RR-ES-V01 Wireless EMG Sensor
“
Specifications
- Component: MCU, Sensing unit, Feedback
unit - Memory: 128MB built-in memory
- Bluetooth: Low-power Bluetooth (version
5.0) - Power: Lithium-battery, charged via USB
Type-C - Electrode extension cable: Collects
electromyographic signals - Fixing device: Can be fixed to different parts
of the body
Product Usage Instructions
1. Bluetooth
Pairing and Connecting Bluetooth for Wireless EMG
Sensor
- Status Indicator Light:
- Pairing Mode: Press and hold the button for
about 7 seconds when the device is off. Release when it vibrates
for the second time to enter pairing mode. - Standby Mode: In off state, press and hold the
button for about three seconds. Release when it vibrates once to
turn on the device. - Streaming Mode: In standby mode, continuous
vibration for 1 second indicates connection to Bluetooth device.
Blue light flashes every second in streaming mode.
FAQ
Q: How do I charge the Wireless EMG sensor?
A: The sensor is powered by a lithium battery which can be
charged using a USB Type-C interface and charging cable
provided.
Q: Can I use the Electrode extension cable and bottom electrode
simultaneously?
A: No, after connecting the extension cable, the system will
collect signals from the extension cord electrode while disabling
the bottom electrode.
Q: How do I fix the Wireless EMG sensor to my body?
A: You can fix the sensor to different parts of the body using
double-sided tape/straps. Ensure electrodes are placed close to the
skin for accurate readings.
“`
Wireless EMG sensor
user manual
Thank you for purchasing Wireless EMG ensor.Model: RR-ES-V01 The product’s manufacture date can be found on the label, and the shelf life is 3 years. Before using Wireless EMG sensor, please carefully read the user manual for proper use. Please keep this user manual properly after reading, so that it can be consulted at any time when needed. Revision date of manual: 2025.06.02, Revision A Wireless EMG sensor firmware:EMGSENSOR-FIRMWARE-1-V1 If you have any questions regarding the product, please contact: After-sales service unit: Vincent Medical (Dongguan) Technology Co., Ltd. Address: Rm101 & 201, Block 10, 1 Taoyuan Road, Songshan Lake Zone, Dongguan City, Guangdong Province, P.R.C. Postal code: 523808 Telephone: 0769-21667506
1. Introduction
Wireless EMG sensor is a device that controls equipment by detecting muscle electrical signals. In terms of control, different voltage signals emitted by muscle contraction are used to control the equipment. Through secondary development, EMG signal control can be achieved, such as toys, home appliances, etc.When purchasing, a testing app will be provided.
Component
Describe
MCU Sensing unit Feedback unit
Built in microcontroller, driving the main functions of the Wireless EMG sens or instrument: data acquisition, processing, storage, and transmission, analyzing t he collected data, and controlling the output of the feedback unit according t o the set feedback scheme.
Built in electromyographic, electrodes/signal, amplifiers/noise, filters and IMU, capable of collecting biological electromyographic, activity data and dynamic
motion data during muscle rest or contraction.
Cooperate with the pre-set feedback, plan to provide feedback on the activity of body muscles and limb movements to the user
Memory
The built-in 128MB, memory can store collected electromyography,exercise data offline.
Bluetooth
Using low-power-
Bluetooth
(version 5.0), to wirelessly transmit data to third-
party developed, adaptive devices via a private data protocol, or to receive
modified settings of electromyography sensors from the adaptive devices.
Power
Powered by lithium-batterie. Charging lithium-batterie using USB, Type-C, interface and charging cable.
Electrode extension cable
After connecting the Electrode extension cable, the system will collect electrom yographic signals from the extension cord electrode, while disabling the
electrode at the bottom of the control box. Both cannot be used simultaneously.
Fixing device
Wireless EMG sensor can be fixed to different, parts of the body with d ouble-
1 / 11
Component
Describe
sided tape/straps, and the electrodes, at the bottom of the electromyography, s ensor can be placed close to the surface of the user’s skin.
