Owner's Manual for MEAN WELL models including: MSP-600-3.3, MSP-600-5, MSP-600-7.5, MSP-600-12, MSP-600-15, MSP-600-24, MSP-600-36, MSP-600-48, MSP-600 Series 600W Single Output Medical Type, MSP-600 Series, 600W Single Output Medical Type, Single Output Medical Type, Output Medical Type, Medical Type, Type

未命名 -1

Shirley Cheng

Блок живлення що вбудовується 600W 36V MSP-600-36 MeanWell за ціною від 8690.00 грн - РКС Компоненти - РАДІОМАГ


File Info : application/pdf, 4 Pages, 638.61KB

PDF preview unavailable. Download the PDF instead.

msp-600-spec
600W Single Output Medical Type

MSP-600 series

 
 https://www.meanwell.com/serviceGTIN.aspx



User's Manual





























SPECIFICATION

MODEL

MSP-600-3.3 MSP-600-5 MSP-600-7.5 MSP-600-12 MSP-600-15 MSP-600-24 MSP-600-36 MSP-600-48

DC VOLTAGE RATED CURRENT

3.3V

5V

7.5V

12V

15V

24V

36V

48V

120A

120A

80A

53A

43A

27A

17.5A

13A

CURRENT RANGE

0 ~ 120A

0 ~ 120A

0 ~ 80A

0 ~ 53A

0 ~ 43A

0 ~ 27A

0 ~ 17.5A

0 ~ 13A

RATED POWER

396W

600W

600W

636W

645W

648W

630W

624W

OUTPUT

RIPPLE & NOISE (max.) Note.2 120mVp-p

VOLTAGE ADJ. RANGE

2.8 ~ 3.8V

VOLTAGE TOLERANCE Note.3 ±2.0%

LINE REGULATION

±0.5%

LOAD REGULATION

±1.0%

150mVp-p
4.3 ~ 5.8V ±2.0% ±0.5% ±1.0%

150mVp-p
6.8 ~ 9V ±2.0% ±0.5% ±1.0%

150mVp-p
10.2 ~ 13.8V ±1.0% ±0.3% ±0.5%

150mVp-p
13.5 ~ 18V ±1.0% ±0.3% ±0.5%

150mVp-p
21.6 ~ 28.8V ±1.0% ±0.2% ±0.5%

200mVp-p
28.8 ~ 39.6V ±1.0% ±0.2% ±0.5%

240mVp-p
40.8 ~ 55.2V ±1.0% ±0.2% ±0.5%

SETUP, RISE TIME HOLD UP TIME (Typ.)

1000ms, 50ms/230VAC 2500ms, 50ms/115VAC at full load 16ms/230VAC 16ms/115VAC at full load

VOLTAGE RANGE Note.5 85 ~ 264VAC

FREQUENCY RANGE

47 ~ 63Hz

120 ~ 370VDC

POWER FACTOR (Typ.)

PF>0.93/230VAC PF>0.99/115VAC at full load

INPUT

EFFICIENCY (Typ.)

78.5%

82%

86%

88%

88%

88%

89%

89%

AC CURRENT (Typ.) INRUSH CURRENT (Typ.) LEAKAGE CURRENT

8.5A/115VAC 5A/230VAC 35A/115VAC 80A/230VAC Earth leakage current < 300A/264VAC , Touch leakage current < 100A/264VAC

OVERLOAD

105 ~ 135% rated output power Protection type : Constant current limiting, recovers automatically after fault condition is removed

PROTECTION OVER VOLTAGE

3.96 ~ 4.62V 6 ~ 7V

9.4 ~ 10.9V 14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V

Protection type : Shut down o/p voltage, re-power on to recover

41.4 ~ 48.6V

OVER TEMPERATURE 5V STANDBY

Shut down o/p voltage, recovers automatically after temperature goes down 5VSB : 5V@0.3A ; tolerance±5%, ripple : 50mVp-p(max.)

