SIGLENT logoSHN900A Series
Vector Network
Analyzer
Data Sheet
EN01ASIGLENT SHN900A Series Vector Network AnalyzerSIGLENT TECHNOLOGIES CO.,LTD

General Description

The SIGLENT SHN900A series of Vector Network Analyzers have a frequency range of 30 kHz to 26.5 GHz, which support 2-port scattering parameter, differential-parameter, and time-domain parameter measurements. The SHN900A series of VNAs are effective instrumentation for determining the Q-factor, bandwidth, and insertion loss of a filter, They feature impedance conversion, movement of measurement plane, limit testing, ripple test, fixture simulation, and adapter removal/insertion adjustments. The VNAs have five sweep types: Linear-Frequency mode, Log-Frequency mode, Power-Sweep mode, CW-Time mode, and Segment-Sweep mode. The SHN900A series VNAs also support scattering-parameter correction of SOLT, SOLR, TRL, Response, and Enhanced Response for increased flexibility in R&D and manufacturing applications.

Key Features

SIGLENT SHN900A Series Vector Network Analyzer - icon Frequency range: 30 kHz – 26.5 GHz
SIGLENT SHN900A Series Vector Network Analyzer - icon Frequency resolution: 1 Hz
SIGLENT SHN900A Series Vector Network Analyzer - icon Level resolution: 0.01 dB
SIGLENT SHN900A Series Vector Network Analyzer - icon Range of IFBW: 10 Hz~3 MHz
SIGLENT SHN900A Series Vector Network Analyzer - icon Setting range of output level:
SIGLENT SHN900A Series Vector Network Analyzer - icon -45 dBm ~ +10 dBm
SIGLENT SHN900A Series Vector Network Analyzer - icon Dynamic range: 110 dB(Typ.)
SIGLENT SHN900A Series Vector Network Analyzer - icon Types of calibration: Response calibration, Enhanced Response calibration, Full-one port calibration, Full-two port calibration, TRL calibration
SIGLENT SHN900A Series Vector Network Analyzer - icon Types of measurement: Scattering-parameter measurement, differential-parameter measurement, receiver measurement, time-domain parameter analysis, limit test, ripple test, impedance conversion, fixture simulation, adapter removal/insertion, spectrum analysis frequency offset, scalar mixer measurement, pulse measurement
SIGLENT SHN900A Series Vector Network Analyzer - icon Internal Bias-Tee connections
SIGLENT SHN900A Series Vector Network Analyzer - icon Support GPS, Time and Location Information Saving
SIGLENT SHN900A Series Vector Network Analyzer - icon Interface: LAN, USB Device, USB Host (USB-GPIB)
SIGLENT SHN900A Series Vector Network Analyzer - icon Remote control: SCPI/ Labview/ IVI based on USB-TMC / VXI-11 / Socket /Telnet / WebServer
SIGLENT SHN900A Series Vector Network Analyzer - icon 8.4-inch touch screen, Mouse, Keyboard

Models and Key Specifications

Model SHN914A SHN920A SHN926A
Frequency range 30kHz-14GHz 30kHz-20GHz 30kHz-26.5GHz
Ports 2
Frequency resolution 1 Hz
Level resolution 0.01 dB
Range of IFBW 10 Hz~3 MHz
Setting range of output level -45 dBm ~ +10 dBm
Dynamic range 110 dB(Typ.)
Types of calibration Response calibration, Enhanced Response calibration, Full-one port calibration, Full-two port calibration, TRL calibration
Types of measurement Scattering-parameter measurement, differential-parameter measurement, receiver measurement, time-domain parameter analysis, limit test, ripple test, impedance conversion, fixture simulation, adapter removal/insertion, enhanced time-domain parameter analysis (TDR), spectrum analysis, frequency offset, scalar mixer measurement, pulse measurement
Bias-Tees Support
Interface LAN, USB Device, USB Host(USB-GPIB)
Remote control SCPI/ Labview/ IVI based on USB-TMC/ VXI-11/ Socket/ Telnet/ WebServer
Display 8.4-inch touch screen
GPS Support

Design Features

Multi-window display: SIGLENT SHN900A Series Vector Network Analyzer - Multi-window display

