Manufacturer Information
ISM Deutschland GmbH, Laubach 30, 40822 Mettmann, Germany
Pioneering Gas Sensing Technology.
Website: ionscience.com
Precautions
Use the instrument ONLY on voltage-free and grounded equipment. Failure to observe this rule may cause fatal injury and/or material damage.
The surface you are going to check for leaks must be free of all liquids like oil or water, and coarse dirt. If necessary, wipe it with a cloth before approaching it with the sensor tip. The ingress of liquids will immediately destroy the SmartSensor, while grease and similar substances, even in small quantities, will obstruct the sample inlet.
Measuring Principle
An advanced high-voltage ionization detector (NIC® - Negative Ion Capture) is used which has been optimised in consideration of the following:
- Sensitivity
- Precision
- Service life
- Reliability
Through miniaturisation, it has been possible to accommodate the sensor directly in the measurement tip, thereby attaining excellent response and recovery timings.
Usage in Practice
Owing to the practice-oriented design of the instruments, work with these is simple and reliable when observing the following:
For Leak Detection
Switch the instrument to Search Mode. Guide the sniffer probe as closely as possible to the suspected leaking locations. Establishing material contact with the sniffer probe will be useful. The velocity at which the sniffer probe is advanced should be at about 20 mm/s. Please do not push the sniffer probe over the test specimen; instead, draw at a 45-degree angle to effectively prevent the ingress of grease and dirt, for example.
For Leak Measurements
Switch the instrument to Measurement Mode. Make sure it is set to the desired units of measurement (cc/s or g/a). Guide the sniffer probe as closely as possible to the suspected leaking locations. Establishing material contact with the sniffer probe will be useful. The magnitude of the leak can be considered as being correctly acquired when the measured value does not change within 2 seconds.
For Concentration Rise Measurements (Integral atm.)
In the set-up, select the unit of measurement “ppm”. Place the test specimen within the chamber free of SF6, and close the chamber. Zero the leak detector in clean ambient air by operating the “Zero” button. For the initial measurement with the P1 XTL leak detector, push the sniffer probe for approximately 10 to 20 seconds through a corresponding connection directly into the chamber. Check the measurement value V1. Remove the sniffer probe from the chamber. After the measurement time defined by the customer has elapsed, Zero the leak detector in clean air by operating the “Zero” button, introduce the sniffer probe into the chamber again, and perform the final measurement for a duration of approximately 10 to 20 seconds. Check the measurement value V2. Remove the sniffer probe from the chamber. The difference between the two measured values (V2-V1) gives the rise in concentration within the measurement time specified by the customer.
Setup and Operation
Opening the Case
Action: Open the Case.
Description: The instrument is housed in a rugged, wheeled case. The lid opens upwards.
Preparing the Instrument
Action: Press firmly on the top of the P1 XT Handgun to release the docking station to the working position.
Description: The handgun unit is secured in a docking station within the case. Pressing a specific point on the handgun releases it from its locked position.
Action: The instrument will slide out in the working position.
Description: Once released, the handgun unit smoothly slides out of the docking station, ready for use.
Action: Wait for any action until working position.
Action: Take one P1 SmartSensor out of the storage bay.
Action: Attach the P1 SmartSensor into the plug at the handgun before you switch the instrument on.
Description: A P1 SmartSensor probe is removed from its storage location within the case and connected to the handgun unit.
Switching On and Modes
Action: To switch the instrument ON, press the left button on the Handgun shortly.
Description: The instrument powers up when the left button on the handgun is pressed briefly.
The P1 XT will start in "SEARCH MODE" or "MEASUREMENT MODE", whichever was used before.
Action: To change from "SEARCH MODE" to "MEASUREMENT MODE" and back again, swipe your finger over the display from left to right.
Action: To set the measured value to Zero, press the right button shortly.
Instrument Overview
Handgun Controls:
- Setup Button: Accesses setup menus.
- Touch Screen: Displays information and allows interaction.
- Hotkey Left: Function assigned via menu.
- Hotkey Right: Function assigned via menu.
Console Features:
- P1 XT Handgun Part Position: Push down to lock the docking station and secure the instrument by closing the lid.
- P1 SmartSensor Storage Area: Compartment for storing spare SmartSensor probes.
- USB Port: For charging only; no data transfer capabilities for mobile devices.
- Mains Power: Connection for power supply (100...240 V 50/60 Hz).
- Storage Shelf: Located for accessories such as the power cable and user manual.
P1 SmartSensor Service Life
The expected service life of the SmartSensors is approximately 200 to 300 hours of continuous operation under average room air conditions. Based on an active measurement time of approximately 2 hours per workday, this gives a utilisation duration of approximately 6 months.
In case of discontinuous measurement operation, the utilisation duration may be optimised by using the standby option. Through the use of standby modes, the SmartSensor is shut down during breaks and is therefore not subjected to wear.
The limiting factor regarding the service life of the SmartSensor is contamination:
- Through small particles taken in with the sample flow air, capable of passing through the front filter (Part No. P1:P-100-0024) and entering the ionization chamber.
