Maxtec MaxBlend2 Oxygen Blender

Specifications

  • Model: MaxBlendTM 2
  • Flow Settings: Low Flow / High Flow
  • Power Specification: 7.5V (MAX), 1.9W/250mA (MAX)
  • Protection Class: II, Type B
  • Protection Against Water: IPX1
  • Mode of Operation: Continuous

Safety Precautions

Before operating the MaxBlend2, ensure you have read and understood the user manual completely to avoid unsafe operating conditions.

Powering On
Connect the device to a power source with the specified voltage (7.5V MAX).

Selecting Flow Settings
Choose between Low Flow or High Flow settings based on your requirements.

Operation
Operate the MaxBlend2 continuously for the desired oxygen concentration output.

Product Disposal
Follow the provided disposal instructions for the sensor, batteries, and circuit board. Return sensor to Maxtec for proper disposal or dispose according to local guidelines.

“`

LOW FLOW / HIGH FLOW
EN Instructions for Use . . . . . . 2 ES Instrucciones de uso . . . . 72 FR Instructions d’utilisation . 16 PT Instruções para utilização . . 86 DE Bedienungsanleitung . . . . 30 FI Käyttöohjeet . . . . . . . . . . 101 NL Gebruiksaanwijzingen . . . 44 ZH . . . . . . . . . . . . . . .. 115 IT Istruzioni per l’uso . . . . . . 58


R229M01 Rev. L

Maxtec 2305 South 1070 West Salt Lake City, Utah 84119 USA

phone: (800) 748.5355 fax: (801) 973.6090 email: sales@maxtec.com web: www.maxtec.com

Emergo Europe Westervoorsedijk 60, 6827 At Arnhem The Netherlands

Conforms to: AAMI STD ES60601-1, ISO STD 80601-2-55, IEC STDS 606011-6, 60601-1-8 Certified to: CSA STD C22.2 No. 60601-1

NOTE: The latest edition of this operating manual can be downloaded from our website at www.maxtec.com

Read this entire manual before attempting to operate or service the MaxBlend2. Attempting to operate the MaxBlend2 without fully understanding its features and functions may result in unsafe operating conditions.

CLASSIFICATION

Protection class ……………………………………………………………………………………………………… II, Type B

Protection against water………………………………………………………………………………………………… IPX1

Mode of operation ……………………………………………………………………………………………….Continuous

Safety of application in the presence of a flammable anesthetic mixture ………..See section 9.4

Power specification …………………………………………………….7.5V(MAX)

1.9W/250mA(MAX)

CAUTION: Federal law restricts this device to sale by, or on the order of, a medical professional.

Product Disposal Instructions: The sensor, batteries, and circuit board are not suitable for regular trash disposal. Return sensor to Maxtec for proper disposal or dispose according to local guidelines. Follow local guidelines for disposal of other components. There are no special considerations for the disposal of the product packaging.
WARRANTY
The MaxBlend2 is designed for air/oxygen delivery. Under normal operating conditions, Maxtec warrants the MaxBlend2 to be free from defects of workmanship or materials for a period of three years from the date of receipt from Maxtec, provided that the unit is properly operated and maintained in accordance with Maxtec’s operating instructions. Based on Maxtec’s product evaluation, Maxtec’s sole obligation under the foregoing warranty is limited to making repairs, or issuing credit, or replacement for equipment found to be defective. This warranty extends only to buyers purchasing new equipment directly from Maxtec, or through Maxtec’s designated distributors and agents. Maxtec warrants the MAX-550E oxygen sensor in the MaxBlend2 to be free from defects in material and workmanship for a period of two years from Maxtec’s date of shipment in a MaxBlend2 unit. Should a sensor fail prematurely, the replacement sensor is warranted for the remainder of the original sensor warranty period. Routine maintenance items, such as batteries, are excluded from warranty. Maxtec, and any other subsidiaries, shall not be liable to the purchaser or other persons for incidental or consequential damages to equipment that has been subject to abuse, misuse, misapplication, alteration, negligence, or accident. THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

NOTE: In order to obtain optimum performance from your MaxBlend2, all operation and maintenance must be performed in accordance with this manual. Please read the manual thoroughly before using the MaxBlend2, and do not attempt any repair or procedure that is not described herein. Maxtec cannot warranty any damage resulting from misuse, unauthor-ized repair, or improper maintenance of the instrument.

EMC Notice This equipment uses, generates, and can radiate radio frequency energy. If not installed and used in accordance with the instructions in this manual, electromagnetic interference may result. The equipment has been tested and found to comply with the limits set forth in IEC 60601-1-2 for medical products. These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments described in this manual.

MRI Notice

This equipment contains electronic and ferrous components, whose operation can be affected by intense electromagnetic fields. Do not operate the MaxBlend 2 in an MRI environment or in the vicinity of high-frequency surgical diathermy equipment, defibrillators, or shortwave therapy equipment. Electromagnetic interference could disrupt the operation of the MaxBlend 2.
WARNINGS
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
The following warnings apply any time you operate or service the MaxBlend2: ALWAYS confirm prescribed flow before administering to patient and monitor flow on a
frequent basis.
Always follow ANSI and CGA standards for medical gas products, flowmeters, and oxygen handling. DO NOT operate the MaxBlend2 unless qualified personnel are in attendance to promptly respond to alarms, inoperative conditions, or sudden malfunctions.
Patients on life-support equipment should be visually monitored at all times. DO NOT ignore audible alarms of the MaxBlend2. Alarms indicate conditions that require your immediate attention. DO NOT use parts, accessories, or options that have not been authorized for use with the MaxBlend2. Using unauthorized parts, accessories, or options may be harmful to the patient or damage the MaxBlend2.
Check all audible and visual alarms periodically to ensure they are operating properly. If an alarm fails to activate, contact your Maxtec Certified Service Technician. DO NOT operate the MaxBlend2 with the monitor powered off or without setting the adjustable alarms. All adjustable alarms must be set to ensure safe operation. DO NOT steam autoclave or otherwise subject the MaxBlend2 to temperatures above 122°F (50°C).
If either the air or oxygen source pressure is reduced or increased sufficiently to create a pressure differential of 20 PSI or more, an audible alarm will sound. This condition may significantly alter the FiO2 and flow output from the MaxBlend2. DO NOT tape, obstruct, or remove the alarm during clinical use. DO NOT occlude the sensor port on the side of the MaxBlend2.
An air inlet/water filter is recommended for use with the MaxBlend2. See section 6.1. DO
NOT use lubricants on the MaxBlend2.
DO NOT use the MaxBlend2 until correct performance has been verified. See section 3.0.
DO NOT use if damaged.
If a condition is detected that could possibly prevent the monitor from continuing to oper-ate safely, it will sound an alarm. If, at any time, EOx (i.e. EO2, EO4, etc.) appears on the LCD refer to section 4.0, or contact a Maxtec Certified Service Technician.
All service should be referred to a Maxtec Certified Service Technician.
Maxtec recommends that the MaxBlend 2 be serviced by Maxtec at a minimum of every three years, or if a leak or other performance problem is suspected.

ENGLISH

2

WWW.MAXTEC.COM · 800-748-5355

If the MaxBlend2 is dropped, follow the procedures outlined in section 3.0 for a performance check before reusing the device.
When the unit is going to be stored (not in use for more than 30 days), always remove the batteries to protect the unit from potential damage due to battery leakage.
Always replace batteries with high-quality, name-brand “AA” alkaline batteries.
The low oxygen alarm can be set below 18%, and as low as 15%. See section 2.3 on setting alarms. This is in accordance with IEC 80601-2-55.
In the event of exposure to an ELECTROMAGNETIC DISTURBANCE the analyzer may display an E06 or E02 error message. If this occurs, refer to Section 4.0 for instructions to resolve the problem.
To prevent risk of burns, fire, or injury to person(s): Mixed gases continuously bleed into the atmosphere at the oxygen concentration rate
set at the control knob. Bleeding oxygen into any closed area could increase the risk of fire or explosion. DO NOT operate this device in the presence of any flame or source of ignition, or when using devices such electrosurgical equipment or defibrillators. DO NOT operate the MaxBlend2 in the presence of flammable anesthetics or in an atmosphere of explosive gases. Operating the MaxBlend2 in flammable or explosive atmospheres may result in fire or explosion.
Galvanic O2 sensor electrolyte gel is acidic and may cause skin or eye irritation and/or burns. Take care when handling or replacing exhausted or damaged disposable O2 sensors. Be sure to dispose of expired sensors in accordance with hospital and/or governmental regulations (O2 Sensor SDS available upon request from Maxtec). DO NOT smoke in an area where oxygen is being used.
The MaxBlend 2 is not suitable for use with oxygen 93
CAUTIONS
Indicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury and property damage.
DO NOT store the MaxBlend2 in hot areas for prolonged periods of time. Temperatures above 80°F (27°C) can shorten battery life.
To minimize the potential for electrostatic shock, DO NOT use antistatic or electrically conductive hoses with the MaxBlend 2.
When cleaning the MaxBlend2: DO NOT use harsh abrasives.
DO NOT immerse the MaxBlend 2 in liquid sterilizing agents or liquids of any kind.
DO NOT spray cleaning solution directly onto the sensor port, bleed muffler, or buzzer opening.
DO NOT allow cleaning solution to pool on the front panel, sensor port, or bleed muffler.
DO NOT sterilize the MaxBlend 2. Standard sterilization techniques may damage the blender.
DO NOT clean or dry the MaxBlend 2 with a high pressure air gun. Applying high pressure air to the MaxBlend 2 may damage components and render the system inoperable.
DO NOT over-clean the MaxBlend 2. Repeated use of a cleaning agent can cause residue build-up on critical components. Excessive residue buildup can affect the MaxBlend 2’s performance.
Dropping or severely jarring the sensor after calibration may shift the calibration point enough to require recalibration.
Always operate the MaxBlend 2 with clean, dry medical-grade gases. Contaminants or moisture can cause defective operation. Oxygen should have a minimum dewpoint of -80°F (-62°C) or moisture content less than 7.9 PPM (0.0059mg/L). Oxygen “purity” should be at least 99.0% and the air used should be medical-grade. Water vapor content must not exceed a dew point of 5°F (-15°C) below the lowest ambient temperature to which the delivery system is exposed.

Particulate content must not exceed that which would be found immediately downstream of a 15 micron absolute filter. Refer to CGA commodity specifications G-4.3 and G7.1 for more information. Water vapor content of medical air or O2 supply to the blender must not exceed 5.63 x 103 milligrams H2O per cubic meter of non-condensable gas. DO NOT disassemble the MaxBlend2. All service should be performed by a Maxtec Certified Service Technician.
Be sure the MaxBlend2 is securely mounted. This device is usually mounted to a hospital rail system or an infusion stand. Dropping the device may cause injury or damage to the device.
The outlets of this device are capable of delivering pressure as high as the inlet pressure. Ensure that the devices that carry the gas from the blender to the patient prevent excessive pressure to the patient. DO NOT use the oxygen sensor if it appears to be damaged or is leaking. The oxygen sensors contain a weak acidic solution encapsulated in a plastic housing. Under normal operating conditions the solution (electrolyte) is never exposed.
NOTES
Indicates supplemental information to assist in the use of the device.
The MaxBlend 2 is tested for compliance with ISO 11195, and meets requirements regarding reverse gas flow as delivered.
Applicable parts used in the MaxBlend2 have been cleaned and degreased for oxygen service. Any lubricants used are designed specifically for the application.
As long as the absolute pressure of the gas mixture being monitored is constant, the MaxBlend 2 will accurately read oxygen concentrations. However, if the absolute pressure varies the reading will fluctuate proportionately, as the sensor actually measures the partial pressure of oxygen in the mixture. The sensor readings will also change propor-tionately with barometric pressure changes; because of this, frequent calibration of thes ensor is recommended.
Users are advised to use pressure regulators which display the outlet pressure.
All specifications assume the following standard environmental conditions, unless specified otherwise: ambient and sample gas temperatures of 77°F (25°C), barometric pressure of 30inHg (760mmHg), sea level altitude, ambient relative humidity of 50%, sample gas relative humidity of 0%.
It is important to note that the oxygen concentration selection scale is provided only as a guideline for selecting O2 concentrations. The clinician should use the MaxBlend 2’s display panel to adjust the O2 concentration to the desired setting.
The alarm limits can be set to levels that would render them useless for atparicular patient’s clinical condition. Ensure that the delivered oxygen level and flow rate are set to values prescribed by the patient’s physician. Also ensure that the high and low alarm limits are set to levels such that they will sound if the oxygen level is outside of safe limits. Be sure to review and, if necessary, re-set the alarm limits when the patient’s clinical con-dition changes, or when the patient’s physician prescribes a change in oxygen therapy.
This device does not contain automatic barometric pressure compensation.
Gas leaks that cause room air to mix with the gas sample may cause inaccurate oxygen readings. Ensure the O-rings on the sensor and flow diverter are in place and intact prior to use
NOTEIOUS incident(s) that occur in relation to the device should be reported to Maxtec
and the competent authority of the Member State in which the user and/or patient is established. Serious Incident(s) is defined as directly or indirectly led, might have led, or might lead to the death of a patient, user, or other person; the temporary or permanent serious deterioration of a patient’s, user’s or other person’s state of health; or a serious public health threat

WWW.MAXTEC.COM · 800-748-5355

3

ENGLISH

TABLE OF CONTENTS

CLASSIFICATION …………………………………………….2
WARRANTY …………………………………………………….2
WARNINGS …………………………………………………2
CAUTIONS ………………………………………………….3
NOTES ……………………………………………………………..3
1.0INTRODUCTION………………………………………… 5 1.1Indication for Use ……………………………………………………………………………………5 1.2MAX-550E Oxygen Sensor……………………………………………………………………….5 1.3Symbol Guide…………………………………………………………………………………………5 1.4Component Identification……………………………………………………………………….6 1.5Display Indicators …………………………………………………………………………………..6 1.6Back View……………………………………………………………………………………………… 7 1.7Requirements for Operating the Blender………………………………………………… 7
2.0OPERATING PROCEDURES ………………………7 2.1Setup and Installation……………………………………………………………………………. 7 2.1.1Battery Installation…………………………………………………………………….. 7 2.1.2MaxBlend 2 Setup……………………………………………………………………… 7 2.1.3Sensor Installation…………………………………………………………………….. 7 2.2Monitoring ……………………………………………………………………………………………. 7 2.3Alarm Setting Procedure………………………………………………………………………..8 2.3.1Low Alarm Setting …………………………………………………………………….8 2.3.2High Alarm Setting ……………………………………………………………………8 2.3.3Smart Alarm Mode…………………………………………………………………….8 2.4Basic Operation……………………………………………………………………………………..8 2.5Alarm Conditions and Priorities………………………………………………………………8 2.6Backlight Operation……………………………………………………………………………….9 2.7External Power Supply Operation …………………………………………………………..9 2.8Calibration Procedures …………………………………………………………………………..9 2.8.1Calibration to 100% Oxygen………………………………………………………..9 2.8.2Calibration to Room Air……………………………………………………………..9 2.9Flowmeter Operation …………………………………………………………………………….9

3.0PERFORMANCE CHECK ……………………………9 3.1Blender Safety Check…………………………………………………………………………… 10
4.0TROUBLESHOOTING ……………………………… 10
5.0CLEANING AND DISINFECTING………………. 11
6.0SERVICE AND MAINTENANCE………………… 11 6.1Maintenance…………………………………………………………………………………………. 11 6.2Replacing O2 Sensor…………………………………………………………………………….. 11 6.3Monitor Alarm Testing………………………………………………………………………….. 11
7.0ABBREVIATION GUIDE …………………………… 12
8.0SPECIFICATIONS ……………………………………. 12 8.1Instrument Specifications……………………………………………………………………..12 8.2Alarm Specifications…………………………………………………………………………….12 8.3O2 Sensor Specifications……………………………………………………………………….12
9.0FACTORS INFLUENCING CALIBRATION … 12 9.1Temperature Effect……………………………………………………………………………….12 9.2Pressure Effect …………………………………………………………………………………….12 9.3Humidity Effect ……………………………………………………………………………………12 9.4Exposure to Anesthetic Gases ………………………………………………………………13
10.0SPARE PARTS AND ACCESSORIES……….. 13
11.0THEORY OF OPERATION ………………………. 13 11.1Operational Diagram ……………………………………………………………………………13 11.2Mixing Operation…………………………………………………………………………………13 11.3Gas Outlet……………………………………………………………………………………………13 11.4Alarm/Bypass Function ……………………………………………………………………….14
12.0FLOW CHARACTERISTICS ……………………. 14
13.0ELECTROMAGNETIC COMPATABILITY …. 14

ENGLISH

4

WWW.MAXTEC.COM · 800-748-5355

1.0INTRODUCTION
The MaxBlend 2 is a compact air/oxygen gas mixing device which incorporates the use of a battery powered oxygen monitor. The gas mixing device (blender) provides precise mixing of medical-grade air and oxygen, while the monitor measures the selected oxygen concentrations from the blender’s gas flow and displays these measured concentrations on a digital display. The monitor provides high and low alarm limits which, when exceeded, cause an audible and visual alarm.
1.1Indication for Use
The MaxBlend 2 is designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life supporting device.