Multi-format display: SIGLENT SHN900A Series Vector Network Analyzer - Multi-format display

Display and compare memory and current data: SIGLENT SHN900A Series Vector Network Analyzer - compare memory and current data

Display data hold: SIGLENT SHN900A Series Vector Network Analyzer - Display data hold

Impedance conversion: SIGLENT SHN900A Series Vector Network Analyzer - Impedance conversion

Equation Editor: SIGLENT SHN900A Series Vector Network Analyzer - Equation Editor

Port Extensions: SIGLENT SHN900A Series Vector Network Analyzer - Port Extensions

Embedding and De-Embedding: SIGLENT SHN900A Series Vector Network Analyzer - Embedding and De-Embedding

CAT SIGLENT SHN900A Series Vector Network Analyzer - CAT

Time-Domain analysis SIGLENT SHN900A Series Vector Network Analyzer - Time-Domain analysis

Enhanced Time-Domain analysis(TDR) SIGLENT SHN900A Series Vector Network Analyzer - Enhanced Time-Domain analysis

Spectrum analysis SIGLENT SHN900A Series Vector Network Analyzer - Spectrum analysis

Definitions

Specifications are valid under the following conditions: The instrument is within the calibration period, has been stored between 0 and 40℃ for at least 2 hours before use, and has been powered on and warmed up for at least 90 minutes. The specifications include the measurement uncertainty unless otherwise noted.
Specifications: All products are guaranteed to meet published specifications at room temperature (approximately 25℃), unless otherwise noted.
Typical: Performance deemed typical implies that 80 percent of the measurement results will meet the typical published performance with a 95th percentile confidence level at room temperature (approximately 25℃). Typical performance is not warranted and does not include measurement uncertainty.
Nominal: This value indicates the expected mean or average performance, or an attribute whose performance is by design, such as the 50 Ohm connector.

Specifications

Dynamic range

Frequency range IFBW Specification(dB) Typical(dB)
30 kHz – 1 MHz 10Hz 90 100
1 MHz – 6 GHz 100 110
6 GHz – 8 GHz 80 110
8 GHz – 14 GHz 100 105
14 GHz – 24 GHz 100 105
24 GHz – 26.5 GHz 95 100

Corrected system performance with calibration kit
User correction: On, system correction: On; Corrected system performance with Keysight 85052D 3.5mm calibration kit, isolation calibration performed.IFBW is 10 Hz, no averaging applied to data, and environmental temperature is 25℃ (± 5℃), with < 1℃ deviation from calibration temperature.

Specification (dB) 30kHz-3 GHz 3GHz-6 GHz 6GHz-14 GHz 9 GHz-20GHz 14GHz-26.5GHz
Directivity 41 39 37 37 37
Source match 36 30 29 29 29
Load match 41 37 35 35 35
Reflect tracking ±0.004 ±0.003 ±0.004 ±0.004 ±0.004
Transmission tracking ±0.06 ±0.09 ±0.11 ±0.11 ±0.11

Reflection uncertainty (Specification, Power: -10 dBm, IFBW:10 Hz):

SIGLENT SHN900A Series Vector Network Analyzer - Reflection uncertainty

Transmission uncertainty (Specification, Power: -10 dBm, IFBW:10 Hz):

SIGLENT SHN900A Series Vector Network Analyzer - Transmission uncertainty

Uncorrected system performance
User correction: Off, system correction: On; IFBW is 10 Hz, no averaging applied to data.

Specification (dB) 30kHz – 300kHz 300kHz – 1 GHz 1GHz-6 GHz 6 GHz-26.5GHz
Directivity 15 15 16 16
Source match 11 16 16 18
Load match 5 5 10 7
Reflect tracking ±1.4 ±1.4 ±1 ±1
Transmission tracking ±1.4 ±1.4 ±1 ±1

Test port output (Source)
SIGLENT SHN900A Series Vector Network Analyzer - icon Test port output frequency