2. Bluetooth
The wireless EMG sensor, connects to paired devices via low-energy Bluetooth to enable da ta transmission. The wireless EMG sensor, needs to enter pairing mode before connecting to a new devic e, in order to pair the low-energy Bluetooth with the device. In pairing mode, the wireless EMG sensor’s low-energy Bluetooth continuously broadcasts signal s, Waiting for new paired devices, to detect the signals and then initiate pairing and subscription. he s ame group of Bluetooth devices only needs to be paired once .The optimal number of de vice pairings is up to 4, which ensures the most stable signal transmission. The more pairi ngs there are, the more delayed the data transmission becomes
2 / 11
Methods for Pairing and Connecting Bluetooth for Wireless EMG Sensor
Status
Status indicator light
Pairing mode
When the device is turned off, press and hold the button for about 7 seconds. Release the button when the device vibrates for the second time. The device will enter pairing mode, during which the blue and orange lights will flash alternately.
After successfully pairing, the device will automatically switch to standby mode. If the device fails to complete pairing, the pairing mode will automatically end and shut down after 2 minutes.
Standby mode
In the off state of the device, press and hold the button for about three seconds. When the device vibrates once, release the button to turn on the device.
In standby mode without a Bluetooth connection, the blue indicator light will flash once every 3 seconds.
Streaming mode
In standby mode, when the equipped device connects to a Bluetooth device, it will vibrate continuously for 1 second.
The blue indicator light flashes once every second, indicating that the device is in streaming mode after being connected via Bluetooth.
Indicator light is flashing.
Indicator light stays continuously lit.
3. Installation and Wearing
When using the wireless EMG sensor, the two electrodes need to make direct contact with the skin surface, and be secured above the muscle to be tested. We usually need to u
se doublesided tape to attach the sensor to the skin surface, or use straps to secure the sensor to the user’s limb.
3 / 11
Figure 1 LFix the Sensor Using Double-sided Tape RUse Strap to Secure the Sensor
Use Scenario 1 When the space around the target muscle is limited and it is not possible to directly attach the wireless EMG sensor to the skin, you can use an electrode extension cable to secure the wireless sensor at a closer position. To minimize the impact on hand gripping activities, thinner and smaller electrodes can be applied to the palm. Use Scenario 2 You can also mount the wireless EMG sensor on the dorsum of the shoe and secure it with the shoelaces. Subsequently, an electrode extension cable can be used to extend the electrode to the skin surface over the tibia in the anterior lower leg, allowing for the measurement of the tibialis anterior muscle that controls dorsiflexion of the foot . Precautions When the electrode extension line is connected to the sensor, the electrodes at the bottom of the sensor will become inactive, so these two types of electrodes cannot be used simultaneously. Only the electrodes on the connected extension line can collect surface electromyography signals.
Figure 2
Figure 3
By using electrode extension lines, the sensor can be placed on the limb, thereby reducing interference and restriction on limb movement.
You can secure the sensor to the shoe using shoelaces, and then use electrode extension wires to measure the muscle electrical activity in other parts of the body.
Note: The electrode extension wires require two additional disposable electrode pads, which can be secured to the electrode port at the end of the extension wire using a button form. Disposable electrode pads are consumables and can be purchased separately from the distributor if needed.
4. Charging
The wireless EMG sensor is equipped with a USB Type-C charging port. Please use USB
4 / 11
Type-C cable to charge the device. This product only includes the wireless EMG sensor. Please prepare a suitable USB Type-C charger and charging cable on your own. The lithium battery specification for the wireless EMG sensor is 300mAh and 3.7V. The working duration with a new battery under different operating conditions is detailed in Table 8.21a.
Type-C
5. Battery
The performance of the device’s indicator lights under different battery level states
Status
Status indicator light
Shutdown
Press and hold the button for about 3 seconds, and the device will vibrate briefly while the blue indicator light flashes.
The battery is After holding the button for about 3 seconds, the device
depleted
showed no response and did not vibrate
The battery is When the device is powered on, the indicator light flashes
too low
orange once every second.
