MPS-65 Series 65W Single Output Medical Type
65W Single Output Medical Type
MPS-65 series
Features :
- Universal AC input / Full range
- Low leakage current <250pA
- Protections: Short circuit / Overload / Over voltage
- Cooling by free air convection
- Medical safety approved (2 x MOPP between primary to secondary)
- 100% full load burn-in test
- Fixed switching frequency at 45KHz
- 3 years warranty
SPECIFICATION
MODEL | MPS-65-3.3 | MPS-65-5 | MPS-65-7.5 | MPS-65-12 | MPS-65-13.5 | MPS-65-15 | MPS-65-24 | MPS-65-27 | MPS-65-48 | |
OUTPUT | DC VOLTAGE | 3.3V | 5V | 7.5V | 12V | 13. W | 15V | 24V | 27V | 48V |
RATED CURRENT | 12A | 12A | 13A | 5.2A | 4.7A | 4.2A | 2.7A | 24A | 1.35A | |
CURRENT RANGE | 0^ 15.2A | 0 – 13.8A | 0^9.6A | 0 – 6A | 0 – 5.1A | 0 – 4.8A | 0^3A | 0 ^ 2.7A | 0^ 1.5A | |
RATED POWER | 39.6W | 60W | 60W | 62.4W | 63.45W | 63W | 641W | 64.8W | 64.8W | |
OUTPUT POWER (max.) | 72W(.3.3V5OW:•5V:69Whviti 18CFM min. Faced air contraction | |||||||||
RIPPLE 6 NOISE (max.) Note.2 | 80mVp.p | 100mVp-p | 100mVp-p | 100mVp-p | 100mVp-p | 100mVp-p | 100mVp-p | 100mVp-p | 100mVp-p | |
VOLTAGE ADJ. RANGE | 2.97 – 3.63V | 4.5^5.5V | 6.75^825V | 10.8^132V | 122 – 14.85‘ | 13,5 – 165V | 21.6-26.4V | 243-29.7V | 43.2^52.8V | |
VOLTAGE TOLERANCE Note.3 | 23.% | ±3.0% | 23.% | ±2.0% | ±2.0% | ±2.0% | ±2.0% | ±2.0% | ±2.0% | |
LINE REGULATION | 21.% | ±1.0% | ±1.0% | 21.% | ±1.0% | ±1.0% | ±1.0% | ±1.0% | ±1.0% | |
LOAD REGULATION | ±3.0% | 23.% | 23.% | ±2.0% | ±2.0% | ±2.0% | ±2.0% | ±2.0% | ±2.0% | |
SETUP. RISE TIME | 800ms. 30ms230VAC 800ms. 30ms/115VAC al Woad | |||||||||
HOLD UP TIME (Typ.) | 50ms1230VAC 16ms/115VAC at full load | |||||||||
INPUT | VOLTAGE RANGE | 93 – 264VAC 127 – 370VDC | ||||||||
FREQUENCY RANGE | 47 -63Hz | |||||||||
EFFICIENCY(Typ.) | 66% 174% 176% 177% 178% 179% 180% 180% 180% | |||||||||
AC CURRENT (Typ) | 1.6A1115VAC 0.9A/230VAC | |||||||||
INRUSH CURRENT (Typ.) | COLD START 174/115VAC 40N230VAC | |||||||||
LEAKAGE CURRENT Note.? | Each leakage current < 25011A/264VAC . Touch current < 600/264VAC | |||||||||
PROTECTION | OVERLOAD | 73 – 105W (3.3V:51 – 75W)(5V:70 – 105W) rated output power | ||||||||
Protection type: Hkcup mode. recovers automatically after the fault condition is removed | ||||||||||
OVERVOLTAGE | 3.8 – 4.46V I 5.75 – 6.75V 1 8.63- 10.1V I 13.8 – 16.2V I 15.5 – 18.2V11725 – 20.25VI 27.6 – 32.4VI 31 – 36.45V I 55.2 -64.8V | |||||||||
Protection type: Hiccup mode. recovers automatically after the fault condition is removed | ||||||||||
EIWIRONVENT | WORKING TEMP. | -10^ +60t (Refer b ‘Derating Curve’) | ||||||||
WORKING HUMIDITY | 20 – 93% RH non-condensing | |||||||||
STORAGE TEMP.. HUMIDITY | -20^ +85t, 10 – 95% FtH | |||||||||
TEMP. COEFFICIENT | ±0.04%!C (0 – 501: ) | |||||||||
VIBRATION | 10 – 500Hz. 2G 10min.71cycle. period Ice 60min. each along X. Y. Z axes | |||||||||
SAFETY 8 EMC (Note 4) |
SAFETY STANDARDS | ANSI/AAMI E560601.1. TUV EN69501.1. IEC60601-1 approved | ||||||||
ISOLATION LEVEL | Primary-Secondary: 2xMOPP. Primary.Earth:14MOPP | |||||||||
WITHSTAND VOLTAGE | UP.O’P:4KVAC lP.FG:2KVAC OP.FG:0.5KVAC | |||||||||
ISOLATION RESISTANCE | VP-0/P. 1/P-FG:10041 Ohms / 500VDC I 25t170% RH | |||||||||
EMC EMISSION | Compliance to EN 55011 (CISPR1I) Class B. 8461000-3-2.-3 | |||||||||
EMC IMMUNITY | Compliance to EN6100041-211551.11. EN60601-1-2, medical level. criteria | |||||||||
OTHERS | MTBF | 359.7Khrs min. MIL-HDBK-217F (251.:) | ||||||||
DIMENSION | 127’76’42mm (L’WH) | |||||||||
PACKING | 0.23Kg: 54pcs/14.6Kg/128CUFT | |||||||||
NOTE | l. Al parameters NOT specially mentioned are measured at 230VAC input rated load end 251:41 ambient temperature. a Ripple 8 noise is measured at 20MHz of banchwth by using an It twisted parties terminated with a Glut & 47uf parallel capacitor. 1 Tolerance: colas set up a tolerance. line regulation and load regulation. 4. The power supply is considered a component that will be instated into the final equipment Al the EMC tests are been executed by mounting the unit on a 360mml6Ornm metal plate with 1mm of thickness. The real equipment must be re-confirmed that it still meets EMC’s duties. For guidance on how to do these EMC tests. refer 4 -EMI testing or component sc./Vies.” (as avaiablo on http:/Mwer ineanwettcom) perform please pont |
|||||||||
S. Mounting holes MI and 142 should be grounded for EMI purposes 6 Heat Sink HSI.HS2 can not be shorted 7 Touch °went vas measured from primary input to DC output ,.• Product Liability Disclaimer: For detailed intonation. Please refer to haps Jiwew meant/ell caWserviceDIsclauner aspx |
Mechanical Specification
Block Diagram
Derating Curve
Static Characteristics
https://www.meanwell.com/Upload/PDF/Medical_EN.pdf
Documents / Resources
![]() |
MEAN WELL MPS-65 Series 65W Single Output Medical Type [pdf] Instruction Manual MPS-65 Series 65W Single Output Medical Type, MPS-65 Series, 65W Single Output Medical Type, Output Medical Type, Medical Type |