NeuroNexus IST-CM Manual Insertion Tool User Guide

NeuroNexus IST-CM Manual Insertion Tool User Guide

IST-CM
A. Place the CM package carefully in the IST-CM Holder.
B. While holding the package in place. Set clamp over the holder and secure it with the two M2 screws.
Note: Always Be Mindful of the probe tip.

NeuroNexus IST-CM Manual Insertion Tool User Guide - IST-CM

IST-Hybrid

  • A. Loosen the M2 Phillips Screw on the backside of IST-Hybrid and lower the IST-holder to the bottom of the IST-Hybrid.
  • B. Carefully Slide the Hybrid package in the holder. One might want to adjust wires at this time. Make sure they don’t run into the probe.
  • C. Again, Loosen the M2 screw, while holding the package in the holder carefully Slide the IST-Holder up the IST-Hybrid, so that the probe is resting on the flat of the IST-Hybrid (see blue arrow). Make Sure only the shanks are hanging off the IST-Hybrid.
  • D. With the desired Adhesive (PEG, or something stronger and biocompatible) and fix the probe to the IST-Hybrid.
  • E. Now the IST- is ready to be attach to the Stereotactic equipment and read for inserted. Note: Always Be Mindful of the probes tip.

NeuroNexus IST-CM Manual Insertion Tool User Guide - IST-Hybrid

IST-X3-16/32/64

  • A. Place the x3-package in the IST-Holder.
  • B. Place the IST-Clamp over the holder and package.
  • C. While holding the package in place. Secure the clamp over the holes and secure it with two M2 screws.
  • D. Make sure that the IST-Holder is in its lowest position on the IST-Hybrid. If one needs to lower the holder, loosen up the M2 screw on the back side of the IST-Hybrid. Lower the Holder in the Implant position and re-tighten the screw.
  • E. Attach to stereotactic equipment and is now ready for insertion.

NeuroNexus IST-CM Manual Insertion Tool User Guide - IST-X3-16

IST-X3-16/32/64-H

  • A. Place the MCM package Carefully in the IST-MCM Holder
  • B. While holding the package in place. Set Clamp over the holes and secure it with two M2 screws.

NeuroNexus IST-CM Manual Insertion Tool User Guide - IST-X3-16IST-AV/I64/128/256

  • A. Place the AV package carefully in the IST-AV Holder
  • B. While holding the package in place. Set Clamp over the holes and secure it with two M2 screws.
  • C. Now one can attach the IST-2mm Rod to the stereotactic equipment and is now ready for insertion.

NeuroNexus IST-CM Manual Insertion Tool User Guide - IST-AV

IST-AV/IH64/128/256

  • A. First, attach the IST-PIM (Probe Inserter Mount), by attaching it to the backside of the IST-Holder, and secure it with the two M2 screws.
  • B. Place the AV/IH package Carefully in the IST-AVH Holder
  • C. While holding the package in place. Set Clamp over the holes and secure it with two M2 screws.
  • D. With the desired Adhesive (PEG, or something stronger and biocompatible) and fix the probe to the IST-Hybrid.
  • E. Now the IST- is ready to be attach to the Stereotactic equipment and read for inserted.

NeuroNexus IST-CM Manual Insertion Tool User Guide - IST-AV 2

NeuroNexus Technologies, Inc. ©2023 | 5200 S. State Street, Suite 200, Ann Arbor, Michigan USA
Telephone: +1.734.913.8858 | Fax: +1.734.786.0069 | support@neuronexus.com

Documents / Resources

NeuroNexus IST-CM Manual Insertion Tool [pdf] User Guide
IST-CM, IST-Hybrid, IST-X3-16-32-64, IST-X3-16-32-64-H, IST-AV-I64-128-256, IST-CM Manual Insertion Tool, Manual Insertion Tool, Insertion Tool
Neuronexus IST-CM Manual Insertion Tool [pdf] User Guide
IST-CM Manual Insertion Tool, IST-CM, Manual Insertion Tool, Insertion Tool, Tool

References

Leave a comment

Your email address will not be published. Required fields are marked *