Description Specification
Frequency range
SHN914A 30 kHz to 14 GHz
SHN920A 30 kHz to 20 GHz
SHN926A 30 kHz to 26.5 GHz
Frequency resolution 1 Hz
CW accuracy
Standard ± 1.0 ppm (23 ± 3℃)
Source stability
Standard ± 1.0 ppm (0 to 40℃)
± 0.5 ppm/year, ± 3.0 ppm/20 year

SIGLENT SHN900A Series Vector Network Analyzer - icon Test port output power

Description Specification
Preset power -10 dBm
Level accuracy ±1.5 dB@-10 dBm 30kHz ~ 20GHz
±2.5 dB@-10 dBm 20GHz ~ 26.5GHz
Level linearity
30 kHz – 100 kHz ±1 dB (-20 dBm to -11 dBm)
100 kHz – 25 MHz ±1 dB (-20 dBm to 0 dBm)
25 MHz – 20 GHz ±1 dB (-20 dBm to -1 dBm)
20 GHz – 26.5 GHz ±2 dB (-20 dBm to -8 dBm)
Range
30 kHz – 100 kHz -45 dBm to -11 dBm
100 kHz – 25 MHz -45 dBm to -5 dBm
25 MHz – 20 GHz -45 dBm to -1 dBm
20 GHz – 26.5 GHz -45 dBm to -8 dBm
Sweep range
30 kHz – 100 kHz -45 dBm to -11 dBm
100 kHz – 25 MHz -45 dBm to -5 dBm
25 MHz – 20 GHz -45 dBm to -1 dBm
20 GHz – 26.5 GHz -45 dBm to -8 dBm
Max leveled power
30 kHz – 100 kHz -11 dBm (Typ.)
100 kHz – 25 MHz -4 dBm (Typ.)
25 MHz – 100 MHz 0 dBm (Typ.)
100 MHz – 6 GHz 2 dBm (Typ.)
6 GHz – 8 GHz 0 dBm (Typ.)
8 GHz – 14 GHz 2 dBm (Typ.)
14GHz – 20 GHz 0 dBm (Typ.)
20 GHz – 26.5 GHz -6 dBm (Typ.)
Level resolution 0.05 dB

SIGLENT SHN900A Series Vector Network Analyzer - icon Test port output signal purity

Description Specification
2nd or 3rd harmonics (0 dBm)
100 kHz to 25 MHz < -10 dBc
25 MHz to 8 GHz < -10 dBc
9 GHz to 26.5 GHz < -10 dBc
Non-harmonic spurious (0 dBm) < -20 dBc

Test port input
SIGLENT SHN900A Series Vector Network Analyzer - icon Test port input levels

Description Specification Typical
Max input level
30 kHz-14 GHz +10 dBm
14 GHz-26.5 GHz +10 dBm
Damage input level
30 kHz-26.5 GHz +27 dBm (RF) or 35 V (DC)
Precision
30 kHz -1 GHz ±1.5 dB@-15 dBm ±0.2 dB@-15 dBm
1 GHz – 20 GHz ±2.0 dB@-15 dBm ±0.2 dB@-15 dBm
20 GHz – 26.5 GHz ±2.5 dB@-15 dBm ±0.5 dB@-15 dBm
Crosstalk
30 kHz -100 kHz -95 dB -105 dB
100 kHz – 6 GHz -75 dB -120 dB
6 GHz – 9 GHz -80 dB -105 dB
9 GHz – 20 GHz -95 dB -105 dB
20 GHz – 26.5 GHz -60 dB -70 dB
Noise floor
30 kHz – 50 kHz -70 dBm/Hz -80 dBm/Hz
50 kHz – 200 kHz -90 dBm/Hz -110 dBm/Hz
200 kHz – 6.2 GHz -100 dBm/Hz -120 dBm/Hz
6.2 GHz – 9 GHz -90 dBm/Hz -100 dBm/Hz
9 GHz – 22 GHz -100 dBm/Hz -115 dBm/Hz
22 GHz – 26.5 GHz -80 dBm/Hz -110 dBm/Hz
Compression level(+10 dBm)
Magnitude
30 kHz- 26.5 GHz 1 dB
Phase
30 kHz- 26.5 GHz 5 deg

