Instruction Manual for MEAN WELL models including: MPS-65 Series 65W Single Output Medical Type, MPS-65 Series, 65W Single Output Medical Type, Output Medical Type, Medical Type

MPS-65-20210915

wangj

65W Single Output MedicalType MPS-65 - MEAN WELL

specification model mps-65-3.3 mps-65-5 mps-65-7.5 mps-65-12 mps-65-13.5 mps-65-15 mps-65-24 mps-65-27 mps-65-48 dc voltage rated current current range rated power output line regulation load regulation setup, rise time hold up time (typ.) voltage range


File Info : application/pdf, 2 Pages, 96.15KB

PDF preview unavailable. Download the PDF instead.

MPS-65-SPEC
65W Single Output Medical Type

MPS-65 series

User's Manual
 Features :
Universal AC input / Full range Low leakage current <250A Protections: Short circuit / Overload / Over voltage Cooling by free air convection Medical safety approved (2 x MOPP between primary to secondary) 100% full load burn-in test Fixed switching frequency at 45KHz 3 years warranty

EN 60601-1 IEC 60601-1

SPECIFICATION

8

MODEL

MPS-65-3.3 MPS-65-5 MPS-65-7.5 MPS-65-12 MPS-65-13.5 MPS-65-15 MPS-65-24 MPS-65-27 MPS-65-48

DC VOLTAGE

3.3V

5V

7.5V

12V

13.5V

15V

24V

27V

48V

RATED CURRENT

12A

12A

8A

5.2A

4.7A

4.2A

2.7A

2.4A

1.35A

CURRENT RANGE

0 ~ 15.2A 0 ~ 13.8A 0 ~ 9.6A 0 ~ 6A

0 ~ 5.4A 0 ~ 4.8A 0 ~ 3A

0 ~ 2.7A 0 ~ 1.5A

RATED POWER

39.6W

60W

60W

62.4W

63.45W

63W

64.8W

64.8W

64.8W

OUTPUT POWER (max.)

72W(+3.3V:50W;+5V:69W)with 18CFM min. Forced air convection

OUTPUT

RIPPLE & NOISE (max.) Note.2 80mVp-p 100mVp-p

VOLTAGE ADJ. RANGE

2.97 ~ 3.63V 4.5 ~ 5.5V

VOLTAGE TOLERANCE Note.3 ±3.0%

±3.0%

LINE REGULATION

±1.0%

±1.0%

LOAD REGULATION

±3.0%

±3.0%

100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p

6.75 ~ 8.25V 10.8 ~ 13.2V 12.2 ~ 14.85V 13.5 ~ 16.5V 21.6 ~ 26.4V 24.3 ~ 29.7V 43.2 ~ 52.8V

±3.0%

±2.0%

±2.0%

±2.0%

±2.0%

±2.0%

±2.0%

±1.0%

±1.0%

±1.0%

±1.0%

±1.0%

±1.0%

±1.0%

±3.0%

±2.0%

±2.0%

±2.0%

±2.0%

±2.0%

±2.0%

SETUP, RISE TIME

800ms, 30ms/230VAC 800ms, 30ms/115VAC at full load

HOLD UP TIME (Typ.)

50ms/230VAC 16ms/115VAC at full load

VOLTAGE RANGE

90 ~ 264VAC 127 ~ 370VDC

FREQUENCY RANGE

47 ~ 63Hz

INPUT

EFFICIENCY(Typ.) AC CURRENT (Typ.)

66%

74%

76%

77%

78%

79%

80%

80%

1.6A/115VAC 0.9A/230VAC

INRUSH CURRENT (Typ.)

COLD START 17A/115VAC 40A/230VAC

LEAKAGE CURRENT Note.7 Earth leakage current < 250A/264VAC , Touch current < 60A/264VAC

OVERLOAD PROTECTION
OVER VOLTAGE

73 ~ 105W (3.3V:51 ~ 75W)(5V:70 ~ 105W) rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed 3.8 ~ 4.46V 5.75 ~ 6.75V 8.63 ~ 10.1V 13.8 ~ 16.2V 15.5 ~ 18.2V 17.25 ~ 20.25V 27.6 ~ 32.4V 31 ~ 36.45V Protection type : Hiccup mode, recovers automatically after fault condition is removed

WORKING TEMP.

-10 ~ +60 (Refer to "Derating Curve")

80% 55.2 ~ 64.8V

WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT VIBRATION SAFETY STANDARDS

20 ~ 90% RH non-condensing -20 ~ +85, 10 ~ 95% RH ±0.04%/ (0 ~ 50) 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved

SAFETY &
EMC (Note 4)

ISOLATION LEVEL WITHSTAND VOLTAGE ISOLATION RESISTANCE EMC EMISSION

EMC IMMUNITY

MTBF

Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25/ 70% RH Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medical level, criteria A 359.7Khrs min. MIL-HDBK-217F (25)

OTHERS NOTE

DIMENSION

127*76*42mm (L*W*H)

PACKING

0.23Kg; 54pcs/14.6Kg/1.28CUFT

1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on
a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 5. Mounting holes M1 and M2 should be grounded for EMI purposes. 6. Heat Sink HS1,HS2 can not be shorted. 7. Touch current was measured from primary input to DC output.

 Product Liability DisclaimerFor detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

File Name:MPS-65-SPEC 2021-09-15

65W Single Output Medical Type

MPS-65 series

Mechanical Specification
M1

5.6

127

115.8

5.6

HS1

Unit:mm

64.8

76

FS1 AC FUSE T2A/250V
1 HS2
2 CN1 FS2 AC FUSE
T2A/250V

CN2
1 2 3 4 5 6

SVR1

LED

4-

3.8

42

3 max.

AC Input Connector (CN1) : Molex 5277-02 or equivalent

Pin No. Assignment Mating Housing

Terminal

1

AC/L

Molex 5195

Molex 5194

2

AC/N

or equivalent

or equivalent

Block Diagram

I/P

EMI

FILTER

RECTIFIERS &
FILTER

POWER SWITCHING

DC Output Connector (CN2) : Molex 5273-06 or equivalent

Pin No. Assignment Mating Housing

Terminal

1,2,3

+V

4,5,6

-V

Molex 5195 or equivalent

Molex 5194 or equivalent

: Grounding Required

1.HS1,HS2 cannot be shorted 2.M1 is safety ground

RECTIFIERS &
FILTER

fosc : 45KHz
+V -V

FG
Derating Curve

O.L.P.

PWM CONTROL

DETECTION CIRCUIT
O.V.P.

Static Characteristics

100

Others

80

60

3.3V,5V

40

20

-10

0

10

20

30

40

50

60 (HORIZONTAL)

AMBIENT TEMPERATURE ()

100 90 80 70 60 50 40
90

100

110

135

155

INPUT VOLTAGE (V) 60Hz

230 264

File Name:MPS-65-SPEC 2021-09-15

LOAD (%) LOAD(%)



References