DC OK SIGNAL FUNCTION REMOTE CONTROL
FAN CONTROL (Typ.) WORKING TEMP.

PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V
RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off Load 35±15% or RTH250 Fan on -40 ~ +70 (Refer to "Derating Curve")

WORKING HUMIDITY

20 ~ 90% RH non-condensing

ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH non-condensing

TEMP. COEFFICIENT

±0.03%/ (0 ~ 50)

VIBRATION

10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes

SAFETY STANDARDS

IEC 60601-1:2005+A1+A2, ANSI/AAMI ES60601-1:2005+A2, CAN/CSA C22.2 No. 60601-1:2014+A2 EAC TP TC 004 approved; Design refer to BS EN/EN60335-1, BS EN/EN 62368-1(by request)

SAFETY &
EMC (Note 4)

ISOLATION LEVEL WITHSTAND VOLTAGE ISOLATION RESISTANCE EMC EMISSION

Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP, Secondary-Earth: 1×MOOP I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25/ 70% RH Compliance to BS EN/EN55011 (CISPR11) Class B, BS EN/EN61000-3-2,-3, EAC TP TC 020

57.6 ~ 67.2V

EMC IMMUNITY MTBF

Compliance to BS EN/EN61000-4-2,3,4,5,6,8,11, BS EN/EN60601-1-2, EAC TP TC 020 138.7K hrs min. MIL-HDBK-217F (25)

OTHERS DIMENSION

218*105*63.5mm (L*W*H)

PACKING

1.57Kg;8pcs/13.6Kg/1.34CUFT

NOTE

 FF 







(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf ) 











Downloaded from Arrow.com.https://www.meanwell.com/serviceDisclaimer.aspx
File Name:MSP-600-SPEC 2024-10-30

600W Single Output Medical Type

Mechanical Specification

(Unit: mm , tolerance ± 1mm)

152.4

20.75

10 8.2

Air flow direction
I/P 1 2 3
12max.

63.5

4-M4 L=6
218 152.4 8-M4 L=4 (Both Sides)

MSP-600 series

32.8 O/P
1 2 3 4 5 6
32.8
18max.

9.2

11

Case No. 977A
19 2 10

105

12.5 38.3 63.5
CN100 LED
(Vo ADJ.)SVR1

AC Input Terminal Pin No. Assignment

Pin No. Assignment

1

AC/L

2

AC/N

3

FG

Block Diagram

I/P

EMI FILTER

FG

Derating Curve

DC Output Terminal Pin No. Assignment

Pin No. Assignment

1~3

-V

4~6

+V

ACTIVE INRUSH CURRENT LIMITING

RECTIFIERS &
PFC
O.T.P.
PFC CONTROL

Connector Pin No. Assignment(CN100) : HRS DF11-10DP-2DS or equivalent

Pin No. Assignment Pin No. Assignment Mating Housing Terminal

1

AUXG

6,8

GND

2

AUX

7

DC-OK

3

RC+

9

+S

HRS DF11-10DS HRS DF11-**SC or equivalent or equivalent

4

RC-

10

-S

5

CS

AUX POWER
POWER SWITCHING

RECTIFIERS &
FILTER
RECTIFIERS &
FILTER

O.L.P.
PWM CONTROL

DETECTION CIRCUIT
O.T.P.
O.V.P.
REMOTE CONTROL

Output Derating VS Input Voltage

PWM fosc : 70KHz
AUX AUXG +S +V -V -S

DC OK

CURRENT SHARING

CS

RC+

RC-

LOAD (%) LOAD(%)

125

100

100

90

80 80
70

50

60

50

20

40

-40

0

10

20

30

40

50

60 70 (HORIZONTAL)

85

100

125

135

155

264

AMBIENT TEMPERATURE ()

INPUT VOLTAGE (V) 60Hz

File Name:MSP-600-SPEC 2024-10-30

Downloaded from Arrow.com.

