LB4071-101 ملٽي ڪائونٽر ٽائمر
“
پيداوار جي ڄاڻ
پيداوار جو نالو: IEC Multi Counter
ماڊل نمبر: LB4071-101
وضاحت: A compact and versatile instrument
for general laboratory timing, counting, measuring frequency or
rate, and performing Geiger counting.
خاص خاصيتون:
- Timing to 0.1 ms
- Crystal locked accuracy of better than 0.01% +/-1 least
significant digit - Microprocessor controlled functions
- LED indication for mode and function selection
وضاحتون:
- طول و عرض: 375mm x 170mm x 107mm
- وزن: 2.4 ڪلو
- طاقت: 220/240V.AC 50/60Hz
پيداوار جي استعمال جون هدايتون
شروعاتي سيٽ اپ
- Plug the unit into a standard 240V.AC power outlet.
- The digital display should illuminate upon powering on.
ياداشت جا ڪم
- MEM UP/DOWN: Scroll through the active memory
اسٽور. - TOTAL: Add all memory values together.
- AVRG: Calculate the average of all memory
قدر. - صاف ڪريو: Remove selected memory values.
- صاف ڪريو: Empty all memory values.
موڊز
- وقت جو طريقو:
- AutoRange: From 0.0001s up to 99.9999 seconds,
then AutoRanges to 999.999 seconds by 0.001s. - Automode: Set by pressing STOP then RESET
buttons sequentially for automatic timing start and stop based on
changes in electrical connections. - افعال:
- شروع / اسٽاپ: Timer runs when START connections
change momentarily; stops and loads memory when STOP connections
change momentarily. - PHOTOGATE: Timer runs when START connections
change; stops and stores value when connections revert to original
status. Provides power for photogate circuits.
اڪثر پڇيا ويندڙ سوال (FAQ)
Q: How do I reset the memory on the Multi Counter?
A: To clear all memory values, press and hold
the CLEAR button until a double beep is heard. The memory store
will be emptied, and the MEM LED will turn off.
Q: What is the power requirement for the Multi Counter?
A: The Multi Counter operates on 220/240V.AC at
50/60Hz پاور سپلائي.
“`
هدايتي ورق
Multi Counter
Timer, Counter, Frequency, Geiger
LB4071-101
وضاحت:
The IEC `MULTI – COUNTER’ is a compact and versatile instrument for general laboratory timing to 0.1 ms, counting, measuring frequency or rate and for performing Geiger counting.
Each of the 3x modes (Timing, Counting/Freq and Geiger) has a set of `Functions’ to select the type of function you want for the mode you selected. All selection is by LED and the indication reminds you always of the mode and function that is operating
Special features are:
· High speed timing to 100 microseconds resolution.
· Large six digit LED display.
· All press button operation with LED indication of functions.
· Automatic loading memory up to a depth of 20 values.
· Memory items can be selectively deleted to remove errors. Memory items can be scrolled, totalled or averaged.
· Sockets for extension speaker.
· Speaker and volume control for all counts and frequency.
· Output sockets for 12V.AC. supply for photogate lamps.
· Accepts both high voltage GM tube and low voltage Alpha detector. Both are available from IEC upon request.
· Start/Stop TIME sockets also operate as remote Start/Stop sockets when running in COUNT, FREQUENCY or GEIGER modes.
ڊگھائي: 375mm
کوٽائي: 170mm
اوچائي: 107mm
وزن: 2.4 ڪلو
LB4071-101(new) no slave display.doc
جون-ايڪسيمڪس
1
3-
هدايتي ورق
وضاحتون:
طاقت:
220/240V.AC 50/60Hz.
ACCURACY: All operations relating to timing and frequency are crystal locked ensuring an accuracy of better than: 0.01% +/-1 least significant digit.
All functions are microprocessor controlled.
شروعاتي پاور آن:
Units are fitted with IEC 3 pin mains socket to accept separate mains cable. Plug into a standard 240V.AC. power outlet. Digital display should illuminate.
· Small LEDs indicate the Mode of operation and Function.
· Press MODE button to select the Mode of operation required.
· Press FUNCTION button to select Function required in that mode.
Press Button Operations:
· START: initiates timing, counting or Geiger counting.
· STOP: stops timing or counting and value is stored in memory.
· RESET: operates after STOP. Zero display and also performs an AutoMode external connection check on START/STOP sockets.
· MEM UP/MEM DOWN scrolls and recalls active memory locations.
ياداشت:
When STOP occurs by either press button or by remote socket, the last value is automatically stored into memory. When any value is stored, the small `MEM’ LED is on. When 20 values are stored (memory full), the memory LED flashes.
MEM UP/DOWN
Buttons scroll through the active memory store. When the first or last stored memory is reached, a
longer beep sounds.
TOTAL
Button adds all memory values together. Press and hold until double beep is heard. Total of memory values will display whilst button is held depressed.
