MEAN WELL MSP-600 Series 600W Single Output Medical Type Owner’s Manual

MEAN WELL logo

600W Single Output Medical Type

MSP-600 series


MEAN WELL MSP-600 Series 600W Single Output Medical Type - 1

■ GTIN CODE

MW Search: https://www.meanwell.com/serviceGTIN.aspx

User’s Manual
MEAN WELL MSP-600 Series 600W Single Output Medical Type - QR Code 1

■ Features
  • Universal AC input / Full range
  • Built-in active PFC function, PF>0.94
  • High efficiency up to 89%
  • Withstand 300VAC surge input for 5 seconds
  • Protections: Short circuit / Overload / Over voltage / Over temperature
  • Medical safety approved (MOOP level)
  • Built-in cooling fan ON-OFF control
  • Built-in DC OK signal
  • Built-in remote ON-OFF control
  • Standby 5V@0.3A
  • Built-in remote sense function
  • No load power consumption<0.8W (Note.7)
  • Current sharing up to 2400W (3+1) (24V,36V,48V)
  • 5 years warranty

MEAN WELL - Labels - 1

SCIFICATION
MODEL MSP-600-3.3 MSP-600-5 MSP-600-7.5 MSP-600-12
OUTPUT DC VOLTAGE 3.3V 5V 7.5V 12V
RATED CURRENT 120A 120A 80A 53A
CURRENT RANGE 0 ~ 120A 0 ~ 120A 0 ~ 80A 0 ~ 53A
RATED POWER 396W 600W 600W 636W
RIPPLE & NOISE (max.) 

Note.2

120mVp-p 150mVp-p 150mVp-p 150mVp-p
VOLTAGE ADJ. RANGE 2.8 ~ 3.8V 4.3 ~ 5.8V 6.8 ~ 9V 10.2 ~ 13.8V
VOLTAGE TOLERANCE 

Note.3

±2.0% ±2.0% ±2.0% ±1.0%
LINE REGULATION ±0.5% ±0.5% ±0.5% ±0.3%
LOAD REGULATION ±1.0% ±1.0% ±1.0% ±0.5%
SETUP, RISE TIME 1000ms, 50ms/230VAC    2500ms, 50ms/115VAC at full load 
HOLD UP TIME (Typ.) 16ms/230VAC    16ms/115VAC at full load
INPUT VOLTAGE RANGE

Note.5

85 ~ 264VAC    120 ~ 370VDC
FREQUENCY RANGE 47 ~ 63Hz
POWER FACTOR (Typ.) PF>0.93/230VAC    PF>0.99/115VAC at full load 
EFFICIENCY (Typ.) 78.5% 82% 86% 88%
AC CURRENT (Typ.) 8.5A/115VAC    5A/230VAC
INRUSH CURRENT (Typ.) 35A/115VAC    80A/230VAC
LEAKAGE CURRENT Earth leakage current < 300 μA/264VAC, Touch leakage current < 100 μA/264VAC
PROTECTION OVERLOAD 105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after fault condition is removed
OVER VOLTAGE 3.96 ~ 4.62V 6 ~ 7V 9.4 ~ 10.9V 14.4 ~ 16.8V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE Shut down o/p voltage, recovers automatically after temperature goes down
FUNCTION 5V STANDBY 5VSB : 5V@0.3A ; tolerance±5%, ripple : 50mVp-p(max.)
DC OK SIGNAL PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V
REMOTE CONTROL RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off
FAN CONTROL (Typ.) Load 35±15% or RTH2≧50℃ Fan on
ENVIRONMENT WORKING TEMP. -40 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT ±0.03%/℃ (0 ~ 50℃)
VIBRATION 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY & EMC (Note 4) SAFETY STANDARDS IEC 60601-1:2005+A1+A2, ANSI/AAMI ES60601-1:2005+A2, CAN/CSA C22.2 No. 60601-1:2014+A2 EAC TP TC 004 approved; Design refer to BS EN/EN60335-1, BS EN/EN 62368-1(by request)
ISOLATION LEVEL Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP, Secondary-Earth: 1×MOOP
WITHSTAND VOLTAGE I/P-O/P:4KVAC    I/P-FG:2KVAC    O/P-FG:0.5KVAC 
ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION Compliance to BS EN/EN55011 (CISPR11) Class B, BS EN/EN61000-3-2,-3, EAC TP TC 020
EMC IMMUNITY Compliance to BS EN/EN61000-4-2,3,4,5,6,8,11, BS EN/EN60601-1-2, EAC TP TC 020
OTHERS MTBF 138.7K hrs min.    MIL-HDBK-217F (25℃)
DIMENSION 218*105*63.5mm (L*W*H)
PACKING 1.57Kg;8pcs/13.6Kg/1.34CUFT
NOTE
  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
  2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1 μ F & 47 μ F parallel capacitor.
  3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. The power supply is considered a component which will be installed into final equipment. Al the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMG tests, please refer to “EM testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf)
  5. Derating may be needed under low input voltages. Please check the derating curve for more details.
  6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
  7. No load power consumption<0.8W when RC+ & RC- (CN100 pin34) 0 ~ 0.8V or short.
  8. When the input voltage is less than 40VAC, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this deviation that does not affect basic safety or essential performance.
  9. The ambient temperature derating of 3.5°C/1000m with fanless models and of 5°C/1000m with fan models for operating altitude higher than 2000m(6500ft). 

