Medtronic 780G System With Simplera Sync Sensor

Medtronic-780G-System-With-Simplera-Sync-Sensor-product

Product Information

Specifications

  • Product Name: MiniMedTM 780G system with Simplera SyncTM sensor
  • Software Update Method: Managed via the MiniMedTM Mobile app
  • Compatibility: Sensor compatibility supports GuardianTM 4 Sensor and Simplera SyncTM sensor
  • Auto Correction Increment: New default is 0.025U
  • Connection Range: Keep phone and pump within 3 meters (10 feet) of each other

Product Usage Instructions

Software Update Process

  • Ensure phone is fully charged and pump battery icon is on green
  • Upload existing pump data to CareLinkTM Personal account before update
  • Access the MiniMedTM Mobile app to complete the update
  • Pair pump and mobile app
  • Maintain a stable Wi-Fi connection
  • Follow in-app instructions for the update process
  • Duration: 45-90 minutes*

Post-Install Instructions

  • Disconnect from the pump temporarily
  • Check insulin delivery has resumed after reconnecting to pump
  • Verify settings are correctly transferred over
  • Re-enter SmartGuardTM feature after 5-hour warm-up period
  • Activate the Suspend Before Low feature if not previously set
  • Duration: 10-20 minutes*

User Interface & Screen Updates

  • Check for sensor update alerts on the device interface
  • New auto-correction increment is 0.025U

Frequently Asked Questions

  • Q: How long does the software update process take?
    • A: The software update process typically takes between 45 to 90 minutes, depending on internet speed.
  • Q: What should I do if there is a disruption in insulin delivery post-install?
    • A: If there is a disruption in insulin delivery post-install, disconnect from the pump temporarily, ensure insulin delivery has resumed after reconnecting, and verify that settings are correctly transferred over.
  • Q: How do I activate the SmartGuardTM feature after the software update?
    • A: After the software update, re-enter the SmartGuardTM feature when prompted to enter blood glucose levels. It is also recommended to activate the Suspend Before Low feature if not previously set.

What’s new?

MiniMed™ 780G system with Simplera Sync™ sensor

Agenda

Stay informed with these updates to ensure the best care for device users

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-1

MiniMed™ 780G system software update

New features and benefits

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-2

Important information for the software update

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-3

Software update process

Device user experience overview

Pump guides the device user via prompts and checklists

Check for updates

  • Device user can use the pump and CGM as usual
  • Device user confirms the device has the latest update
  • Device user to access the main menu in the MiniMed™ mobile app and tap the ‘Update pump’ option to start update

Download

Device user can use the pump and CGM as usual

  • In-app instructions
    Device user to follow in-app instructions
  • Regular use of the device
    Device user can use pump and CGM as usual
  • Duration*
    45-90 minutes

Install

The device user prompted to disconnect from pump

  • Disconnects from pump
    Temporary disruption in insulin delivery
  • No SG or notifications
    A temporary disruption in data flow in the MiniMed™ Mobile app and CareLink™ Connect app
  • Duration*
    10-20 minutes

Post-install

The device user was prompted to reconnect to pump

  • Reconnect to pump
    Check insulin delivery has resumed
  • Verify settings
    Ensure settings are correctly transferred over
  • 5-hour warm-up
    Re-enter the SmartGuard™ feature when prompted to enter BG
  • Activate safety
    If not previously set, recommended to activate the Suspend Before Low feature

Device user interface & screen updates

System updates

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-4

MiniMed™ 780G system

Device user interface/screen updates

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-5

  • Sensor update alerts
    • Provide time the senso update may last – such as ‘can take up to 30 minutes’ or ‘can take up to 90 minutes’
  • Bolus Wizard™ and SmartGuard™ features
    • Allow BG entry
  • Count-down timer displays during sensor warm-up
    • Simplera Sync™ sensor and Guardian™ 4 Sensors: starts at 2 hours; counts down to 0 minutes
  • Three options for changing reservoir/set
    • New Reservoir and Set
    • New Reservoir Only
    • New Set Only
  • Set change reminders range
    • 2-7 days
  • New Simplera Sync™ sensor pairing process
    • After insertion, the pump requires a “CODE” for pairing