AVRG
Button calculates the average of all the memory values. Press and hold until double beep is heard. Average will display whilst button is held depressed.
پورهيت
Button removes selected memory values. Scroll to select the unwanted value. Press and hold button until double beep is heard. Selection is now erased from memory leaving the other values untouched. Display shows ‘——‘.
صاف ڪريو
Button empties all memory values. Press and hold button until double beep is heard. Memory store will be empty and the small `MEM’ LED will be off.
طريقا:
Three different modes of operation are selectable: · Timing · Counting & Frequency · Geiger counting.
LB4071-101(new) no slave display.doc
جون-ايڪسيمڪس
2
3-
هدايتي ورق
وقت:
AutoRange:
0.0001s up to 99.9999 seconds, then AutoRanges to 999.999 seconds by 0.001s.
Automode:
This function is set by pressing STOP then RESET buttons sequentially. When set, the starting and stopping of timing will occur upon any change of state of the START / STOP electrical connections. This automatic feature can save classroom time and difficulty by eliminating the necessity of creating specific ‘making’ or ‘breaking’ external connections for experiments.
There are four different functions of timing:
شروع / اسٽاپ:
When status of the START connections is changed momentarily the timer runs. The start connections then have no effect. When the status of the STOP connections is changed momentarily the timer stops and memory is loaded.
PHOTOGATE:
When status of the START connections is changed the timer runs. When same sockets revert to original status the timer stops and the value is stored in memory. The sockets also provide the power required to run most photogate circuits.
مدو:
When the status of the START connections is changed the timer runs. When same sockets revert to original status there is no effect. When same sockets are changed again, the value is stored in memory, the timer is reset and then starts timing the next period. To stop the timing press STOP.
PENDULUM:
When the status of the START connections is changed the timer runs. When same sockets revert to original status there is no effect. When same sockets are changed again, there is no effect. Upon the fourth change, the value is stored in memory, the timer is reset and then starts timing the next pendulum period. To stop timing press STOP. Effectively this is a double `PERIOD’.
Counting & Frequency:
The START and STOP buttons or the joining of the TIME START/STOP sockets permits the counting and frequency measurement to Start or Stop. When stopped, the last value is stored into memory.
Input Response:
Pulses of 20mv P/P to 100V can be counted. The sensitivity of the counting input can be adjusted between these limits. For low level pulses, increase SENSITIVITY until steady and reliable counting occurs.
There are four different functions of counting and frequency:
مسلسل:
Counting continues until Stop button pressed or Stop sockets change in state. Value is stored
خودڪار.
100 سيڪنڊ:
Counts for 100 seconds. After this time has expired, counting stops and the total is displayed. Value is automatically stored into memory.
10 سيڪنڊ:
Counts for 10 seconds. After this time has expired, counting stops and the total is displayed. Value is automatically stored into memory.
تعدد:
The pulses applied are counted per second and displayed as frequency to a maximum of 999,999Hz. Starting and stopping of the frequency function is performed by the buttons or the sockets in the TIME mode section. Each time the frequency is updated, the last value is automatically stored in memory.
LB4071-101(new) no slave display.doc
جون-ايڪسيمڪس
3
3-
هدايتي ورق
Geiger Counting:
The setting of the GM VOLTS should suit the tube type being used. The normal wide range Alpha, Beta & Gamma halogen quenched GM tube (type MX168 or similar), the voltage should be about 450V.DC. for best reliability and sensitivity.
There are four different functions of Geiger counting:
مسلسل:
Counting continues until STOP button pressed or STOP sockets change state. Each Geiger count applied to the socket is counted. The voltage applied to the GM tube can be adjusted from 200 to 600 V.DC. for optimum sensitivity and for experiments involving ‘Plateau Voltages’. In addition to the normal high voltage GM tube system, IEC manufactures a special solid state ALPHA particle detector, with inbuilt amplifier, which can be used for low level Alpha Particle detection.
TOTAL
Counts over a 10 or a 100 second period: After this time has expired, Geiger counting stops and the total is displayed. Value is automatically stored into memory.
شرح:
The pulses detected are counted per second and displayed as frequency or rate to a maximum of 999,999Hz. Starting and stopping of the frequency function is performed at the buttons or the sockets in the TIME mode section. Each time the function is stopped, the last value is stored in memory.
اسپيڪر:
The instrument has an inbuilt speaker to monitor the `GM clicks’ together with sockets for an extension speaker (8 ohm impedance). A speaker volume control is provided.
Lamp پيداوار:
Output sockets provide 12V.AC. at 1 amp for Photogate lamps وغيره.
ريموٽ:
Duplicates the RESET button function. Using a long cable, this socket can be joined to the common or `GRND’ socket by a switch or press button to create a remote RESET control.
اختياري لوازمات:
· Photogates for experiments. · Geiger Muller Tube with tube holder and lead. · Solid state ALPHA particle detector with holder and lead. · Extension speaker, 8 ohm impedance.