TSS COM_DEV_081 Ver TempMonitor BLE Kit
Specifications
- Product: TempMonitor BLE Kit
- Gateway: TSS Gateway
- Included in the package: Adapter, cable, TempMonitor BLE, 1 connected probe with damper
- Recommended distance: Less than 30 meters from TempMonitor BLE, free line of sight
- Compatibility: TM BLE FW 1.2.0 and later
Product Usage Instructions
Installation Guide
- Attach the country adapter to the TSS Gateway and connect it to a power outlet.
- Place the TSS Gateway in the same room as the TempMonitor, ensuring it is less than 30 meters away and has a free line of sight.
- Check for a green light on the TSS Gateway indicating power and connection status (this may take up to 15 minutes).
- Place the TempMonitor BLE outside the fridge and insert the probe with damper inside the fridge.
Starting the Logger
- Ensure the damper is in place at least 1 hour before starting the logger to prevent temperature excursions.
- Quickly press the START button once to turn on the logger.
- Continue pressing the START button for at least 5 seconds until the display changes from Strt to dELY.
- If a start delay is configured, ‘dELY’ will be displayed for the set time before recording data begins.
Probe and Damper Placement
The recommended placement for the probe is inside the fridge with the damper. The damper serves to ensure accurate temperature readings by stabilizing the probe’s environment within the fridge.
Acknowledging Temperature Alarms
To acknowledge a temperature alarm, press the START button once to remove the warning symbol. If the temperature is within the set threshold, the cross mark (X) will disappear.
Introduction
The TSS TempMonitor BLE is a wireless storage logger with an internal temperature sensor and support for up to two probes, that can communicate with TSS Gateway and TSS Mobile App.
Key Features
The TempMonitor BLE, Bluetooth low energy temperature data logger can be paired with TSS GW, smartphones and transmit data via Bluetooth 5. It records up to 50,000 data points readings and monitors drugs across a range of -50°C to 85°C using it’s supported probes.
Logger Specifications
FEATURES
Type | Temperature Data Logger |
Operating Range | -20˚C to +50˚C, -50˚C to +85˚C (probe) |
Temperature Accuracy | ± 0,5°C |
Memory Capacity | 50 000 Data Points |
Temperature Resolution | 0,0625˚C (1/16th of a Degree) |
Compliance | RoHs, CE, FCC, DO-160G |
Calibration | Each unit is delivered with calibration certificate from accredited ISO-17025 laboratory |
Battery | 3 x AAA Alkaline 1.5V (removable) |
Alarms/Events | Temperature Alarms, Calibration Expiry, Battery, Probe Disconnect |
Audible Alarm/Events | Yes |
Buttons | Start, Toggle Temperature, Mute |
Sample Rate | 1 min to 18 hours |
Start Delay | 0 min to 30 days |
IP Classification | IP54 |
Security | Highly secure utilizing BLE security features |
Weight | Approx. 118 g |
Measurement | Approx. 100 mm x 80 mm x 30 mm |
Housing | Blistered rugged plastic |
Traceability | Unique 16-character ID number that allows for absolute traceability |
Display Interface
LCD Symbol meaning
LCD text meaning
Buttons
Button action are displayed below:
Start up Logger
Screen is lit with symbols.
Quick Press START button once for to start up the logger.
NOTE: Before starting the logger ensure the probe(s) are placed at the given location.
Start Logging
- Press and Hold START button until screen shows ‘dELY’¹
¹ Note, when pressing START button a beep will be heard indicating that the button is activated, ensure that button is continuously pressed until the display changes from “Strt” to “dELY”. - The logger is successfully started but has configured start delay, ‘dELY’ will be displayed for the configured time and then start to record the data.
- Logger records temperature data.
Connection and signal strength
To verify that the logger has connection to TSS Gateway the Bluetooth symbol shall be visible on the LCD screen.
Logger has Bluetooth connection to TSS Gateway.
In case the logger cannot establish a Bluetooth connection to the Gateway the symbol will be marked with an X.
Logger does not have Bluetooth connection to TSS Gateway.
The signal strength between the logger and the Gateway is indicated by the signal bar strength on the LCD. The more bars the stronger the signal strength.
