Instruction Manual for RICE LAKE models including: MSI-7001, PN 229258, MSI-7001 Tran Send SendIt, MSI-7001, Tran Send SendIt, Send SendIt, SendIt
1. Connect the receiver unit to the computer through the serial port. See the TranSend/SendIt technical manual for more information on RS-232 serial connections ...
1 день назад — 1. Connect the receiver unit to the computer through the serial port. See the TranSend/SendIt technical manual for more information on RS-232 serial connections ...
If the TranSend/SendIt pair is not calibrated, the indicator may have unexpected range limits and distorted weight values. Manuals are available from Rice Lake ...
File Info : application/pdf, 4 Pages, 358.27KB
DocumentDocumentTranSend/SendIt Wireless mV Scale Calibration Wireless TranSend/SendIt transmitter and receiver pairs must be calibrated to the connected load cells before an indicator can be calibrated to the load cells. If the TranSend/SendIt pair is not calibrated, the indicator may have unexpected range limits and distorted weight values. i Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals Warranty information is available at www.ricelake.com/warranties System Requirements Specifications subject to change without notice. Windows® Operating System Display: 800 x 600 or greater Built-in serial port or USB to serial port adapter FTDI chip set required (Tripp-Lite USB/Serial Adapter (PN 153603) recommended - install driver) JAVA JRE 1.7 or newer To download and install JAVA JRE: https://java.com/en/download/manual.jsp Install Software To install MSI mV Scale Calibration program: 1. Open the Rice Lake website and navigate to the MSI-7001 TranSend product page. 2. Download the MSI mV Scale Calibration software to the computer. 3. Extract the MSI mV Scale Calibration folder. 4. Open the folder and double click on ScaleCoreConnect_XX-XX.exe. Calibrate TranSend/SendIt Perform the following instructions to calibrate an TranSend/SendIt pair. 1. Connect the receiver unit to the computer through the serial port. See the TranSend/SendIt technical manual for more information on RS-232 serial connections. 2. Power on the transmitter and receiver units. 3. Navigate to and launch the installed MSI mV Scale Calibration program on the computer. 4. Select Calibration from the setup type drop-down menu. 5. Select . May 29, 2025 Figure 1. Calibration PN 229258 Rev A MSI ScaleCore mV/V Crane Scale Calibration 6. Select the COM port the TranSend/SendIt receiver is connected to from the drop-down menu. NOTE: All other terminal applications must be closed. If no Com Port selection is available, ensure the serial cables are properly connected, the drivers are installed properly and that the TranSend/SendIt pair is powered up. Once a connection is made, restart the calibration program. Figure 2. COM Port Selection 7. Select . 8. Select the Advanced Mode checkbox. 9. If necessary, change upper and lower percents detected by the TranSend/SendIt system. 10. Select . Figure 3. Advanced Mode 2 Visit our website www.RiceLake.com 11. Enter the total scale capacity. MSI ScaleCore mV/V Crane Scale Calibration 12. Select . 13. Remove all weight from the scale. Figure 4. Scale Capacity Figure 5. Zero Calibration 14. When scale is stable, select . 15. Load test weights onto the scale. 16. Enter the value of the test weights in the weight dialogue box. Figure 6. Span Calibration © Rice Lake Weighing Systems All Rights Reserved 3 17. Select . Calibration Complete displays. Figure 7. Calibration Complete 18. Select to return to the calibration start display. 19. Calibrate the indicator, see the indicator manual for instructions. © Rice Lake Weighing Systems Content subject to change without notice. 230 W. Coleman St. · Rice Lake, WI 54868 · USA USA: 800-472-6703 · International: +1-715-234-9171 May 29, 2025 www.ricelake.com PN 229258 Rev A