1. Bubuka
The Metrel MI 3290 Earth Analyzer is a portable, robust, and versatile test instrument designed for comprehensive earth resistance, specific earth resistance, and earth potential measurements. It is suitable for various energetic and non-energetic objects, offering a wide range of measurement methods from classic 3-wire earth resistance to advanced 4-clamp methods for pylons. Its high electrical noise immunity makes it ideal for industrial environments.

Figure 1: The MI 3290 GL Earth Analyzer housed within its durable, protective carrying case.
2. Fitur konci
The MI 3290 GL Earth Analyzer offers a robust set of features for accurate and reliable earth measurements:
- Comprehensive Earth Resistance Testing: Supports 2, 3, and 4-pole earth resistance measurements, selective earth resistance with one clamp, and earth resistance with two iron clamps.
- Specific Earth Resistance: Utilizes Wenner and Schlumberger methods for detailed analysis.
- High-Frequency (HF) Earth Resistance: Measures at 25 kHz, compliant with IEEE Std 81.
- Pylon Earth Resistance: Capable of measuring earth resistance of mono pylons with 10m flex clamps and multi-leg pylons with up to four flex clamps.
- Pangukuran ayeuna: Supports measurements using both iron and flex clamps.
- Low Ohm Measurement: Performs measurements at 7 mA and 200 mA.
- Earth Potential, Step, and Contact Measurements: Provides comprehensive analysis of earth systems.
- Impulse Earth Measurement: Simulates lightning strikes with 10/350 µs impulses.
- Wide Frequency Band Analysis: Analyzes earth impedance as a function of frequency (55 Hz to 15 kHz).
- Tampilan canggih: Features a 4.3" color LCD with backlight and touch screen for intuitive operation.
- Desain mantap: Offers IP 65 protection when the case is closed and IP 54 when open, ensuring durability in various environments.
- PC Software Integration: Includes Metrel ES Manager PC software for test structure preparation, result download, data analysis (tree-view, méja view, graphical view), storing, and printing.

Gambar 2: Detil view of the MI 3290 GL's user interface and connection ports.
3. Kaasup Asesoris
The MI 3290 GL Earth Analyzer comes with a comprehensive set of accessories to facilitate various measurement tasks:
3.1 Standard Included Accessories
- Instrumén MI 3290
- Current earth spike 90 cm, 2 pcs
- Potential earth spike 42 cm, 2 pcs
- Connection lead black 2 m
- Test lead 5 m blue
- Test lead 5 m red
- Test lead 50 m black on cable reel
- Test lead 50 m green on cable reel
- Test lead 50 m blue on cable reel
- Shielded test lead 75 m on reel
- G clamp
- Set of test probes, crocodile clip and 2 m test lead, 4 pcs
- Kantong pikeun asesoris
- kabel USB
- Kabel listrik
- Sertipikat kalibrasi
- Manual instruksi (dokumen ieu)
- PC SW Metrel ES Manager
3.2 GL Set Specific Inclusions
- Licence GL
- Beusi clamp A 1018-EA with 3.5 m test lead
- Beusi clamp A 1019

Gambar 3: Leuwihview of the MI 3290 GL Earth Analyzer and its comprehensive accessory kit.