SIGLENT SHN900A Series Vector Network Analyzer - icon Trace noise

Description Specification Typical
Note:Setting max output power
Transmission trace noise magnitude
30 kHz- 50 kHz (IFBW=30 Hz) 0.003 dB rms 0.0015 dB rms
50 kHz- 1 MHz (IFBW=30 Hz) 0.003 dB rms 0.0015 dB rms
1 MHz- 9 GHz (IFBW=10 kHz) 0.003 dB rms 0.0015 dB rms
9 GHz-14 GHz (IFBW=10 kHz) 0.005 dB rms 0.0025 dB rms
14GHz-26.5 GHz (IFBW=10 kHz) 0.005 dB rms 0.0025 dB rms
Reflection trace noise magnitude
30 kHz- 50 kHz (IFBW=30 Hz) 0.003 dB rms 0.0005 dB rms
50 kHz- 1 MHz (IFBW=30 Hz) 0.003 dB rms 0.0007 dB rms
1 MHz- 9 GHz (IFBW=10 kHz) 0.003 dB rms 0.0015 dB rms
9 GHz-14 GHz (IFBW=10 kHz) 0.004 dB rms 0.002 dB rms
14GHz-26.5 GHz (IFBW=10 kHz) 0.004 dB rms 0.002 dB rms
Transmission trace noise phase
30 kHz- 50 kHz (IFBW=30 Hz) 0.03 deg rms 0.015 deg rms
50 kHz- 1 MHz (IFBW=30 Hz) 0.03 deg rms 0.015 deg rms
1 MHz- 9 GHz (IFBW=10 kHz) 0.04 deg rms 0.004 deg rms
9 GHz-14 GHz (IFBW=10 kHz) 0.04 deg rms 0.004 deg rms
14GHz-26.5 GHz (IFBW=10 kHz) 0.06 deg rms 0.006 deg rms
Reflection trace noise phase
30 kHz- 50 kHz (IFBW=30 Hz) 0.03 deg rms 0.015 deg rms
50 kHz- 1 MHz (IFBW=30 Hz) 0.03 deg rms 0.015 deg rms
1 MHz- 9 GHz (IFBW=10 kHz) 0.04 deg rms 0.004 deg rms
9 GHz-14 GHz (IFBW=10 kHz) 0.04 deg rms 0.004 deg rms
14GHz-26.5 GHz (IFBW=10 kHz) 0.06 deg rms 0.006deg rms

SIGLENT SHN900A Series Vector Network Analyzer - icon Stability

Description Specification Typical
Magnitude
30 kHz- 9 GHz ± 0.01 dB/°C
9 GHz- 26.5 GHz ± 0.05 dB/°C
Phase
30 kHz- 9 GHz ± 0.1 deg/°C
9 GHz- 26.5 GHz ± 0.3 deg/°C

SIGLENT SHN900A Series Vector Network Analyzer - icon Dynamic accuracy

Description Specification
Relative to -10 dBm input power
Magnitude
-10 dBm ± 0.5 dB
-30 dBm ± 0.5 dB
-55 dBm ± 2.5 dB
Phase
-10 dBm ± 4.5 deg
-30 dBm ± 5 deg
-55 dBm ± 16.5 deg

SA Mode

Description Specification Typical
Power Range -70dBm – +15dBm
Noise Floor 110 dBm/Hz – 130dBm/Hz 130dBm/Hz
Phase Noise ≤ -98dBc/Hz (1GHz@100kHz)
Max Input Level Without Damaged 27dBm
Residual Response ≤ -80dBm -100dBm

CAT Mode

Description Specification
DTF Range: Velocity Factor ×Light Velocity×(Points Number- 1)/BW×2 Resolution: Range/(Points Number- 1)
Return Loss Range(dB) -6k dB – +6k dB
VSWR Resolution 1mdB – 1k dB
VSWR Range 1.001 – 1G
VSWR Resolution 0.001