1.2MAX-550E Oxygen Sensor
The MAX-550E is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two electrodes (a cathode and an anode), a teflon membrane and an electrolyte. Oxygen diffuses through the teflon membrane and immediately reacts electrochemically at a gold cathode. Concurrently, oxidation occurs electrochemically at a lead anode, generating an electrical current and providing a voltage output. Electrodes are immersed in a unique gelled weak acid electrolyte, which is responsible for the sensor’s long life and motion insensitive characteristic. Since the sensor is specific to oxygen, the current generated is proportional to the amount of oxygen present in the sample gas. When no oxygen is present, there is no electrochemical reaction, and therefore, negligible current is produced. In this sense, the sensor is self-zeroing.
CAUTION: The MAX-550E oxygen sensor is a sealed device containing a mild acid electrolyte and lead (Pb). These materials are hazardous waste constituents and should be disposed of properly, or returned to Maxtec for proper disposal or recovery.
CAUTION: Dropping or severely jarring the sensor after calibration may shift the calibration point enough to require recalibration.

1.3Symbol Guide
The following symbols and safety labels are found on the MaxBlend 2 and/or labeling:

ON/OFF Key

Silent Key

Backlight Key

Unlock Key

Smart Alarm Key

CAL (Calibration Key)

Sleep Mode Indicator

High Alarm Indicator

Low Alarm Indicator

Calibration Reminder

Smart Alarm Mode Indicator

Alarm Silence Indicator

Below 18% Alarm Indicator

Low Battery Indicator

DOWN (Low Alarm Key)

Do Not
Corrosive Attention, consult accompanying documents Catalog Number Inlet

MR Unsafe Manufacturer Ingress Protection Rating Serial Number Outlet

Caution
Federal law (USA) restricts this device to sale by or on order of a physician.
Liters Per Minute
Authorized Representative in the European Community
Read Flow at Center of Ball

Do not throw away. Follow local guidelines for disposal

Meets ETL standards

50°C
-15°C (122°F) Storage Temperature Range
(5°F)

Date of Manufacture

Medical Device

CH

Authorized Representative UK

in Switzerland

POWER SUPPLY SYMBOL GUIDE The following symbols and safety labels are found on the MaxBlend 2 power supply (sold separately):

Two means of patient protection (double insulated)

Conforms to EU requirements. Combined UL / CSA Mark

For use in dry indoor locations.

UP (High Alarm Key) Warning Type B Applied Parts Oxygen Percent Oxygen Enriched Air Direct Current Use No Oil Responsible Person in the UK
Combined UL / CSA Mark

WWW.MAXTEC.COM · 800-748-5355

5

ENGLISH

12

p

i

3

u

o

4

y t r e
w

%O2

60

50

70

40

80

30 21

90 100

5 6

q 98 7

1.4Component Identification
1 LOW ALARM LED — In a low alarm condition, the “LOW ALARM” LED will flash twice a
second, accompanied by the audio buzzer.

2 HIGH ALARM LED — In a high alarm condition, the “HIGH ALARM” LED will flash twice a
second, accompanied by the audio buzzer.

3 ON/OFF KEY — This key is used to turn the device on or off. To turn the device OFF,
the button must be held while a rapid 3-2-1 countdown takes place to prevent accidental
power-off.

4 UNLOCK KEY — The unlock key is used to lock and unlock the instrument.

5 UP (ALARM HIGH) KEY —

The up key is used in setting the high FiO2 alarm limit.

The device must be in the unlocked state for the key to operate. See section 2.3.2 for

instructions on setting the high FiO2 alarm limit.

6 CALIBRATION KEY — This key is used to calibrate the device. The device must be in
the unlocked state for the key to operate. See section 2.8 for instructions on calibrating.

7 O2 INLET FITTING — An O2 fitting for connection to the O2 inlet hose from the gas source.

8 OXYGEN CONCENTRATION SELECTOR CONTROL — A knob which allows for selections of
mixed oxygen concentrations from 21% to 100%.

9 MEDICAL AIR INLET CONNECTOR — An air fitting for connection to an air inlet hose from
the air gas source.

q FLOWMETER OUTLET — A fitting for connection to the patient delivery tubing.

w BLEED TOGGLE SWITCH — The Bleed Toggle Switch turns on the auxiliary bleed. For the
blender to accurately mix the oxygen and air, the bleed must be turned on whenever
the total flow delivered to the patient is less than 15 LPM for high flow model, or less than
3 LPM for low flow model.

l

k

a

j

s

hf gd

e FLOW CONTROL KNOB — The flow control knob controls the flow of gas exiting the
flowmeter.

r OXYGEN FLOWMETER — Measures the flow of mixed gas exiting the flowmeter outlet.

t SMART ALARM KEY — The Smart Alarm key is used to help set the High-Low Alarm win-
dow quickly. See section 2.3.3 for instructions on using the Smart Alarm setting.

y DOWN (LOW ALARM) KEY —

The down key is used in setting the low FiO2 alarm

limit. The device must be in the unlocked state for the key to operate. See section 2.3.1

for instructions on setting the low FiO2 alarm limit.

u BACKLIGHT KEY — The backlight key will manually activate the backlight for 30 seconds.
See section 2.6 for more information on backlighting operation.

i SILENT KEY — In an alarm condition, pressing the SILENT key will deactivate the
audio alarm for 2 minutes.

o LCD DISPLAY — The digit liquid crystal display (LCD) provides direct readout of oxygen
concentrations. The digits also display error codes, alarm set modes, and calibration
codes as necessary.

1.5Display Indicators
p OXYGEN CONCENTRATION — Current oxygen concentration percentage from the oxygen
sensor.

a HIGH ALARM INDICATOR — The high alarm setting is displayed at all times just
below the “HIGH” icon on the LCD readout. The indicated value represents the oxygen
percentage at which the high alarm will be activated.

s OXYGEN HIGH ALARM LIMIT — High oxygen alarm setpoint. Audible and visual alarms
will trigger when this limit is exceeded.

d CALIBRATION REMINDER — The calibration reminder symbol is located at the bot-
tom of the display. This symbol will display after one week has elapsed from the previ-
ous calibration.

f SLEEP MODE INDICATOR — The sleep mode indicator is used to help with battery
consumption.

g LOW BATTERY INDICATOR —

The Low Battery Indicator is located at the middle

of the display and is only activated when the voltage on the batteries is below a normal

operating level.

h ALARM SILENCE/SMART ALARM INDICATOR — When the Silent key is pressed the indica-
tor will display with cross bars to alert the condition. When Smart Alarm key is
pressed the indicator will display with T-bars to alert the condition.

ENGLISH

6

WWW.MAXTEC.COM · 800-748-5355

v

b

W n

Q

m

/

,

.
j OXYGEN LOW ALARM LIMIT — Low oxygen alarm setpoint. Audible and visual alarms will
trigger when this limit is exceeded.
k LOW ALARM INDICATOR — The low alarm setting is displayed at all times just below
the “LOW” icon on the LCD readout. The indicated value represents the oxygen percentage at which the low alarm will be activated.
l <18% ALARM INDICATOR — The <18% alarm indicator is located above the Low Alarm
Indicator digits. When the low alarm setting is set below <18%, the indicator will flash each second to alert the operator of this special condition. See section 2.3.1 for setting this low alarm condition.

1.7Requirements for Operating the Blender
All operator-detachable inlet pressure hoses supplied with the gas mixer comply with ASTM/ ISO 5359.
PRESSURIZED OXYGEN: The compressed oxygen source must provide clean, dry, medicalgrade oxygen at the pressure specified in Section 8.0.
PRESSURIZED AIR: The compressed air source must provide clean, dry, medical-grade air at the pressure specified in Section 8.0.
2.0OPERATING PROCEDURES
2.1Setup and Installation
2.1.1Battery Installation
All MaxBlend 2 units are powered by four “AA” alkaline batteries (4 x 1.5 Volts) and are shipped without the batteries installed. Only name-brand batteries should be used. Batteries should be replaced by trained service personnel.
To install the batteries: Open the battery drawer by squeezing inward on both tabs as shown in the figure below. If you have difficulty squeezing the tabs in with your fingers, use two flat screwdrivers or two coins. Remove the battery drawer completely from the MaxBlend 2. Install four new, “AA” alkaline batteries into the unit, observing the orientation shown on the plastic inside the drawer. Slide the drawer back in with the batteries facing upward. Press in on the drawer until both tabs latch into place.
When batteries are replaced, the unit will start up and perform an automatic calibration. Ensure calibration is performed as directed in Section 2.8.
WARNING: Battery replacement by inadequately trained personnel could result in a safety hazard.
WARNING: Electrical shock or damage to the equipment may occur if an inappropriate external power supply is used. Maxtec recommends using only the Maxtec MaxBlend 2 External Power Supply–R230P10.

1.6Back View
v BATTERY COMPARTMENT — The MaxBlend 2 is powered by four “AA” alkaline batteries.

b SENSOR CABLE — The cable connects the MaxBlend 2 to the MAX-550E Sensor.
n SENSOR WITH DIVERTER — The sensor with flow diverter is designed to fit into a port
behind the flowmeter.
m O2 SENSOR PORT — A sampling port for the oxygen sensor. It allows mixed gas from the
blender to flow over the sensor membrane.
, PRESSURE DIFFERENTIAL REED ALARM — An audible alarm which, when activated,
indicates that an unacceptable pressure differential exists between the two gas source pressures.
. AUXILIARY MIXED GAS OUTLET — The outlet can be used as power take off.
/ AUXILIARY MIXED GAS OUTLET — The outlet can be used to add additional flowmeters.
Q MOUNT ADAPTER — An adapter which allows the MaxBlend 2 to be mounted onto a
bracket for attaching to a rail or ventilator system.
W EXTERNAL POWER SUPPLY PORT — The port provides connection for the external power
adapter. See section 2.7 for more information on the power supply.

2.1.2MaxBlend 2 Setup
1. Connect the pressurized air source to the Air Inlet Connector. 2. Connect the pressurized oxygen source to the O2 Inlet Fitting. 3. Flush gas at the highest possible flow rate through the blender for at least one minute
to eliminate any particulate that may have been introduced into the system during handling and installation.
2.1.3Sensor Installation
1. Attach the flow diverter onto the oxygen sensor. 2. Place the sensor into the sensor port located behind the flowmeter. 3. Attach the sensor cable directly to the sensor and the sensor jack on the back of the
monitor enclosure. Ensure the cable is fully inserted into both connections. 4. Calibrate the sensor prior to use according to the calibration procedures in section 2.8
2.2Monitoring
Before use on a patient, the oxygen concentration of the delivered gas should be checked at the setting intended for use.
1. To begin monitoring, press the ON/OFF key located on the front panel. Monitoring will begin immediately.

WWW.MAXTEC.COM · 800-748-5355

7

ENGLISH

2. Should oxygen level exceed either the HIGH or LOW alarm set points, the alarm indicator on the front panel will illuminate indicating either a high or low oxygen condition. The alarm will cease when oxygen levels are within limits, or the limits are adjusted.
NOTE: The MaxBlend 2 will monitor the oxygen concentration of the gas delivered from the blender regardless of which outlet port is in use. Therefore, the integrated flowmeter has ultimately no effect on the oxygen sensing.

2.3Alarm Setting Procedure

2.3.1Low Alarm Setting

To adjust the low alarm setting:

1. Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL, and

HIGH icons will begin to flash indicating the SET OPERATING MODE.

2. Press the DOWN (Low Alarm) key

on the keypad.

NOTE: The Low Alarm digits begin to flash indicating the Low Alarm manual setting.

3. Use the UP

and DOWN

keys to set the low alarm to the desired value.

Pressing the arrow keys changes the value in 1% increments. If the keys are held down

for more than 1 second, the display will scroll at a rate of 1% per second.

NOTE: If 30 seconds elapse between key actuations, the system will store the latest low alarm setting and will revert to normal operation. If this occurs inadvertently, simply repeat the alarm setting procedure.

There is a special condition that allows the low oxygen alarm to be set below 18%. To access this condition, press the DOWN arrow key for three seconds while the low alarm reading displays 18%. The alarm setting can now be adjusted to 17, 16, or 15%. A bar will blink above the setting to provide further indication that the alarm has been set to this special <18% condition.

The low alarm value cannot be set lower than 15%, nor can it be set closer than 1% from the high alarm value. For example, if the high alarm is set at 25%, the system will not accept a low alarm setting greater than 24%.

When the low alarm value is set, press the Unlock key again to accept the low alarm setting and return to normal operation.

NOTE: The default low alarm setting is 18% O2. Removing the batteries, or shutting the unit OFF will reset the low alarm limit to 18%.

2.3.2High Alarm Setting
To adjust the high alarm setting: 1. Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL and HIGH icons will begin to flash indicating the SET OPERATING MODE. 2. Press the UP (High Alarm) key on the keypad.
NOTE: The High Alarm digits begin to flash indicating the High Alarm manual setting.
3. Use the UP and DOWN keys to set the high alarm to the desired value. Pressing the arrow keys changes the value in 1% increments. If the keys are held down for more than 1 second the display will scroll at a rate of 1% per second.
NOTE: If 30 seconds elapse between key actuations, the system will store the latest high alarm setting and will revert to normal operation. If this occurs inadvertently, simply repeat the alarm setting procedure.
4. When the high alarm setting is set above 100%, the high alarm will indicate two dashes ­ ­. This special condition turns off, or deactivates, the high alarm.
5. When the high alarm value is set, press the Unlock again to accept the high alarm setting and return to normal operation.
NOTE: The default high alarm setting is 50% O2. Removing the batteries will reset the high alarm limit to 50%.

2.3.3Smart Alarm Mode

1. Press the Unlock Key to unlock the keypad. Note the LOW, SMART ALARM, CAL, and

HIGH icons will begin to flash indicating the SET OPERATING MODE.

