UniLog Pro Temperature Data Logger

Operation Manual
This brief manual is primarily meant for quick reference to wiring connections and parameter searching. For more details on operation and application; please log on to www.ppiindia.net

 

OPERATOR PARAMETERS
Parameters Settings (Default Value)
‘Start’ Command for Batch Recording

(Available if Batch Recording is selected)

No Yes
BATCH START>> NO  
‘Stop’ Command for Batch Recording (Available if Batch Recording is selected)  No Yes
BATCH STOP>> NO  

SUPERVISORY CONFIGURATION

PPI UniLog Pro Temperature Data Logger

Note: AII the other Parameters are under SUPERVISORY CONFIGURATION

ALARM SETTING
Parameters Settings (Default Value)
Channel Name for Alarm Settings

SELECT CHANNEL>>

Channel-1

User defined or default names for channel-1 to channel-8 / 16 (Default : NA)
Select Alarm

SELECT ALARM>> AL1

AL1, AL2, AL3, AL4

(The actual available options depends on the numbers   of Alarms set per channel on

Alarm configuration page)

Alarm Type

AL1 TYPE>> None

None Process Low Precess High (Default : None)
Alarm Setpoint

AL1 SETPOINT>> 0

Min. to Max. of selected input type range (Default : 0)
Alarm Hysteresis

AL1 HYSTERESIS>> 2

1 to 3000 or

0.1 to 3000.0

(Default : 2 or 2.0)

Alarm Inhibit

AL1 INHIBIT>> Yes

No Yes

(Default : No)

DEVICE CONFIGURATION
Parameters Settings (Default Value)
Channel Update Time in Auto Scan Mode

SCAN RATE>>

3

 

1 Sec. to 99 Sec. (Default : 3 Sec.)

Device Identification Number

RECORDER ID>>

2

 

1 to127

(Default : 1)

Select Total Numbers of Channels

TOTAL CHANNELS>>

16

8

16

(Default : 16)

Erase All Stored Records

DELETE RECORDS>>

No

 

No Yes

(Default : No)

 
CHANNEL CONFIGURATION
Parameters Settings (Default Value)
Select Channel Name

SELECT CHANNEL>> Channel-1

User defined or default names for channel-1 to channel – 8 / 16

(Default : NA)

Skip Channel for Display

SKIP>> No

No Yes

(Default : Yes)

Signal Input Type

INPUT TYPE>> Type K (Cr-Al)

Refer Table 2 (Default : Type K (Cr-Al)
Display Resolution for Measured PV

RESOLUTION>> 0.1 Unit

1 Unit

0.1 Unit

0.01 Unit *

0.001 Unit * (Default : 0.1 Unit)

(* for 4-20mA)

Display Units for Measured PV

UNITS>> °C

 

Refer Table 1 (Default : °C)

Range Low (for 4-20mA) RANGE LOW>> 0 -19999 to 30000 Counts with Selected Resolution (Default : 0.0)
Range High

(for 4-20mA) RANGE HIGH>> 1000

-19999 to 30000 Counts with Selected Resolution (Default : 100.0)
Apply Lower Clip on Displayed PV

(for 4-20mA) LOW CLIPPING>> Disable

 

Disable Enable

(Default : Disable)

Preset Lower Clip Level

(for 4-20mA) LOW CLIP VAL>> 0.0

 

-19999 to 30000

(Default : 0)

Apply Upper Clip on Displayed PV

(for 4-20mA) HIGH CLIPPING>> Disable

 

Disable Enable

(Default : Disable)

Preset Upper Clip Level

(for 4-20mA) HIGH CLIP VAL>> 100.0

 

-19999 to 30000

(Default : 100.0)

Zero Offset

ZERO OFFSET>> 0

-1999 / 3000 or

-1999.9 / 3000.0

(Default : 0)

ALARM CONFIGURATION
Parameters Settings (Default Value)
Alarms Per Channel

ALARMS / CHAN >> 4

1 to 4

(Default : 4)

Relay-1 Logic

RELAY-1 LOGIC >> Normal

Normal   Reverse (Default : Normal)
Relay-2 Logic

RELAY-2 LOGIC >> Normal

Normal   Reverse (Default : Normal)
 
RECORDER CONFIGURATION
Parameters Settings (Default Value)
Normal Recording Interval

NORMAL INTERVAL>> 0:00:30

0:00:00 (H:MM:SS)

to 2:30:00 (H:MM:SS)

(Default : 0:00:30)

Zoom Recording Interval

ZOOM INTERVAL>> 0:00:01

0:00:00 (H:MM:SS)

to 2:30:00 (H:MM:SS)

(Default : 0:00:01)

Record Generation on Alarm Status Toggle

ALARM TOGGLE REC>> Disable

 

Disable Enable

(Default : Enable)

Select Recording Mode

RECORDING MODE>> Continuous

Continuous Batch

(Default : Continuous)

Time Interval for Batch Recording (for Batch mode only)

