RICE LAKE MSI-7001 Tran Send SendIt Instruction Manual

MSI-7001 Tran Send SendIt

Specifications:

  • Display: 800 x 600 or greater
  • Built-in serial port or USB to serial port adapter
  • FTDI chip set required (Tripp-Lite USB/Serial Adapter (PN
    153603) recommended – install driver)
  • JAVA JRE 1.7 or newer

Product Usage Instructions:

Install Software:

  1. Open the Rice Lake website and navigate to the MSI-7001
    TranSend product page.
  2. Download the MSI mV Scale Calibration software to the
    computer.
  3. Extract the MSI mV Scale Calibration folder.
  4. Open the folder and double click on
    ScaleCoreConnect_XX-XX.exe.

Calibrate TranSend/SendIt:

  1. Connect the receiver unit to the computer through the serial
    port. Refer to the technical manual for RS-232 serial
    connections.
  2. Power on the transmitter and receiver units.
  3. Launch the installed MSI mV Scale Calibration program on the
    computer.
  4. Select Calibration from the setup type drop-down menu.
  5. Select the COM port the TranSend/SendIt receiver is connected
    to from the drop-down menu.
  6. Select the Advanced Mode checkbox.
  7. If needed, adjust upper and lower percents detected by the
    system.
  8. Enter the total scale capacity.
  9. Remove all weight from the scale and wait for stability.
  10. Load test weights onto the scale and enter their values in the
    dialogue box.
  11. Select Calibration Complete when done.
  12. Follow indicator manual instructions to calibrate the
    indicator.

FAQ:

Q: Where can I download the manuals for TranSend/SendIt?

A: Manuals can be downloaded from Rice Lake Weighing Systems at
www.ricelake.com/manuals.

Q: How can I get warranty information for TranSend/SendIt?

A: Warranty information is available at www.ricelake.com/warranties.

“`

TranSend/SendIt
Wireless mV Scale Calibration
Wireless TranSend/SendIt transmitter and receiver pairs must be calibrated to the connected load cells before an indicator can be calibrated to the load cells. If the TranSend/SendIt pair is not calibrated, the indicator may have unexpected range limits and distorted weight values.
i Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals Warranty information is available at www.ricelake.com/warranties
System Requirements Specifications subject to change without notice. Windows® Operating System
Display: 800 x 600 or greater Built-in serial port or USB to serial port adapter
FTDI chip set required (Tripp-Lite USB/Serial Adapter (PN 153603) recommended – install driver) JAVA JRE 1.7 or newer
To download and install JAVA JRE: https://java.com/en/download/manual.jsp
Install Software To install MSI mV Scale Calibration program:
1. Open the Rice Lake website and navigate to the MSI-7001 TranSend product page. 2. Download the MSI mV Scale Calibration software to the computer. 3. Extract the MSI mV Scale Calibration folder. 4. Open the folder and double click on ScaleCoreConnect_XX-XX.exe.
Calibrate TranSend/SendIt Perform the following instructions to calibrate an TranSend/SendIt pair.
1. Connect the receiver unit to the computer through the serial port. See the TranSend/SendIt technical manual for more information on RS-232 serial connections.
2. Power on the transmitter and receiver units. 3. Navigate to and launch the installed MSI mV Scale Calibration program on the computer. 4. Select Calibration from the setup type drop-down menu.

5. Select

.

May 29, 2025

Figure 1. Calibration

PN 229258 Rev A

MSI ScaleCore mV/V Crane Scale Calibration
6. Select the COM port the TranSend/SendIt receiver is connected to from the drop-down menu. NOTE: All other terminal applications must be closed. If no Com Port selection is available, ensure the serial cables are properly connected, the drivers are installed properly and that the TranSend/SendIt pair is powered up. Once a connection is made, restart the calibration program.

Figure 2. COM Port Selection

7. Select

.

8. Select the Advanced Mode checkbox. 9. If necessary, change upper and lower percents detected by the TranSend/SendIt system.

10. Select

.

Figure 3. Advanced Mode

2

Visit our website www.RiceLake.com

11. Enter the total scale capacity.

MSI ScaleCore mV/V Crane Scale Calibration

12. Select

.

13. Remove all weight from the scale.

Figure 4. Scale Capacity

Figure 5. Zero Calibration

14. When scale is stable, select

.

15. Load test weights onto the scale. 16. Enter the value of the test weights in the weight dialogue box.

Figure 6. Span Calibration

© Rice Lake Weighing Systems All Rights Reserved

3

17. Select

. Calibration Complete displays.

Figure 7. Calibration Complete

18. Select

to return to the calibration start display.

19. Calibrate the indicator, see the indicator manual for instructions.

© Rice Lake Weighing Systems Content subject to change without notice. 230 W. Coleman St. · Rice Lake, WI 54868 · USA USA: 800-472-6703 · International: +1-715-234-9171

May 29, 2025

www.ricelake.com

PN 229258 Rev A

Documents / Resources

RICE LAKE MSI-7001 Tran Send SendIt [pdf] Instruction Manual
MSI-7001, PN 229258, MSI-7001 Tran Send SendIt, MSI-7001, Tran Send SendIt, Send SendIt, SendIt

References

Leave a comment

Your email address will not be published. Required fields are marked *