Sweep time

Start frequency: 30 kHz, Stop frequency: 14 GHz; IFBW: 500 kHz
Points 201 401 1601 6401
2-port cal 28 ms 28 ms 75 ms 300 ms
Start frequency: 30 kHz, Stop frequency: 14 GHz; IFBW: 100 kHz.
Points 201 401 1601 6401
2-port cal 30 ms 30 ms 85 ms 340 ms
Start frequency: 30 kHz, Stop frequency: 14 GHz; IFBW: 10 kHz.
Points 201 401 1601 6401
2-port cal 60 ms 70 ms 350 ms 1400 ms
Start frequency: 30 kHz, Stop frequency: 14 GHz; IFBW: 1 kHz.
Points 201 401 1601 6401
2-port cal 300 ms 500 ms 2500 ms 10000 ms
Start frequency: 30 kHz, Stop frequency: 26.5 GHz; IFBW: 500 kHz.
Points 201 401 1601 6401
2-port cal 28 ms 28 ms 75 ms 300 ms
Start frequency: 30 kHz, Stop frequency: 26.5 GHz; IFBW: 100 kHz.
Points 201 401 1601 6401
2-port cal 30 ms 30 ms 85 ms 340 ms
Start frequency: 30 kHz, Stop frequency: 26.5 GHz; IFBW: 10 kHz.
Points 201 401 1601 6401
2-port cal 60 ms 70 ms 350 ms 1400 ms
Start frequency: 30 kHz, Stop frequency: 26.5 GHz; IFBW: 1 kHz.
Points 201 401 1601 6401
2-port cal 300 ms 500 ms 2500 ms 10000 ms

General information

Description Characteristics
Operating environment
Temperature 0 to 40 ℃
Humidity 85%: 40 ℃, 24 hours
Altitude 0 to 3000 m
Non-operating storage environment
Temperature -20 ℃ to 70 ℃
Humidity 85%: 65 ℃, 24 hours
Altitude 0 to 15000 m
Size W×H×D=310 mm × 215 mm × 78.5 mm
Weight 3.2 kg
EMC
Conducted disturbance: CISPR 11/EN 55011 CLASS A group 1, 150 kHz – 30 MHz
Radiated disturbance: CISPR 11/EN 55011 CLASS A group 1, 30 MHz -1 GHz
Electrostatic discharge(ESD): IEC61000-4-2/EN61000-4-2 4.0 kV (contact), 8.0 kV (air)
Radio-frequency electromagnetic field Immunity: IEC 61000-4-3/EN 61000-4-3 10 V/m (80 MHz to 1 GHz);
3 V/m (1.4 GHz to 2 GHz);
1 V/m (2.0 GHz to 2.7 GHz)
Electrical fast transients (EFT): IEC 61000-4-4/EN 61000-4-4 2 kV (AC power ports)
Surges: IEC 61000-4-5/EN 61000-4-5 1 kV (Line to line) ; 2 kV (Line to ground)
Radio-frequency continuous conducted Immunity: IEC 61000-4-6/EN 61000-4-6 3 V, 0.15-80 MHz
Voltage dips and interruptions: IEC 61000-4-11/EN 61000-4-11 Voltage dips:
0% UT during 1 cycle;
40% UT during 10/12 cycles; 70% UT during 25/30 cycles;
Voltage interruptions: 0% UT during 250 cycles
Safety
UL 61010-1:2012/R: 2018-11; CAN/CSA-C22.2 No. 61010-1:2012/A1:2018-11. UL 61010-2-030:2018; CAN/CSA-C22.2 No. 61010-2-030:2018.

Panel information

Top Panel
RF connectors 3.5mm NMD (male), 50Ω
USB Host USB-A 2.0
USB Device USB-C 2.0
LAN LAN (VXI11), 10/100 Base, RJ-45
GPS Antenna SMA Female, 3.3V, 50Ω
Bias Out SMB Female, 12V-32V, step 0.1V
External Trigger Input 1 kΩ, 5V TTL, BNC Female
10M Reference Input 10 MHz, -5 dBm~+10 dBm, BNC Female, 50Ω
Remote control
Hardware connectors LAN, USB-TMC, GPIB (USB-GPIB adaptor)
Remote control interfaces SCPI/ Labview/ IVI based on USB-TMC/ VXI-11/ GPIB/ Socket/Telnet NI-MAX Web Browser (HTML 5 Supported)