BATCH TIME>> 1.00

0:01 (HH:MM)

to 250:00 (HHH:MM)

(Default : 1:00)

 
RTC SETTING
Parameters Settings
Set Clock Time (HH:MM)

TIME (HH:MM)>>

15:53

 

0.0 to 23:59

Set Calendar Date

DATE>> 23

 

1 to 31

Set Calendar Month

MONTH>> 11

 

1 to 12

Set Calendar Year

YEAR>> 2011

 

2000 to 2099

 
UTILITIES
Parameters Settings
Master Lock Enable Disable

LOCK>> No UNLOCK>> No

 

 

No Yes

UIM Default

UIM DEFAULT>> No

No Yes
CIM Default

CIM DEFAULT>> No

 No Yes
Make CIM & UIM Compatible

CPY CIM TO UIM>> No CPY UIM TO CIM>> No

No Yes
TABLE- 1
Option Description
°C Degree Centigrade
°F Degree Fahrenheit
(none) No Unit (Blank)
°K Degree Kelvin
EU Engineering Units
% Percentage
Pa Pascals
Mpa Mpascals
kPa Kpascals
bar Bar
mbar Milli bar
psi PSI
kg/sq.cm kg/cm2
mmH2O mm water gauge
inH2O Inches water gauge
mmHg mm mercury
Torr Torr
litre/hr Litres per hour
litre/min Litres per minute
%RH % Relative Humidity
%O2 % Oxygen
%CO2 % Carbon di-oxide
%CP % Carbon Potential
V Volts
A Amps
mA Milli Amps
mV Milli Volts
ohm Ohms
ppm Parts per million
rpm Revolutions per minute
mSec Milli seconds
Sec Seconds
min Minutes
hrs Hours
PH PH
%PH %PH
miles/hr Miles per hour
mg Milli grams
g Grams
kg Kilo grams
TABLE- 2
Option Range (Min. to Max.) Resolution
PPI UniLog Pro Temperature Data Logger - icon  

0 to +960°C / +32 to +1760°F

 

 

 

 

 

 

 

 

 

Fixed 1°C / 1°F

PPI UniLog Pro Temperature Data Logger - icon 1  
PPI UniLog Pro Temperature Data Logger - icon 3  

PPI UniLog Pro Temperature Data Logger - icon 5

 

0 to +1771°C / +32.0 to +3219°F

PPI UniLog Pro Temperature Data Logger - icon 6  

0 to +1768°C / +32 to +3214°F

PPI UniLog Pro Temperature Data Logger - icon 8  
PPI UniLog Pro Temperature Data Logger - icon 9  
PPI UniLog Pro Temperature Data Logger - icon 10 Reserved for customer specific Thermocouple type not listed above.
PPI UniLog Pro Temperature Data Logger - icon 11 -199 to +600°C / -328 to +1112°F

199.9 to             or          to 1112.0°F

600.0°C / -328.0

User settable 1°C / 1°F

or   0.1°C / 0.1°F

 

 

 

PPI UniLog Pro Temperature Data Logger - icon14

 

 

 

 

19999 to +30000 units User settable 1 / 0.1 / 0.01/

0.001 unts

ID SETTING FOR MORE THAN 1 CIM
Applicable only for UNILOG PRO PLUS

PPI UniLog Pro Temperature Data Logger - fig

FRONT PANEL LAYOUT

PPI UniLog Pro Temperature Data Logger - fig 1

Symbol

Key Function
PPI UniLog Pro Temperature Data Logger - icon 15 PAGE Press to enter or exit set-up mode.
PPI UniLog Pro Temperature Data Logger - icon 17  

DOWN

Press to decrease the parameter value. Pressing once decreases the value by one count; holding pressed speeds up the change.
PPI UniLog Pro Temperature Data Logger - icon 18  

UP

Press to increase the parameter value. Pressing once increases the value by one count; holding pressed speeds up the change.
PPI UniLog Pro Temperature Data Logger - icon 20 ENTER Press to store the set parameter value and to scroll to the next parameter on the PAGE.

ELECTRICAL CONNECTIONS
USER INTERFACE MODULE (UIM) 

PPI UniLog Pro Temperature Data Logger - fig 20

COMMUNICATION PORT FOR INTERFACING WITH CIM(S)
Applicable only for UNILOG PLUS
PPI UniLog Pro Temperature Data Logger - fig 10

JUMPER SETTINGS
INPUT-CHANNEL INTERFACE MODULE (CIM)
PPI UniLog Pro Temperature Data Logger - fig 14PPI UniLog Pro Temperature Data Logger - fig 15

JUMPER SETTINGS
INPUT-CHANNEL INTERFACE MODULE (CIM)
PPI UniLog Pro Temperature Data Logger - fig 25

ELECTRICAL CONNECTIONS
CHANNEL INTERFACE MODULE (CIM) 
PPI UniLog Pro Temperature Data Logger - fig 18

 

 

 

 

Documents / Resources