- By larger particles which block the filter element, thereby reducing its filtering capacity.
Thus, the service life is strongly dependent on ambient conditions.
Any contamination in the ionization chamber results in a more noisy output signal from the SmartSensor, making itself felt through fluctuations and sudden changes of the Zero line. The SmartSensor must be considered worn out when these fluctuations, compared to the set-up limit, become too great. However, measurement sensitivity is maintained at an unchanged level over the entire utilisation duration. In case of extreme contamination, it can happen that the SmartSensor cannot be initialised anymore when starting the system.
The decaying air permeability of the filter element is compensated for by increasing the operating vacuum within a wide range. However, in case of severe contamination, the vacuum increases to such a level that the discharge within the SmartSensor becomes unstable or is extinguished without any apparent reason. The filter element must be considered worn out when this occurs frequently.
From the 300th operating hour onwards, the operator is reminded upon starting the system that the SmartSensor must be replaced. Up to a maximum of 320 hours, the SmartSensor can still be operated after confirming the reminder. Thereafter, it must be replaced.
Exchanging the P1 SmartSensor
Before replacing the SmartSensor, the instrument must always be switched off first.
To remove the SmartSensor, simply take hold of it at the chequered section of the plug's movable section and pull the sensor straight out of the socket.
When inserting the new SmartSensor, make sure that the marks at the plug and socket line up, and that the movable section of the plug engages in the locking position by resting flush against the outer collar of the socket.
Instrument Firmware Log
Instrument Firmware | Amendement | Manual Version | PC Software |
---|---|---|---|
1.0.08 | Launch Version – P1 XTL First Edition | V1 | n.a. |
1.0.11 | Display Units ppm and g/year added Technical Data Screen added | V2 | n.a. |
1.0.12 | Battery drain Issue improved Calibration stabilisation time improved Passed calibration screen message improved SmartSensor lifetime message The firmware update process via the radio has been stabilised | V3 | n.a. |
1.0.13 | Option „C4-FN" added Reworked parameters for HV generation Reworked Zero Tracking Reworked filter for sensor noise Reworked signal cleardown Streaming selectable parameters added | V4 | 1.0.01 |
Technical Data
Parameter | Value |
---|---|
Detection principle | NIC (Negative Ion Capture) |
Detects | SF6 and gasmixtures with C4-FN (option) |
SF6 Sensitivity | Standard 1.0E-7 cc/s - 1,0 ppm - 0,01 g/a HIGHsens 1.0E-8 cc/s - 0,1 ppm - 0,001 g/a |
Response time t90 | approx. 0,5 s |
Response time t10 | approx. 0,5 s |
Alarm | Selectable Handset Vibration |
Audio | Audio LED-Projection |
Display | Frequency / repeat rate coupled to leak rate condition signals 2.8" Touch Display on handset |
Self diagnostics | Sample flow, sensor condition, battery condition, hardware fault |
Battery | Handset Li-ion, inductive charge at docking station, shelf life 12 months if charged fully |
Storage conditions | -10°C ... + 60°C |
Operating temperature | 0°C ... 50°C |
Power supply | 100 - 240V 50/60Hz |
Dimension | Handset 300x105x90 mm (H,D,W) Console 420x240x470 mm (H,D,W) |
Weight | Handset 750 g Console 9,8 kg |
EU Declaration of Conformity
According to Decision No. 90/2016 Sb. of the European Parliament and of the Council
SF6 LEAKCHECK P1 XTL
The manufacturer stated below declares that the characteristics of the product meet the required technical standards, directives and specification and that it conforms to the respective European Union harmonisation standards. Furthermore the manufacturer declares the product to be safe whilst adhering to the conditions for its correct installation, maintenance and use. This declaration of conformity is issued under the sole responsibility of the manufacturer.
Manufacturer:
ISM Deutschland GmbH, Laubach 30, 40822 Mettmann, Germany - www.ism-d.de -
Notification of quality assurance:
DIN EN ISO 9001:2015 Certificate No.: DE011836-1
Product description:
SF6 Leak detector is sensitive instrument for determining SF6 or gasmixture with C4-FN leaks.
Conformity assessment procedure:
The product's conformity was assessed with respect to the following requirements:
- EMC Directive 2014/30/EU, LVD Directive 2014/35/EU.
- It was compared with the submitted documentation.
- Issued on fundamentals of declaration of conformity of the producer – it was tested according standards.
List of standards:
- EN 55011 ed. 4:2017 + A1:2017 +A11:2020 +A2:2021
- EN 61000-3-3 ed. 3:2014 + A1:2019 +A2:2022
- EN 61000-4-2 ed. 2:2009
- EN 61000-4-4 ed. 3:2013
- EN 61000-4-6 ed. 4:2014
- EN IEC 61000-3-2 ed. 5:2019 +A1:2021
- EN IEC 61000-4-11 ed. 3:2020
- EN IEC 61000-4-3 ed. 4:2021
- EN IEC 62368-1 ed. 2+ A11:2021
Date and Place of Issue:
Mettmann 12.04.2024
Authorised representative:
Clemens A. VERLEY
CE Mark [symbol]