SPECIFICATIONS
NI 5753
16-Bit, 120 MS/s, 12 Channel DC/16 Channel AC Receiver Adapter Module
NI 5753 12 Channel DC/16 Channel AC Receiver Adapter Module
This document lists specifications for the NI 5753 adapter module. Pair these specifications with the specifications listed in your FlexRIO FPGA module specifications document or your Controller for FlexRIO specifications document.
Caution The protection provided by the NI 5753 can be impaired if it is used in a manner not described in this document.
Caution To avoid permanent damage to the NI 5753, disconnect all signals connected to the NI 5753 before powering down the module, and only connect signals after the module has been powered on by the FlexRIO FPGA module or the Controller for FlexRIO.
Note All numeric specifications are typical unless otherwise noted. All graphs illustrate the performance of a representative module.
Specifications are subject to change without notice. For the most recent device specifications, visit ni.com/manuals.
FlexRIO Documentation
Table 1. FlexRIO Documentation Locations and Descriptions
Document | Location | Description |
Getting started guide for your FlexRIO FPGA module or Controller for FlexRIO |
Available from the Start menu and at ni.com/manuals. |
Contains installation instructions for your FlexRIO system. |
Specifications document for your FlexRIO FPGA module or Controller for FlexRIO | Available from the Start menu and at ni.com/manuals. |
Contains specifications for your FlexRIO FPGA module or Controller for FlexRIO. |
Getting started guide for your adapter module |
Available from the Start menu and at ni.com/manuals. |
Contains signal information, examples, and CLIP details for your adapter module. |
Specifications document for your adapter module |
Available from the Start menu and at ni.com/manuals. | Contains specifications for your adapter module. |
LabVIEW FPGA Module Help |
Embedded in LabVIEW Help and at ni.com/manuals. | Contains information about the basic functionality of the LabVIEW FPGA Module. |
Real-Time Module Help | Embedded in LabVIEW Help and at ni.com/manuals. |
Contains information about realtime programming concepts, stepby-step instructions for using LabVIEW with the Real-Time Module, reference information about Real-Time Module VIs and functions, and information about LabVIEW features on real-time operating systems. |
FlexRIO Help | Available from the Start menu and at ni.com/manuals. |
Contains information about the FPGA module front panel connectors and I/O, controller for FlexRIO front panel connectors and I/O, programming instructions, and adapter module component-level IP (CLIP). |
FlexRIO Adapter Module Development Kit User Manual |
Available from the Start menu at Start» All Programs» National Instruments» NI FlexRIO» NI FlexRIO Adapter Module Development Kit» Documentation. |
Contains information about how to create custom adapter modules for use with FlexRIO FPGA modules. |
LabVIEW Examples | Available in NI Example Finder. In LabVIEW, click Help» Find Examples» Hardware Input and Output»FlexRIO. |
Contains examples of how to run FPGA VIs and Host VIs on your device. |
IPNet | Located at ni.com/ipnet. | Contains LabVIEW FPGA functions and intellectual property to share. |
FlexRIO product page | Located at ni.com/flexrio. | Contains product information and data sheets for FlexRIO devices. |
General Characteristics
Number of channels | |
DC-coupled | 12 analog input |
AC-coupled | 16 analog input |
Input impedance | 50 Ω |
Sample rate | 120 MS/s |
Frequency range | |
DC-coupled | DC to 250 MHz |
AC-coupled | 0.1 MHz to 250 MHz |
ADC type | AD9653 1 |
Related Information
Refer to www.analog.com for more information about the AD9653.
Typical Specifications
Analog Input Levels
ADC offset error (2 Vpp input) 2 | -6 mV |
Uncalibrated DC coupling offset error 3 | ±10 mV |
ADC gain error | -5% full scale |
DC offset control (DC-coupled option) | |
Range (at input) | -0.5 V to 0.5 V |
Resolution | 500 µV/step |
Full-scale input (0 dBFS, 50 Ω, 20.1 MHz) | |
AC-coupled build | |
Number of channels | 16 |
Input voltage | 1.8 Vpp |
Input power | +9.1 dBm |
- The AC-coupled version uses four AD9653 to implement 16 channels. The DC-coupled version uses three AD9653 to implement 12 channels.
- ADC offset dominates when AC-coupled, does not include offset compensation.
- Typical DC offset due to component tolerance, does not include system DC variation.
