KERN ORM 2UN Digital Refractometer

Digital Refractometer

INSTRUCTION MANUAL

Carefully read through the operating manual even if you have prior experience with KERN refractometers

1. General information

1.1 Intended use

The refractometer is a measuring instrument for determining the refractive index of transparent substances in liquid or in some cases also in the solid state. It is used to observe the behaviour of light as it passes from a prism with known properties to the substance being tested.
Use of the refractometer for other purposes is contrary to its intended use and may be hazardous. The manufacturer shall not be liable for any damages caused by improper use.

1.2 Warranty

The warranty shall be void in the event of:

▸ Failure to observe the instructions in the operating manual
▸ Use for purposes other than those described
▸ Modifications or opening the device housing
▸ Mechanical damage and/or damage resulting from media, liquids, natural wear and tear

This digital refractometer cannot measure any liquid that is highly corrosive to metal or glass. When measuring liquids that are corrosive to plastics or react chemically with plastics, be careful not to drop the measured liquid onto the shell. Otherwise it will corrode the shell.

2. Introduction

2.1 Description

Digital Refractometer

 

 

Digital Refractometer

 

2.2 Scope of delivery
1x Strorage box | 1x Digital refractometer | 1x Operating manual | 1x AAA Battery 1.5 V | 1x Pipette | 1x Screwdriver

3. Display & operating buttons

3.1 Description display & operating buttons

Digital Refractometer

Note:Please replace the battery when the BATTERY  is displayed.

4. Preparing before operating

4.1 Install the battery

Digital Refractometer

5. Turn on & measure

5.1 Turn on

 

Digital Refractometer

Note :

1. When used outdoors, please avoid strong light so as not to affect the measurement accuracy.
2. Please keep the instrument in a stable and still statement and position.

5.2 Measure

After turning on, clean the sample tank with distilled water and then dry it. Now fill the sample up to the mark, close the cover and press “READ”.

Digital Refractometer

5.3 Average value measurement

Press “READ” for 2 seconds. The device starts an automatic measurement series of 15 measurements and shows the average value.
Afterwards, the device automatically turns back to the normal measuring mode.

 

Digital Refractometer

6. Calibration

The refractometer can only be calibrated with distilled water. To do this, fill the sample tank with distilled water up to the mark and close the cover.

 

Digital Refractometer

While “CAL” is flashing in the display, press “CAL” again to start the calibration. When the calibration is finished, the display shows “End”.
After approx. 10 seconds, the device automatically returns to normal mode.

Digital Refractometer

Digital Refractometer

We recommend calibrating the refractometer,

  • when commissioning
  • after a strong shock
  • after longer transport
  • after a change of location with a large temperature difference
  • if the device has not been used for a long time

Always use distilled water and make sure that the refractometer, the water and the environment are at the same temperature.

7. Changing scale & termperature unit

Digital Refractometer

7.1 Changing scale

Digital Refractometer

7.2 Changing termperature unit

If exceed the temperature limitations, the signs “HHH” or “LLL” would show

Digital Refractometer

8. Turning off

If without any operations for 1 minute, the instrument would be automatically turned off

9. Cleaning & maintenance

1. To avoid damages to the prism and the sample tank, clean them with distilled water after each use.
2. Dry it with a soft cloth afterwards.
3. Do not use hard or abrasive objects for cleaning.
4. Do not leave any residue in the sample tank.
5. If the refractometer is not going to be used for a longer time, remove the battery and store it at a We recommend calibrating the refractometer, cool and dry place.

10. Disposal

The packaging consists of environmentally friendly materials which can be disposed of via local recycling facilities.
The device and storage box should be disposed

11. Technical data

Scale +accuracy+ resolution Oepents to the model
Temperature o,o -40,o•c / 32,o-  104,o’F
AutomaticTemperature Compensation Yes
Minimumsample volume 0.2 – 0.3 ml (Marking ring)
AUTO-OFF 60 seconds
Averagingmeasurement 15measurements
Banery 1 xAAA 1.5 V
Lifetime of thebattery Approx. 10.000 measurements
Overall dimensions LxWxH 12Sx6Sx30 mm
Net weight 140  g (without battery)

