1.0 INTRODUCTION
1.1 Technical parameters
  • 6-digit 1.6-inch LCD, various indicating lamps, long service life and good shock resistance
  • 7 buttons, simple operation
  • Protection level: IP5x
  • Excitation voltage: +5VDC
  • Load capacity of sensor: at most four 350Ω simulation sensors
  • Input signal range at null point: 0~5mV
  • Input signal range at full scale: 1~10 mV
  • Inner resolution: 1 million
  • Display division: 1000~30000
  • A/D sampling rate: 120times/sec
  • Power supply mode
    Battery: 7.4V/4AH lithium battery
    Adaptor: input voltage 100-240VAC; output voltage 8.4V/1.2A; frequency: 50-60Hz
  • RS232
  • Working temperature: -10℃-40℃, relative humidity: below 85 %
  • Storage temperature: -20℃-60℃, relative humidity: below 85 %
1.2 Main Functions
  • Basic functions: zero and tare
  • Totalization, counting
  • Auto power-save
  • Parameter redundancy backup
  • Real-time clock
  • Auto power-off
1.3 Dimension

VETEK HL318 Plus SSC LCD Display Weight Indicator - 2

1.4 Model

HL318 Plus SSC

2.0 Interface
2.1 Power supply
VETEK HL318 Plus SSC LCD Display Weight Indicator - 3 1Pin(inner core): DC+      2Pin(outer core): DC-
2.2 RS232
VETEK HL318 Plus SSC LCD Display Weight Indicator - 4 1Pin: TXD(红) 2Pin: RXD(绿) 3Pin: GND(黑)
2.3 Load cell
VETEK HL318 Plus SSC LCD Display Weight Indicator - 5 1Pin: +V(红) 2Pin: +SN(蓝) 3Pin: +S(绿)
4Pin: 屏蔽(黑) 5Pin: -S(白) 6Pin: -V(黑)
7Pin: -SN(黄)
3.0 Operation
3.1 Indicating Lamps
Sign Meaning Specification
Dynamic/static When the scale is dynamic the lamp is on; otherwise, the lamp is off.
VETEK HL318 Plus SSC LCD Display Weight Indicator - 6 Zero-center When the absolute weight value is less than ±d/4 the lamp is on; otherwise, the lamp is off.
Net Gross/net weight The lamp is on at net weight and off at gross weight.
kg Weight unit The lamp indicates the weight unit in use.
Total Totalization When the totalization function is in use the lamp is on.
3.2 Buttons
Button Name Normal Setting
VETEK HL318 Plus SSC LCD Display Weight Indicator - 7 〖TOTAL〗 Totalization function in use
Short press: add weights to total
Long press: check/clear
None
VETEK HL318 Plus SSC LCD Display Weight Indicator - 8 〖COUNT〗 Counting function in use
Short press: switch between quantity and weight
Long press: no definition
Return/exit
VETEK HL318 Plus SSC LCD Display Weight Indicator - 9 〖PRINT〗 Short press: print Move cursor to left
VETEK HL318 Plus SSC LCD Display Weight Indicator - 10 〖GROSS〗 At gross weight: no definition
At net weight: add skin weight to net weight (net weight—>gross weight)
Move cursor to right
VETEK HL318 Plus SSC LCD Display Weight Indicator - 11 〖TARE〗 Remove skin weight (gross weight —> net weight) Decrease
VETEK HL318 Plus SSC LCD Display Weight Indicator - 12 〖ZERO〗 Zeroing Increase
VETEK HL318 Plus SSC LCD Display Weight Indicator - 13 〖ON/OFF〗 Power on/off Confirm
4.0 Setting
4.1 Setting Entry

Normally press 〖CAL〗 until VETEK HL318 Plus SSC LCD Display Weight Indicator - 27 shows. Press〖ON/OFF〗 to enter into the menu setting interface to set parameters from F1~F5. Normally, press 〖GROSS〗 until VETEK HL318 Plus SSC LCD Display Weight Indicator - 27 shows. Press 〖ON/OFF〗 to enter into the menu setting interface to set parameters from F2~F5. Below shows where the 〖CAL〗 button is.

Open the back cover → VETEK HL318 Plus SSC LCD Display Weight Indicator - 14

4.2 Detailed Parameter Setting

F1 Scale Parameter Setting

F1.1 Capacity
Selectable parameters:3~200000(default: 6)

F1.2 Decimal
Selectable parameters: 0——-no decimal
1—— 1 decimal
2——–2 decimals
3——-3 decimals (default)
4——-4 decimals

F1.3 Division
Selectable parameters: 1 (default), 2, 5, 10, 20, 50
Note: when F1.2, F1.3 or F1.4 is being set, division value should not be beyond 10000.

