PCE Instruments PCE-CT 80 Material Thickness Gauge

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-product

Product Information

The product is a multi-functional instrument that offers various measurement capabilities. It comes with a user manual that provides detailed instructions on how to use the product effectively.

Specifications

  • Technical specifications: The technical specifications of the product are listed in section 2.1 of the user manual.
  • Delivery contents: The items included in the package when purchasing the product are listed in section 2.2 of the user manual.
  • Optional accessories: Additional accessories that can enhance the functionality of the product are listed in section 2.3 of the user manual.

Safety notes

Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.

  • The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
  • The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
  • Do not expose the device to shocks or strong vibrations.
  • The case should only be opened by qualified PCE Instruments personnel.
  • Never use the instrument when your hands are wet.
  • You must not make any technical changes to the device.
  • The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
  • The device must only be used with accessories from PCE Instruments or equivalent.
  • Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
  • Do not use the instrument in explosive atmospheres.
  • The measurement range as stated in the specifications must not be exceeded under any circumstances.
  • Non-observance of the safety notes can cause damage to the device and injuries to the user.

We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of business. If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.

Specifications

Technical specifications

Specifications of the meter PCE-CT 80

Specification Description
Measurement range Depending on sensor (see sensor chart)
Accuracy Depending on sensor (see sensor chart)
Resolution 0.1 µm (<100 µm)

1 µm (>100 µm)

Measurable materials Non-magnetic layers on Fe (steel, iron)

Non-electrically conductive layers on NFe (aluminum, copper)

Min. radius of curvature convex 5 mm
Min. radius of curvature concave 25 mm
Min. measuring surface Ø 17 mm
Min. substrate thickness 0.2 mm (magnetic materials)

0.05 mm (non-magnetic materials)

Probe mode Automatic mode with material recognition (Fe + NFe) Magnetic mode (Fe)

Eddy current (NFe)

Measure mode Single measurement
Calibration Multi-point calibration (1 … 4 points per group) Zero-point calibration
Units µm / mm / mils
Data transfer Data transfer via USB 2.0
Memory One group of measurements (DIR mode) that is lost when meter is turned off

Four groups of measurements with auto save and max. 2000 readings (GEN mode)

Note: Each group of measurements has individual statistics, alarm thresholds and calibration

Statistic functions Number of readings, mean value, minimum, maximum, standard deviation
Alarm Indication when set upper and lower alarm thresholds are exceeded or fallen below
Operating time Auto Power Off function
Power supply 3 x 1.5 V AAA batteries
Display 128 x 128 LCD
Displayed information Battery level Fault detection
Operating conditions 0 °C… +50 °C

20 %… 90 % RH non-condensing

Storage conditions -10 °C… +60 °C

20 %… 90 % RH non-condensing

Dimensions 143 x 71 x 37 mm (L x W x H)
Weight incl. probe approx. 271 g

Specifications of the selectable probes

Specification Description
Type Ø [mm] Measurement range [µm] Accuracy
PCE-CT 80-FN0.5* 17 Fe: 0 … 500

NFe: 0 … 500

±(1 % of rdg. + 1 µm)
PCE-CT 80-FN1.5* 17 Fe: 0 … 1500

NFe: 0 … 1500

±(2 % of rdg. + 1 µm)
PCE-CT 80-FN2* 17 Fe: 0 … 2000

NFe: 0 … 2000

±(2 % of rdg. + 1 µm)
PCE-CT 80-FN2.5* 17 Fe: 0 … 2500

NFe: 0 … 2500

±(2 % of rdg. + 1 µm)
PCE-CT 80-FN3* 17 Fe: 0 … 3000

NFe: 0 … 3000

±(2 % of rdg. + 1 µm)
PCE-CT 80-F5N3* 17 Fe: 0 … 5000

NFe: 0 … 3000

±(2 % of rdg. + 1 µm)
  • Fe and NFe: for ferromagnetic and non-ferromagnetic substrates