Figure 4: The A 1018-EA and A 1019 Iron Clamps, part of the GL set.
4. Setup Guide
Proper setup is crucial for accurate measurements and safe operation. Follow these steps before beginning any testing:
4.1 Initial Inspection and Power
- Unpack the Instrument: Carefully remove the MI 3290 GL from its carrying case.
- Pariksa Karusakan: Check the instrument and all accessories for any visible signs of damage. Do not use if damaged.
- Daya Hurung: The device can be powered by its internal battery or via the mains cable. Ensure the battery is sufficiently charged or connect the mains cable to a suitable power outlet (90-260 VAC / 45-65 Hz).
- Uji diri: Upon powering on, the instrument will perform a self-test. Wait for this process to complete.
4.2 Connecting Accessories for Measurement
The connection method depends on the specific measurement type. Refer to the on-screen diagrams and the detailed operating instructions for precise wiring. General connection points are:
- Current (C1) and Potential (P1, P2) Terminals: For connecting earth spikes and potential leads in 3-pole, 4-pole, and specific earth resistance measurements.
- Earth (E) Terminal: For connecting the earth electrode.
- Clamp Terminals (F1-F4): For connecting current clamps (iron or flex) for selective earth resistance or multi-leg pylon measurements.
- Port USB: For connecting to a PC for data transfer and software interaction.
Always ensure all connections are secure and correctly polarized before initiating a measurement.
5. Parentah Operasi
The MI 3290 GL offers a variety of measurement functions accessible via its touch screen interface. Navigate through the menus to select the desired test method.
5.1 Measurement Methods Overview
- Earth Resistance (2, 3, 4-pole): Standard methods for measuring the resistance of an earthing system. The 4-pole method provides the most accurate results by eliminating the resistance of the test leads.
- Selective Earth Resistance (1 x Clamp): Allows measurement of individual earth electrodes in parallel systems without disconnecting them, using one current clamp.
- Earth Resistance (2 x Iron Clamps): Measures earth resistance using two iron clamps, suitable for specific applications.
- Specific Earth Resistance (Wenner and Schlumberger): Used to determine soil resistivity, crucial for designing earthing systems. The instrument guides the user through the probe placement for these methods.
- HF-Earth Resistance (25 kHz): Measures earth resistance at high frequencies, relevant for lightning protection systems and high-frequency applications.
- Earth Resistance of Pylons: Specialized functions for measuring earth resistance of mono-leg and multi-leg pylons, utilizing flex clamps for accurate readings on large structures.
- Pangukuran ayeuna: Measures AC current using iron or flex clamps, useful for identifying stray currents or load currents in earthing systems.
- Low Ohm Measurement (7 mA and 200 mA): For measuring low resistance values, such as continuity of protective conductors.
- Earth Potential, Step, and Contact Measurements: Assesses potential differences on the ground surface, critical for safety in substations and industrial sites.
- Impulse Earth Measurement (10/350 µs): Simulates lightning strike conditions to evaluate the earthing system's response to transient overvoltages.
5.2 Ngalakukeun Pangukuran
- Select Test: From the main menu, select the desired measurement function using the touch screen.
- Sambungkeun Arah: Follow the on-screen wiring diagram to connect the appropriate test leads, earth spikes, and/or clamps to the instrument and the system under test.
- Setél Parameter: Adjust any necessary parameters (e.g., test frequency, current range) as prompted by the instrument or as required by the test standard.
- Mimitian Tés: Press the START button to begin the measurement.
- Review Hasil: The measurement results will be displayed on the screen. The instrument can store these results for later analysis.
- Simpen Data: Use the save function to store the measurement data in the instrument's internal memory.
For detailed step-by-step instructions for each specific measurement method, refer to the full technical manual provided with the instrument or the Metrel ES Manager software documentation.
6. Pangropéa sarta perawatan
To ensure the longevity and accuracy of your MI 3290 GL Earth Analyzer, adhere to the following maintenance guidelines:
- beberesih: Wipe the instrument's exterior with a soft, damp lawon. Entong nganggo pembersih atanapi pangleyur anu abrasif. Pastikeun sadaya palabuhan garing sateuacan dianggo.
- Panyimpenan: Store the instrument in its protective case in a dry, cool environment, away from direct sunlight and extreme temperatures.