During charging
During battery charging, the orange indicator light stays continuously lit.
The battery charging is complete
When the battery charging is complete, the purple indicator light stays continuously lit.
The device is in standby mode
After powering on, the blue indicator light flashes every 3 seconds, indicating that the device is in Bluetooth pairing mode. If the device remains in standby mode without a Bluetooth connection for 2 minutes, it will automatically power off.
5 / 11
Indicator light is flashing.
Indicator light stays continuously lit.
6. Data collection mode
After successfully connecting the Bluetooth device, you can press “Next” to enter the EMG. measurement page. On this page, the mobile software will continuously monitor and display the changes
in the target EMG, signal in real time, as well as display the current root mean square (RMS), value of the EMG, signal in real time. You can press a button to toggle, the display of the surface electromyography (sEMG). signals. In th e signal chart, the sEMG will be represented by a blue line. You can press a button to toggle, the display of the root mean square (RMS), of the surface electromyography (sEMG), signals. In the signal chart, the RMS, will be represented by a red line. T he root mean square (RMS), of surface EMG signals is defined as calculating the average of the s quared values. of a set of surface EMG signals, and then taking the square root of that average. The root mean square (RMS), results in a smooth linear envelope used to estimate the average ampl itude of signals composed of positive and negative oscillations. As shown in Figure 4, the red ro ot mean square curve illustrates the variation of the measured, surface electromyography signal’s effective value. The following is the calculation formula for the root mean square (RMS) of electromyography signals.
Figure 4 Data display page
7. Data Description
The following are the data contents and applicable settings that can be collected by wireless EMG sensor.
Date
Sensor
Description/Specification
sEMG
EMG electrodes, signal amplifiers, noise filters
sampling frequency1000 Hz Sampling bits (resolution)16 Voltage amplitude of microprocessor3V Computable square meanRoot Mean SquareRMS
6 / 11
Rotational angular velocity gyroscope (three-axis)*XYZ
Mobile acceleration (three- accelerometer axis) * XYZ
Magnetic induction intensity (three-axis) * XYZ
Time Stamp
magnetometer Real Time Clock
=
1
2
Optional sampling frequency1-255 Hz Optional total-range±250dps,±500dps,±1000dps,±2000dps
Optional sampling frequency1-255 Hz Optional total-range2g,±4g,±8g,±16g Optional elimination of gravitational acceleration effects
Optional sampling frequency1-70 Hz
Using Coordinated Universal Time (UTC), adjustable/calibrated.
Does the feedback data
/
exceed the threshold
Exceeding/falling below the threshold. Determine whether the monitored data stream value exceeds the threshold based on the pre-set feedback plan.
sEMG monitoring record /
According to the established monitoring plan, record the user’s electromyographic signal activity
* Based on the IMU, the three-dimensional rotation vector can be calculated.
8. Specification
Specification Model
Size
Sampling rate of surface electromyography signals Variable value of sampling rate for rotational angular velocity Variable range of rotational angular velocity across the entire range Variable value of mobile acceleration sampling rate Variable range of mobile acceleration over the full range Variable sampling rate of magnetic induction intensity Battery Endurance Input
Value RR_ES_V01 Length52.5 mm±0.5mm Width38.5 mm±0.5mm Height13.8 mm±0.5mm 1000 Hz 1-255 Hz ±250dps, ±500dps, ±1000dps, ±2000dps 1-255 Hz ±2g, ±4g, ±8g, ±16g 1-70 Hz 300mAh, 3.7V 4 hours USB Type-C, 5V dc, 500mA
9. Other function/setting
The following list provides all the functions that users can operate.