Delivery contents

  • 1 x coating thickness gauge PCE-CT 80
  • 1 x probe FN1.5
  • 3 x AAA batteries
  • 1 x carrying case
  • 1 x set of standard foils
  • 1 x calibration plates (Fe and NFe) 1 x ISO-calibration certificate
  • 1 x user manual

Optional accessories

  • Probe PCE-CT 80-FN0.5
  • Probe PCE-CT 80-FN2
  • Probe PCE-CT 80-FN2.5
  • Probe PCE-CT 80-FN3
  • Probe PCE-CT 80-F5N3
  • PC software with data cable

System description

Device

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-2

  1. LC display
  2. LED red
  3. LED green
  4. Membrane keypad
  5. Probe

Interfaces

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-3

  1. Probe socket
  2. Micro USB 2.0 connector

Display

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-4

  1. Reading
  2. No. of readings
  3. Memory group
  4. Limit value display
  5. Probe mode
  6. Material setting
  7. Unit
  8. Statistics (Mean, minimum, maximum value)
  9. Battery level
  10. USB communication status
  11. Auto Power Off

Function keys

 

Key

 

Designation

Functions
Measure mode Menu mode
PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-5
 

On/off

 

On/off (2 s)

 

On/off (2 s)

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-6  

Calibration

 

Start and stop calibration

 

/

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-7  

Zero point calibration

Start and stop zero point calibration Back to measuring mode directly
Press and hold during device start-up to reset meter to factory settings
PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8  

Enter

 

Open menu

 

Save, select, confirm

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9  

Back

 

Backlight on/off

 

Cancel, back, close

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10  

Up

 

Delete readings

 

Up, increase value

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11  

Down

 

Refresh screen

 

Down, decrease value

Getting started

Power supply

For power supply, three AAA Alkaline batteries are required. Before replacing the batteries, turn off the meter. The battery compartment is located at the rear side of the meter and fixed with two screws. Loosen the screws, remove the cover, insert the batteries as marked and close the battery compartment by tightening the screws.

Preparation

To turn on the meter, press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-5until the display reacts. After charging, the meter will show the measuring screen. To turn off the meter, pressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-5. Connect the included probe FN1.5 to the probe socket of the meter. Make sure that the plug is in the right position. The arrow must be visible from the front side of the meter.

Options

To change the standard settings, navigate to the submenu “Options“. Follow these steps:

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-12

  1. Press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8to enter the “Root“ menu.
  2. Use the PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11key to go to “Options“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. Navigate to the desired settings with thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  4. Select an element withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
Setting Element Description
Measure mode Single Individual measurements
Group mode DIR, GEN

General 1,

General 2,

General 3,

General 4

DIR (Direct):

Data are lost when the meter is turned off. When the memory is full, the last readings are deleted.

GEN (General 1-4):

The data are saved to one of the 4 groups and are not lost when turning off the meter. When the memory is full,

measurements are still possible but not saved and ”fl“ is displayed.

Probe mode Auto, FE, N- FE Auto:

The substrate material is recognised automatically. FE:

You can measure on magnetic materials only. N-FE:

You can measure on non-magnetic materials only.

When the meter recognises a magnetic material, it displays an “F“. If the material is non-magnetic, it displays an ”N“.

Unit settings µm, mils, mm Selection of measuring unit
Speed Normal, fast The higher the speed, the lower the accuracy will be.
Language German, English,

Russian, Chinese

Selection of one out of four languages
Auto Poweroff Deactivate Activate After inactivity for a certain period, the device turns off automatically. Before powering off, the meter will beep several times. By pressing any key, you can cancel this

process.

Operation

Before making a measurement or calibration, you should be aware of the factors that can influence the measuring accuracy. These influences are listed in the following chart.