2. Press the Smart Alarm key on the keypad. Note the LOW digits, Alarm Mode and

HIGH digits begin a slow flash indicating SMART ALARM MODE. The high alarm will now

be set equal to the current oxygen reading +3% (rounded to the nearest integer). The

low alarm will now be set equal to the current oxygen reading -3% (rounded to the

nearest integer, but never lower than 18%.

3. Pressing of the UP key

will add 1% to the high alarm setting and subtract 1%

from the low alarm setting. Pressing the DOWN key

will subtract 1% from the

high alarm setting and add 1% to the low alarm setting. In other words, the UP key

widens the alarm band and the DOWN key tightens the alarm band. This feature will

not set the alarm levels above 100% or below 18%.

4. Once the desired alarm settings are attained, press the Unlock key again to save

the settings and return to normal operation mode. If 30 seconds elapse without a

key press by the user, the device will automatically save the latest alarm settings and

return to normal operation mode.

2.4Basic Operation
To check the oxygen concentration of the delivered gas: 1. Ensure the sensor is connected to the flow diverter and inserted completely in the sensor port behind the flowmeter. 2. Set the oxygen blender control knob to the desired oxygen setting. 3. Using the ON/OFF key , make sure the unit is in the power on mode. 4. Allow the oxygen reading to stabilize. 5. Adjust the flowmeter to the desired flow rate. Read the flow rate at the center of the float ball.

2.5Alarm Conditions and Priorities
In the event of either a low alarm or high alarm condition, the corresponding LED will begin to flash, accompanied by the audio buzzer. Pressing the Silent key will deactivate the buzzer, but the LED and the alarm value digits on the display will continue to flash until the alarm condition has been rectified. If the alarm condition still exists 120 seconds after silencing the audio buzzer, the beeper will start to sound again.

ALARM
Line Power Plugged In
Line Power Unplugged
External DC Power Supply Voltage Out
of Range
Battery Voltage too low for device to operate (E04)
Oxygen/ pressure level above the high alarm setting
Oxygen/pressure level below the low alarm setting
Oxygen level below the low oxygen alarm setting and lower than 18%

ALARM PRIORITY Informational Informational Informational
Medium
Medium
Medium
High

LOW ALARM LED
Off Single Yellow Pulse
Solid Yellow
Pulsing Yellow
Off
Pulsing Yellow
Pulsing Red

HIGH ALARM LED
Off Single Yellow Pulse
Solid Yellow
Pulsing Yellow
Pulsing Yellow
Off
Off

AUDIBLE ALARM 2 Pulses 2 Pulses 2 Pulses
3 Pulses
3 Pulses
3 Pulses
5+5 Pulses

AUDIBLE ALARM REPEAT
No Repeat No Repeat Every 15 Sec.
Every 25 Sec.
Every 25 Sec.
Every 25 Sec.
Every 15 Sec.

ENGLISH

8

WWW.MAXTEC.COM · 800-748-5355

· A low alarm condition will remain until the actual concentration is 0.1% higher than the low alarm setting.
· A high alarm condition will remain until the actual concentration is 0.1% lower than the high alarm setting.
To differentiate the level of severity, the monitor provides three unique sound bursts.
2.6Backlight Operation
To turn on the backlighting: 1. When the unit is on, pressing the Backlight button will turn the backlighting on for 30 seconds. An additional press will turn off the backlighting. 2. If the device is being used in a dark location, any button press will activate the backlight.
CAUTION: Excessive use of the backlight can reduce the life of the batteries.
2.7External Power Supply Operation
To extend the life of the batteries, an external Maxtec approved 7.5V DC external power supply can be purchased. Once connected to the unit, total power is supplied by the power supply. Batteries are still required in the unit and will provide emergency power in the event main AC power is lost.
NOTE: Use only the Maxtec external power supply described in section 10.0. NOTE: The power supply is not a battery charger.
DO NOT use rechargeable batteries.
2.8Calibration Procedures
2.8.1Calibration to 100% Oxygen
The MaxBlend 2 should be calibrated before being placed into clinical use. Thereafter, Maxtec recommends calibration of the unit on a weekly basis. Frequent calibration will have no adverse effect on the performance of the MaxBlend 2. Calibration should also be performed upon replacement of a sensor.
The sensor is best calibrated while mounted in the MaxBlend 2 sensor port. As in normal operation, the oxygen sensor responds best when installed in a vertical position with the sensor facing down.
Changes in barometric pressure can affect the oxygen reading. A 1% change in the barometric pressure results in an error of 1% of the actual reading. (Example: If you are reading a 50% oxygen mix and the barometric pressure drops from 1000mbar to 990mbar the reading will drop to 50% x (990/1000) = 49.5%). Maxtec recommends that you recalibrate after changing point-of-use elevation by more than 500 feet (150m).
It is best to calibrate the MaxBlend 2 using the sensor port with a technical grade oxygen standard (99.0% or better). Calibration of the unit with room air is less accurate over the full FiO2 operating range.
To calibrate to 100% oxygen: 1. Connect the oxygen supply line (pressure differential alarm may sound). Verify the sensor is plugged into the O2 sensor port and connected to the sensor cable. DO NOT connect air supply line at this time. 2. Using the ON/OFF key, make sure the MaxBlend 2 is in the power on mode. 3. Rotate the FiO2 control knob to the 100% stop. Allow a few minutes for the reading to stabilize. 4. Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL, and HIGH icons will begin to flash indicating the SET OPERATING MODE. 5. Press the CAL (Calibration) key on the keypad. The word “CAL” will appear on the display for approximately 5 seconds and then finish with 100.0%. 6. The unit is now calibrated and in the normal operating mode.

2.8.2Calibration to Room Air
The MaxBlend 2 can quickly be calibrated to room air (20.9%).
To calibrate to room air: 1. Connect the air supply line (pressure differential alarm may sound). Verify the sensor is plugged into the O2 sensor port and connected to the sensor cable DO NOT connect oxygen supply line at this time. (If preferred, room-air calibration may be performed by removing the sensor from the O2 sampling port and detaching the flow diverter. If using this method, the gas-supply lines may remain attached.) 2. Using the ON/OFF key, make sure the MaxBlend 2 is in the power on mode. 3. Rotate the FiO2 control knob to the 21% stop. Allow a few minutes for the reading to stabilize. 4. Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL and HIGH icons will begin to flash indicating the SET OPERATING MODE. 5. Press the CAL (Calibration) key on the keypad. The word “CAL” will appear on the display for approximately 5 seconds and then finish with 20.9%. 6. The unit is now calibrated and in the normal operating mode.
2.9Flowmeter Operation
1. Adjust the flowmeter to the desired setpoint as read by the center of the float ball. · To increase flow–turn knob counter-clockwise · To decrease flow–turn knob clockwise
2. Check the toggle bleed switch to ensure it is in the appropriate position. · For a low flow blender the bleed toggle switch should be turned on (up position) if the total combined flow from the blender is less than 3LPM. · For a high flow blender the bleed toggle switch should be turned on (up position) if the total combined flow from the blender is less than 15LPM. · At delivered flows greater than the above limits, the bleed toggle switch can be turned off (down position) to conserve gas.
CAUTION: Failure to activate the bleed as described above may result in inaccurate oxygen concentrations from the blender. However, the MaxBlend 2 will always display the actual delivered concentration.
CAUTION: The outlets of this device are capable of delivering pressure as high as the inlet pressure. Ensure that the devices that carry the gas from the blender to the patient prevent excessive pressure to the patient.
3.0PERFORMANCE CHECK
Prior to placing the MaxBlend 2 into clinical use, perform the following tests.
WARNING: If the MaxBlend 2 does not function as described on the next page, contact your Maxtec Distributor or contact a Maxtec Certified Service Technician at:
Maxtec 2305 South 1070 West Salt Lake City, UT 84119 385-549-8000 or 800-748-5355
DO NOT use the MaxBlend 2 until correct performance has been verified.

WWW.MAXTEC.COM · 800-748-5355

9

ENGLISH

3.1Blender Safety Check
NOTE: Before proceeding, ensure that the high alarm set point control is OFF [display reads (–)], and that the low alarm set point control is below 20%.

PROCEDURE

BLENDER RESPONSE

1. Connect the blender to 50 ±5 PSIG air/oxygen gas sources. Adjust the blender control knob to 60%. Turn the bleed toggle switch on (up position) and adjust the flowmeter to a minimum of 2 LPM.

No response. Monitor display reads 60% ±3%.

2. Disconnect the 50 PSIG AIR source from the MaxBlend 2. NOTE: The blender must be flowing gas for the alarm to activate.

Audible alarm sounds. Monitor display reads 100% ±3%.

3. Reconnect 50 PSIG AIR source to the MaxBlend 2.

Audible alarm stops. Verify the monitor display reads 60% ±3%.

4. Disconnect 50 PSIG OXYGEN source from the MaxBlend 2.

Audible alarm sounds. Monitor display reads 20.9% ±3%.

5. Reconnect 50 PSIG OXYGEN to the MaxBlend 2.

Audible alarm stops. Verify the monitor display reads 60% ±3%.

6. Adjust both air and oxygen inlet regulators to 0 PSIG. No response.

7. Remove air inlet hose at regulator and insert end into beaker of water.

No response.

8. Slowly raise pressure of oxygen regulator to 50 PSIG and back to 0 PSIG while observing air hose end in beaker.

No bubbles should be observed. Audible alarm sounds.

9. Dry and reattach air inlet hose to regulator.

No response.

10. Remove oxygen inlet hose at regulator, and insert end into beaker of water.

No response.

11. Slowly raise pressure of air regulator to 50 PSIG and back to 0 PSIG while observing oxygen hose end in beaker.

No bubbles should be observed. Audible alarm sounds.

12. Dry and reattach oxygen inlet hose to regulator.

No response.

PROBLEM: Selected oxygen concentration accurate only when gas pressures are equal. POTENTIAL CAUSES AND SOLUTIONS:
· MaxBlend 2 balance module not functioning properly. Contact Maxtec for repair.
PROBLEM: Blank display. POTENTIAL CAUSES AND SOLUTIONS:
· Batteries not installed. Install batteries. Refer to section 2.1.1, Battery Installation. · Battery completely dead. Replace batteries. Refer to section 2.1.1, Battery Installation. · Monitor defective. Contact Maxtec for repair.
PROBLEM: Partial or distorted display. POTENTIAL CAUSES AND SOLUTIONS:
· Monitor damaged. Contact Maxtec for repair.
PROBLEM: Sensor will not calibrate. POTENTIAL CAUSES AND SOLUTIONS:
· Sensor cell exhausted. Replace sensor. Refer to section 6, Replacing O2 sensor. · Sensor cable defective. Return to Maxtec. · Monitor defective. Contact Maxtec for repair.
PROBLEM: Sensor will calibrate, but takes too long or does not return to 21% ±2% oxygen in air (2 to 5 minutes) when performing calibration. POTENTIAL CAUSES AND SOLUTIONS:
· Disposable oxygen sensor damaged or defective. Replace sensor. Refer to section 6.2, Replacing O2 sensor.
PROBLEM: Sensor will calibrate, but reading at any constant level drifts more than ±3% over a 24 hour period. POTENTIAL CAUSES AND SOLUTIONS:
· Barometric pressure change since last calibration. Recalibrate. Refer to section 2.8, Calibration Procedures.
· Room or gas temperature went below 59°F (15°C) or above 104°F (40°C). Correct temperature and recalibrate.
PROBLEM: Low battery icon. POTENTIAL CAUSES AND SOLUTIONS:
· If, at any time, the low battery icon is displayed on the LCD readout, the batteries should be replaced as quickly as possible.

4.0TROUBLESHOOTING
PROBLEM: Oxygen concentration discrepancy between oxygen concentration selection knob and actual reading on display, greater than 3%. POTENTIAL CAUSES AND SOLUTIONS:
· Bleed is turned off. Turn bleed toggle switch on. Refer to section 2.9, Flowmeter Operation.
· Monitor out of calibration. Calibrate. Refer to section 2.8, Calibration Procedure. · Sensor exhausted. Replace sensor. Refer to section 6.2. · Gas supply contaminated. Contact Maxtec for repair of the MaxBlend 2. · Blender out of calibration. Contact Maxtec for repair.
PROBLEM: Pressure differential alarm sounding. POTENTIAL CAUSES AND SOLUTIONS:
· Inlet pressure differences of 20 PSI or more. Correct pressure difference. · Pressure alarm not calibrated properly. Contact Maxtec for repair. · MaxBlend 2 blender operation out of calibration. Contact Maxtec for repair.
PROBLEM: Inlet pressure has supply loss, no audible pressure differential alarm. POTENTIAL CAUSES AND SOLUTIONS:
· Reed alarm cap damaged or defective. Contact Maxtec for repair.

PROBLEM: E01: Sensor voltage is too low to perform a valid calibration. POTENTIAL CAUSES AND SOLUTIONS:
· Manually attempt a new calibration. Refer to section 2.8, Calibration Procedures. · If unit repeats this error more than three times, contact Maxtec’s Customer Service
Department for possible sensor replacement.
PROBLEM: E02: No sensor attached. POTENTIAL CAUSES AND SOLUTIONS:
· Disconnect the sensor and reconnect, making sure the male plug is fully inserted into the receptacle. The analyzer should now perform a new calibration with the error cleared.
· If the error still persists, remove the batteries, wait 30 seconds, then reinstall, to perform a factory reset and diagnostic on the analyzer. The analyzer should again perform a new calibration with the error cleared.
· Contact Maxtec Customer Service Department if the error code cannot be cleared.
PROBLEM: E03: No valid calibration data available. POTENTIAL CAUSES AND SOLUTIONS:
· Make sure unit has reached thermal equilibrium and perform a calibration routine. Refer to section 2.8, Calibration Procedures.
PROBLEM: E04: Battery below minimum operating voltage. POTENTIAL CAUSES AND SOLUTIONS:

ENGLISH

10

WWW.MAXTEC.COM · 800-748-5355

· Replace batteries. A medium priority alarm will sound every 25 seconds until the batteries are replaced or become too dead to sound the alarm. Refer to section 2.1.1, Battery Installation.

PROBLEM: E05: Sensor voltage is too high to perform a valid calibration. POTENTIAL CAUSES AND SOLUTIONS:
· Manually attempt a new calibration. Refer to section 2.8, Calibration Procedures. · If unit repeats this error more than three times, contact Maxtec’s Customer Service
Department for possible sensor replacement.