Ordering Information

Items Description Order number
Products 2 ports, 14G Vector Network Analyzer SHN914A
2 ports, 20G Vector Network Analyzer SHN920A
2 ports, 26.5G Vector Network Analyzer SHN926A
standard fittings Quick Start, USB Type C Line, Rechargeable lithium battery, AC-DC adapter, Portable bag
TDA Option Time Domain Analysis SHN900-TDA
TDR Option Enhanced Time Domain Analysis SHN900-TDR
SA Option Spectrum analysis SHN900-SA
3.5mm, Male, 50Ω Calibration Kit, 0-4.5GHz F603ME
3.5mm, Female, 50Ω Calibration Kit, 0-4.5GHz F603FE
3.5mm, Male, 50Ω Calibration Kit, 0-9GHz F604MS
3.5mm, Female, 50Ω Calibration Kit, 0-9GHz F604FS
3.5mm, Male and Female, 50Ω Calibration Kit, 0-9GHz F604TS
3.5mm, Male and Female, 50Ω Calibration Kit, 0-26.5GHz F606TS
Electronic Calibration Kit SEM5000A
RF Test Demo Board SNA-TB01
Adjustable Differential TDR probe DC-18 GHz ADP-18
Adjustable Differential TDR probe DC-26.5 GHz ADP-26
Adjustable Differential TDR probe DC-18 GHz ASP-18
Adjustable Differential TDR probe DC-26.5 GHz ASP-26
SMA(M)-SMA(M) cable DC-18 GHz, 1000 mm SMA-SMA-18L
SMA(M)-SMA(M) cable DC-26.5 GHz, 1000 mm SMA-SMA-26L
SMA(F)-SMA(M) cable DC-26.5 GHz, 1000 mm SMAF-SMA-26L
NMD 3.5 female-NMD 3.5 Male DC-26.5 GHz, 635 mm V26-N35MN35F-25IN
NMD 3.5 female-APC 3.5 female DC-26.5 GHz, 635 mm V26-N35FA35F-25IN
USB-GPIB Adaptor USB-GPIB
GPS antenna, SMA(M), 1000 mm ANT-GPS1

Headquarters:
SIGLENT Technologies Co., Ltd
Add: Bldg No.4 & No.5, Antongda Industrial
Zone, 3rd Liuxian Road, Bao’an District,
Shenzhen, 518101, China
Tel: + 86 755 3688 7876
Fax: + 86 755 3359 1582
Email: sales@siglent.com
Website: int.siglent.com

North America:
SIGLENT Technologies America, Inc
6557 Cochran Rd Solon, Ohio 44139
Tel: 440-398-5800
Toll Free: 877-515-5551
Fax: 440-399-1211
Email: info@siglentna.com
Website: www.siglentna.com

Europe:
SIGLENT Technologies Germany GmbH
Add: Staetzlinger Str. 70
86165 Augsburg, Germany
Tel: +49(0)-821-666 0 111 0
Fax: +49(0)-821-666 0 111 22
Email: info-eu@siglent.com
Website: www.siglenteu.com

About SIGLENT
SIGLENT is an international high-tech company, concentrating on R&D, sales, production and services of electronic test & measurement instruments.
SIGLENT first began developing digital oscilloscopes independently in 2002.
After more than a decade of continuous development, SIGLENT has extended its product line to include digital oscilloscopes, isolated handheld oscilloscopes, function/arbitrary waveform generators, RF/MW signal generators, spectrum analyzers, vector network analyzers, digital multimeters, DC power supplies, electronic loads and other general purpose test instrumentation. Since its first oscilloscope was launched in 2005, SIGLENT has become the fastest growing manufacturer of digital oscilloscopes. We firmly believe that today SIGLENT is the best value in electronic test & measurement.

Follow us on
Facebook: SiglentTechSIGLENT SHN900A Series Vector Network Analyzer - QR Codehttps://www.facebook.com/SiglentTech

Documents / Resources

SIGLENT SHN900A Series Vector Network Analyzer [pdf] Owner's Manual
SHN900A Series Vector Network Analyzer, SHN900A Series, Vector Network Analyzer, Network Analyzer, Analyzer

References

Leave a comment

Your email address will not be published. Required fields are marked *