 

Influencing factor

Measuring principle  

Recommendation

Magnetic Eddy current
Magnetic properties x Calibrate meter
Electrical properties x Calibrate meter
Material radius x x Chapter 2.1 Calibrate meter
Substrate thickness x x Chapter 2.1 Calibrate meter
Size of measuring surface x x Chapter 2.1 Calibrate meter
Surface roughness x x
Position and shape x x
Sample deformation x x Do not measure objects which are too soft or too thin
Glue x x Clean probe and measuring surface
Strong magnetic fields x Avoid strong magnetic fields
Temperature and air humidity x x Re-calibrate under the same environmental conditions
Measuring processes x x Chapter
Low battery x x Replace battery
Probe wear x x Contact PCE Instruments

Calibration

The meter should be calibrated (zero point / multi-point calibration) each time a new sample is measured. The material properties and the radius of curvature of the uncoated calibration surface should be as similar as possible to those of the sample. Make sure to choose similar calibration/measuring points, particularly when measuring corners and edges of small parts. The radius of curvature, the minimum material thickness and minimum measurement range stated in chapter 2.1 should be observed. The meter offers zero-point and multi-point calibration. Each memory group has its individual zero-point and multi-point calibration. The values previously measured are not affected by a new calibration.

Zero-point calibration

A zero-point calibration must be made on an uncoated calibration surface. If it is not possible to calibrate the meter on the sample, please use the calibration plates that come with the meter. A zero-point calibration is recommended where the measuring accuracy is ±(2 % + 1 μm). To carry out a zero-point calibration, follow these steps:

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-13

  1. Press and hold PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-7until you hear three beep sounds and ”ZERO“ is displayed.
  2. Place the probe on the uncoated calibration surface and lift it as soon as you hear a beep sound.
  3. Repeat that step several times in order to get a mean value.
  4. PressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-7 to leave zero-point calibration.

Multi-point calibration

The meter’s multi-point calibration feature allows up to 4 calibration points to be considered. When making a single-point calibration, the calibration value should be as similar as possible to the expected measurement value. When making a multiple-point calibration, the expected measurement values should be somewhere between the calibration values. The permissible maximum measurement error is ±(2 % + 1 μm).

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-14

  1. If you have already made a calibration, you have to delete the existing calibration values first, as described in chapter 5.1.3. If not, start with the next step.
  2. In measure mode, press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-6to open the calibration dialogue which can be enabled or disabled in the calibration settings.
  3. If you have previously made a zero-point calibration, the zero point will be used as the first calibration point. If not, follow the steps described in chapter 5.1.1.
  4. Place one of the included standard foils on the uncoated calibration surface.
  5. Place the sensor on the foil and remove it when you hear a beep sound. The reading will now be displayed.
  6. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to adapt the reading to the actual foil thickness (flashing value) and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8. The calibration point will then be displayed in the dialogue box (non-flashing value). You can cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9.
  7. Repeat these steps if you wish to get all four calibration points. Use a different standard foil for each calibration point.
  8. PressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-6 to exit the multi-point calibration.

Delete calibration

To start a new calibration, you can delete the calibration points recorded during zero-point calibration and during multi-point calibration individually. Follow these steps:

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-15

  1. PressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to enter the “Root“ menu.
  2. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to navigate to “Calibration“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. Select according to the following subsections (a, b, c) what values you would like to delete.

Delete multi-point calibration values

  1. Navigate to “Point Cal“ by using thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  2. Use the PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11keys to navigate to “Clear Magnetic“ or ”Clear Magnetic“ as required and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. A dialogue box opens. Confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to delete the calibration values or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9.

Delete zero-point calibration values

  1. Use the PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11keys to navigate to “Zero Cal“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  2. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to navigate to “Clear Magnetic“ or ”Clear Magnetic“ as required and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. A dialogue box opens. Confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to delete the calibration values or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.

Delete all calibration values

  1. Use the keys PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11to navigate to “Clear All“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  2. A dialogue box opens. Confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to delete all calibration values or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.

