MANUAL_WB_EPWN_ENG_V1.04
1.0 INTRODUCTION
1.1 Technical parameters
- 6-digit 1.6-inch LCD, various indicating lamps, long service life and good shock resistance
- 7 buttons, simple operation
- Protection level: IP5x
- Excitation voltage: +5VDC
- Load capacity of sensor: at most four 350Ω simulation sensors
- Input signal range at null point: 0~5mV
- Input signal range at full scale: 1~10 mV
- Inner resolution: 1 million
- Display division: 1000~30000
- A/D sampling rate: 120times/sec
- Power supply mode
Battery: 7.4V/4AH lithium battery
Adaptor: input voltage 100-240VAC; output voltage 8.4V/1.2A; frequency: 50-60Hz - RS232
- Working temperature: -10℃-40℃, relative humidity: below 85 %
- Storage temperature: -20℃-60℃, relative humidity: below 85 %
1.2 Main Functions
- Basic functions: zero and tare
- Totalization, counting
- Auto power-save
- Parameter redundancy backup
- Real-time clock
- Auto power-off
1.3 Dimension
1.4 Model
HL318 Plus SSC
2.0 Interface
2.1 Power supply
![]() |
1Pin(inner core): DC+ 2Pin(outer core): DC- |
2.2 RS232
![]() |
1Pin: TXD(红) 2Pin: RXD(绿) 3Pin: GND(黑) |
2.3 Load cell
![]() |
1Pin: +V(红) 2Pin: +SN(蓝) 3Pin: +S(绿) 4Pin: 屏蔽(黑) 5Pin: -S(白) 6Pin: -V(黑) 7Pin: -SN(黄) |
3.0 Operation
3.1 Indicating Lamps
Sign | Meaning | Specification |
~ | Dynamic/static | When the scale is dynamic the lamp is on; otherwise, the lamp is off. |
![]() |
Zero-center | When the absolute weight value is less than ±d/4 the lamp is on; otherwise, the lamp is off. |
Net | Gross/net weight | The lamp is on at net weight and off at gross weight. |
kg | Weight unit | The lamp indicates the weight unit in use. |
Total | Totalization | When the totalization function is in use the lamp is on. |
Button | Name | Normal | Setting |
![]() |
〖TOTAL〗 | Totalization function in use Short press: add weights to total Long press: check/clear |
None |
![]() |
〖COUNT〗 | Counting function in use Short press: switch between quantity and weight Long press: no definition |
Return/exit |
![]() |
〖PRINT〗 | Short press: print | Move cursor to left |
![]() |
〖GROSS〗 | At gross weight: no definition At net weight: add skin weight to net weight (net weight—>gross weight) |
Move cursor to right |
![]() |
〖TARE〗 | Remove skin weight (gross weight —> net weight) | Decrease |
![]() |
〖ZERO〗 | Zeroing | Increase |
![]() |
〖ON/OFF〗 | Power on/off | Confirm |
4.0 Setting
4.1 Setting Entry
Normally press 〖CAL〗 until shows. Press〖ON/OFF〗 to enter into the menu setting interface to set parameters from F1~F5. Normally, press 〖GROSS〗 until
shows. Press 〖ON/OFF〗 to enter into the menu setting interface to set parameters from F2~F5. Below shows where the 〖CAL〗 button is.
Open the back cover →
4.2 Detailed Parameter Setting
F1 Scale Parameter Setting
F1.1 Capacity
Selectable parameters:3~200000(default: 6)
F1.2 Decimal
Selectable parameters: 0——-no decimal
1—— 1 decimal
2——–2 decimals
3——-3 decimals (default)
4——-4 decimals
F1.3 Division
Selectable parameters: 1 (default), 2, 5, 10, 20, 50
Note: when F1.2, F1.3 or F1.4 is being set, division value should not be beyond 10000.
F1.4 Zero Calibration
remove weights off scale and press 『ON/OFF』. The indicator shows
and it decreases to
. Finally,
appears for one second and zero calibration ends.
F1.5 Load Calibration
add weights
Add weights to scale and make sure: full capacity *50% ≤ weights ≤ full capacity. Press 『ON/OFF』. Enter the same value as the weights. Wait until the scale is stable and press 『ON/OFF』. The indicator shows
and it decreases to
.
Finally, appears for one second and load calibration ends.
F1.6 Auto zero-tracking rate
Selectable parameters: OFF, 0.5d (default)
At net weight status, zero-tracking is ineffective.
F1.7 Auto zeroing range at power-on
Selectable parameters: OFF (default), ±2 %, ±10 %
F1.8 Zeroing range by button
Selectable parameters: OFF (default), ±2 %
F1.9 Digital filter
Selectable parameters: 0~9, default: 5
F1.10 Scale function setting
Selectable parameters: 0——–Totalization (default)
1——–Counting
F 1.11 Restoring factory default
Selectable parameters: 0——–No restore
1——–Restore parameters from F1 to F4 while scale parameters unaffected
F 2 Applicable Parameter Setting
F 2.1 Sampling method
Selectable parameters: 0—-weight on scale sampling (default)
1—–manual input
F 3 Indicator Parameter Setting
F 3.1 Date format setting
Selectable parameters: 0—–year.month.day (default)
1—–month.day.year
2—–day.month.year
F 3.2 Date setting (see F 3.1)
F 3.3 Time setting (format: hour.minute.second)
F 3.4 Overtime backlight shutdown time setting
Selectable parameters: 0~999 second (default: 0)
If 0 is set, this function is shutdown.