PROBLEM: E06: Non-compatible oxygen sensor. POTENTIAL CAUSES AND SOLUTIONS:
· Disconnect the sensor and reconnect, making sure the male plug is fully inserted into the receptacle. The analyzer should now perform a new calibration with the error cleared.
· If the error still persists, remove the batteries, wait 30 seconds, then reinstall, to perform a factory reset and diagnostic on the analyzer. The analyzer should again perform a new calibration with the error cleared.
· Contact Maxtec Customer Service Department if the error code cannot be cleared.
PROBLEM: E07: Sensor signal is not stable enough to perform a valid calibration. POTENTIAL CAUSES AND SOLUTIONS:
· Wait for displayed oxygen reading to stabilize when calibrating the device at 100% oxygen.
· Wait for unit to reach thermal equilibrium. Please note that this can take up to one half hour, if the device is stored in temperatures outside the specified operating temperature range.
PROBLEM: E08: Battery voltage is too low to perform a valid calibration. POTENTIAL CAUSES AND SOLUTIONS:
· Replace batteries. Refer to section 2.1.1, Battery Installation.
NOTE: Use only the Maxtec-approved Max-550E sensor described in section 10.0, Spare Parts and Accessories. The Max550E sensor is equipped with an authentication chip to ensure the monitor is used with an approved sensor. NOTE: The operator must be facing the device and positioned within 4 meters to distinguish the visual alarm indicators. Audible alarms can be distinguished as long as the operator is in the same room, and the ambient noise level is typical for a clinical setting.
5.0CLEANING AND DISINFECTING
The external surfaces of the device, and its accessories, can be cleaned and disinfected using the process detailed below. Under normal use conditions, the sensing surfaces of the sensor should not become contaminated. If you suspect that the sensing face of the sensor or internal surfaces of the flow diverter have become contaminated, these items should be discarded and replaced. Store the device in a clean, dry location when not in use.
To clean and disinfect: 1. Ensure battery drawer is closed and sensor/diverter are inserted into their port. 2. Using disposable Super Sani-Cloth germicidal wipes (medical-grade 2-in-1 cleaning/ disinfecting wipes), remove all visible contamination from the external surfaces of the device and its accessories. Be sure to closely inspect and remove contamination from seams and recesses on the device that may trap contaminants. Wipe with clean paper towel to remove debris and bioburden. 3. After all visible contamination is removed, use a second germicidal wipe to thoroughly wet the surfaces of the device and accessories. Allow to remain wet for 4 minutes. Use additional wipes, if needed, to assure surfaces are wetted continuously for 4 minutes. 4. Allow device to air dry completely. 5. Visually inspect the device for visible contamination. Repeat cleaning/disinfection process if visible soil remains. DO NOT allow the liquid or spray to penetrate the device.
6. Repeat sections 2.2 through 2.4 prior to cleaning.

DO NOT spray cleaning solution directly onto the sensor port, bleed muffler or buzzer openings.
Be sure to thoroughly clean and disinfect the areas depicted in the images below. These regions are contacted during normal use and may contribute to cross-contamination if not disinfected properly.
CAUTION: Excessive rubbing of labels may cause them to become illegible.
DO NOT immerse the device or sensor into liquid decontamination agents.
DO NOT use strong solvent cleaners.
DO NOT allow cleaning liquids to contact the face of the sensor as this may impair the readings of the sensor.
DO NOT Attempt to sterilize the device with steam, ethylene oxide or irradiation.
6.0SERVICE AND MAINTENANCE
6.1Maintenance
Maxtec recommends that the Performance Check listed in section 3.0 is performed at least once a year.
When using the MaxBlend 2 with a medical-grade compressed air source, it is recommended that an air inlet watertrap/filter be attached to the air inlet of the MaxBlend 2 prior to use. Contaminants from hospital air lines may compromise the function of the MaxBlend 2.
Maxtec recommends that the MaxBlend 2 be overhauled and serviced at a minimum of every three years.
Repair of this equipment must be performed by a Maxtec Certified Service Technician experienced in the repair of this device.
6.2Replacing O2 Sensor
The oxygen sensor is designed to operate for two years under normal use conditions. The oxygen sensor should be replaced whenever any of the problems listed in section 4.0, Troubleshooting dictate the need to do so.
To replace the O2 sensor: 1. Remove the sensor from the sensor monitor port. 2. Remove the sensor from the sensor cable. 3. Install a new O2 sensor with flow diverter and attach to the sensor cable. 4. Calibrate the sensor following the instructions for calibration listed in section 2.8.
6.3Monitor Alarm Testing
· Testing of alarms should be performed on a yearly basis. · To check the low alarm, adjust the low alarm setting to 23%, or higher, and expose the
sensor to room air (20.9%). The low alarm LED should flash with the alarm sound. · To check the high alarm, adjust the low alarm setting to 17%, or lower, and the high
alarm setting to 18% and expose the sensor to room air (20.9%). The high alarm LED should flash with the alarm sound. If one, or both alarms malfunction, contact a Maxtec Certified Service Technician.

WWW.MAXTEC.COM · 800-748-5355

11

ENGLISH

7.0ABBREVIATION GUIDE

TERM

DESCRIPTION

Air/O2

Mixture of compressed air and oxygen

°C

Degrees Celsius

CGA

Compressed Gas Association

DISS

Diameter Indexed Safety System

°F

Degrees Fahrenheit

FiO2

Fractional Concentration of Inspired Oxygen

O2

Oxygen

LPM

Liters Per Minute

PSIG

Pounds Per Square Inch Gauge

8.0SPECIFICATIONS
8.1Instrument Specifications
Weight (unpackaged)…………………………………………………………………………………… 5.3 lbs. (2.4 kg.) Power Source……………………………………………………………..Four “AA” alkaline batteries, 1.5 V each Battery Life………………………………………………….5000 hours (continuous operation, no alarming) Oxygen Measurement Range……………………………………………………………………0% to 100% oxygen Display Resolution………………………………………………………………………………………………0.1% oxygen O2 Concentration Adjustment Range……………………………………………………………… 21% to 100% O2 Gas Supply Pressure…………………..The gas supplies must provide clean, dry, medical-grade air
and oxygen at a pressure of 30 to 75 PSIG (2.0 to 5.2 BAR). Air and oxygen must be within 20 PSI (1.3 BAR). Optimal performance is achieved at 50 PSIG inlet pressures.
Pressure Drop ………………………………………………..Less than 6 PSIG (0.4 BAR) @ 50 PSIG (3.4 BAR) supply pressures and 10 LPM flow rate
Sensor Bleed Flow………………………………………………………………………. 0.1LPM @ 50 PSIG (3.4 BAR) Bleed Flow (toggle ON) ……………..3 LPM for low flow blender and 13 LPM for high flow blender Outlet Flow Range………………………. 0-30 LPM for low flow blender and 2-100 LPM for high flow blender with inlet pressures at 50 PSIG (3.4 BAR) Mixed Gas Stability* Ambient Operating Conditions…………………………………………….. ±1% oxygen Operating Temperature Range……………………………………………………..59°F to 104°F (15°C to 40°C) Relative Humidity Range…………………………………………………………………..0-95%, non-condensing Ambient Storage Conditions Temperature Range………………………….. 5°F to 122°F (-15°C to 50°C) Flowmeter accuracy** ………………….. +/-10% of indicated value or 0.5 LPM whichever is greater,
with inlet pressure set to 50PSIG. Flush Flow***……………………………………………………………………………. 3 LPM Flowmeter: 20-30 LPM
15 LPM Flowmeter: 20-30 LPM 30 LPM Flowmeter: 45 LPM 70 LPM Flowmeter: 80 LPM
*The delivered oxygen concentration will remain constant within ±1% of the set point value with constant inlet pressures. The displayed value may vary more than this due to sensor accuracy, age, environmental conditions, and the length of time since last sensor calibration.
**Position the device such that the flowmeters are vertical to ensure accuracy.
***Any flow beyond the last calibrated line on the Flow Tube with unrestricted flow is Flood/Flush flow. The maximum flow (flood/flush flow) is indicated above. The above flush flows are based on 50 psi pressure, 70 F (21 C), at standard atmospheric pressure. The Specifications are subject to change without prior notice.

8.3O2 Sensor Specifications
Total accuracy* ……………………. ±3% actual oxygen level over full operating temperature range Oxygen Measurement Accuracy ………………………………………………………………………….. ±1% oxygen Linearity …………………………………………………………….. ± 1% at constant temperature and pressure Error Over Operating Temp Range ……………………………………………………. ±3% oxygen, maximum Response Time to 90% of Final Reading*…………………………………….. @ 77°F (25°C) 20 seconds Storage Temperature Range…………………………………………………………. 5°F to 122°F (-15°C to 50°C) Expected Useful Life………………………………..1,500,000 O2% hours (approx. 2 years average use) *The accuracy of the oxygen monitor is not affected by supply gas inlet pressure to the blender, however, pressures below 50 PSIG may result in a longer response time.
NOTE: All specifications assume the following standard environmental conditions, unless specified otherwise:
· Ambient and sample gas temperatures of 77°F (25°C) · Barometric pressure of 30 inHg (102 kPa) · Ambient relative humidity of 50% · Sample gas relative humidity of 0%
9.0FACTORS INFLUENCING CALIBRATION
9.1Temperature Effect
The MaxBlend 2 Monitor will hold calibration and read correctly within +/-3% when in thermal equilibrium within the operating temperature range.
The device accuracy will be better than +/-3% if operated at the same temperature at which it was calibrated. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before reading is accurate. For these reasons, the following is recommended:
1. Allow adequate time for the sensor to equilibrate to a new ambient temperature. Please note that this can take up to one half hour, if the device is stored in temperatures outside the specified operating temperature range.
2. For best results, perform the calibration procedure at a temperature close to the temperature where analysis will occur.
9.2Pressure Effect
Changes in barometric pressure can affect the oxygen reading. A 1% change in the barometric pressure results in an error of 1% of actual reading (Example: If you are reading a 50% oxygen mix and the barometric pressure drops from 30kPa to 29kPa, the reading will drop to: 50% x (29/30) = 48.3%. Maxtec recommends that you recalibrate after changing point-of-use elevation by more than 500 feet (150m).
9.3Humidity Effect
Humidity in the sample gas will affect the oxygen reading. Maxtec recommends that the gas delivered to the MaxBlend 2 be medical-grade, clean, and dry. Refer to ISO 7396-1 for further details.

8.2Alarm Specifications
Pressure Differential Alarm Activation ……………………….When supply pressures differ by 20 PSI (1.3 BAR) or more alarm is activated
Low Oxygen Alarm Range…………………………………………….15%-99% (>1% lower than high alarm) High Oxygen Alarm Range……………………. 16%-100% (>1% higher than low alarm) (according to
IEC 60601-1-8 Audible Alarms in Medical Equipment) Alarm Volume (all priorities)………………………………………………………72 dB(A) ± 7 dB(A) at 1 meter

ENGLISH

12

WWW.MAXTEC.COM · 800-748-5355

9.4Exposure to Anesthetic Gases
Because of the unique chemistry of the oxygen sensors provided with the MaxBlend 2, there are no significant effects when exposed to commonly used anesthetic gases; however, the monitor is not designed for exposure to flammable gas mixtures (See WARNING page 2).

INTERFERENT

VOLUME % DRY

INTERFERENCE IN O2%

Nitrous Oxide

60%, balance O2

<1.5%

Halothane

4%

<1.5%

Enflurane

5%

<1.5%

Isoflurane

5%

<1.5%

Helium

50%, balance O2

<1.5%

Sevoflurane

5%

<1.5%

Desflurane

15%

<1.5%

NOTE: Balance mixture 30% O2/70%N2O, unless otherwise specified.

10.0SPARE PARTS AND ACCESSORIES

DESCRIPTION

PART NUMBER

MAX550E Oxygen Sensor ACCESSORIES DESCRIPTION

R140P02-001

Monitor Cable

R228P49

Rail Mount Bracket

R100P09

Adjustable Pole Mount Bracket

R100P22

Pole Mount Bracket

R100P26

Compact Wall Mount

RP05P07

Maxtec-Approved Power Supply

R230P10

Wall Mount Large Bracket

RP05P09

10′ Dual Blender hose (DISS)

R129P01

Repair of this equipment must be performed by a Maxtec Certified Service Technician, experienced in the repair of this device.

Equipment in need of repair shall be sent to: Maxtec Service Department 2305 South 1070 West Salt Lake City, Ut 84119 1.800.748.5355 (Include RMA number issued by Customer Service)

11.0THEORY OF OPERATION
11.1Operational Diagram

1st Stage Balance Module

2nd Stage Balance Module

Alarm/Bypass

Proportioning Valve
O 2

Oxygen Inlet Audible Alarm Air Inlet Mixed Gas Outlet

Mixed Gas Outlet to Flowmeter Sensor Bleed
Manifold

11.2Mixing Operation
The MaxBlend 2 is designed to utilize two 50 PSIG (3.4 BAR) gas sources. The two gas sources enter through the air and oxygen inlet connectors located on the bottom of the MaxBlend 2. Each inlet connector incorporates a 30 micron particulate filter. Once through the filters, each gas passes through a duckbill check valve, which prevents possible reverse gas flow from either the air or the oxygen supply systems.
The two gases then pass through a two-stage balance regulator. The purpose of this regulator is to equalize the operating pressures of the air and oxygen gas sources. Once these pressures have been balanced, the gases are proportioned according to the oxygen concentration selected on the oxygen concentration selection knob. The oxygen concentration knob allows the clinician to select a desired oxygen concentration from 21% to 100% O2. From this point, the mixed gas flows to the outlet port.
11.3Gas Outlet
There are three gas outlets on the MaxBlend 2. One is a fixed acrylic flowmeter, the other two are auxiliary ports (one on the bottom of the unit, and one on the right side). These outlets are capable of delivering combined metered flows of 0-30 LPM for the low flow model and 0-120 LPM for the high flow model. The auxiliary outlet ports are fitted with an automatic shut-off valve. The flow of gas from either outlet port is automatically initiated by attaching a pneumatic device (such as a flowmeter) to the outlet port.
Regardless of whether or not the outlet has any device connected to it, a minimal gas bleed flow of 0.1 LPM flows from the MaxBlend 2 sensor port on the left side of the blender. It is from this bleed flow that the gas is analyzed by the oxygen sensor. In addition, a toggle switch is provided allowing the user to activate an additional gas bleed, which ensures the blender has sufficient flow to function accurately when the total flow delivered to the patient is below a certain minimum threshold. For a low flow model, this additional bleed should be activated if the total flow delivered to the patient is less than 3 LPM.
For a high flow model, the additional bleed should be activated if the total flow delivered to the patient is less than 15 LPM. At delivered flows greater than these limits, the bleed toggle can be deactivated to conserve oxygen.
CAUTION: Failure to activate the bleed as described above may result in significant drift in the oxygen concentration delivered to the patient.

WWW.MAXTEC.COM · 800-748-5355

13

ENGLISH

11.4Alarm/Bypass Function
The MaxBlend 2 includes a pressure differential alarm which provides an audible alarm if gas source pressures differ by 20 PSI (1.3 BAR) (nominal) or more, or if there is a gas supply failure of one of the source gases. This alarm is generated by a reed alarm located in a cap on the bottom of the MaxBlend 2.
The primary purpose of the alarm is to audibly warn the operator of an excessive pressure drop or depletion of either source gas pressure. Should both gas pressures increase or decrease simultaneously, an alarm will not activate.
If either source gas pressure drops, the outlet pressure will drop similarly, as the mixed gas is always balanced to that of the lower gas source.
The gas bypass function operates in unison with the alarm. Once the pressure alarm is activated, the bypass function is actuated and the gas with the higher pressure flows directly to the outlet port, bypassing the mixing function of the MaxBlend 2. The oxygen concentration flowing out of the MaxBlend 2 will be that of the gas with the higher pressure. In the alarm/ bypass mode, the blender will deliver oxygen (100%) or medical air (21%) until pressures have been restored to a differential of 6 PSI or less (0.4 BAR).
If the MaxBlend 2 is set to deliver 21% and the OXYGEN source pressure is reduced enough to produce a 20 PSI (1.3 BAR) differential, the unit may not alarm because it will continue to deliver 21% concentration according to the setting. If the setting is moved slightly from 21%, the pressure differential alarm will sound. Similarly, if the MaxBlend 2 is set to deliver 100% and the AIR source pressure is reduced or lost, the unit may not alarm because it will continue to deliver 100% concentration.

