UNI-T UT331 Plus Digital Temperature Humidity Meters Instruction Manual

UT331 Plus Digital Temperature Humidity Meters

Specifications:

  • Model: UT331+/UT332+
  • P/N: 110401109799X
  • Features: High precision sensor, super-low power consumption
    microprocessor, manual and auto storage, USB data transfer, USB
    power supply, backlight alarm, real-time display, low power
    consumption
  • Applications: Factories, laboratories, research institutes,
    food and health, environmental protection

Product Usage Instructions:

Meter Description:

The UT330+ series temperature humidity meters feature a high
precision sensor and super-low power consumption microprocessor.
The meter has various functions including real-time display, manual
and auto storage, USB data transfer, USB power supply, backlight
alarm, and more.

Operating Instructions:

  1. Press the power button to turn on/off the meter.
  2. Use the buttons to navigate and select different functions such
    as saving data, freezing displayed readings, setting alarms, and
    more.
  3. Short press the SETUP button to enter setting mode and navigate
    through different setup options.
  4. Long press certain buttons for specific functions like turning
    on/off the backlight or alarm.

Setting Meters:

Follow the below steps to perform the setups:

  1. Setting Temperature Unit:
    • Short press SETUP to access temperature unit setup mode.
    • Press (MAX/MIN) to select Celsius or Fahrenheit unit.
  2. Setting Time Unit:
    • Short press SETUP to switch to time interval mode.
    • Press to change the interval for automatic data storage.
    • Long press to increase/decrease the time quickly.

FAQ:

Q: What are the applications of the UT330+ series temperature
humidity meters?

A: These meters can be used in factories, laboratories, research
institutes, food and health, environmental protection, and other
fields.

Q: How many pieces of batteries are included in the unpacking
inspection?

A: Four pieces of batteries are included.

P/N:110401109799X

UT331+/UT332+ Digital Temperature Humidity Meters
Overview
UT330+ series temperature humidity meters (hereafter referred to as “Meters”) adopts high precision sensor and super-low power consumption microprocessor in its design, featuring high accuracy, manual and auto storage, USB data transfer, USB power supply, backlight alarm, real-time display, low power consumption, etc. These Meters can be used in factories, laboratories, research institutes, food and health, environmental protection and other fields.
Unpacking Inspection
Operating Manual ——————————— 1 piece Battery ———————————————– 4 pieces USB Data Cable ———————————– 1 piece
Safety Information
Warning A warning identifies conditions and actions that pose hazards to the use.rTo avoid electrical shock or personal injury, follow these guidelines:
Inspect the case before using Meters. Do not use Meters if it appears damaged. Look for cracks or missing plastic. Pay particular attention to the insulation around the connectors. Replace the batteries as soon as the battery indicator ” ” appears. Do not use Meters if it operates abnormally. Protection may be impaired. When in doubt, have Meters serviced. Do not operate Meters around explosive gas, vapor, or dust. When servicing Meters, use only specified replacement parts. Do not use Meters with any part of the case or cover removed. Do not clean the sensor with water. Turn to close the protective sensor cover when not in use. To avoid damage to the device, please use with care. Never charge the battery. Use the approved alkaline battery (AAA 1.5V) only. Pay attention to “+” and “-” marks when installing the batteries. Close the sensor cover and remove the batteries if Meters is not used for a long time.
Meter Description
Meter Structure (See Figure 1)

No.

Description

Protective cover of sensor
1 (turn to open or close)

2 Display screen

3 Power on/off

Press to display Max/Min value among
4 current readings. Long press to exit the
function (also used as button).

5 Press to save data. Refer to “Store and
Recall Data” for details

6

To set Meters. It is used together with buttons. See “Setting Meters” for details.

Short press to turn on/off the backlight,
7 long press to turn on/off alarm function.

Figure 1

8

Press to freeze the displayed readings, press again to exit the function. (also used as button)

9

Press to view data, it is used together with Recall Data” for details.

buttons. See “Store and

10

Dew-point temperature and wet bulb temperature selection key (also used as confirm button).

Display Screen

Figure 2

No.

Description

1 HI: Upper limit. LO: Lower limit

2 Dew-point temperature

3 Wet-bulb temperature

4 Storage group number

5 Maximum value

6 Minimum value

7 Humidity display zone

8 Data hold

9 Humidity unit

No.

