ANKROYU DO9100

ANKROYU DO9100 Dissolved Oxygen Meter Instruction Manual

Model: DO9100

1. Introduction

The ANKROYU DO9100 is a portable digital dissolved oxygen meter designed for accurate measurement of dissolved oxygen concentration (mg/L), dissolved oxygen saturation (%), and temperature in various water environments. It features a built-in microcomputer processor for precise data calculation and compensation, a clear digital display with backlight, and a detachable probe for ease of use and maintenance. This instrument is suitable for applications in aquariums, swimming pools, freshwater aquaculture, marine aquaculture, school experiments, laboratory settings, and scientific research.

Key features include:

2. Safety Information

Please read and understand all safety instructions before operating the device. Failure to follow these instructions may result in injury or damage to the instrument.

3. Package Contents and Product Overview

Verify that all items are present in the package:

Product Diagram

ANKROYU DO9100 Dissolved Oxygen Meter with probe and cap
Figure 3.1: ANKROYU DO9100 Dissolved Oxygen Meter with its protective cap removed, showing the probe connection area.

The main unit features a digital display, power button, HOLD/TEMP button, and MODE/CAL button. The detachable probe connects to the bottom of the main unit.

ANKROYU DO9100 Dissolved Oxygen Meter and separate probe
Figure 3.2: The ANKROYU DO9100 main unit alongside its separate dissolved oxygen probe, connected by a cable.
Close-up of ANKROYU DO9100 display and control buttons
Figure 3.3: A close-up view of the ANKROYU DO9100's digital display and control buttons, including Power, HOLD/TEMP, and MODE/CAL.

4. Setup

4.1 Battery Installation

The ANKROYU DO9100 requires batteries (not included). To install:

  1. Locate the battery compartment cover on the back of the main unit.
  2. Open the cover.
  3. Insert the required batteries, ensuring correct polarity (+/-).
  4. Close the battery compartment cover securely to prevent water damage.

4.2 Electrode Connection

Connect the dissolved oxygen electrode to the main unit:

  1. Remove the protective cap from the electrode connector on the main unit.
  2. Align the electrode cable connector with the port on the main unit.
  3. Gently push and twist to secure the connection. Ensure it is firmly attached to prevent signal loss.
Close-up of the probe connection point on the ANKROYU DO9100 meter
Figure 4.1: A detailed view of the probe connection area at the bottom of the ANKROYU DO9100 meter, where the dissolved oxygen electrode attaches.

4.3 Membrane Group Installation (Initial Setup/Replacement)

The dissolved oxygen electrode uses a membrane group. Ensure it is properly installed before use.

  1. Unscrew the outer cap of the dissolved oxygen electrode.
  2. Carefully place a new membrane group onto the electrode tip, ensuring it is flat and free of air bubbles.
  3. Screw the outer cap back onto the electrode, tightening it gently until snug. Avoid overtightening.

5. Operating Instructions

5.1 Powering On/Off

5.2 Measurement Procedure

Before taking measurements, ensure the electrode is properly connected and a membrane group is installed.

  1. Power on the meter.
  2. Immerse the dissolved oxygen electrode into the water sample you wish to test. Ensure the membrane tip is fully submerged and free of air bubbles.
  3. Allow a few moments for the reading to stabilize. The display will show the dissolved oxygen concentration (mg/L), dissolved oxygen saturation (%), and temperature (℃/℉).
  4. For best accuracy, gently stir the probe in the water sample to ensure fresh water flows past the membrane.
ANKROYU DO9100 Dissolved Oxygen Meter measuring in a fishbowl
Figure 5.1: The ANKROYU DO9100 meter in use, with its probe submerged in a fishbowl to measure dissolved oxygen levels.
ANKROYU DO9100 Dissolved Oxygen Meter measuring in a natural water body
Figure 5.2: The ANKROYU DO9100 meter being used to measure dissolved oxygen in a natural outdoor water environment.

5.3 HOLD/TEMP Button

5.4 MODE/CAL Button (Calibration)

The MODE/CAL button is used for calibration. Refer to the detailed calibration procedure in the full manual for precise instructions. Regular calibration ensures measurement accuracy.

Close-up of the control buttons on the ANKROYU DO9100 meter
Figure 5.3: A close-up of the ANKROYU DO9100's control buttons, highlighting the HOLD/TEMP and MODE/CAL functions.

6. Maintenance

6.1 Electrode Care and Cleaning

6.2 Membrane Replacement

The dissolved oxygen membrane group is a consumable part and needs periodic replacement to maintain accuracy. Replace the membrane if readings become unstable, inaccurate, or if the membrane is visibly damaged.

  1. Unscrew the outer cap of the dissolved oxygen electrode.
  2. Carefully remove the old membrane group.
  3. Install a new membrane group as described in Section 4.3.
  4. After replacing the membrane, it is recommended to recalibrate the meter.

6.3 Storage

7. Troubleshooting

ProblemPossible CauseSolution
Meter does not power onDead or incorrectly installed batteriesCheck battery polarity, replace batteries.
Inaccurate or unstable readings
  • Dirty or damaged membrane
  • Electrode not calibrated
  • Air bubbles on membrane
  • Interference
  • Clean or replace membrane
  • Perform calibration
  • Gently stir probe to remove bubbles
  • Move away from strong electrical devices
Low battery warning displayedBatteries are lowReplace batteries immediately.
No temperature readingFaulty thermistor or connectionEnsure probe is securely connected. If problem persists, contact support.

8. Specifications

ParameterValue
ModelDO9100
Measuring Range (Oxygen concentration)0.0-40.0 mg/L
Measuring Range (Dissolved Oxygen Saturation)0.0-300.0%
Measuring Range (Temperature)0.0-40.0℃ (32.0-104.0℉)
Resolution (Oxygen concentration)0.1 mg/L
Resolution (Dissolved Oxygen Saturation)0.10%
Resolution (Temperature)0.1℃ / 0.1℉
Accuracy (Oxygen concentration)±1.5% F.S
Accuracy (Dissolved Oxygen Saturation)±1.5% F.S
Accuracy (Temperature)±0.5℃
Response Time45 seconds (at 20℃)
Temperature Compensation RangeAutomatic (0-40℃)
Height CompensationWithout
Battery PowerLow battery warning
Automatic Shut DownApproximately 15 minutes
Use Environment0.1-40.0℃, RH maximum 90%
MaterialABS
Package Dimensions7.87 x 7.09 x 1.97 inches
Item Weight11.25 ounces

9. Warranty & Support

This ANKROYU product is covered by a standard manufacturer's warranty. Please refer to the specific warranty terms provided at the time of purchase or contact your retailer for details regarding warranty coverage and duration.

For technical support, troubleshooting assistance, or inquiries about replacement parts, please contact the seller or manufacturer directly. Keep your purchase receipt or order information handy when seeking support.

Related Documents - DO9100

Preview Портативный оксиметр Yinmik DO9100, BLE9100: Руководство пользователя
Полное руководство пользователя для портативного оксиметра Yinmik моделей DO9100 и BLE9100. Охватывает технические характеристики, калибровку, эксплуатацию, обслуживание и гарантию.