12.0FLOW CHARACTERISTICS
The outlet pressure of the MaxBlend 2 decreases as the total flow rate increases. The total flow rate is the measurement of the total flow from all outlet ports. The charts below indicate the pressure drop that occurs for both low flow, and high flow, models at 3 inlet pressure settings; 30 PSIG (2.07 BAR), 50 PSIG (3.45 BAR), and 75 PSIG (5.17 BAR).

The fixed acrylic flowmeter on the left side of the MaxBlend 2 has been pressure compensated to accommodate for the pressure loss through the blender at each flow rate, using an inlet pressure of 50 PSIG.

Flow Rate -vs- Pressure Drop:

For High Flow Blender
25 50 psig

20 30 psig

15

75 psig

10

5

0

0

20

40

60

80

100

120

Flow Rate (LPM)

For Low Flow Blender
25

20

30 psig

15

50 psig

10

75 psig

5

0

0

5

10

15

20

25

30

35

Flow Rate (LPM)

13.0ELECTROMAGNETIC COMPATABILITY
The information contained in this section (such as separation distances) is in general specifically written with regard to the MaxBlend 2. The numbers provided will not guarantee faultless operation but should provide reasonable assurance of such. This information may not be applicable to other medical electrical equipment; older equipment may be particularly susceptible to interference.

NOTE: Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in this document and the remainder of the instructions for use this device.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Cables and accessories not specified within the instructions for use are not authorized. Using other cables and/or accessories may adversely impact safety, performance and electromagnetic compatibility (increased emission and decreased immunity).
Care should be taken if the equipment is used adjacent to or stacked with other equipment; if adjacent or stacked use is inevitable, the equipment should be observed to verify normal operation in the configuration in which it will be used.

ELECTROMAGNETIC EMISSIONS

This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure that it is used in such an environment.

EMISSIONS

COMPLIANCE ACCORDING TO

ELECTROMAGNETIC ENVIRONMENT

RF Emissions (CISPR 11)

Group 1

The MaxBlend 2 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

CISPR Emissions Class A Classification
Harmonic Emissions Class A (IEC 61000-3-2)
Voltage Fluctuations Complies

The MaxBlend 2 is suitable for use in all establishments other than domestic and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic purposes. NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radiofrequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.

Pressure drop (PSIG)

Pressure drop (PSIG)

ENGLISH

14

WWW.MAXTEC.COM · 800-748-5355

ELECTROMAGNETIC IMMUNITY

This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure that it is used in such an environment.

IMMUNITY AGAINST

IEC 60601-1-2: (4TH EDITION) TEST LEVEL

ELECTROMAGNETIC ENVIRONMENT

Professional

Home

Healthcare Facility Healthcare

Environment Environment

Electrostatic Contact discharge: ±8 kV discharge, ESD Air discharge: ±2 kV, ±4 (IEC 61000-4-2) kV, ±8 kV, ±15 kV

Electrical fast transients / bursts (IEC 61000-4-4)

Power supply lines: ±2 kV Longer input / output lines: ±1 kV

Surges on AC Common mode: ±2 kV mains lines Diferential mode: ±1 kV (IEC 61000-4-5)

3 A/m power frequency magnetic field 50/60 Hz (IEC 61000-4-8)

30 A/m 50 Hz or 60 Hz

Voltage dips and short interruptions on AC mains input lines (IEC 61000-4-11)

Dip>95%, 0.5 periods Dip 60%, 5 periods Dip 30%, 25 periods Dip >95%, 5 seconds

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be kept at levels to reduce electrostatic charge to suitable levels. Mains power quality should be that of a typical commercial or hospital environment. Equipment which emits high levels of power line magnetic fields (in excess of 30A/m) should be kept at a distance to reduce the likelihood of interference. If user requires continued operation during power mains interruptions, ensure that batteries are installed and charged. Ensure that battery life exceeds longest anticipated power outages or provide an additional uninterruptible power source.

Recommended separation distances between portable and mobile RF communications equipment and the equipment

RATED MAXIMUM OUTPUT POWER OF
TRANSMITTER
W

Separation distance according to frequency of transmitters in meters

150 kHz to 80 MHz 80 MHz to 800 MHz

d=1.2/V1] P

d=1.2/V1] P

800MHz to 2.5 GHz d=2.3 P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propogation is affected by absorption and reflection from structures, objects, and people.

This equipment is intended for use in the electromagnetic environment specified below. The customer or the user of this equipment should assure that it is used in such an environment.

IMMUNITY IEC 60601-1-2: 2014 (4TH TEST EDITION) TEST LEVEL

ELECTROMAGNETIC ENVIRONMENT – GUIDANCE

Professional Healthcare Facility Environment

Home Healthcare Environment

Conducted RF coupled into lines (IEC 61000-4-6)

3V (0.15 – 80 MHz) 6V (ISM bands)

Radiated RF immunity (IEC 61000-4-3)

3 V/m
80 MHz – 2.7 GHz 80% @ 1 KHz AM Modulation

3V (0.15 – 80 MHz) 6V (ISM & Amateur bands)
10 V/m
80 MHz – 2.7 GHz 80% @ 1 KHz AM Modulation

Portable and mobile RF communications equipment (including cables) should be used no closer to any part of the equipment than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter as below.
Recommended sparation distance: d=1.2 P d=1.2 P 80 MHz to 800 MHz d=2.3 P 800 MHz to 2.7 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range b.

Interference may occur in the vicinity of equipment marked with the following symbol:

The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.

WWW.MAXTEC.COM · 800-748-5355

15

ENGLISH

Maxtec 2305 South 1070 West Salt Lake City, Utah 84119 États-Unis

Téléphone : (800) 748.5355 Fax : (801) 973.6090 E-mail : sales@maxtec.com internet : www.maxtec.com

Emergo Europe Westervoorsedijk 60, 6827 At Arnhem The Netherlands

REMARQUE : La dernière édition de ce mode d’emploi peut être téléchargée depuis notre site internet à l’adresse www.maxtec.com

Conforme à : AAMI STD ES60601-1, ISO STD 80601-2-55, IEC STDS 606011-6, 60601-1-8 Certifié selon : CSA STD C22.2 No. 60601-1

Lire l’intégralité de ce manuel avant d’utiliser ou d’entretenir le MaxBlend 2. L’utilisation de MaxBlend 2 sans avoir complètement compris ses caractéristiques et ses fonctions peut entraîner des conditions d’utilisation non sécurisées.
CLASSIFICATION
Classe de protection ……………………………………………………………………………………………………………….. II, type B Protection contre l’eau ……………………………………………………………………………………………………………………. IPX1 Mode de fonctionnement ………………………………………………………………………………………………………….Continu Sécurité de l’application en présence de mélange anesthésique inflammable……….Voir la section 9.4 Caractéristiques d’alimentation ……………………………………………7,5 V(MAX) 1,9 W/250 mA(MAX)
MISE EN GARDE : Selon la loi fédérale américaine, ce dispositif ne peut être vendu que par un professionnel de la santé ou sur ordonnance.
Instructions relatives à l’élimination du produit : Le capteur, les batteries et le circuit imprimé ne doivent pas être jetés avec les ordures ménagères. Retourner le capteur à Maxtec pour une mise au rebut adaptée ou l’éliminer conformément aux recommandations en vigueur. Suivre les recommandations en vigueur pour la mise au rebut des autres composants.
GARANTIE
Le MaxBlend 2 est conçu pour l’administration d’air/oxygène. En conditions normale d’utilisation, Maxtec garantit que le MaxBlend 2 est exempt de défauts de fabrication ou de matériaux pendant une période de trois ans à compter de sa date de réception par Maxtec, sous réserve que l’appareil soit correctement utilisé et entretenu conformément au mode d’emploi de Maxtec. Sur la base de l’évaluation du produit par Maxtec, la seule obligation de Maxtec au titre de la garantie susmentionnée se limite à effectuer des réparations ou des remplacements ou à émettre un avoir pour le matériel identifié comme étant défectueux. Cette garantie s’applique uniquement aux acheteurs qui achètent un matériel neuf directement auprès de Maxtec ou par l’intermédiaire de distributeurs et agents désignés par Maxtec. Maxtec garantit que le capteur d’oxygène MAX-550E dans le MaxBlend 2 est exempt de défauts de fabrication ou de matériaux pendant une période de deux ans à compter de la date d’expédition par Maxtec d’une unité MaxBlend 2. En cas de défaillance prématurée d’un capteur, le capteur de remplacement est garanti pour le restant de la période de garantie du capteur d’origine. Les éléments soumis à un entretien courant, tels que les piles, sont exclus de la garantie. Maxtec et ses autres filiales ne sauraient être tenus responsables vis-à-vis de l’acheteur ou d’autres personnes des dommages accidentels ou consécutifs, ni des appareils ayant fait l’objet d’une utilisation abusive, d’une mauvaise utilisation, d’une mauvaise application, d’une modification, d’une négligence ou d’un accident. CES GARANTIES SONT EXCLUSIVES ET REMPLACENT TOUTES LES AUTRES GARANTIES, EXPLICITES OU IMPLICITES, NOTAMMENT TOUTE GARANTIE DE QUALITÉ MARCHANDE ET D’ADÉQUATION À UN USAGE PARTICULIER.
REMARQUE : Afin que votre MaxBlend 2 fonctionne au mieux de ses performances, son utilisation et sa maintenance doivent être effectuées conformément au présent manuel Lire attentivement le manuel avant d’utiliser le MaxBlend 2 et ne pas tenter d’effectuer une réparation ou une autre procédure qui n’y serait pas décrite. Maxtec ne peut garantir tout dommage résultant d’une mauvaise utilisation, d’une réparation non approuvée ou d’une maintenance incorrecte de l’instrument.
Remarque concernant la CEM Cet équipement utilise, génère et peut rayonner de l’énergie en radiofréquence. S’il n’est pas installé conformément aux instructions figurant dans le présent manuel, des interférences électromagnétiques pourraient survenir. L’équipement a été testé et est conforme aux limites définies dans la IEC 60601-1-2 relative aux produits médicaux. Ces limites fournissent une protection raisonnable contre les interférences électromagnétiques en cas d’utilisation dans les environnements d’utilisation prévus décrits dans le présent manuel.

Remarque concernant l’IRM Cet équipement contient des composants électroniques et ferreux dont le fonctionnement peut être altéré par des champs électromagnétiques intenses. Ne pas utiliser le MaxBlend 2 dans un environnement IRM ou à proximité d’équipement de diathermie chirurgicale à haute fréquence, de défibrillateurs ou d’équipement de thérapie par ondes courtes. Les interférences électromagnétiques pourraient perturber le fonctionnement de MaxBlend 2.
AVERTISSEMENTS
Indiquent une situation potentiellement dangereuse qui, si elle n’est pas évitée, peut occasionner des blessures graves, voire mortelles.
Les avertissements suivants s’appliquent en permanence lors de l’entretien et de l’utilisation du MaxBlend 2 : Toujours confirmer le débit prescrit avant de l’administrer au patient et surveiller le débit
fréquemment.
Toujours respecter les normes ANSI et CGA pour la manipulation des produits gazeux à usage médical, des débitmètres et de l’oxygène.
NE PAS utiliser le MaxBlend 2 à moins que du personnel formé soit prêt à répondre rapidement aux alarmes, aux blocages ou aux dysfonctionnements soudains.
Les patients sous équipement de maintien en vie doivent être surveillés visuellement à tout moment.
NE PAS ignorer les alarmes sonores du MaxBlend 2. Elles signalent des états qui nécessitent une attention immédiate.
NE PAS utiliser de pièces, accessoires ou options dont l’utilisation avec MaxBlend 2 n’a pas été approuvée. L’utilisation de pièces, accessoires ou options non autorisées peut nuire au patient ou endommager le MaxBlend 2.
Vérifier régulièrement toutes les alarmes sonores et visuelles pour s’assurer de leur bon fonctionnement. Si une alarme ne s’active pas, contacter le technicien de service agréé par Maxtec.
NE PAS utiliser le MaxBlend 2 avec le moniteur éteint ou sans configurer les alarmes réglables. Toutes les alarmes réglables doivent être configurées de manière à garantir un fonctionnement sécurisé.
NE PAS faire bouillir l’autoclave et ne pas soumettre le MaxBlend 2 à des températures supérieures à 50 °C ou (122 °F).
Si la pression d’air ou d’oxygène est réduite ou augmentée suffisamment pour créer un différentiel de pression de 20 PSI ou plus, une alarme sonore retentit. Cet état peut altérer significativement la FiO2 et le débit de sortie du MaxBlend 2.
NE PAS mettre de ruban adhésif, obstruer ou retirer l’alarme pendant l’utilisation clinique.
NE PAS bloquer l’orifice du capteur sur le côté du MaxBlend 2.
Un filtre d’eau/d’entrée d’air est recommandé pour l’utilisation du MaxBlend 2. Voir la section 6.1
NE PAS utiliser de lubrifiants sur le MaxBlend 2.
NE PAS utiliser le MaxBlend 2 avant que ses bonnes performances n’aient été vérifiées. Voir la section 3.0.
Si un état est détecté qui pourrait éventuellement empêcher le moniteur de poursuivre son fonctionnement en toute sécurité, une alarme retentit. Si, à tout moment, EOx (c.-à-d. EO2, EO4,

FRANÇAIS

16

WWW.MAXTEC.COM · 800-748-5355

etc.) apparaît sur l’écran LCD, se reporter à la section 4.0 ou contacter un technicien de service agréé par Maxtec.
Tous les services doivent être confiés à un technicien de service agréé par Maxtec.
Maxtec recommande l’entretien du MaxBlend 2 par Maxtec tous les trois ans au moins, ainsi qu’en cas de suspicion de fuite ou d’autre problème de performance.
En cas de chute du MaxBlend 2, suivre les procédures décrites à la section 3.0 pour effectuer un contrôle des performances avant de réutiliser l’appareil.
Lorsque l’unité est sur le point d’être stockée (pas d’utilisation pendant au moins 30 jours), toujours retirer les piles pour protéger l’unité contre tout risque potentiel lié à une fuite d’une pile.
Toujours remplacer les piles par des piles alcalines « AA » de marque et de haute qualité.
L’alarme d’oxygène bas peut être réglée à moins de 18 % et jusqu’à 15 %. Voir la section 2.3 sur le réglage des alarmes. Ceci est conforme à la norme IEC 80601-2-55.
Encasd’expositionàunePERTURBATIONÉLECTROMAGNÉTIQUE,l’analyseurpourraafficherunmessage d’erreur E06 ou E02. Le cas échéant, se reporter à la section 4.0 pour résoudre le problème.
Afin de prévenir les risques de brûlures, d’incendie ou de blessures : Lemélangegazeuxestpurgéencontinudansl’atmosphèreautauxdeconcentrationd’oxygène
défini avec le bouton de commande. L’oxygène purgé dans une zone fermée pourrait augmenter le risque d’incendie ou d’explosion.
NE PAS utiliser ce dispositif en présence de flammes ou de sources d’inflammation, ou en cas d’utilisation de dispositifs, tels que les équipements d’électrochirurgie ou les défibrillateurs.
NE PAS utiliser le MaxBlend 2 en présence d’anesthésiques inflammables ou dans une atmosphère de gaz explosifs. L’utilisation de MaxBlend 2 dans une atmosphère inflammable ou explosive peut entraîner un incendie ou une explosion.
Le gel électrolyte du capteur O2 galvanique est acide et peut irriter et/ou brûler la peau ou les yeux. Faire attention lors de la manipulation ou du remplacement des capteurs O2 jetables endommagés ou usés. S’assurer de mettre les capteurs éliminés au rebut conformément aux réglementations hospitalières et/ou gouvernementales (FDS du capteur O2 disponible sur demande auprès de Maxtec).
NE PAS fumer dans une zone où l’oxygène est utilisé.
Le MaxBlend 2 ne convient pas à une utilisation avec de l’oxygène 93
MISES EN GARDE
Indiquent une situation potentiellement dangereuse qui, si elle n’est pas évitée, peut occasionner des blessures mineures ou modérées et des dommages matériels.
NE PAS conserver le MaxBlend 2 dans des zones chaudes pendant des périodes prolongées. Des températures supérieures à 27 °C (80 °F) peuvent raccourcir la durée de vie des batteries. Afin de réduire le risque de choc électrostatique, NE PAS utiliser de flexibles antistatiques ou conducteurs électriques avec le MaxBlend 2.
Lors du nettoyage du MaxBlend 2 : NE PAS utiliser d’abrasifs agressifs.
NE PAS immerger le MaxBlend 2 dans des agents de stérilisation liquides ou dans des liquides de tout type.
NE PAS pulvériser de solution de nettoyage directement sur l’orifice du capteur, le silencieux de purge ou l’ouverture de l’avertisseur sonore.
NE PAS laisser de solution de nettoyage s’accumuler sur le panneau avant, l’orifice du capteur ou le silencieux de purge.
NE PAS stériliser le MaxBlend 2. Les techniques de stérilisation standards pourraient endommager le mélangeur.
NE PAS nettoyer ou sécher le MaxBlend 2 avec un pistolet à air haute pression. L’application d’air sous haute pression sur le MaxBlend 2 pourrait endommager ses composants et rendre le système inutilisable.
NE PAS nettoyer excessivement le MaxBlend 2. Un usage répété d’agents nettoyants peut entraîner une accumulation de résidus sur des composants essentiels. Cette accumulation excessive de résidus peut altérer les performances de MaxBlend 2.