- Perawatan batré: If storing for extended periods, ensure the battery is partially charged (around 50%) and recharge it periodically to prevent deep discharge.
- Inspeksi aksésori: Regularly inspect test leads, spikes, and clamps for wear, damage, or insulation integrity. Replace any damaged accessories immediately.
- Kalibrasi: The instrument requires periodic calibration to maintain its measurement accuracy. Refer to the calibration certificate for recommended intervals. Contact authorized Metrel service centers for calibration services.
7. Cara ngungkulan
This section provides solutions to common issues you might encounter with the MI 3290 GL Earth Analyzer.
| Masalah | Mungkin Cukang lantaranana | Solusi |
|---|---|---|
| Alatna teu bisa dihurungkeun. | Low or depleted battery; loose mains cable connection. | Charge the battery or connect the mains cable securely. |
| Bacaan anu teu akurat atanapi teu stabil. | Poor test lead connections; high noise environment; incorrect measurement setup; dry soil conditions. | Ensure all connections are secure. Move away from strong electromagnetic fields. Verify the correct test method and parameters. Moisten the soil around earth spikes if too dry. |
| "Open Circuit" or "High Resistance" error. | Broken test lead; poor contact with earth spike/electrode; test lead not connected to the correct terminal. | Check continuity of test leads. Ensure earth spikes are firmly inserted into the ground. Verify all connections match the selected measurement diagram. |
| Display is unresponsive or frozen. | Software glitch. | Perform a soft reset by holding the power button until the device restarts, or disconnect power and reconnect. |
| Teu tiasa mindahkeun data ka PC. | Incorrect USB cable; driver issues; Metrel ES Manager software not running or configured correctly. | Use the supplied USB cable. Ensure Metrel ES Manager is installed and running. Check device manager for driver issues. Refer to the software manual. |
If the problem persists after attempting these solutions, contact Metrel customer support or an authorized service center.
8. Spésifikasi Téknis
The following table outlines the key technical specifications of the Metrel MI 3290 GL Earth Analyzer:
| Parameter | Nilai / Rentang | Inpormasi Tambahan |
|---|---|---|
| Earth Resistance (2,3,4-pole) | 0.00 Ω ... 19.99 kΩ | Test Frequencies: 55 Hz, 329 Hz, 1.6 kHz, 3.29 kHz, 15 kHz; Test Voltage: 20 or 40 VAC; Short-circuit current > 220 mA |
| Specific Earth Resistance | 0.00 Ωm ... 19.99 kΩm | Wenner and Schlumberger methods supported. |
| HF Earth Resistance | 0.0 Ω ... 299 Ω | Frékuénsi Uji: 25 kHz |
| Impulse Earth Measurement | 0.0 Ω ... 199.9 Ω | Impulse: 10/350 µs |
| DC Resistance (RLDW, CONT) | 0.0 Ω ... 19.99 kΩ | Test Current: 200 mA (RLDW) |
| Current RMS (Flex/Iron Clamp) | 0.0 mA ... 49.9 A | Nominal Frequency: 10 Hz ... 1.5 kHz |
| Earth Potential | 0.0 mV ... 49.99 V | Calculated value (consider uncertainty of 2-pole measurement) |
| Sumber Daya | Didukung ku batré / listrik utama | Battery: 14.4 V DC (4.4 Ah Li-ion); Mains: 90-260 VAC / 45-65 Hz, 100 VA (200 V CAT II) |
| tampilan | 4.3" color LCD | With backlight and touch screen |
| Komunikasi | USB, Bluetooth | For PC connectivity and data transfer. |
| Mémori | >8 GB | Internal storage for measurement results. |
| Gelar panyalindungan | IP 65 (kasus ditutup), IP 54 (kasus muka) | Protection against dust and water ingress. |
| Ukuran (W x H x D) | 330 x 280 x 160 mm | Approximate dimensions of the instrument. |

Figure 5: Excerpt from the MI 3290 GL Technical Specifications datasheet.
9. Garansi jeung Rojongan
The Metrel MI 3290 GL Earth Analyzer comes with a manufacturer's warranty and dedicated support services.
- Mangsa Garansi: The product is covered by a two-year (24-month) warranty from the manufacturer, starting from the date of purchase. This warranty covers defects in materials and workmanship under normal use.
- Klaim Garansi: For warranty service, please contact your authorized Metrel distributor or the manufacturer directly. Proof of purchase will be required.
- Bantosan Téknis: For technical assistance, operational queries, or troubleshooting beyond what is covered in this manual, please reach out to Metrel's customer support. Contact information can typically be found on the Metrel official websitus atanapi dina bungkusan.
- Jasa Kalibrasi: Regular calibration is recommended to ensure continued accuracy. Metrel offers professional calibration services through its authorized service network.
For the most up-to-date support information and resources, please visit the official Metrel websitus.