Recommended decontamination procedures for all models. Please note that these instructions are intended as a GUIDE ONLY and all pipettes requiring decontamination must be decontaminated according to your institution’s own guidelines and the type of contaminant present.

GENER AL DAILY CLE ANING/DECONTAMINATION

  • 1:10 dilution is generally recommended,
  • Spray or wipe handle and shaft (insuring the tip ejector button and volumeter are also cleaned)
  • Leave for a minimum contact time of 10 minutes

ORDERING INFORMATION

starlab-HLD4L-Laboratory-Disinfectant-fig-1

COMPLETE DECONTAMINATION

Pipettes requiring full decontamination can be dismantled and cleaned:

  • 1:10 dilution is generally recommended,
  • Use a container that is deep enough to immerse the parts completely.
  • (NOTE: Electronic pipettes must not be submerged in any liquids)
  • Dismantle the pipette (as specified by the manufacturer) and place only parts that can be immersed, eg tip holder, piston, tip ejector etc completely into the solution for 30 minutes
  • Remove parts and rinse with de-mineralised water
  • Lie on a lint-free cloth and place in a dry oven at 60°C for at least one hour or until dry. Alternatively, place on the bench on a lint-free cloth and air dry for approximately 4 hours
  • Reassemble the pipette following the manufacturer’s instructions
  • If complete decontamination is required prior to servicing, please ensure all parts are completely dry before reassembly and presenting for servicing

NOTE: All pipettes for servicing must be accompanied by a completed Decontamination Declaration stating that the pipette is safe for us to work on, irrespective of whether the pipette required decontamination or not.