OPERATOR PARAMETERS

OPERATOR PARAMETERS
Parameters Settings (Default Value)
Filter Repair Ack >> Yes No

(Default : No)

Cascade HPLP Ack >> Cabinet HPLP Ack >> Yes No

(Default : No)

Ctrl Setpoint >> Setpoint Low Limit to Setpoint High Limit

(Default : -80.0)

Ctrl Lo Deviation >> 0.2 to 99.9

(Default : 2.0)

Ctrl Hi Deviation >> 0.2 to 99.9

(Default : 2.0)

Ctrl Zero Offset >> -50.0 to 50.0

(Default : 0.0)

Filter Zero Offset >> or

Coil Zero Offset >>

-50.0 to 50.0

(Default : 0.0)

Defrost Start >> Defrost Stop >> Start Stop

(Default : NA)

Change Password >> 1 to 100

(Default : 0)

SUPERVISORY SERIAL PARAMETERS
Parameters Settings (Default Value)
Device Id >> 1 to 127

(Default : 1)

Baud Rate >> 2400, 4800,

9600, 19200,

38400, 57600

(Default : 9600)

Parity >> None Even Odd

(Default : Even)

Serial Write >> No Yes

(Default : No)

SUPERVISORY     CONTROL
Parameters Settings (Default Value)
Setpoint Lo Limit >> -100.0 to Setpoint HI Limit (Default : -100.0)
Setpoint Hi Limit >> ‘Single’ or ‘Single with Filter’ Compressor Mode

Setpoint LO Limit to 100.0

‘Cascade’ or ‘Cascade with Filter’ or ‘Cascade with Coil’ Compressor Mode Setpoint LO Limit to Cascade SP

(Default : -10.0)

Comp On Offset >> 0.1 to 99.9

(Default : 1.0)

Comp Off Offset >> 0.1 to 99.9

(Default : 0.1)

PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 8 PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 10 PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 11

FACTORY    ALARM PARAMETERS
Parameters Settings (Default Value)
Low Alarm Hyst >> 0.1 to 99.9

(Default : 0.2)

Low Alarm Inhibit >> Disable Enable

(Default : Enable)

High Alarm Hyst >> 0.1 to 99.9

(Default : 0.2)

High Alarm Inhibit >> Disable Enable

(Default : Enable)

FACTORY    COMPRESSOR OPERATION
Parameters Settings (Default Value)
Operation Mode>> Single Single with Filter

Cascade Cascade with Filter Cascade with Coil

(Default : Cascade with Coil)

Cascade Sp>> -70.0 to 10.0

(Default : -30.0)

Off To On Delay>> 0 to 1000 Sec (Default : 200 Sec)
Switching Delay>> 0 to 1000 Sec (Default : 15 Sec)
Cascade Comp Hyst >> 1 to 50.0

(Default : 20)

FACTORY     FILTER OPERATION
Parameters Settings (Default Value)
Filter Alarm Sp >> -100.0 to 100.0

(Default : 55.0)

Filter Alarm Hyst >> 0.1 to 99.9

(Default : 0.2)

Filter Cut-off Sp >> Alarm SP to 100.0 (Default : 60.0)
FACTORY    DEFROST OPERATION
Parameters Settings (Default Value)
Operation >> Disable Automatic Manual

(Default : Disable)

Defrost Time [M:S] >> 1 to 99.59 (Min:Sec) (Default : 5 Sec)
Defrost Interval [H:M] >> 1 to 99.59 (Hour:Min) (Default : 8 Min)
Parameters Settings (Default Value)
Switch Logic >> Open : Door Open Close : Door Open

(Default : Close : Door Open)

Door Alrm Dly (sec) >> 0 to 1000 Sec (Default : 60 Sec)
FACTORY     CASCADE COMP HPLP

(Available only for Cascade Operation Mode)

Parameters Settings (Default Value)
Enable >> No Yes

(Default : No)

Switch Logic >> Open : HP-LP Close : HP-LP

(Default : Close : HP-LP)

FACTORY CABINET COMP HPLP
Parameters Settings (Default Value)
Enable >> No Yes

(Default : No)

Switch Logic >> Open : HP-LP Close : HP-LP

(Default : Close : HP-LP)

FACTORY FACTORY DEFAULT
Parameters Settings (Default Value)
Set To Default >> No Yes

(Default : No)

FACTORY     PASSWORD
Parameters Settings (Default Value)
Change Password >> 2000 to 2999

(Default : 321)

FACTORY     EXIT
Parameters Settings (Default Value)
Exit Setup Mode >> No Yes

(Default : No)

Table 1
What it means Range (Min. to Max.) Resolution
3-wire, RTD Pt100 -199.9 to 600.0°C Fixed 0.1°C
0 to 20mA DC current  

-199.9 to 999.9 units

 

Fixed

0.1 unit

4 to 20mA DC current
0 to 5.0V DC voltage
0 to 10.0V DC voltage
1 to 5.0V DC voltage

FRONT PANEL LAYOUT

Keys OperationPPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 1

PV Error Indications

Message Error Type Cause
PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 2 Sensor Open RTD Pt100 Sensor/Cable is Broken / Open
PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 3 Over-range Temperature above Max. Specified Range
PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 4 Under-range Temperature below Min. Specified Range

ELECTRICAL CONNECTIONSPPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 5

CASCADE OPERATION MODE

* For more Information refer User Manual

Cascade Compressors Operation Mode (with Coil Sensor)PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 6

Cascade Compressors Operation Mode (without Coil Sensor)PPI-FrizCon-Deep-Freezer-Controller-with-Graphic-Display-fig 6

This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

101, Diamond Industrial Estate, Navghar, Vasai Road (E), Dist. Palghar – 401 210.
Sales : 8208199048 / 8208141446
Support : 07498799226 / 08767395333
E: sales@ppiindia.net,
support@ppiindia.net