600W Single Output Medical Type

MSP-600 series

Function Description of CN100

Pin No. Function Description

1 AUXG Auxiliary voltage output ground. The signal return is isolated from the output terminals (+V & -V).

2

AUX

Auxiliary voltage output, 4.75~5.25V, referenced to pin 1(AUXG). The maximum load current is 0.3A. This output is not controlled by the "remote ON/OFF control".

3

RC+ Turns the output on and off by electrical or dry contact between pin 4 (RC-), Short: Power OFF, Open: Power ON.

4

RC- Remote control ground.

5

CS

Current sharing signal. When units are connected in parallel, the CS pins of the units should be connected to allow current balance between units.

6,8 GND This pin connects to the negative terminal(-V). Return for DC-OK signal output.

7 DC-OK DC-OK signal is a TTL level signal, referenced to pin8(DC-OK GND). High when PSU turns on.

9

+S

Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V.

10

-S

Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V.

Function Manual
1.Remote Sense The remote sensing compensates voltage drop on the load wiring up to 0.5V.

CN100
21 10 9

CN100 2 AUX AUXG 1

RC- RC+

GND CS

GND DC-OK

10 -S

9 +S

+S (Pin9)

+LOAD

-S (Pin10)

2.DC-OK Signal DC-OK signal is a TTL level signal. High when PSU turns on.

Between DC-OK(pin7) and GND(pin6,8) Output Status

3.3 ~ 5.6V

ON

0 ~ 1V

OFF

CN100
21 10 9

Fig 1.1

Sense lines should be twisted in pairs
CN100 2 AUX AUXG 1
RC- RC+ GND CS GND DC-OK 10 -S +S 9

Fig 2.1

Downloaded from Arrow.com.

File Name:MSP-600-SPEC 2024-10-30

600W Single Output Medical Type

MSP-600 series

3.Remote Control The PSU can be turned ON/OFF by using the "Remote Control" function.

Between RC+(pin3) and RC-(pin4) Output Status

SW ON (Short)

OFF

SW OFF (Open)

ON

CN100
21 10 9

CN100 2 AUX AUXG 1
RC- RC+ GND CS GND DC-OK 10 -S +S 9 SW

Fig 3.1

4.Current Sharing with Remote Sensing (Only for 24V, 36V and 48V) MSP-600 has the built-in active current sharing function and can be connected in parallel to provide higher output power : (1)Parallel operation is available by connecting the units shown as below. (+S,-S,CS and GND are connected mutually in parallel). (2)Difference of output voltages among parallel units should be less than 2%. (3)The total output current must not exceed the value determined by the following equation. (output current at parallel operation)=(Rated current per unit)×(Number of unit)×0.9 (4)In parallel operation 4 units is the maximum, please consult the manufacturer for applications of more connecting in parallel. (5)The power supplies should be paralleled using short and large diameter wiring and then connected to the load.

PSU1

PSU2

PSU3

PSU4

CN100

-V +V
CN100

-V +V
CN100

-V +V
CN100

-V +V

PSU1 CN100 1 AUXG RC+ CS DC-OK +S 9
2 AUX RC- GND GND -S 10

PSU2 CN100 1 AUXG RC+ CS DC-OK +S 9
2 AUX RC- GND GND -S 10

PSU3 CN100 1 AUXG RC+ CS DC-OK +S 9
2 AUX RC- GND GND -S 10

PSU4 CN100 1 AUXG RC+ CS DC-OK +S 9
2 AUX RC- GND GND -S 10

+V -V LOAD

Sense lines should be twisted in pairs

Fig 4.1
Note : 1.In parallel connection, maybe only one unit (master) operate if the total output load is less than 2% of rated load condition. The other PSU (slave) may go into standby mode and its output LED and relay will not turn on.
2.2% min. of dummy load is required.

File Name:MSP-600-SPEC 2024-10-30