Toggle Temperature Channels
Probe selection
Press TOGGLE to switch between temperature channels if the logger is configured for measuring more than one channel.
TOGGLE is used for toggling between temperature sensors.
T is the internal temperature sensor. Note that only two channels can be active during recording.
Max and Min Temperature
The highest (Max) and lowest (Min) recorded temperature values are displayed on the screen. It is based on the values for the Last 30 days.
Max and Min temperature for the Last 30 days
Alarms and Events
Temperature Alarms
Temperature Alarm symbol is activated
General Warning symbol is activated
Acknowledge Alarms
Quick Press START button once for acknowledging alarms on the logger.
By pressing the START-button, the General Warning symbol will be removed, indicating that the alarm has been acknowledged. If the temperature is within set threshold the cross mark (X) will be changed to check (✓).
If the temperature is still out of range during the time of acknowledgement, will the cross mark (X) remain until the temperature is within range. The buzzer will be muted by the acknowledgement. If a new alarm is triggered will the buzzer continue to beep as per configuration.
Mute/Unmute
If an alarm has been triggered the buzzer can be muted by pressing the mute button once. It will be muted for 5 minutes. To unmute the logger, press the mute button once more. The icon will be removed indicating that the logger is no longer muted.
MUTE is used for muting/unmuting the buzzer
Battery
Battery Level – Always shown to indicate battery level in 1 of 3 states: Full, Low and Critical.
Battery alarm levels can be configured and in case a battery alarm is triggered the general warning sign will be activated on the LCD.
General Warning symbol is activated
Battery replacement
The logger is equipped with user replaceable batteries, see section 3.2 Battery for details. When a low battery warning appears, it is time to change batteries. The logger requires three user-replaceable AAA 1.5V alkaline or optional lithium batteries.
Before starting the procedure:
- make it sure that there is up-to-date data in your system from the logger.
To replace the batteries simply remove the battery lid and replace the existing batteries with new. Close the lid.
The logger will automatically continue logging temperature after the new batteries have been inserted¹. If there was any unread temperature measurement, event, probe event or alarm the logger display will show
Upload. This indicates that the logger is waiting for the Gateway to read out the unread events before it can start logging temperature again. Ensure the Gateway is online.
Note, if the probe is removed before battery replacement and connected again after batteries have been inserted the logger must be manually started on start button.
Calibration Expiry
When the sensor(s) have less than 60 days until they expire a General Warning symbol will be lit. Additional notifications related to the Calibration Expiry are handled by TSS Cloud.
General Warning symbol is activated
Note, if a sensor has expired it is not possible to start the logger.
EXPR calibration expiry has passed, logger can’t be started
Memory
In case the logger has not been connected to the Gateway for a long period of time a low memory symbol can appear. This is to indicate that the logger’s memory is starting to get full. During normal operations with the logger connected to a Gateway this will not happen. In case this this message appears ensure that the Gateway is connected to power and is online.
Low Memory symbol is activated
General Warning symbol is activated
General Logger Information
Logger and Probe Identification
Each TempMonitor BLE is identified with a unique 16-character id, located on its back.
Each Probe TMBLE100, is identified with a unique 16-chracter id. The Expiry Date of the probe is stated too.
Logger Components
Battery
The logger requires three user-replaceable AAA 1.5 V alkaline or optional lithium batteries for operation at the extreme ends of the logger operating range. Expected battery life varies based on the ambient temperature where the logger is deployed, the frequency of connecting to the phone or tablet and downloading reports, the duration of audible alarms, and battery performance. New batteries typically last at least 1 year with logging intervals of 5 minutes. Deployments in cold or hot temperatures or a logging interval faster than 5 minutes can impact battery life. Estimates are not guaranteed due to uncertainties in initial battery conditions and operating environment.
WARNING: Do not cut open, incinerate, heat above 85°C (185°F), or recharge the lithium batteries. The batteries may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case. Do not dispose of the logger or batteries in fire. Do not expose the contents of the batteries to water. Dispose of the batteries according to local regulations for lithium batteries.
Troubleshooting
Should you experience issues with any of the above steps, or with the TSS TempMonitor BLE in general, please reach out to us at customer.service@tss.se.