Measurement

Procedure:

  1. First connect the desired probe as described in chapter 4.2.
  2. Make the required device settings (measure mode, group mode, probe mode, speed, etc.) as described in chapter 4.3.
  3. Make a zero-point calibration and a multi-point calibration before the measurement (chapter 5.1).
  4. Place the probe vertically on the surface to be measured. Lift the probe when you hear the beep sound.
Further functions

Delete data

Delete last reading

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-16

  • In measure mode, press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10to delete the last recorded reading.
  1. PressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to enter the “Root“ menu.
  2. Use the PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11keys to navigate to “Delete Data“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to navigate to “Current Data“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  4. A dialogue box opens. Confirm with PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8to delete the last reading or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9.

Delete current memory group

  • In measure mode, press and holdPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 until you hear another beep sound if you want to delete all values saved in a memory group.
  1. Press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8to enter the “Root“ menu.
  2. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to navigate to “Delete Data“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to navigate to “Current Group“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  4. A dialogue box opens. Confirm with PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8to delete all readings saved in the memory group or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9.

To delete other memory groups and values, change the memory group as described in chapter 4.3 and follow the steps above again.

Delete values

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-17

Press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8to enter the “Root“ menu. Navigate to “Recall“ with the PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8. In the upper left part of the display, you will now see what value of a memory group has been selected (1/4). Select the desired reading withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11. You can delete that reading by pressingPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.

Statistics

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-18

When making a measurement in measure mode, the current mean value, the maximum and the minimum values will be displayed below the current reading. To view the complete statistics, pressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 first. In the menu that is now displayed, go to “Browse Stat.“ by using thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8. You can now see the statistical values of the currently selected memory group. To view the statistics of the other memory groups, change the memory group as described in chapter 4.3 and go back to the statistics function.

The meter creates individual statistics based on the saved readings of each memory group. In the statistics, the number of readings, the mean value, the minimum and maximum values as well as the standard deviation will be displayed. When the memory is full in direct mode, the first reading will be deleted and the new reading will be saved. The statistics will be updated accordingly. When the memory is full in general mode, no new readings will be saved and the statistics will not be updated. If you delete the current reading or all readings of the group, the statistics will update.

Limit settings

Each memory group has its own upper and lower limit value. When changing the memory group, the limit values will be switched automatically. When a limit value is exceeded/fallen below, the red LED on the left will flash. Moreover, aPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-20 will be displayed when the upper limit has been exceeded or PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-21a if the lower limit has been fallen below.

To enter the upper and lower limit value, follow these steps:

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-19

  1. PressPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to enter the “Root“ menu.
  2. Navigate to “Limit“ with thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. Navigate to “Settings“ with thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  4. Use thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys to navigate to “High Limit“ or “Low Limit“ and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  5. Set the limit value withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 and confirm with PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9.

To enter the upper and lower limit value, follow these steps:

  1. Press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8to enter the “Root“ menu.
  2. Navigate to “Limit“ with thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  3. Navigate to “Clear“ with thePCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-10 PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-11 keys and confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8.
  4. A dialogue box opens. Confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 to delete the readings of the memory group or cancel the procedure by pressingPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9.

To enter or delete the limit values of other memory groups, change the group as described in chapter 4.3 and follow the steps described above again.

USB communication (only possible with optional software)

The meter can be connected to a PC for communication via the micro USB 2.0 connector. To do so, first install the software and the USB driver for the meter. Then connect the meter to your computer by using the USB cable. If the connection has been successful, you will see the 5 icon at the bottom of the display. You can now organise, format, print or copy the measured data on your PC.

System reset

To reset the meter to its factory default settings, follow these steps:

  1. Press PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-5to turn off the meter.
  2. Press and hold PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-7while turning on the meter withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-5.
  3. Release the PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-7key when the dialogue box for the system reset is displayed.
  4. To reset the meter, confirm withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-8 or cancel the procedure withPCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-9. The meter will restart automatically.

Warranty

You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms.

Disposal

For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose. In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law. For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations. If you have any questions, please contact PCE Instruments.

www.pce-instruments.com

Contact Information

Germany

United Kingdom

The Netherlands

United States of America

Hong Kong

China

www.pce-instruments.com

Scan

PCE-Instruments-PCE-CT-80-Material-Thickness-Gauge-fig-1

© PCE Instruments