Une chute ou une forte secousse du capteur après l’étalonnage peut entraîner le déplacement du point d’étalonnage et nécessiter un nouvel étalonnage.
Toujours utiliser le MaxBlend 2 avec des gaz de qualité médicale propres et secs. Les contaminants et l’humidité peuvent entraîner des dysfonctionnements. L’oxygène doit avoir un point de rosée minimal de -62 °C (-80 °F) ou une teneur en humidité inférieure à 7,9 ppm (0,0059 mg/l). La « pureté » de l’oxygène doit être d’au moins 99,0 % et l’air utilisé doit être de qualité médicale. La teneur en vapeur d’eau ne doit pas dépasser le point de rosée de -15 °C (5 °F) en dessous de la température ambiante la plus basse à laquelle le système d’administration est exposé.
La teneur en particules ne doit pas dépasser celle que l’on peut trouver immédiatement en aval d’un filtre absolu de 15 microns. Se reporter aux caractéristiques des produits CGA G-4.3 et G7.1 pour plus d’informations. La teneur en vapeur d’eau de l’air de qualité médicale ou de l’O2 fourni au mélangeur ne doit pas dépasser 5,63 x 103 milligrammes H2O par mètre cube de gaz non condensable.
NE PAS démonter le MaxBlend 2. Tous les services doivent être réalisés par un technicien de service agréé par Maxtec.
S’assurer que le MaxBlend 2 est solidement fixé. Ce dispositif est généralement monté sur un système de rail hospitalier ou un pied à perfusion. Toute chute de ce dispositif peut occasionner des blessures ou endommager l’appareil.
Les sorties de cet appareil sont capables de fournir des pressions aussi élevées que la pression d’admission. S’assurer que les dispositifs acheminant le gaz du mélangeur au patient empêchent la pression d’être excessive.
NE PAS utiliser le capteur d’oxygène s’il semble endommagé ou s’il présente des fuites. Le capteur à oxygène contient une solution faiblement acide encapsulée dans un logement en plastique. En conditions normales de fonctionnement, cette solution (électrolyte) n’est jamais exposée.
REMARQUES
Indiquent des informations supplémentaires facilitant l’utilisation du dispositif.
Le MaxBlend 2 a été testé conformément à la norme ISO 11195 et satisfait aux exigences relatives à l’administration de gaz en sens inverse.
Les pièces concernées utilisées dans MaxBlend 2 ont été nettoyées et dégraissées pour un service d’oxygène. Tous les lubrifiants utilisés sont spécifiquement conçus dans ce but.
Tant que la pression absolue du mélange gazeux est sous surveillance constante, le MaxBlend 2 lira précisément les concentrations en oxygène. Cependant, si la pression absolue varie, le relevé variera proportionnellement, puisque le capteur mesure en réalité la pression partielle d’oxygène dans le mélange. Les relevés du capteur changeront également proportionnellement aux changements de la pression barométrique. C’est pourquoi un étalonnage fréquent du capteur est recommandé.
Il est conseillé aux utilisateurs d’avoir recours à des régulateurs de pression qui affichent la pression de sortie.
Toutes les caractéristiques reposent sur les conditions environnementales standards suivantes, sauf indication contraire : températures ambiantes et des gaz d’échantillonnage de 25 °C (77 °F), pression barométrique de 760 mmHg (30 inHg), altitude par rapport au niveau de la mer, humidité relative ambiante de 50 %, humidité relative des gaz d’échantillonnage de 0 %.
Il est important de noter que l’échelle de sélection de la concentration en oxygène est fournie uniquement à titre indicatif pour la sélection de la concentration en O2. Le praticien devrait utiliser le panneau d’affichage du MaxBlend 2 pour ajuster la concentration en O2 au réglage souhaité.
Les seuils d’alarme peuvent être réglés à des niveaux qui les rendraient inutiles pour l’état clinique d’un patient particulier. S’assurer que le niveau et le débit d’oxygène fourni sont réglés aux valeurs prescrites par le médecin du patient. S’assurer également que les seuils des alarmes haute et basse sont réglés à des niveaux tels que les alarmes retentiront si le niveau d’oxygène est en dehors des limites de sécurité. S’assurer de réviser et, si nécessaire, de régler à nouveau les seuils des alarmes lorsque l’état clinique du patient change ou lorsque le médecin du patient prescrit un changement dans le traitement par oxygène.
Cet appareil ne contient pas de compensation de pression barométrique automatique.
Desfuitesdegazquientraînentunmélangedel’airambiantavecl’échantillondegazpourraient entraîner des relevés d’oxygène inexacts. S’assurer que les joints toriques du capteur et du système de dérivation de débit sont en place et intacts avant utilisation.
REMARQUE : Le(s) incident(s) grave(s) se produisant regardant le dispositif devraient être signalés à Maxtec et à l’autorité compétente de l’état membre dans lequel l’utilisateur et/ ou le patient réside. Le(s) incident(s) grave(s) est/sont défini(s) comme ayant directement ou indirectement conduit, pu conduire ou être la cause du décès d’un patient, d’un utilisateur ou d’une autre personne ; de la détérioration sévère temporaire ou permanente de l’état de santé d’un patient, d’un utilisateur ou d’une autre personne ; ou d’une sérieuse menace pour la santé publique

WWW.MAXTEC.COM · 800-748-5355

17

FRANÇAIS

TABLE DES MATIÈRES

CLASSIFICATION ………………………………………….. 16
GARANTIE …………………………………………………….. 16
AVERTISSEMENTS ……………………………………. 16
MISES EN GARDE …………………………………….. 17
REMARQUES…………………………………………………. 17
1.0INTRODUCTION………………………………………. 19 1.1Indications d’utilisation …………………………………………………………………………19 1.2Capteur d’oxygène MAX-550E ……………………………………………………………….19 1.3Guide des symboles ……………………………………………………………………………..19 1.4Identification des composants…………………………………………………………….20 1.5Témoins de l’écran………………………………………………………………………………20 1.6Vue arrière …………………………………………………………………………………………..21 1.7Exigences relatives au fonctionnement du mélangeur…………………………..21
2.0PROCÉDURES DE FONCTIONNEMENT ….. 21 2.1Configuration et installation………………………………………………………………….21 2.1.1Installation des piles …………………………………………………………………21 2.1.2Configuration du MaxBlend 2 ……………………………………………………21 2.1.3Installation du capteur ……………………………………………………………..21 2.2Surveillance ………………………………………………………………………………………… 21 2.3Procédure de réglage des alarmes ………………………………………………………22 2.3.1Réglage de l’Alarme basse ……………………………………………………..22 2.3.2Réglage de l’Alarme haute………………………………………………………22 2.3.3Mode Alarme intelligente………………………………………………………..22 2.4Fonctionnement de base…………………………………………………………………….22 2.5États et priorités des alarmes………………………………………………………………22 2.6Fonctionnement du rétroéclairage…………………………………………………….. 23 2.7Fonctionnement sur alimentation électrique externe…………………………. 23 2.8Procédures d’étalonnage…………………………………………………………………… 23 2.8.1Étalonnage à 100 % d’oxygène ………………………………………………. 23 2.8.2Étalonnage à l’air ambiant…………………………………………………….. 23 2.9Fonctionnement du débitmètre…………………………………………………………. 23

3.0VÉRIFICATION DES PERFORMANCES …….23 3.1Contrôle de sécurité du mélangeur …………………………………………………….. 24
4.0DÉPANNAGE …………………………………………. 24
5.0NETTOYAGE ET DÉSINFECTION……………..25
6.0ENTRETIEN ET MAINTENANCE …………….. 26 6.1Maintenance………………………………………………………………………………………. 26 6.2Remplacement du capteur O2……………………………………………………………. 26 6.3Test des alarmes de surveillance ……………………………………………………….. 26
7.0GUIDE DES ABRÉVIATIONS …………………… 26
8.0CARACTÉRISTIQUES …………………………….. 26 8.1Caractéristiques de l’instrument…………………………………………………………. 26 8.2Caractéristiques des alarmes……………………………………………………………… 26 8.3Caractéristiques du capteur O2…………………………………………………………… 26
9.0FACTEURS INFLUENÇANT L’ÉTALONNAGE . 27
9.1Effet de la température………………………………………………………………………..27 9.2Effet de la pression……………………………………………………………………………..27 9.3Effet de l’humidité………………………………………………………………………………27 9.4Exposition aux gaz anesthésiques……………………………………………………….27 10.0PIÈCES DE RECHANGE ET ACCESSOIRES…………………………………………..27
11.0PRINCIPE DE FONCTIONNEMENT ………….27 11.1Schéma de fonctionnement…………………………………………………………………27 11.2Opération de mélange………………………………………………………………………..27 11.3Sortie de gaz………………………………………………………………………………………27 11.4Fonction d’alarme/de dérivation……………………………………………………….. 28
12.0CARACTÉRISTIQUES DU DÉBIT …………….28
13.0COMPTABILITÉ ÉLECTROMAGNÉTIQUE .28

FRANÇAIS

18

WWW.MAXTEC.COM · 800-748-5355

1.0INTRODUCTION
Le MaxBlend 2 est un dispositif de mélange air/oxygène compact qui intègre l’utilisation d’un moniteur d’oxygène alimenté par pile. Le dispositif de mélange des gaz (mélangeur) fournit un mélange précis d’air de qualité médicale et d’oxygène, tandis que le moniteur mesure les concentrations en oxygène sélectionnées à partir du débit de gaz du mélangeur et affiche ces concentrations mesurées sur un écran numérique. Le moniteur est doté de seuils d’alarmes haute et basse qui, lorsqu’ils sont dépassés, déclenchent une alarme sonore et visuelle.
1.1Indications d’utilisation
Le MaxBlend 2 est conçu pour fournir un mélange gazeux air/oxygène continu et pour surveiller en continu la concentration d’oxygène fournie à un patient en pédiatrie ou adulte. Cet appareil est un dispositif médical dont l’utilisation est restreinte et réservée au seul personnel qualifié et dûment formé, sous la direction d’un médecin, dans un établissement de santé professionnel, c’est-à-dire des hôpitaux, des établissements de soins subaigus et des établissements de soins infirmiers où l’administration et la surveillance de mélanges air/oxygène sont nécessaires. Il ne s’agit pas d’un appareil de maintien des fonctions vitales.

1.2Capteur d’oxygène MAX-550E
Le MAX-550E est un capteur de pression partielle galvanique qui est spécifique à l’oxygène. Il consiste en deux électrodes (une cathode et une anode), une membrane en téflon et un électrolyte. L’oxygène se diffuse à travers la membrane en téflon et réagit immédiatement de façon électrochimique à une cathode en or. En parallèle, l’oxydation survient électro-chimiquement au niveau d’une anode en plomb, générant un courant électrique et fournissant une tension de sortie. Les électrodes sont immergées dans un électrolyte gélifié unique à faible teneur en acide, qui est responsable de la longue durée de vie du capteur et de sa caractéristique d’insensibilité au mouvement. Puisque le capteur est spécifique à l’oxygène, le courant généré est proportionnel à la quantité d’oxygène présent dans les gaz d’échantillonnage. En l’absence d’oxygène, il n’y a pas de réaction électro-chimique et donc un courant négligeable est produit. En ce sens, le capteur se met à zéro lui-même.
MISE EN GARDE : Le capteur d’oxygène MAX-550E est un dispositif étanche contenant un électrolyte acide doux et du plomb (Pb). Ces matériaux constituent des déchets dangereux et doivent être éliminés correctement ou retournés à Maxtec pour une élimination correcte ou un recyclage.
MISE EN GARDE : Une chute ou une forte secousse du capteur après l’étalonnage peut entraîner le déplacement du point d’étalonnage et nécessiter un nouvel étalonnage.

1.3Guide des symboles
Les symboles et les étiquettes de sécurité suivants sont disponibles sur le MaxBlend 2 et/ou l’étiquetage :

Touche Marche/Arrêt

Touche Silencieux

Touche Rétroéclairage

Touche Alarme intelligente

CAL (Touche d’étalonnage)

Touche Mode veille

Témoin Alarme basse
Témoin d’Alarme inférieur à 18 %

Rappel d’étalonnage Témoin de Batterie faible

Témoin de Mode alarme intelligente
DOWN (Touche Alarme basse)

Ne pas

Non sûr pour RM

Mise en garde

Corrosif
Attention, consulter la documentation jointe Référence de catalogue
Admission Ne pas jeter. Suivre les recommandations en vigueur pour la mise au rebut.
Date de fabrication

Fabricant

Selon la loi fédérale américaine, ce dispositif ne peut être vendu que par un professionnel de la santé ou sur son ordonnance.

Indice de protection à l’entrée

Litres par minute

Numéro de série Sortie Conforme aux normes ETL

Représentant agréé dans l’Union européenne
Lire le débit au centre du flotteur

50 °C (122 °F)

Plage

de

température

-15 °C (5 °F)

de stockage

Appareil médical

Représentant officiel

CH

en Suisse

UK

GUIDE DES SYMBOLES D’ALIMENTATION ÉLECTRIQUE Les symboles et étiquettes de sécurité suivants sont situés sur l’alimentation électrique du MaxBlend 2 (vendue séparément) :

Deux moyens de protection pour le patient (double isolation)

Conforme aux exigences UE Marquage UL/CSA combiné

Pour utilisation en intérieur sec.