Simplera Sync™ sensor and inserter

Components

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-6

The Simple Sync™ sensor is an all-in-one glucose sensor and transmitter

  • The inserter provides a sterile barrier and is disposable
  • Sensor adheres to skin with an adhesive patch
  • Two-step insertion can be inserted one-handed
  • Remove cap
  • Push inserter against skin
  • No over tape required

Simplera Sync ™ sensor

Sensor insertion

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-7

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-8

  • Related
  • A blood glucose (BG) reading is needed when entering the SmartGuard ™ feature. If glucose alerts and CGM readings do not match your symptoms, use a BG meter to make diabetes treatment decisions. Refer to System User Guide SmartGuard ™ feature. Some device user interaction

Simplera Sync™ sensor insertion

Tips to manage bleeding

Bleeding may occur at the insertion site and is not unusual.

Here are 3 tips to help you manage bleeding:

Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-9

  1. If bleeding persists, apply steady pressure with a sterile gauze pad or clean cloth on top of the sensor for up to 3 minutes.
  2. If bleeding continues after 3 minutes and is significantly visible, or there is excessive pain or discomfort, then remove the sensor and continue to apply steady pressure until bleeding stops. Discard sensor and insert a new Simplera Sync™ sensor in a different spot.
  3. Check the site for redness, bleeding, irritation, pain, tenderness, or inflammation. If you notice any of these, contact a healthcare professional.

Simplera Sync™ sensor

Sensor insertion

  • Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-10Key sensor precautions
    • Only use if cap label is undamaged and the tamper band is intact.
    • For data security, it only accepts recognised radio frequency (RF) signals.
  • Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-11Important considerations
    • Magnetic field exposure: avoid MRI, diathermy, and radiation-emitting devices.
    • Air travel: use alternative screening to X-rays and full-body scanners.
    • Skincare product contact: avoid getting product onthe  sensor; clean the sensor immediately if contacted.
  • Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-12Medication alerts
    • Acetaminophen interaction: stop medication usage before using sensor glucose values to make treatment decisions; use BG meter readings to confirm glucose levels; program temp target for up to 8 hours, or the duration recommended by HCP.
    • Hydroxyurea caution: don’t use CGM with hydroxyurea.
  • Medtronic-780G-System-With-Simplera-Sync-Sensor-fig-13Sensor life of use
    • The sensor is single-use only and has a maximum life of 170 hours (7 days + 2-hour sensor warm-up).
    • All sensor warm-up periods will be 2 hours.

Disclaimer

MiniMed™ 780G system

  • This material does not replace or supersede the instructions for use.
  • It should not be considered the exclusive source of information and should be used in conjunction with the User Guide.
  • See the User Guide for detailed information regarding the instructions for use, indications, contraindications, warnings, precautions, and potential adverse events. For further information, contact your local Medtronic representative and/or consult the Medtronic website at Medtronic.eu.
  • EMEA –CGM-2400034 © 2024 Medtronic. All rights reserved. Medtronic and the Medtronic logo are Medtronic’s trademarks.
  • ™* Third-party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.
  • ACCU-CHEK, ACCU-CHEK GUIDE, ACCU-CHEK FASTCLIX, and ACCU-CHEK SOFTCLIX are trademarks of Roche Diabetes Care.
  • DreaMed Diabetes is a trademark of DreaMed Diabetes, Ltd. The MiniMed™ 780G system algorithm includes technology developed by DreaMed Diabetes.

Documents / Resources

Medtronic 780G System With Simplera Sync Sensor [pdf] Owner's Manual
780G System With Simplera Sync Sensor, 780G, System With Simplera Sync Sensor, Simplera Sync Sensor, Sync Sensor, Sensor

References

Leave a comment

Your email address will not be published. Required fields are marked *