Description

10 Auto power off

11 Low battery/USB power supply

12 Alarm

13 Humidity icon

14 Temperature icon

15 Celsius unit

16 Temperature display zone

17 Fahrenheit unit

18 Time display zone

Setting Meters
SETUP Options Short press SETUP button to enter into setting mode, first enter into unit setting page, and then press recyclable to switch the setting state until the setting is storage and exit. On any setting page, the setting mode will exit and be back to measuring page if no any button is pressed for 15 seconds. The cycle sequence:

Measuring stateTemperature unit setup Time interval setupTemperature & Humidity Upper Limit Alarm SetupTemperature & Humidity Lower Limit Alarm SetupSleep time setupSystem time setupTemperature & Humidity Offset setup Measuring state.

Performing the Setups
1. Setting Temperature Unit Short press SETUP to access temperature unit setup mode, press
(MAX/MIN) to select or unit.

(HOLD)/

2. Setting Time Unit Short press SETUP to switch to time interval mode, “int” shows on LCD. Press to change the interval for automatic data storage. Long press to increase /decrease the time quickly. The interval can be set up to 59:59 and down to 00:00 (auto data storage disabled at 00:00).

3. Upper-limit temperature & humidity setting Short press SETUP for 3 times to access upper limit setup mode, “HI” shows on LCD, short press ENTER/DP to switch upper-limit temperature & humidity, upper-limit dew-point temperature, upper-limit wet-bulb temperature, press to change the value.

4. Lower-limit temperature & humidity setting Short press SETUP for 4 times to switch to lower limit setup mode, “LO” shows on LCD, short press ENTER/DP to switch following setups: lower-limit temperature & humidity, lower-limit dew-point temperature, lower-limit wet-bulb temperature, press to change the value.

5. Sleep Mode Setting Short press SETUP for 5 times to switch to sleep setting mode, “SLP” icon shows, press to set up sleep time (5~60mins), long press to increase/decrease
the time quickly. When the set time is 5mins, “OFF” icon shows and sleep function is disabled.

6. System Time Setting Short press SETUP for 6 times to switch to system time setting mode, short press ENTER/DR to switch Year/Month/Day/Hour/Minute, press to the time value, long press to increase/decrease the time quickly. Only when the minute setting is completed and short press ENTER/DP again can the system time is stored.

7. Setting Temperature & Humidity Offset Short press SETUP for 7 times to switch to temperature & humidity offset setting mode, “t-o” shows on LCD, short press ENTER/DP to switch temperature offset setting (t-o), humidity offset setting (h-o), dew-point temperature offset setting (d-o), wet bulb temperature ofset setting (b-o), press to change ofset value
that within -6~6.

Using Meters
Turn on/off Meters 1. Long press yellow button to turn on Meters. 2. With Meters already turned on, long pressing yellow button again will turn off
Meters. 3. After the sleep time is set, Meters will auto turn of f if no button is active within
the set time. 4. To ensure an accurate result, the user is recommended to wait for a while before
taking the readings if the measured humidity & temperature change dramatically.
Data Hold 1. Press HOLD button to freeze the reading, “HOLD” appears on LCD. 2. Press HOLD again to quit the function.

Turn on/off Backlight Short press backlight button to turn on the backlight, press again to turn it off.
Long press backlight button to activate alarm, the alarm icon appears, long
press backlight button again to inactivate alarm, the alarm icon disappears.

With the alarm function activated, the red backlight will flash every 500ms if Meters alarms due to upper limit, the blue backlight will flash every 500ms if Meters alarms due to lower limit. If Meters alarms due to upper & lower limit, the red/blue backlight will flash every 500ms in turn and the corresponding icon will light up, for example, if Meters alarms due to upper temperature limit, the red backlight, temperature icon and upper icon (HI) will flash at the same time.
View Max/Min Reading
1. Short press MAX/MIN button to check Maximum reading, minimum reading and normal test value.
2. Long press MAX/MIN to exit the mode.

View dew-point temperature and wet bulb temperature Short press ENTER to view dew point temperature, wet bulb temperature and
normal temperature. For dew point temperature, “DEW POINT” is displayed on
the screen, for wet bulb temperature, “WET BULB” is displayed on the screen.