Touche Déverrouillage Témoin Alarme haute Témoin d’Alarme silencieuse UP (Touche Alarme élevée) Avertissement Pièces appliquées de type B Pourcentage d’oxygène Air enrichi en oxygène Courant continu Ne pas utiliser d’huile Personne responsable au Royaume-Uni
Marquage UL/CSA combiné

WWW.MAXTEC.COM · 800-748-5355

19

FRANÇAIS

12

p

i

3

u

o

4

y t r e
w

%O2

60

50

70

40

80

30 21

90 100

5 6

q 98 7

1.4Identification des composants
1 LED D’ALARME BASSE — En état d’alarme basse, la LED « ALARME BASSE » clignote deux fois et s’accompagne d’un avertisseur sonore.

2 LED D’ALARME HAUTE — En état d’alarme haute, la LED « ALARME HAUTE » clignote deux fois et s’accompagne d’un avertisseur sonore.

3 TOUCHE MARCHE/ARRÊT — Cette touche sert à activer et désactiver l’appareil. Pour désactiver l’appareil, le bouton doit être maintenu pendant un décompte rapide 3-2-1 pour prévenir une désactivation accidentelle.

4 TOUCHE DÉVERROUILLAGE — La touche de déverrouillage sert à verrouiller et à déverrouiller l’instrument.

5 TOUCHE HAUT (ALARME HAUTE) —

La touche Up sert à régler le seuil d’alarme

haute de FiO2. L’appareil doit être en mode Déverrouillé pour que cette touche puisse

être utilisée. Voir la section 2.3.2 pour connaître les instructions de réglage du seuil

d’alarme haute de FiO2.

6 TOUCHE D’ÉTALONNAGE — Cette touche est utilisée pour étalonner l’appareil. L’appareil doit être en mode Déverrouillé pour que cette touche puisse être utilisée. Voir la section 2.8 pour connaître les instructions d’étalonnage.

7 RACCORD D’ENTRÉE D’O2 — Raccord O2 permettant le branchement au flexible d’admission d’O2 provenant de la source de gaz.

8 SÉLECTEUR DE CONCENTRATION EN OXYGÈNE — Bouton permettant de sélectionner les concentrations en oxygène mélangé allant de 21 à 100 %.

9 CONNECTEUR D’ADMISSION D’AIR DE QUALITÉ MÉDICALE — Raccord permettant de brancher un tuyau d’admission d’air provenant de la source de gaz.

q SORTIE DU DÉBITMÈTRE — Raccord permettant le branchement au système de tubage d’administration au patient.

w CONTACTEUR DE PURGE À BASCULE — Le contacteur de purge à bascule active la purge auxiliaire. Pour que le mélangeur puisse précisément mélanger l’oxygène et l’air, la

l

k

a

j

s

hf gd

purge doit être activée dès que le débit total administré au patient est inférieur à 15 l/ min pour les modèles à haut débit et inférieur à 3 l/min pour les modèles à bas débit.

e BOUTON DE COMMANDE DE DÉBIT — Le bouton de commande de débit contrôle le débit de gaz sortant du débitmètre.

r DÉBITMÈTRE D’OXYGÈNE — Il mesure le débit de mélange gazeux sortant de la sortie du débitmètre.

t TOUCHE D’ALARME INTELLIGENTE — La touche d’alarme intelligente sert à régler plus rapidement la fenêtre d’alarme haute-basse. Voir la section 2.3.3 pour connaître les instructions d’utilisation du réglage de l’Alarme intelligente.

y TOUCHE DOWN (ALARME BASSE) —

La touche Basse sert à régler le seuil

d’alarme basse de FiO2. L’appareil doit être en mode Déverrouillé pour que cette touche

puisse être utilisée. Voir la section 2.3.1 pour connaître les instructions de réglage du

seuil d’alarme basse de FiO2.

u TOUCHE DE RÉTROÉCLAIRAGE — La touche de rétroéclairage active manuellement le rétroéclairage pendant 30 secondes. Voir la section 2.6 pour plus d’informations sur le fonctionnement du rétroéclairage.

i TOUCHE SILENCIEUX — En état d’alarme, appuyer sur la touche SILENCE désactive l’alarme sonore pendant 2 minutes.

o ÉCRAN LCD — L’écran à affichage à cristaux liquides (LCD) fournit un relevé direct des concentrations en oxygène. L’écran affiche également les codes d’erreur, les modes de réglage des alarmes et les codes d’étalonnage au besoin.

1.5Témoins de l’écran
p CONCENTRATION EN OXYGÈNE — Concentration actuelle en oxygène, en pourcentage, provenant du capteur d’oxygène.

a TÉMOIN D’ALARME HAUTE — Le réglage de l’Alarme haute est affiché constamment juste sous l’icône « HAUTE » sur le relevé LCD. La valeur indiquée représente le pourcentage d’oxygène auquel l’alarme haute sera activée.

s SEUIL D’ALARME HAUTE D’OXYGÈNE — Point de consigne de l’alarme d’oxygène haute. Des alarmes sonores et visuelles se déclenchent lorsque ce seuil est dépassé.

d RAPPEL D’ÉTALONNAGE — Le symbole du rappel d’étalonnage est situé en bas de l’écran. Il s’affiche une semaine après le dernier étalonnage.

f TÉMOIN DU MODE VEILLE — Le témoin du mode veille sert à gérer la consommation de la batterie.

g TÉMOIN DE BATTERIE FAIBLE —

Le témoin de batterie faible est situé au milieu

de l’écran et n’est activé que lorsque la tension des piles est inférieure au niveau de

fonctionnement normal.

h TÉMOIN D’ALARME SILENCIEUSE/ALARME INTELLIGENTE — Lorsque la touche Silence est activée, le témoin s’affiche avec une croix pour signaler cet état. Lorsque la touche Alarme silencieuse est enfoncée, le témoin s’affiche, avec des T, pour signaler l’état.

FRANÇAIS

20

WWW.MAXTEC.COM · 800-748-5355

v W

b

1.7Exigences relatives au fonctionnement

du mélangeur

Tous les flexibles de pression d’admission amovibles par l’opérateur fournis avec le mélangeur de

gaz sont conformes à la norme ASTM/ISO 5359.

OXYGÈNE SOUS PRESSION : La source d’oxygène comprimé doit fournir un oxygène propre, sec de qualité médicale à la pression indiquée à la section 8.0.

n AIR SOUS PRESSION : La source d’air comprimé doit fournir un air propre, sec de qualité médicale à la pression indiquée à la section 8.0.

Q

m

/

,

.
j SEUIL D’ALARME BASSE D’OXYGÈNE — Point de consigne de l’alarme d’oxygène basse. Des alarmes sonores et visuelles se déclenchent lorsque ce seuil est dépassé.
k TÉMOIN D’ALARME BASSE — Le réglage de l’Alarme basse est affiché constamment juste sous l’icône « BASSE » sur le relevé LCD. La valeur indiquée représente le pourcentage d’oxygène auquel l’alarme basse sera activée.
l TÉMOIN D’ALARME <18 % — Le témoin d’alarme <18 % est situé dans l’écran du témoin d’alarme basse. Lorsque le réglage d’alarme basse est défini à moins de 18 %, le témoin clignote toutes les secondes pour avertir l’utilisateur de cet état particulier. Voir la section 2.3.1 pour connaître le réglage de cet état d’alarme basse.

2.0PROCÉDURES DE FONCTIONNEMENT
2.1Configuration et installation
2.1.1Installation des piles
Toutes les unités MaxBlend 2 sont alimentées par quatre piles alcalines AA (4 x 1,5 V) et sont expédiés sans les piles installées. Seules des piles de marque doivent être utilisées. Les piles doivent être remplacées par du personnel de service formé.
Pour installer les piles : Ouvrir le tiroir des piles en serrant les deux languettes vers l’intérieur comme indiqué sur la figure ci-dessous. En cas de difficultés à serrer les languettes avec les doigts, utiliser deux tournevis plats ou deux pièces. Retirer le tiroir de piles complètement du MaxBlend 2. Installer quatre piles alcalines AA neuves dans l’unité, en respectant le sens indiqué sur le plastique à l’intérieur du tiroir. Glisser le tiroir dans son emplacement avec les piles tournées vers le haut. Appuyer sur le tiroir jusqu’à ce que les deux languettes se bloquent.
Lorsque les piles sont remplacées, l’unité démarre et réalise un étalonnage automatique. Vérifier que l’étalonnage est effectué conformément à la section 2.8.
AVERTISSEMENT : Le remplacement des piles par du personnel insuffisamment formé peut entraîner des risques pour la sécurité.
AVERTISSEMENT : Un choc électrique ou des dommages de l’équipement peuvent survenir si une alimentation électrique externe inadaptée est utilisée. Maxtec recommande l’utilisation de l’alimentation électrique externe Maxtec MaxBlend 2 — R230P10 uniquement.

1.6Vue arrière
v COMPARTIMENT DES PILES — Le MaxBlend 2 est alimenté par quatre piles alcalines AA. b CÂBLE DU CAPTEUR — Ce câble connecte le MaxBlend 2 au capteur MAX-550E. n CAPTEUR AVEC SYSTÈME DE DÉRIVATION — Le capteur avec système de dérivation de
débit est conçu pour s’insérer dans un orifice derrière le débitmètre.
m ORIFICE DE CAPTEUR O2 — Orifice d’échantillonnage pour le capteur d’oxygène. Il permet au mélange gazeux provenant du mélangeur de circuler par la membrane du capteur.
, ALARME À ANCHE DE DIFFÉRENTIEL DE PRESSION — Alarme sonore qui, lorsqu’elle est activée, signale l’existence d’un différentiel de pression inacceptable entre les deux pressions des sources de gaz.
. SORTIE DE MÉLANGE GAZEUX AUXILIAIRE — La sortie peut être utilisée comme prise de force. / SORTIE DE MÉLANGE GAZEUX AUXILIAIRE — La sortie peut être utilisée pour ajouter des débit-
mètres supplémentaires.
Q ADAPTATEUR DE MONTAGE — Adaptateur qui permet au MaxBlend 2 d’être monté sur un support en vue d’une fixation sur un rail ou un système de ventilation.
W ORIFICE D’ALIMENTATION ÉLECTRIQUE EXTERNE — L’orifice permet le branchement d’un adaptateur d’alimentation externe. Voir la section 2.7 pour plus d’informations sur l’alimentation électrique.

2.1.2Configuration du MaxBlend 2
1. Brancher la source d’air sous pression au connecteur d’admission d’air. 2. Brancher la source d’oxygène sous pression au raccord d’admission d’O2. 3. Rincer les gaz au débit le plus élevé possible à travers le mélangeur pendant au moins une
minute pour éliminer toute particule qui aurait pu s’introduire dans le système au cours de la manipulation et de l’installation.
2.1.3Installation du capteur
1. Fixer le système de dérivation de débit sur le capteur d’oxygène. 2. Placer le capteur sur l’orifice du capteur situé derrière le débitmètre. 3. Fixer le câble du capteur directement au capteur et à la prise du capteur sur l’arrière
du compartiment du moniteur. Vérifier que le câble est complètement inséré dans les deux raccordements. 4. Étalonner le capteur avant utilisation conformément à la procédure d’étalonnage décrite à la section 2.8.
2.2Surveillance
Avant l’utilisation de l’appareil sur un patient, la concentration en oxygène du gaz administré doit être contrôlée au réglage prévu pour l’utilisation.
1. Pour commencer la surveillance, appuyer sur la touche Marche/Arrêt située sur le panneau avant. La surveillance commence immédiatement.

WWW.MAXTEC.COM · 800-748-5355

21

FRANÇAIS

2. Si le niveau d’oxygène dépasse les points de consigne d’alarme HIGH ou LOW témoin d’alarme sur le panneau avant s’allume, signalant un état d’oxygène élevé ou bas. L’alarme s’arrête lorsque les niveaux d’oxygène reviennent dans les limites ou que les seuils sont ajustés.
REMARQUE : Le MaxBlend 2 surveille la concentration en oxygène du gaz fourni par le mélangeur indépendamment de l’orifice de sortie utilisé. Ainsi, le débitmètre intégré n’a finalement pas d’effet sur la détection de l’oxygène.
2.3Procédure de réglage des alarmes 2.3.1Réglage de l’Alarme basse
Pour ajuster le réglage de l’Alarme basse : 1. Appuyer sur la touche Déverrouiller pour déverrouiller le clavier. Noter que les icônes LOW, ALARME INTELLIGENTE, CAL et HIGH commencent à clignoter indiquant le MODE RÉGLAGE D’UTILISATION. 2. Appuyer sur la touche DOWN (Alarme basse) sur le clavier.
REMARQUE : Les chiffres de l’Alarme basse commencent à clignoter et indiquent que le réglage manuel de l’Alarme basse est en cours.
3. Utiliser les touches UP et DOWN pour paramétrer l’alarme basse à la valeur souhaitée. Appuyer sur les touches fléchées pour modifier la valeur par incrément de 1 %. Si les touches sont maintenues enfoncées pendant plus d’une seconde, l’affichage avance à une vitesse de 1 % par seconde.
REMARQUE : Après 30 secondes sans activation de touche, le système conserve les derniers réglages d’alarme basse et retourne en fonctionnement normal. Si cela arrive par inadvertance, répéter simplement la procédure de réglage des alarmes. Il existe un état particulier qui permet à l’alarme d’oxygène basse d’être définie à moins de 18 %. Pour accéder à cet état, appuyer sur la touche fléchée BASSE pendant trois secondes pendant que le relevé d’alarme basse affiche 18 %. Le réglage de l’alarme peut maintenant être défini à 17, 16 ou 15 %. Une barre clignotera au-dessus du réglage pour indiquer plus clairement que l’alarme a été réglée à cet état particulier <18 %. La valeur d’alarme basse ne peut être paramétrée à moins de 15 %, elle ne peut pas non plus être paramétrée à moins de 1 % de la valeur d’alarme haute. Par exemple, si l’alarme haute est réglée à 25 %, le système n’acceptera pas un réglage d’alarme basse supérieur à 24 %. Lorsque la valeur d’alarme basse est définie, appuyer sur la touche Déverrouiller à nouveau pour valider le réglage d’alarme basse et retourner en mode utilisation normale. REMARQUE : L’alarme basse est réglée par défaut à 18 % de O2. Le retrait des piles ou la désactivation de l’unité réinitialise le seuil d’alarme basse à 18 %.
2.3.2Réglage de l’Alarme haute
Pour ajuster le réglage de l’Alarme haute : 1. Appuyer sur la touche Déverrouiller pour déverrouiller le clavier. Noter que les icônes LOW, ALARME INTELLIGENTE, CAL et HIGH commencent à clignoter indiquant le MODE RÉGLAGE D’UTILISATION. 2. Appuyer sur la touche UP (Alarme haute) sur le clavier.
REMARQUE : Les chiffres de l’Alarme haute commencent à clignoter pour indiquer que le réglage manuel de l’Alarme haute est en cours.
3. Utiliser les touches UP et DOWN pour paramétrer l’alarme haute à la valeur souhaitée. Appuyer sur les touches fléchées pour modifier la valeur par incrément de 1 %. Si les touches sont maintenues enfoncées pendant plus d’une seconde, l’affichage avance à une vitesse de 1 % par seconde.
REMARQUE : Après 30 secondes sans activation de touche, le système conserve les derniers réglages d’alarme haute et retourne en fonctionnement normal. Si cela arrive par inadvertance, répéter simplement la procédure de réglage des alarmes.
4. Lorsque l’alarme haute est paramétrée à plus de 100 %, l’alarme haute indique deux tirets. — –. Cet état spécial désactive l’alarme haute.
5. Lorsque la valeur d’alarme haute est paramétrée, appuyer sur la touche Déverrouiller à nouveau pour valider le réglage d’alarme haute et retourner en mode utilisation normale.
REMARQUE : L’alarme haute est paramétrée par défaut à 50 % de O2. Le retrait des piles réinitialise le seuil d’alarme haute à 50 %.
2.3.3Mode Alarme intelligente
1. Appuyer sur la touche Déverrouiller pour déverrouiller le clavier. Noter que les icônes LOW, ALARME INTELLIGENTE, CAL et HIGH commencent à clignoter indiquant le MODE RÉGLAGE D’UTILISATION.

