tool
Langkah 2: Selehake sensor pH menyang beaker sing ngemot pH = 7.01 buffer nanging durung dikalibrasi
Langkah 3: Enteni kira-kira 3 menit kanggo maca saka sensor kanggo stabil, lan nulis 7 mudhun menyang saran Register 300 lan nilai=1 kanggo phcalibEnb Reg 299 nggunakake Func 16
Langkah 4: Copot elektroda saka buffer. Cuci nganggo banyu sing disuling lan blot nganggo lap lab banjur lebokake standar pH = 10.01
Langkah 5: Enteni kira-kira 3 menit kanggo maca stabil, tulisake umpan balik kanthi 10 menyang Reg 300
Langkah 6: Tulis pHcalibEnb kanthi 0 menyang Reg 299 kanggo mungkasi kalibrasi. Banjur mbilas elektroda nganggo banyu sing disuling lan usap kanthi teliti nganggo lap lab sing resik.
Maca Sample karo Elektroda
- Cuci elektroda nganggo banyu sing disuling lan mbilas nganggo lap lab. Selehake elektroda ing beaker ngemot sample lan bar nglakoake. Ing sample kudu ing suhu padha buffer digunakake kanggo standarisasi elektroda. Nglakoake kaya sadurunge. Rekam pH nalika maca stabil.
- Copot elektroda saka sample, lan mbilas elektroda karo banyu disaring liwat beaker "sampah". Blot elektroda garing karo ngusap lab. Elektroda saiki siap kanggo maca pH saka s liyaneamples.
Nyimpen Elektroda
Term Urut
- Ing antarane pangukuran, simpen elektroda pH ing beaker sing ngemot buffer pH 4.01.
Jangka Panjang
- Nalika nyimpen kanggo wektu maneh, simpen elektroda pH ing botol panyimpenan utawa boot protèktif sing teka karo elektroda. Priksa manawa umpluk ing botol panyimpenan utawa bal katun ing boot protèktif wis wetted karo solusi panyimpenan pH kanggo njaga lingkungan teles watara bohlam pH lan prapatan. Njaga lingkungan teles ing botol panyimpenan utawa ing boot protèktif sak panyimpenan.
- Yen elektroda pH dilengkapi tutup bolongan pangisi, tutup tutup ing bolongan pangisi.
Reresik Elektroda
Aja nggunakake pelarut sing kuwat (contone, aseton, karbon tetraklorida, lsp.) kanggo ngresiki elektroda pH. Dadi manawa kanggo kalibrasi ulang elektroda sawise ngresiki.
- Yen elektroda wis dilapisi lenga utawa pelumas, cuci elektroda kanthi ati-ati ing banyu sing adhem kanthi nggunakake deterjen pencuci piring. Cuci sak tenane nganggo banyu seger banjur mbilas nganggo banyu suling. Rendhem elektroda ing larutan panyimpenan elektroda pH suwene 30 menit sawise prosedur reresik iki. Kalibrasi ulang elektroda sadurunge digunakake.
- Yen elektroda wis kena protein utawa bahan sing padha, rendhem ing pepsin asam suwene 5 menit. Cuci sak tenane nganggo banyu sing disaring. Rendhem ing solusi panyimpenan kanggo 30 menit sadurunge kalibrasi maneh.
- Yen prosedur reresik sadurunge gagal mulihake respon, rendhem elektroda ing 0.1 N HCI kanggo 30 menit. Cuci sak tenane nganggo banyu sing disaring. Recalibrate sadurunge digunakake.
- Yen respon elektroda ora dibalèkaké, ngganti elektroda.
Kontak
- Produsen
- Daviteq Technologies Inc
- No.11 Street 2G, Nam Hung Vuong Res., An Lac Ward, Binh Tan Dist., Ho Chi Minh City, Vietnam.
- Tel: +84-28-6268.2523/4 (ext.122)
- Email: info@daviteq.com | www.daviteq.com
- Revisi #5
- Digawe Jum, 9 Jul 2021 8:51 dening Kiệt Anh Nguyễn
- Dianyari Sen, 13 Desember 2021 2:53 dening Kiệt Anh Nguyễn
Dokumen / Sumber Daya
![]() |
Daviteq MBRTU-PHFLAT Datar pH Sensor Modbus Output [pdf] Pandhuan pangguna MBRTU-PHFLAT Sensor pH Datar Modbus Output, MBRTU-PHFLAT, Datar pH Sensor Modbus Output, Sensor Modbus Output, Modbus Output, Output |