DC-coupled build | |
Number of channels | 12 |
Input voltage | 0.9 Vpp |
Input power | +3.2 dBm |
AC-Coupled Performance
Bandwidth | |
-3 dB | 1 MHz to 176 MHz |
-1 dB | 1 MHz to 85 MHz |
SNR 4 | |
20.17 MHz input | 75.86 dBFS |
70.17 MHz input | 70.75 dBFS |
124.17 MHz input | 66.09 dBFS |
SINAD 4 | |
20.17 MHz input | 75.8 dBFS |
70.17 MHz input | 70.56 dBFS |
124.17 MHz input | 65.95 dBFS |
SFDR 4 | |
20.17 MHz input | -94.03 dBc |
70.17 MHz input | -84.84 dBc |
124.17 MHz input | -79.52 dBc |
RX channel to RX channel isolation 5 | |
1 MHz | 86 dB |
50 MHz | 77 dB |
100 MHz | 71 dB |
150 MHz | 68 dB |
200 MHz | 64 dB |
4 Measured with a 120 MHz onboard Sample Clock and a -1 dBFS input signal.
5 Worst adjacent channel with -1 dBFS input.
Figure 1. AC-Coupled Passband
Figure 2. Analog Input One-Tone Spectral Measurement (20.17 MHz, -1 dBFS)
DC-Coupled Performance
Bandwidth | |
-3 dB | 1 MHz to 117 MHz |
-1 dB | 1 MHz to 64 MHz |
SNR 4 | |
20.17 MHz | 69.82 dBFS |
70.17 MHz | 66.95 dBFS |
124.17 MHz | 63.27 dBFS |
SINAD 4 | |
20.17 MHz | 69.66 dBFS |
70.17 MHz | 64.67 dBFS |
124.17 MHz | 54.64 dBFS |
SFDR 4 | |
20.17 MHz | 85.66 dBc |
70.17 MHz | 69.19 dBc |
122.17 MHz | 56.38 dBc |
RX channel to RX channel isolation 6 | |
1 MHz | 99 dB |
50 MHz | 78 dB |
100 MHz | 72 dB |
150 MHz | 68 dB |
200 MHz | 66 dB |
Figure 3. DC-Coup led Passband
Figure 4. Analog Input One-Tone Spectral Measurement (20.17 MHz, -1 dBFS)
DC Power Requirements
Current Requirements
VEEPROM | 2 mA |
AC-coupled build | |
VccoA | <1 mA |
VccoB | <1 mA |
P33V | 870 mA |
P12V | 211 mA |
DC-coupled build | |
VccoA | <1 mA |
VccoB | <1 mA |
P33V | 866 mA |
P12V | 231 mA |
Voltage Requirements
VccoA, VccoB | 2.50 V |
VEEPROM | 3.30 V |
P33V | 3.3 V |
P12V | 12 V |
CLK IN
The CLK IN connector provides access to the external Reference Clock and the external Sample Clock.
External Reference Clock (CLK IN) | |
Input impedance | 50 Ω |
Coupling | AC |
Input voltage (50 Ω) | 1.4 Vpp to 4.4 Vpp |
Input power (50 Ω) | +7 dBm to +16.8 dBm |
Frequency | 10 MHz |
External Sample Clock (CLK IN) | |
Input impedance | 50 Ω |
Coupling | AC |
Input voltage (50 Ω) | 1.4 Vpp to 2.5 Vpp |
Input power (50 Ω) | +7 dBm to +12 dBm |
Frequency | 120 MHz |
Internal Clock
Internal Sample Clock | |
Frequency (default) | 120 MHz |
Reference spurs | 60 dBc |
Phase Noise | |
10 kHz offset | -95 dBc/Hz |
100 kHz offset | -115 dBc/Hz |
Internal Reference | |
Frequency (default) | 10 MHz |
Stability | ±1.0 ppm |
Phase Noise | |
1 kHz offset | -137 dBc/Hz |
10 kHz offset | -150 dBc/Hz |
100 kHz offset | -155 dBc/Hz |
AUX I/O (Port 0 DIO <0..3>, Port 1 DIO <0..3>, and PFI <0..3>)
Number of channels | 12 bidirectional (8 DIO and 4 PFI) |
Connector type | HDMI |
Interface standard | 3.3 V LVCMOS |
Interface logic | |
Maximum VIL | 0.8 V |
Minimum VIH | 2.0 V |
Maximum VOL | 0.4 V |
Minimum VOH | 2.7 V |
Maximum VOH | 3.6 V |
Zout | 50 Ω ± 20% |
Iout (DC) | ±2 mA |
Pull-down resistor | 150 kΩ |
Recommended operating voltage | -0.3 V to 3.6 V |
Overvoltage protection | ±10 V |
Maximum toggle frequency | 100 MHz |
+5 V maximum power | 10 mA |
+5 V voltage tolerance | 4.2 V to 5 V |
Power Draw
AC-coupled build | 5.40 W |
DC-coupled build | 5.62 W |
Physical
Dimensions | 11.7 x 2.00 x 10.2 cm (4.6 x 0.8 x 4.0 in.) |
Weight | 317.5 g (11.2 oz) |
Front panel connectors | One ganged MCX connector, one SMB connector, and one HDMI connector |
Environment
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only. |
Operating Environment
Ambient temperature range | 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.) |
Relative humidity range | 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Storage Environment
Ambient temperature range | -40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.) |
Relative humidity range | 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Shock and Vibration
Operating shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) |
Random vibration | |
Operating | 5 Hz to 500 Hz, 0.3 grms |
Nonoperating | 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.
U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2016 National Instruments. All rights reserved.
375442A-01 Dec16
COMPREHENSIVE SERVICES
We offer competitive repair and calibration services, as well as easily accessible documentation and free downloadable resources.
SELL YOUR SURPLUS
We buy new, used, decommissioned, and surplus parts from every NI series.
We work out the best solution to suit your individual needs.
My Sell For Cash
Get Credit
Receive a Trade-In Deal
OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Refurbished, and Reconditioned NI Hardware.
Bridging the gap between the manufacturer and your legacy test system.
1-800-915-6216
www.apexwaves.com
sales@apexwaves.com
All trademarks, brands, and brand names are the property of their respective owners.
Documents / Resources
![]() |
NATIONAL INSTRUMENTS NI 5753 12 Channel DC/16 Channel AC Receiver Adapter Module [pdf] Owner's Manual NI 5753, NI 5753 12 Channel DC 16 Channel AC Receiver Adapter Module, NI 5753, 12 Channel DC 16 Channel AC Receiver Adapter Module, 12 Channel AC Receiver Adapter Module, DC 16 Channel AC Receiver Adapter Module, 16 Channel AC Receiver Adapter Module, AC Receiver Adapter Module, Receiver Adapter Module, Receiver Module, Adapter Module, Module |