12. Error codes

A01 Beyond the scope of calibration temperature.  (0.0°C~40.0°C)
A02 During calibration, no solution or solution wrong.
A03 This instrument has a hardware failure.
Model Scale No. Range Unit Resolution Accuracy
ORM 50BM Brix S01 0.0~50.0 % 0.1% ±0.2%
Refractive Index S02 1.3330~1.4200 nD 0.0001nD ±0.0003nD
ORM 1RS Brix S01 0.0~90.0 % 0.1% ±0.2%
Refractive Index S02 1.330~1.5177 nD 0.0001nD ±0.0003nD
Fructose ORM 1SU Fructose S01 0.0~68.9 % 0.1% ±0.2%
Glucose S02 0.0~59.9 % 0.1% ±0.2%
Brix S03 0.0~90.0 % 0.1% ±0.2%
Refractive Index S04 1.3330~1.5177 nD 0.0001nD ±0.0003nD
ORM 2SU Lactose S01 0.0~16.5 % 0.1% ±0.2%
Maltose S02 0.0~15.6 % 0.1% ±0.2%
Dextran S03 0.0~10.6 % 0.1% ±0.2%
Brix S04 0.0~50.0 % 0.1% ±0.2%
Honey ORM 1HO Honey Water S01 5.0~38.0 % 0.1% ±0.2%
Honey Baume S02 33.0~48.0 °Bé 0.1 ±0.2
Brix S03 0.0~90.0 % 0.1% ±0.2%
Refractive Index S04 1.3330~1.5177 nD 0.0001nD ±0.0003nD
Salinity ORM 1NA Salinity (NaCl) % S01 0.0~28.0 % 0.1% ±0.2%
Salinity (NaCl) ‰ S02 0~280 1‰ ±2‰
Specific Weight S03 1.000~1.220 0.001 ±0.002
Brix S04 0.0~50.0 % 0.1% ±0.2%
Refractive Index S05 1.3330~1.4200 nD 0.0001nD ±0.0003nD
ORM 1SW Salinity Seawater S01 0~100 1‰ ±2‰
Chlorinity Seawater S02 0~57 1‰ ±2‰
Specific Weight S03 1.000~1.070 0.001 ±0.002
Brix S04 0.0~50.0 % 0.1% ±0.2%
Refractive Index S05 1.3330~1.4200 nD 0.0001nD ±0.0003nD
Alcohol ORM 1AL Alcohol Mass. S01 0~72 % 1% ±1%
Alcohol Vol. S02 0~80 % 1% ±1%
Brix S03 0.0~50.0 % 0.1% ±0.2%
Refractive Index S04 1.3330~1.4200 nD 0.0001nD ±0.0003nD
Beer ORM 1BR Plato S01 0.0~30.5 °P 0.1 ±0.3
SG Wort S02 1.000~1.130 0.001 ±0.002
Brix S03 0.0~50.0 % 0.1% ±0.2%
Refractive Index S04 1.3330~1.4200 nD 0.0001nD ±0.0003nD
Wine ORM 1WN Oechsle S01 0~150 °Oe 1 ±2
Vol% S02 0.0~22.0 % 0.1% ±0.2%
KMW (Babo) S03 0.0~25.0 0.1 ±0.2
Brix S04 0.0~50.0 % 0.1% ±0.2%
ORM 2WN Oechsle France S01 0~230 °Oe 1 ±2
Vol% S02 0.0~22.0 % 0.1% ±0.2%
KMW (Babo) S03 0.0~25.0 0.1 ±0.2
Brix S04 0.0~50.0 % 0.1% ±0.2%
Coffee ORM 1CO Coffee TDS 1 S01 0.0~25.0 0.1 ±0.2
Brix S02 0.0~50.0 % 0.1% ±0.2%
Refractive Index S03 1.3330~1.4200 nD 0.0001nD ±0.0003nD
ORM 2CO Coffee TDS 2 S01 0.00~25.00 0.01 ±0.20
Brix S02 0.00~30.00 % 0.01% ±0.20%
Refractive Index S03 1.3330~1.4200 nD 0.0001nD ±0.0003nD
Urine ORM 1UN Urine Human S01 1.000~1.050 0.001 ±0.002
Serum Protein S02 0.0~12.0 g/100ml 0.1 ±0.2
Brix S03 0.0~50.0 % 0.1% ±0.2%
Refractive Index S04 1.3330~1.4200 nD 0.0001nD ±0.0003nD
ORM 2UN Urine Cat S01 1.000~1.060 0.001 ±0.002
Urine Dog S02 1.000~1.060 0.001 ±0.002
Brix S03 0.0~50.0 % 0.1% ±0.2%
Refractive Index S04 1.3330~1.4200 nD 0.0001nD ±0.0003nD
Car / Industry ORM 1CA Cleaner S01 (-60.0)~0.0 °C 0.1℃ ±0.5℃
AdBlue® S02 0.0~51.0 % 0.1% ±0.2%
Battery Fluid S03 1.000~1.500 0.001 ±0.005
Brix S04 0.0~50.0 % 0.1% ±0.2%
Refractive Index S05 1.3330~1.4200 nD 0.0001nD ±0.0003nD
Ethylenglycol (%) S01 0.0~100.0 % 0.1% ±0.5%
Ethylenglycol (°C) S02 (-50.0)~0.0 0.1℃ ±0.5℃
Propylenglycol (%) S03 0.0~100.0 % 0.1% ±0.5%
ORM 2CA Propylenglycol (°C) S04 (-60.0)~0.0 °C 0.1℃ ±0.5℃
Brix S05 0.0~90.0 % 0.1% ±0.2%

 

Specifications

  • Measuring instrument: Refractometer
  • Measurement type: Refractive index of transparent substances
  • Display: LCD Multi-function display
  • Material: Stainless steel sample tank
  • Power source: 1.5V AAA Battery

Frequently Asked Questions (FAQ)

Q: What should I do if an error code is displayed?

A: Check the error code in the manual to identify the issue. If needed, recalibrate the refractometer following the calibration instructions.

Q: Can the refractometer measure highly corrosive liquids?

A: No, the digital refractometer cannot measure liquids highly corrosive to metal or glass. Be cautious with liquids corrosive to plastics to avoid damaging the device.

Documents / Resources

KERN ORM 2UN Digital Refractometer [pdf] Instruction Manual
ORM 2UN, P -V1.5.240906, ORM 2UN Digital Refractometer, ORM 2UN, Digital Refractometer, Refractometer

References