MPS-65 Series 65W Single Output Medical Type

MEAN Logo65W Single Output Medical Type
MPS-65 series MEAN WELL MPS 65 Series 65W Single Output Medical Type

Features :

  • Universal AC input / Full range
  • Low leakage current <250pA
  • Protections: Short circuit / Overload / Over voltage
  • Cooling by free air convection
  • Medical safety approved (2 x MOPP between primary to secondary)
  • 100% full load burn-in test
  • Fixed switching frequency at 45KHz
  • 3 years warranty

MEAN WELL MPS 65 Series 65W Single Output Medical Type - icon

SPECIFICATION

MODEL MPS-65-3.3 MPS-65-5 MPS-65-7.5 MPS-65-12 MPS-65-13.5 MPS-65-15 MPS-65-24 MPS-65-27 MPS-65-48
OUTPUT DC VOLTAGE 3.3V 5V 7.5V 12V 13. W 15V 24V 27V 48V
RATED CURRENT 12A 12A 13A 5.2A 4.7A 4.2A 2.7A 24A 1.35A
CURRENT RANGE 0^ 15.2A 0 – 13.8A 0^9.6A 0 – 6A 0 – 5.1A 0 – 4.8A 0^3A 0 ^ 2.7A 0^ 1.5A
RATED POWER 39.6W 60W 60W 62.4W 63.45W 63W 641W 64.8W 64.8W
OUTPUT POWER (max.) 72W(.3.3V5OW:•5V:69Whviti 18CFM min. Faced air contraction
RIPPLE 6 NOISE (max.) Note.2 80mVp.p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p
VOLTAGE ADJ. RANGE 2.97 – 3.63V 4.5^5.5V 6.75^825V 10.8^132V 122 – 14.85‘ 13,5 – 165V 21.6-26.4V 243-29.7V 43.2^52.8V
VOLTAGE TOLERANCE Note.3 23.% ±3.0% 23.% ±2.0% ±2.0% ±2.0% ±2.0% ±2.0% ±2.0%
LINE REGULATION 21.% ±1.0% ±1.0% 21.% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0%
LOAD REGULATION ±3.0% 23.% 23.% ±2.0% ±2.0% ±2.0% ±2.0% ±2.0% ±2.0%
SETUP. RISE TIME 800ms. 30ms230VAC 800ms. 30ms/115VAC al Woad
HOLD UP TIME (Typ.) 50ms1230VAC 16ms/115VAC at full load
INPUT VOLTAGE RANGE 93 – 264VAC 127 – 370VDC
FREQUENCY RANGE 47 -63Hz
EFFICIENCY(Typ.) 66% 174% 176% 177% 178% 179% 180% 180% 180%
AC CURRENT (Typ) 1.6A1115VAC 0.9A/230VAC
INRUSH CURRENT (Typ.) COLD START 174/115VAC 40N230VAC
LEAKAGE CURRENT Note.? Each leakage current < 25011A/264VAC . Touch current < 600/264VAC
PROTECTION OVERLOAD 73 – 105W (3.3V:51 – 75W)(5V:70 – 105W) rated output power
Protection type: Hkcup mode. recovers automatically after the fault condition is removed
OVERVOLTAGE 3.8 – 4.46V I 5.75 – 6.75V 1 8.63- 10.1V I 13.8 – 16.2V I 15.5 – 18.2V11725 – 20.25VI 27.6 – 32.4VI 31 – 36.45V I 55.2 -64.8V
Protection type: Hiccup mode. recovers automatically after the fault condition is removed
EIWIRONVENT WORKING TEMP. -10^ +60t (Refer b ‘Derating Curve’)
WORKING HUMIDITY 20 – 93% RH non-condensing
STORAGE TEMP.. HUMIDITY -20^ +85t, 10 – 95% FtH
TEMP. COEFFICIENT ±0.04%!C (0 – 501: )
VIBRATION 10 – 500Hz. 2G 10min.71cycle. period Ice 60min. each along X. Y. Z axes
SAFETY 8
EMC (Note 4)
SAFETY STANDARDS ANSI/AAMI E560601.1. TUV EN69501.1. IEC60601-1 approved
ISOLATION LEVEL Primary-Secondary: 2xMOPP. Primary.Earth:14MOPP
WITHSTAND VOLTAGE UP.O’P:4KVAC lP.FG:2KVAC OP.FG:0.5KVAC
 ISOLATION RESISTANCE VP-0/P. 1/P-FG:10041 Ohms / 500VDC I 25t170% RH
EMC EMISSION Compliance to EN 55011 (CISPR1I) Class B. 8461000-3-2.-3
EMC IMMUNITY Compliance to EN6100041-211551.11. EN60601-1-2, medical level. criteria
OTHERS MTBF 359.7Khrs min. MIL-HDBK-217F (251.:)
DIMENSION 127’76’42mm (L’WH)
PACKING 0.23Kg: 54pcs/14.6Kg/128CUFT
NOTE l. Al parameters NOT specially mentioned are measured at 230VAC input rated load end 251:41 ambient temperature.
a Ripple 8 noise is measured at 20MHz of banchwth by using an It twisted parties terminated with a Glut & 47uf parallel capacitor. 1 Tolerance: colas set up a tolerance. line regulation and load regulation.
4. The power supply is considered a component that will be instated into the final equipment Al the EMC tests are been executed by mounting the unit on a 360mml6Ornm metal plate with 1mm of thickness. The real equipment must be re-confirmed that it still meets EMC’s duties. For guidance on how to do these EMC tests. refer 4 -EMI testing or component sc./Vies.” (as avaiablo on http:/Mwer ineanwettcom) perform please pont
S. Mounting holes MI and 142 should be grounded for EMI purposes
6 Heat Sink HSI.HS2 can not be shorted
7 Touch °went vas measured from primary input to DC output
,.• Product Liability Disclaimer: For detailed intonation. Please refer to haps Jiwew meant/ell caWserviceDIsclauner aspx

Mechanical Specification

MEAN WELL MPS 65 Series 65W Single Output Medical Type - Mechanical

Block Diagram

MEAN WELL MPS 65 Series 65W Single Output Medical Type - Diagram

Derating Curve

MEAN WELL MPS 65 Series 65W Single Output Medical Type - Derating Curve

Static Characteristics

MEAN WELL MPS 65 Series 65W Single Output Medical Type - Derating Curve 1

MEAN WELL MPS 65 Series 65W Single Output Medical Type - QR Codehttps://www.meanwell.com/Upload/PDF/Medical_EN.pdf

Documents / Resources

MEAN WELL MPS-65 Series 65W Single Output Medical Type [pdf] Instruction Manual
MPS-65 Series 65W Single Output Medical Type, MPS-65 Series, 65W Single Output Medical Type, Output Medical Type, Medical Type

References