※Product Liability Disclaimer – For detailed information, please refer to  https://www.meanwell.com/serviceDisclaimer.aspx

MODEL MSP-600-15 MSP-600-24 MSP-600-36 MSP-600-48
OUTPUT DC VOLTAGE 15V 24V 36V 48V
RATED CURRENT 43A 27A 17.5A 13A
CURRENT RANGE 0 ~ 43A 0 ~ 27A 0 ~ 17.5A 0 ~ 13A
RATED POWER 645W 648W 630W 624W
RIPPLE & NOISE (max.) 

Note.2

150mVp-p 150mVp-p 200mVp-p 240mVp-p
VOLTAGE ADJ. RANGE 13.5 ~ 18V 21.6 ~ 28.8V 28.8 ~ 39.6V 40.8 ~ 55.2V
VOLTAGE TOLERANCE 

Note.3

±1.0% ±1.0% ±1.0% ±1.0%
LINE REGULATION ±0.3% ±0.2% ±0.2% ±0.2%
LOAD REGULATION ±0.5% ±0.5% ±0.5% ±0.5%
SETUP, RISE TIME 1000ms, 50ms/230VAC    2500ms, 50ms/115VAC at full load 
HOLD UP TIME (Typ.) 16ms/230VAC    16ms/115VAC at full load
INPUT VOLTAGE RANGE

Note.5

85 ~ 264VAC    120 ~ 370VDC
FREQUENCY RANGE 47 ~ 63Hz
POWER FACTOR (Typ.) PF>0.93/230VAC    PF>0.99/115VAC at full load 
EFFICIENCY (Typ.) 88% 88% 89% 89%
AC CURRENT (Typ.) 8.5A/115VAC    5A/230VAC
INRUSH CURRENT (Typ.) 35A/115VAC    80A/230VAC
LEAKAGE CURRENT Earth leakage current < 300 μA/264VAC, Touch leakage current < 100 μA/264VAC
PROTECTION OVERLOAD 105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after fault condition is removed
OVER VOLTAGE 18.8 ~ 21.8V 30 ~ 34.8V 41.4 ~ 48.6V 57.6 ~ 67.2V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE Shut down o/p voltage, recovers automatically after temperature goes down
FUNCTION 5V STANDBY 5VSB : 5V@0.3A ; tolerance±5%, ripple : 50mVp-p(max.)
DC OK SIGNAL PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V
REMOTE CONTROL RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off
FAN CONTROL (Typ.) Load 35±15% or RTH2≧50℃ Fan on
ENVIRONMENT WORKING TEMP. -40 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT ±0.03%/℃ (0 ~ 50℃)
VIBRATION 10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY & EMC (Note 4) SAFETY STANDARDS IEC 60601-1:2005+A1+A2, ANSI/AAMI ES60601-1:2005+A2, CAN/CSA C22.2 No. 60601-1:2014+A2 EAC TP TC 004 approved; Design refer to BS EN/EN60335-1, BS EN/EN 62368-1(by request)
ISOLATION LEVEL Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP, Secondary-Earth: 1×MOOP
WITHSTAND VOLTAGE I/P-O/P:4KVAC    I/P-FG:2KVAC    O/P-FG:0.5KVAC 
ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION Compliance to BS EN/EN55011 (CISPR11) Class B, BS EN/EN61000-3-2,-3, EAC TP TC 020
EMC IMMUNITY Compliance to BS EN/EN61000-4-2,3,4,5,6,8,11, BS EN/EN60601-1-2, EAC TP TC 020
OTHERS MTBF 138.7K hrs min.    MIL-HDBK-217F (25℃)
DIMENSION 218*105*63.5mm (L*W*H)
PACKING 1.57Kg;8pcs/13.6Kg/1.34CUFT
NOTE
  1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
  2. Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1 μ F & 47 μ F parallel capacitor.
  3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. The power supply is considered a component which will be installed into final equipment. Al the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMG tests, please refer to “EM testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf)
  5. Derating may be needed under low input voltages. Please check the derating curve for more details.
  6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
  7. No load power consumption<0.8W when RC+ & RC- (CN100 pin34) 0 ~ 0.8V or short.
  8. When the input voltage is less than 40VAC, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this deviation that does not affect basic safety or essential performance.
  9. The ambient temperature derating of 3.5°C/1000m with fanless models and of 5°C/1000m with fan models for operating altitude higher than 2000m(6500ft). 