F1.4 Zero Calibration
VETEK HL318 Plus SSC LCD Display Weight Indicator - 33 remove weights off scale and press 『ON/OFF』. The indicator shows VETEK HL318 Plus SSC LCD Display Weight Indicator - 34 and it decreases to VETEK HL318 Plus SSC LCD Display Weight Indicator - 35. Finally, VETEK HL318 Plus SSC LCD Display Weight Indicator - 28 appears for one second and zero calibration ends.

F1.5 Load Calibration
VETEK HL318 Plus SSC LCD Display Weight Indicator - 26 add weights
Add weights to scale and make sure: full capacity *50% ≤ weights ≤ full capacity. Press 『ON/OFF』. VETEK HL318 Plus SSC LCD Display Weight Indicator - 36 Enter the same value as the weights. Wait until the scale is stable and press 『ON/OFF』. The indicator shows VETEK HL318 Plus SSC LCD Display Weight Indicator - 34 and it decreases to VETEK HL318 Plus SSC LCD Display Weight Indicator - 35.
Finally, VETEK HL318 Plus SSC LCD Display Weight Indicator - 28 appears for one second and load calibration ends.

F1.6 Auto zero-tracking rate
Selectable parameters: OFF, 0.5d (default)
At net weight status, zero-tracking is ineffective.

F1.7 Auto zeroing range at power-on
Selectable parameters: OFF (default), ±2 %, ±10 %

F1.8 Zeroing range by button
Selectable parameters: OFF (default), ±2 %

F1.9 Digital filter
Selectable parameters: 0~9, default: 5

F1.10 Scale function setting
Selectable parameters: 0——–Totalization (default)
1——–Counting

F 1.11 Restoring factory default
Selectable parameters: 0——–No restore
1——–Restore parameters from F1 to F4 while scale parameters unaffected

F 2 Applicable Parameter Setting

F 2.1 Sampling method
Selectable parameters: 0—-weight on scale sampling (default)
1—–manual input

F 3 Indicator Parameter Setting

F 3.1 Date format setting
Selectable parameters: 0—–year.month.day (default)
1—–month.day.year
2—–day.month.year

F 3.2 Date setting (see F 3.1)

F 3.3 Time setting (format: hour.minute.second)

F 3.4 Overtime backlight shutdown time setting
Selectable parameters: 0~999 second (default: 0)
If 0 is set, this function is shutdown.

F 3.6 Auto power-off time setting
Selectable parameters: 0~60min (default: 0)
If 0 is set, this function is shutdown.

F 4 Serial Communication

F 4.1 Method
Selectable parameters:0—–no output (default)
1—–serial output (only when the scale is stable)
2——printing output (see Appendix Ⅲ)

F4.2 Data and verification setting
Selectable parameters: 8_N_1 —-8 digits, no verification (default)
7_E_1—–7 digits, odd verification
7_O_1 —-7 digits, even verification
8_E_1 —-8 digits, odd verification
8_O_1—–8 digits, even verification

F 4.3 Baud rate
Selectable parameters: 2400, 4800, 9600 (default), 19200

F 4.4 New line
Selectable parameters: 0~9 (default: 0)

F 5 Maintenance and Service

F 5.1 Button test
When the indicator displays VETEK HL318 Plus SSC LCD Display Weight Indicator - 37, press〖ON/OFF〗〖ZERO〗〖TARE〗〖GROSS〗 〖PRINT〗and〖COUNT〗in order and the indicator displays “on.off” “Zero” “Tare” “Gross” “Print” and “Count”. Press 〖TOTAL〗to exit.

F 5.2 Display test
The indicator conducts an auto-inspection to make sure there is no lack of strokes. Press 〖TOTAL〗or〖ON/OFF〗to exit.

F 5.3 Inner code
The indicator shows the after-filtering inner code. Press 〖COUNT〗or〖ON/OFF〗to exit.