Use offset value to adjust Meters error. Users can compensate the error in temperature & humidity offset setting of SETUP.
Data Access
Store Data Manual storage: time interval is set at 00:00, in test page, short press ST ORE to store data manually, Meters flashes and show the storage area. Take LOG 001 as an example, it represents the data is stored at 001. UT331+ can store 299 groups of data, UT332+can store 999 groups of data, the record data is the data shown in the main display screen. Auto storage: time interval is not set at 00:00, in test page, short press STORE to store data manually, long press STORE to store data automatically, the storage area, i.e. LOG 001, will flash at intervals, Meters will automatically save data in circle at the set interval, the record data is the data shown in the main display screen.

Recall Data Short press RECALL to check recorded data, press again to exit. 1. Short press RECALL to check recorded data, the screen displays the storage area, i.e. LOG 001, press to scroll through recorded readings, Meters defaults at the reading saved last time you exit the STORE mode. 2. Short pressing ENTER under RECALL mode can quickly review from 01 to 299 /999 data and also allow you to transfer data to PC through USB, after quick review, the number stops at 299/999. On data reviewing status, short press RECALL to exit and go back to main display screen. 3. Long press RECALL under RECALL MODE to clear the recorded data, the screen will keep displaying dELE until the data is completely cleared, and then go back to main display screen. Note: The data will be completely cleared once the function is enabled. Step-by -Step clearing is not allowed. To avoid any data loss, please think it twice before you use the function.

Transfer Data Meters will transfer data to PC under following two circumstances: 1. Enter into RECALL status, short pressing ENTER button will not only allow you to review the data quickly but also transfer the data automatically to PC. 2. When PC is running the supplied software and has been already connected to Meters, the measured data will be automatically transferred to PC.

USB power supply Connect USB cable to PC USB interface or 5V DC adaptor, Meters will automatically switch to USB power supply. Disconnect the USB cable to switch to battery power supply. Note: The DC voltage of USB power supply is 5V, voltage over 5V may damage Meters.

USB Communication
1. Please download PC software according to the attached general file, and install step by step.
2. After running the master computer software, connect the USB cable to the computer.
3. Browse and analyze data through the master computer software, for the usage of master computer software, please refer “Software Manual” from the Help Option of Operation Page.

Maintenance

Replacing the battery (See Figure 3) When Meters indicates ” “, please replace the

battery timely. Follow the steps as below:

1.Unscrew the battery cover and remove it from

Meters. 2.Replace the batteries with new 4×AAA batteries.

3.Rejoin the battery cover and tight up the screw.

Figure 3

Cleaning the casing

When it becomes necessary to clean the surface of Meters, clean it with soft

cloth or sponge dampened with just a little of water, soap water or mild detergent.

Prohibit washing Meters with water directly, which may damage the circuits inside.

Technical Index
Function
Measurement range

UT 331+
Measurement accuracy

UT 332+
Measurement accuracy

Temperature
Dew point temperature
Wet bulb temperature

-20.0°C~ 70.0°C
-78.2°C ~ 70.0°C
-21.6°C ~ 70.0°C

±0.5 ±1 ±0.5

±0.3 ±0.8 ±0.3

Humidity

0~99.9%RH

10%~90%RH, ±3%RH. 10%~80%RH, ±2%RH. For others: ±3.5%RH. For others: ±2.5%RH.

Resolution

Temperature: 0.1°C, Humidity: 0.1%RH

Temperature measurement Manual storage

Measurement temperature/Dew point temperature/Wet bulb temperature

Auto storage

Users can turn on/off auto storage(It defaults at “turn off”) and adjust the storage interval 0s60min.

Storage group number Data viewing Unit setting

299 groups

999 groups

View storage data

°C or °F

Alarm setting

Users can turn on/off the alarm (It defaults at “turn off”). Backlight alarm (For upper-limit alarm, the red light flashes every 500ms. For lower-limit
alarm, the blue light flashes every 500ms.)

Max./Min. value measurement

Data hold

Data export Auto power off Real time

Users can view the real-time data through the software of master computer and export the
storage data.
Users can turn on/off the function of auto power off (It defaults at “turn on”) and adjust the auto
power off time (5min60min).
Set the real time of system.

Offset setting

Temperature, dew point temperature, wet bulb temperature, humidity offset value (-66).

Backlight USB power supply

Users can turn it on/off (It defaults at “turn off”).
Connect USB and switch to USB power supply, USB voltage is 5V.

EMC Standard: EN61326-1 2013 Reference Standard:
JJG205-2005 Mechanical Hygro-Thermometer JJF1076-2001 Calibration Specification of Humidity Sensor