※Product Liability Disclaimer – For detailed information, please refer to  https://www.meanwell.com/serviceDisclaimer.aspx

■ Mechanical Specification

(Unit: mm , tolerance ± 1mm)

Case No.977A

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a1 MEAN WELL MSP-600 Series 600W Single Output Medical Type - a2

  1. Airflow direction

AC Input Terminal Pin No. Assignment

Pin No. Assignment
1 AC/L
2 AC/N
3 FG Earthing symbol 5

DC Output Terminal Pin No. Assignment

Pin No. Assignment
1~3 -V
4~6 +V

Connector Pin No. Assignment(CN100) : HRS DF11-10DP-2DS or equivalent

Pin No. Assignment Mating Housing Terminal
1 AUXG HRS DF11-10DS or equivalent HRS DF11-**SC or equivalent
2 AUX
3 RC+
4 RC-
5 CS
6,8 GND
7 DC-OK
9 +S
10 -S
■ Block Diagram

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a3

■ Derating Curve

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a4

■ Output Derating VS Input Voltage

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a5

■ Function Description of CN100
Pin No. Function Description
1 AUXG Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V).
2 AUX Auxiliary voltage output, 4.75~5.25V, referenced to pin 1(AUXG). The maximum load current is 0.3A. This output is not controlled by the “remote ON/OFF control”.
3 RC+ Turns the output on and off by electrical or dry contact between pin 4 (RC-), Short: Power OFF, Open: Power ON.
4 RC- Remote control ground. 
5 CS Current sharing signal. When units are connected in parallel, the CS pins of the units should be connected to allow current balance between units.
6,8 GND This pin connects to the negative terminal(-V). Return for DC-OK signal output.
7 DC-OK DC-OK signal is a TTL level signal, referenced to pin8(DC-OK GND). High when PSU turns on.
9 +S Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
10 -S Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V.
■ Function Manual

1. Remote Sense

The remote sensing compensates voltage drop on the load wiring up to 0.5V.

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a6

Fig1.1

  1. Sense lines should be twisted in pairs

2. DC-OK Signal

DC-OK signal is a TTL level signal. High when PSU turns on.

Between DC-OK(pin7) and GND(pin6,8) Output Status
3.3 ~ 5.6V ON
0 ~ 1V OFF

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a7       MEAN WELL MSP-600 Series 600W Single Output Medical Type - a8

Fig 2.1

3. Remote Control

The PSU can be turned ON/OFF by using the “Remote Control” function.

Between RC+(pin3) and RC-(pin4) Output Status
SW ON (Short) OFF
SW OFF (Open) ON

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a9      MEAN WELL MSP-600 Series 600W Single Output Medical Type - a10

Fig 3.1

4.Current Sharing with Remote Sensing (Only for 24V, 36V and 48V)

MSP-600 has the built-in active current sharing function and can be connected in parallel to provide higher output power :
(1)Parallel operation is available by connecting the units shown as below.
(+S,-S,CS and GND are connected mutually in parallel).
(2)Difference of output voltages among parallel units should be less than 2%.
(3)The total output current must not exceed the value determined by the following equation.
(output current at parallel operation)=(Rated current per unit)×(Number of unit)×0.9
(4)In parallel operation 4 units is the maximum, please consult the manufacturer for applications of more connecting in parallel.
(5)The power supplies should be paralleled using short and large diameter wiring and then connected to the load.

MEAN WELL MSP-600 Series 600W Single Output Medical Type - a11

Fig 4.1

  1. Sense lines should be twisted in pairs

Note :

  1. In parallel connection, maybe only one unit (master) operate if the total output load is less than 2% of rated load condition.
    The other PSU (slave) may go into standby mode and its output LED and relay will not turn on.
  2. 2% min. of dummy load is required.

File Name:MSP-600-SPEC  2024-10-30

Downloaded from Arrow.com.