USING INSTRUCTION
- Check that following is included in the package
- Place the gateway with LED facing upward. Antennas mostly vertical, a slight angle out is recommended.
- Place TempMonitor BLE outside the fridge and the probe in its damper inside the fridge.
- Quick Press START
button once to TURN ON logger
- Press and Hold START button until screen shows “dELY”¹
Note, when pressing START button a beep will be heard indicating that the button is activated, ensure that button is continuously pressed for at least 5 seconds until the display changes from “Strt” to “dELY”. - TOGGLE is used for toggling between (probe) temperature sensors
FAQ
Question | Answer |
What does it mean when the display of the TempMonitor shows “Strt”? | The logger is ready to be started, refer to Quick Installation Guide. |
Where is the recommended placement of the TempMonitor? | Recommend placing it outside the refrigerator, using its magnets on the backside. If used in Ambient temperature place logger at appropriate place and put the probe with damper at desired
location. |
Where is the recommended placement of the probe and the damper? | Placement of damper shall be as close as possible to where the products are stored, if unsure
please contact your study manager. |
What purpose does the damper serve? | The primary purpose of the damper is to protect against sudden and sharp air temperature changes – such as fluctuations due to the opening or closing of refrigerator doors or normal
defrost cycles. |
What does Max °C/Min °C mean? | It shows the highest and lowest measured temperature for the last 30 days. |
How do I acknowledge a temperature alarm? | By pressing the START button once the Warning symbol will be removed, indicating that the alarm has been acknowledged. If the temperature is within set threshold the cross mark (X) will be changed to check (✓). If the temperature is still out of range during the time of acknowledgement the cross mark (X) will remain until the temperature is within range. The buzzer will muted by the acknowledgement. If a new alarm is triggered the buzzer will continue
to beep as per configuration. |
The logger keeps beeping although it is back within threshold? | The logger has recorded a temperature alarm and has not been acknowledged. Press START
button once to acknowledge the temperature alarm. |
The display on the logger shows X, indicating temperature alarm, but the
logger is within range? |
The logger has recorded a temperature alarm and has not been acknowledged. Press START
button once to acknowledge the temperature alarm. |
Which one is “probe slot 1” and “probe slot 2”? | The Temp Monitor BLE always comes with one mounted (attached) probe, this probe is attached
to “probe slot 1” and the other (available) “probe slot” is “probe slot 2”. (If a second probe has been ordered, please connect it to the available “probe slot 2″) |
I’m pressing the start button, but the logger will not start. What do I need to do?
First of all, verify that the logger is in a state that allow it to be started.
- The display should show “Strt”, indicating the logger is ready to be started
- Ensure that the START button is firmly pressed and held down until the screen changes. This can take several seconds.
I’m being told there is no data in the system, although the logger seems to be working. What do I need to do?
First, verify that the Gateway and the TempMonitor BLE is close to each other, we recommend keeping them in the same room and less than 30 meters apart.
- Check that the Gateway is online.
- The Gateway should show a solid green light. If the Gateway is blinking yellow, it means that it can’t connect to the Internet. This can be resolved by changing location of the Gateway, typically closer to a window where it has better cellular connectivity. A tip is to check on your mobile phone, if you have cellular coverage with your phone chances are good that the Gateway will work there too.
Tip: check the cellular strength on your mobile phone
If the above is unsuccessful, please report back to TSS Customer service.
I need the Calibration Certificate, where can I find it?
The probe is calibrated, and its calibration certificate can be found here* https://certificates.tssgeneral.com/
- Click on SEARCH CERTIFICATE
- Scan or Enter the Probe ID from the label attached to the probe cable and click search.
- Click download to access the certificate.
*In TSS Clinical Module the calibration certificate is located under TSS > Administration
Loggers > Overview > Logger Details
Still having issues or looking for more information? Contact TSS Customer service
Please reach out to us at customer.service@tss.se.
Documents / Resources
![]() |
TSS COM_DEV_081 Ver TempMonitor BLE Kit [pdf] Installation Guide COM_DEV_081 Ver C, COM_DEV_021 verA, COM_DEV_081 Ver TempMonitor BLE Kit, COM_DEV_081 Ver, TempMonitor BLE Kit, BLE Kit, Kit |