F 3.6 Auto power-off time setting
Selectable parameters: 0~60min (default: 0)
If 0 is set, this function is shutdown.
F 4 Serial Communication
F 4.1 Method
Selectable parameters:0—–no output (default)
1—–serial output (only when the scale is stable)
2——printing output (see Appendix Ⅲ)
F4.2 Data and verification setting
Selectable parameters: 8_N_1 —-8 digits, no verification (default)
7_E_1—–7 digits, odd verification
7_O_1 —-7 digits, even verification
8_E_1 —-8 digits, odd verification
8_O_1—–8 digits, even verification
F 4.3 Baud rate
Selectable parameters: 2400, 4800, 9600 (default), 19200
F 4.4 New line
Selectable parameters: 0~9 (default: 0)
F 5 Maintenance and Service
F 5.1 Button test
When the indicator displays , press〖ON/OFF〗〖ZERO〗〖TARE〗〖GROSS〗 〖PRINT〗and〖COUNT〗in order and the indicator displays “on.off” “Zero” “Tare” “Gross” “Print” and “Count”. Press 〖TOTAL〗to exit.
F 5.2 Display test
The indicator conducts an auto-inspection to make sure there is no lack of strokes. Press 〖TOTAL〗or〖ON/OFF〗to exit.
F 5.3 Inner code
The indicator shows the after-filtering inner code. Press 〖COUNT〗or〖ON/OFF〗to exit.
5.0 Functions
5.1 Totalization
F 1.10=0
Operation: Normally add weights to scale and press 〖TOTAL〗. The screen shows “Add-” and returns to the main weighing interface. Remember to zero the scale before putting things on scale every time; otherwise, there is no totalization.
Check and clear: Normally long press 〖TOTAL〗and “TOTAL ” shows for one second.
Then the total quantity interface “Cn xxx” and the total weight interface “t xx.xx” show.
Press〖Print〗or 〖GROSS〗 to switch between the above two interfaces. Press 〖ZERO〗 to clear the total quantity or the total weight value. Press 〖ON/OFF〗to confirm and press 〖COUNT〗to exit.
5.2 Counting
F 1.10=1
Operation: Normally press 〖COUNT〗to switch between the weight and quantity display.
Sampling: Long press 〖COUNT〗 until “SAMPLE” shows. Press 〖ON/OFF〗 and as “Sn XXX” shows, put in the correct quantity. If F 2.4=0, put the corresponding quantity on scale and press 〖ON/OFF〗to confirm the sample quantity and weight; if F 2.4=1, press 〖ON/OFF〗 and as “XXXXXX” shows, put in the corresponding weight and press 〖ON/OFF〗 to confirm the sample quantity and weight.
Appendix Ⅰ Indicator Prompt Message
Normally the indicator is stable and reliable. If the indicator fails, restart it first. Find out what the error is before repairing it. Repair the indicator according to the error codes.
No. | Code | Error | Countermeasure |
1 | ![]() ![]() |
Unable to zero after startup |
|
2 | ![]() |
load above upper limit | Reduce load |
3 | ![]() |
load less than lower limit | Press『Zero』to reset |
4 | ![]() ![]() |
Out of zeroing range | Remove load on scale if any. |
5 | ![]() |
Invalid | |
6 | ![]() |
EEPROM verification and error | Press 〖ON/OFF〗to restore and then restart. If the code shows again, return the indicator to factory; if there’s no code re-calibrate the indicator. Note: Restoring factory default includes restoring all parameters. |
7 | ![]() |
The scale is dynamic during calibration | Check the scale |
8 | ![]() |
Wrong date or time | Set date and time as required |
9 | ![]() |
Wrong AD initialization | Restart. If the error shows again return the indicator to factory. |
10 | ![]() |
load calibration: load weight to scale | Add weights to scale as required |
11 | ![]() |
Enter into menu setting | Press 〖ON/OFF〗to continue setting. |
12 | ![]() |
The end of zero/load calibration | |
13 | ![]() |
Load default value |
Appendix Ⅱ Serial output format
Serial output format
HEAD1:OL overload, underload or no zeroing at start-up
ST the scale is stable
US the scale is unstable
HEAD2:GS gross weight
NT net weight
DATA: data display
UNIT: kg/lb
CR/LF: New line
Example 1: stable, gross weight: 18.000kg, sp = space.
Example 2: unstable, net weight: -0.200kg, sp = space.
Appendix Ⅲ Printing Output Format
LIST
Date 2018/04/14
Time 15:08:46
———————–
Gross 2.061kg
Tare 0.000kg
Net 2.061kg
LIST
Date 2018/04/14
Time 15:10:18
———————–
Count 206PCS