Some functions can be operated using the device’s buttons, while access to other functions requires the
use of Bluetooth command sets, in conjunction with third-party developed applications and compatible
devicesIf software developers need a detailed Bluetooth instruction set, please request the software development manual from the
vendor for reference
Funtion
Operate
Bluetooth Command Set
Power on
7 / 11
Power off Enter pairing mode Adjust feedback threshold Start/Stop Online Streaming Mode
Initiate data acquisition of sEMG
Initiate data collection for rotational angular velocity
Initiate data collection for mobile acceleration
Initiate data collection of magnetic induction intensity Adjust the sampling frequency of the rotational angular velocity Adjust the sampling frequency of the moving acceleration Adjust the sampling frequency of magnetic induction intensity Adjust the total-range of rotational angular velocity Adjust the total-range of mobile acceleration Adjust the sampling frequency of the 3D rotation vector Adjust real-time clock Obtain real-time clock values Obtain the percentage value of battery power Revise feedback plan Set a timed feedback prompt schedule Start/Stop LED
Raw data Calibration data Root mean square Raw data Calibration data Raw data Calibration data Eliminate the influence of gravity acceleration Raw data Calibration data
Fault Analysis and Troubleshooting
Issue
Reason
Check Device
Solution
Check 1:
Press and hold the button for 3
The device is not Not powered Please check the device’s indicator lights. If the system seconds until the device vibrates
responding
on
is powered on, the signal blue light on the host should briefly, and then the device will
be lit or blinking.
turn on.
8 / 11
He device is completely out of battery
Check 2: Please press and hold the button for more than 3 seconds, then release it. If the device is not powered on, it will vibrate briefly and start up. If the device is already powered on, the indicator light will flash orange 3 times and then turn off. If there is no response, the device may be out of battery.
Insert a USB Type-C power source to charge the device.
Low battery
Check 3: You can check if the orange indicator light is on and flashing, which indicates that the battery is low.
Insert a USB Type-C power source to charge the device.
The device to be connected has Bluetooth turned off.
Check 5: Check the Bluetooth settings of the paired device.
Please ensure that the Bluetooth on the compatible device is turned on and functioning properly.
Cannot connect to the Bluetooth device
Unpaired Bluetooth device.
Check 5: Check the Bluetooth settings of the paired device.
Hold down the button for 7 seconds while the device is powered off to enter pairing mode. The Bluetooth settings of the compatible device should be able to search for and pair with the wireless EMG sensor. (See section pairing Bluetooth devices.)
The adapter has selected the wrong device number.
Check 6: Check the device number on the surface of the wireless EMG sensor and compare it with the Bluetooth settings to ensure that the correct device number is selected on the paired device client.
Please make sure to pair and connect with the correct Bluetooth device and equipment number.
Other Issue
Please consult technical support.
10.Maintenance
Clean: If the surface of the product is dirty, it can be wiped with 75% alcohol system maintenance: If there is any abnormal data in the product, please contact Vincent Medical (Dongguan) Technology Co., Ltd. for after-sales service. Please do not disassemble it without authorization.
11.Operational environment
Environmental conditions for use
General environment Equipment used in ventilated indoor environments.
Operating environment
5 °C 40 °C, humidity80%
precaution
Allow for general vibrations and impacts during use to facilitate the movement of the device.
12.Transportation and storage
Transport: Handle with care, Keep dry ,Not cool. Storage: Stored in -20 °C 55°Ctemperatureand humidity80%, Indoor environment with no corrosive gases and good ventilation.
13.Unpacking and checking
Unpacking : Please check if the accessories are complete. If you encounter packaging damage or water immersion marks, please contact the supplier.
9 / 11
Packing list Accessory Wireless EMG sensor Electrode extension cable EMG strap Manual
Qty. 1 PC 2 PCS 4 PCS 1 PC
Warning and precautions: Do not disassemble, impact or squeeze, do not put into fire or water, and do not place in high temperature environments
14. Symbol Definition
Symbol
Meaning
Symbol
Meaning
Waste Electrical and Electronic Equipment
keep dry
CE Marking
Recyclable
Direct Current Temperature extremes Limit of the atmosphere Humidity limit
Handle with care This side up
Keep out of the direct sun
10 / 11
Wireless EMG sensor function Table 11 / 11
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Any Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help. FCC Radiation Exposure statement The device has been evaluatec to meel general RF exposure requirement. The device can be used in portable exposure condition without restriction.