2. Appuyer sur la touche Alarme intelligente sur le clavier. Noter que les chiffres LOW,

le mode Alarme et les chiffres HAUTE commencent à clignoter lentement, indiquant que

le MODE ALARME INTELLIGENTE est en cours. L’alarme haute est maintenant paramétrée au

même niveau que le relevé d’oxygène actuel +3 % (arrondi à l’entier le plus proche). L’alarme

basse est maintenant paramétrée au même niveau que le relevé d’oxygène actuel -3 %

(arrondi à l’entier le plus proche, mais jamais inférieur à 18 %).

3. Appuyer sur la touche UP permet d’ajouter 1 % au réglage de l’alarme haute et de

soustraire 1 % au réglage de l’alarme basse. Appuyer sur la touche DOWN

permet de

soustraire 1 % au réglage de l’alarme haute et d’ajouter 1 % au réglage de l’alarme basse. En

d’autres termes, la touche UP élargit la plage de l’alarme et la touche DOWN la resserre. Cette

fonction ne permet pas de définir des niveaux d’alarme supérieurs à 100 % ou inférieurs à 18 %.

4. Une fois les réglages d’alarme souhaités atteints, appuyer sur la touche Déverrouiller à

nouveau pour sauvegarder les réglages et revenir au mode d’utilisation normal. S’il s’écoule

plus de 30 secondes entre deux activations de touche par l’utilisateur, l’appareil sauvegarde

automatiquement les derniers réglages d’alarme et revient au mode d’utilisation normal.

2.4Fonctionnement de base
Pour vérifier la concentration en oxygène du gaz fourni : 1. S’assurer que le capteur est branché au système de dérivation de débit et inséré complètement dans l’orifice du capteur derrière le débitmètre. 2. Placer le bouton de commande du mélangeur à oxygène sur le réglage d’oxygène souhaité. 3. À l’aide de la touche Marche/Arrêt , s’assurer que l’unité est activée. 4. Laisser la valeur d’oxygène se stabiliser. 5. Ajuster le débitmètre au débit souhaité. Lire le débit au centre du flotteur.

2.5États et priorités des alarmes
En cas d’alarme haute ou basse, la LED correspondante commence à clignoter et s’accompagne d’un avertisseur sonore. Appuyer sur la touche Silence désactive l’avertisseur mais la LED et les chiffres de la valeur d’alarme sur l’écran continuent à clignoter jusqu’à ce que l’état d’alarme ait été rectifié. Si l’état d’alarme persiste 120 secondes après la mise en silence de l’avertisseur sonore, celui-ci commence à retentir à nouveau.

ALARME
Câble d’alimentation branché
Câble d’alimentation débranché
Tension d’alimentation c.c. externe
hors plage
Tension de batterie trop basse pour le fonctionnement de l’appareil (E04)
Niveau d’oxygène/de pression
supérieur au réglage d’alarme haute
Niveau d’oxygène/de pression
supérieur au réglage d’alarme basse
Niveau d’oxygène inférieur au
réglage d’alarme d’oxygène basse et inférieur à 18 %

PRIORITÉ D’ALARME Informative Informative Informative
Moyenne
Moyenne
Moyenne
Haute

LED D’ALARME
BASSE Désactivée Clignotement jaune unique
Jaune fixe
Jaune clignotant
Désactivée
Jaune clignotant
Rouge clignotant

LED D’ALARME
HAUTE Désactivée Clignotement jaune unique
Jaune fixe
Jaune clignotant
Jaune clignotant
Désactivée
Désactivée

ALARME SONORE
2 clignotements

RÉPÉTITION DE L’ALARME
SONORE
Pas de répétition

2 clignotements

Pas de répétition

2 clignotements

Toutes les 15 s

3 clignotements

Toutes les 25 s

3 clignotements

Toutes les 25 s

3 clignotements

Toutes les 25 s

5+5 clignotements

Toutes les 15 s

FRANÇAIS

22

WWW.MAXTEC.COM · 800-748-5355

· Un état d’alarme basse persiste jusqu’à ce que la concentration actuelle soit de 0,1 % supérieure au réglage d’alarme basse actuelle.
· Un état d’alarme haute persiste jusqu’à ce que la concentration actuelle soit de 0,1 % inférieure au réglage d’alarme haute actuelle.
Pour différencier le niveau de gravité, le moniteur émet trois bips sonores uniques.
2.6Fonctionnement du rétroéclairage
Activer le rétroéclairage : 1. Lorsque l’unité est allumée, appuyer sur le bouton Rétroéclairage pour allumer le rétroéclairage pendant 30 secondes. Appuyer de nouveau pour désactiver le rétroéclairage. 2. Si l’appareil est en utilisation dans un endroit sombre, tout bouton enfoncé active le rétroéclairage.
MISE EN GARDE : L’utilisation excessive du rétroéclairage peut réduire la durée de vie des piles.
2.7Fonctionnement sur alimentation électrique externe
Pour prolonger la durée de vie des piles, une alimentation électrique externe de 7,5 V c.c. approuvée par Maxtec peut être achetée. Une fois connectée à l’unité, toute l’alimentation est fournie par l’alimentation électrique. Les piles sont toujours requises dans l’unité et fournissent une alimentation d’urgence au cas où l’alimentation c.a. principale est interrompue.
REMARQUE : Utiliser uniquement l’alimentation électrique Maxtec décrite à la section 10.0. REMARQUE : L’alimentation électrique n’est pas un chargeur de piles.
NE PAS utiliser de piles rechargeables.
2.8Procédures d’étalonnage
2.8.1Étalonnage à 100 % d’oxygène
Le MaxBlend 2 doit être étalonné avant d’être placé en mode utilisation clinique. Ainsi, Maxtec recommande un étalonnage hebdomadaire de l’unité. Un étalonnage fréquent n’a pas d’effet indésirable sur les performances du MaxBlend 2. L’étalonnage doit également être réalisé lors du remplacement d’un capteur.
Le capteur est étalonné au mieux lorsqu’il est monté dans l’orifice du capteur du MaxBlend 2. Comme en fonctionnement normal, le capteur d’oxygène répond le mieux lorsqu’il est installé en position verticale avec le capteur vers le bas.
Les changements de la pression barométrique peuvent altérer le relevé d’oxygène. Un changement de 1 % de la pression barométrique entraîne une erreur de 1 % du relevé réel. (Exemple : en cas d’un relevé de mélange d’oxygène de 50 % et d’une chute de pression barométrique de 1000 mbar à 990 mbar, la valeur chutera à 50 % x (990/1000) = 49,5 %). Maxtec recommande un étalonnage après chaque changement d’élévation du point d’utilisation de plus de 150 mètres (500 pi).
Il est mieux d’étalonner le MaxBlend 2 en utilisant l’orifice du capteur avec un oxygène de qualité technique standard (99,0 % ou mieux). L’étalonnage de l’unité avec l’air ambiant est moins précis sur toute la plage de fonctionnement de FiO2.
Étalonner à 100 % d’oxygène : 1. Brancher la conduite d’alimentation en oxygène (l’alarme de différentiel de pression pourra retentir). Vérifier que le capteur est branché dans l’orifice du capteur à O2 et raccordé au câble du capteur. Ne PAS brancher la conduite d’alimentation en air à ce moment. 2. À l’aide de la touche Marche/Arrêt, s’assurer que le MaxBlend 2 est activé. 3. Tourner le bouton de commande de la FiO2 jusqu’à l’arrêt à 100 %. Attendre quelques minutes que le relevé se stabilise. 4. Appuyer sur la touche Déverrouiller pour déverrouiller le clavier. Noter que les icônes LOW, ALARME INTELLIGENTE, CAL et HIGH commencent à clignoter indiquant le MODE RÉGLAGE D’UTILISATION. 5. Appuyer sur la touche CAL (Étalonnage) sur le clavier. Le mot « CAL » apparaît sur l’écran pendant environ 5 secondes puis se termine par 100,0 %. 6. L’unité est maintenant étalonnée et en mode de fonctionnement normal.

2.8.2Étalonnage à l’air ambiant
Le MaxBlend 2 peut rapidement être étalonné à l’air ambiant (20,9 %).
Étalonner à l’air ambiant :
1. Brancher la conduite d’alimentation en air (l’alarme de différentiel de pression pourra retentir). Vérifier que le capteur est branché dans l’orifice du capteur à O2 et raccordé au câble du capteur NE PAS brancher la conduite d’alimentation en oxygène à ce moment. (Il est également possible de réaliser l’étalonnage à l’air ambiant en retirant le capteur de l’orifice d’échantillonnage d’O2 et en détachant le système de dérivation de débit). Le recours à cette méthode permet de laisser les conduites d’alimentation en gaz attachées.)
2. À l’aide de la touche Marche/Arrêt, s’assurer que le MaxBlend 2 est activé. 3. Tourner le bouton de commande de la FiO2 jusqu’à l’arrêt 21 %. Attendre quelques
minutes que le relevé se stabilise. 4. Appuyer sur la touche Déverrouiller pour déverrouiller le clavier. Noter que les icônes
LOW, ALARME INTELLIGENTE, CAL et HIGH commencent à clignoter indiquant le MODE RÉGLAGE D’UTILISATION. 5. Appuyer sur la touche CAL (Étalonnage) sur le clavier. Le mot « CAL » apparaît sur l’écran pendant environ 5 secondes puis se termine par 20,9 %. 6. L’unité est maintenant étalonnée et en mode de fonctionnement normal.
2.9Fonctionnement du débitmètre
1. Ajuster le débitmètre au point de consigne souhaité tel qu’il est lu par le centre du flotteur. · Augmenter le débit–tourner le bouton dans le sens inverse des aiguilles d’une montre · Diminuer le débit–tourner le bouton dans le sens des aiguilles d’une montre
2. Vérifier que le contacteur de purge à bascule est dans la bonne position. · Pour les mélangeurs à bas débit, le contacteur de purge à bascule doit être activé (position haute) si le débit total combiné du mélangeur est inférieur à 3 l/min. · Pour les mélangeurs à haut débit, le contacteur de purge à bascule doit être activé (position haute) si le débit total combiné du mélangeur est inférieur à 15 l/min. · À des débits fournis supérieurs à ces seuils, le contacteur de purge à bascule peut être désactivé (position basse) pour conserver le gaz.
MISE EN GARDE : La non-activation de la purge telle que décrite ci-dessus peut entraîner des concentrations d’oxygène inexactes dans le mélangeur. Cependant, le MaxBlend 2 continuera à afficher la concentration fournie réelle.
MISE EN GARDE : Les sorties de cet appareil sont capables de fournir des pressions aussi élevées que la pression d’admission. S’assurer que les dispositifs acheminant le gaz du mélangeur au patient empêchent la pression d’être excessive.
3.0VÉRIFICATION DES PERFORMANCES
Avant l’utilisation clinique de MaxBlend 2, réaliser les tests suivants :
AVERTISSEMENT : Si le MaxBlend 2 ne fonctionne pas tel que décrit sur la page suivante, contacter votre distributeur Maxtec ou un technicien de service agréé de Maxtec :
Maxtec 2305 South 1070 West Salt Lake City, UT 84119 385-549-8000 or 800-748-5355
NE PAS utiliser le MaxBlend 2 avant que ses bonnes performances n’aient été vérifiées.

WWW.MAXTEC.COM · 800-748-5355

23

FRANÇAIS

3.1Contrôle de sécurité du mélangeur
REMARQUE : Avant de poursuivre, s’assurer que la commande du point de consigne d’alarme haute est DÉSACTIVÉE [l’écran affiche (–)] et que la commande du point de consigne d’alarme basse est inférieure à 20 %.

PROCÉDURE

RÉPONSE DU MÉLANGEUR

1. Brancher le mélangeur à des sources d’oxygène / d’air de 50 ±5 PSIG. Ajuster le bouton du mélangeur à 60 %. Activer le contacteur de purge à bascule (position haute) et ajuster le débitmètre à un minimum de 2 l/min.

Absence de réponse. Le moniteur affiche 60 % ±3%.

2. Débrancher la source d’air 50 PSIG du MaxBlend 2. REMARQUE : Le mélangeur doit acheminer le gaz pour que l’alarme soit active.

L’alarme sonore retentit. Le moniteur affiche 100 % ±3 %.

3. Rebrancher la source d’air 50 PSIG au MaxBlend 2.

L’alarme sonore s’arrête. Vérifier que le moniteur affiche 60 % ±3 %.

4. Débrancher la source d’oxygène 50 PSIG du MaxBlend 2.

L’alarme sonore retentit. Le moniteur affiche 20,9 % ±3 %.

L’alarme sonore s’arrête. 5. Rebrancher la source d’oxygène 50 PSIG au MaxBlend 2. Vérifier que le moniteur
affiche 60 % ±3 %.

6. Ajuster les deux régulateurs d’admission d’air et d’oxygène à 0 PSIG.

Absence de réponse.

7. Débrancher le tuyau d’admission d’air du régulateur et insérer l’extrémité dans le bécher d’eau.

Absence de réponse.

8. Élever lentement la pression du régulateur d’oxygène à 50 PSIG puis la ramener à 0 PSIG tout en observant l’extrémité du flexible d’air dans le bécher.

Aucune bulle ne doit être constatée. L’alarme sonore retentit.

9. Sécher et fixer à nouveau le flexible d’admission d’air au régulateur.

Absence de réponse.

10. Débrancher le tuyau d’admission d’oxygène du régulateur et insérer l’extrémité dans le bécher d’eau.

Absence de réponse.

11. Élever lentement la pression du régulateur d’air à 50 PSIG puis la ramener à 0 PSIG tout en surveillant l’extrémité du flexible d’oxygène dans le bécher.

Aucune bulle ne doit être constatée. L’alarme sonore retentit.

12. Sécher et fixer à nouveau le flexible d’admission d’oxygène au régulateur.

Absence de réponse.

4.0DÉPANNAGE
PROBLÈME : Écart dans la concentration d’oxygène entre le bouton de sélection de la concentration en oxygène et la valeur affichée sur l’écran supérieure à 3 % CAUSES POSSIBLES ET SOLUTIONS :
· La purge est désactivée. Activer le contacteur de purge à bascule. Se reporter à la section 2.9, Fonctionnement du débitmètre.
· Surveillance mal étalonnée. Étalonner. Se reporter à la section 2.8, Procédure d’étalonnage.
· Capteur usé. Remplacer le capteur. Se reporter au chapitre 6.2. · Alimentation en gaz contaminée. Contacter Maxtec pour la réparation du MaxBlend 2. · Mélangeur mal étalonné. Contacter Maxtec pour la réparation.
PROBLÈME : Déclenchement de l’alarme de différentiel de pression. CAUSES POSSIBLES ET SOLUTIONS :

· Différences de pression d’admission de 20 PSI ou plus. Corriger la différence de pression.
· Alarme de pression incorrectement étalonnée. Contacter Maxtec pour la réparation. · Fonctionnement du mélangeur de MaxBlend 2 mal étalonné. Contacter Maxtec pour la
réparation.
PROBLÈME : La pression d’admission a une perte d’alimentation, aucune alarme sonore de différent