5.0 Functions
5.1 Totalization

F 1.10=0
Operation: Normally add weights to scale and press 〖TOTAL〗. The screen shows “Add-” and returns to the main weighing interface. Remember to zero the scale before putting things on scale every time; otherwise, there is no totalization.
Check and clear: Normally long press 〖TOTAL〗and “TOTAL ” shows for one second.
Then the total quantity interface “Cn xxx” and the total weight interface “t xx.xx” show.
Press〖Print〗or 〖GROSS〗 to switch between the above two interfaces. Press 〖ZERO〗 to clear the total quantity or the total weight value. Press 〖ON/OFF〗to confirm and press 〖COUNT〗to exit.

5.2 Counting

F 1.10=1
Operation: Normally press 〖COUNT〗to switch between the weight and quantity display.
Sampling: Long press 〖COUNT〗 until “SAMPLE” shows. Press 〖ON/OFF〗 and as “Sn XXX” shows, put in the correct quantity. If F 2.4=0, put the corresponding quantity on scale and press 〖ON/OFF〗to confirm the sample quantity and weight; if F 2.4=1, press 〖ON/OFF〗 and as “XXXXXX” shows, put in the corresponding weight and press 〖ON/OFF〗 to confirm the sample quantity and weight.

Appendix Ⅰ Indicator Prompt Message

Normally the indicator is stable and reliable. If the indicator fails, restart it first. Find out what the error is before repairing it. Repair the indicator according to the error codes.

No. Code Error Countermeasure
1 VETEK HL318 Plus SSC LCD Display Weight Indicator - 15 VETEK HL318 Plus SSC LCD Display Weight Indicator - 16 Unable to zero after startup
  1. Make sure there’s no load on scale at startup.
  2. Redo zero-calibration.
2 VETEK HL318 Plus SSC LCD Display Weight Indicator - 17 load above upper limit Reduce load
3 VETEK HL318 Plus SSC LCD Display Weight Indicator - 18 load less than lower limit Press『Zero』to reset
4 VETEK HL318 Plus SSC LCD Display Weight Indicator - 19 VETEK HL318 Plus SSC LCD Display Weight Indicator - 20 Out of zeroing range Remove load on scale if any.
5 VETEK HL318 Plus SSC LCD Display Weight Indicator - 21 Invalid
6 VETEK HL318 Plus SSC LCD Display Weight Indicator - 22 EEPROM verification and error Press 〖ON/OFF〗to restore and then restart. If the code shows again, return the indicator to factory;
if there’s no code re-calibrate the indicator.
Note: Restoring factory default includes restoring all parameters.
7 VETEK HL318 Plus SSC LCD Display Weight Indicator - 23 The scale is dynamic during calibration Check the scale
8 VETEK HL318 Plus SSC LCD Display Weight Indicator - 24 Wrong date or time Set date and time as required
9 VETEK HL318 Plus SSC LCD Display Weight Indicator - 25 Wrong AD initialization Restart. If the error shows again return the indicator to factory.
10 VETEK HL318 Plus SSC LCD Display Weight Indicator - 26 load calibration: load weight to scale Add weights to scale as required
11 VETEK HL318 Plus SSC LCD Display Weight Indicator - 27 Enter into menu setting Press 〖ON/OFF〗to continue setting.
12 VETEK HL318 Plus SSC LCD Display Weight Indicator - 28 The end of zero/load calibration
13 VETEK HL318 Plus SSC LCD Display Weight Indicator - 29 Load default value

Appendix Ⅱ Serial output format

Serial output format

VETEK HL318 Plus SSC LCD Display Weight Indicator - 30

HEAD1:OL   overload, underload or no zeroing at start-up
ST   the scale is stable
US   the scale is unstable

HEAD2:GS   gross weight
NT   net weight

DATA: data display

UNIT: kg/lb

CR/LF: New line

Example 1: stable, gross weight: 18.000kg, sp = space.

VETEK HL318 Plus SSC LCD Display Weight Indicator - 31

Example 2: unstable, net weight: -0.200kg, sp = space.

VETEK HL318 Plus SSC LCD Display Weight Indicator - 32

Appendix Ⅲ Printing Output Format

LIST

Date    2018/04/14
Time   15:08:46
———————–
Gross   2.061kg
Tare     0.000kg
Net       2.061kg

LIST

Date     2018/04/14
Time    15:10:18
———————–
Count   206PCS

Documents / Resources

VETEK HL318 Plus SSC LCD Display Weight Indicator [pdf] Instruction Manual
HL318 Plus SSC, HL318 Plus SSC LCD Display Weight Indicator, LCD Display Weight Indicator, Display Weight Indicator, Weight Indicator

References

Leave a comment

Your email address will not be published. Required fields are marked *