UNI-T UTS3000T+ Series Spectrum Analyzer
Datasheet
Product Features
- Frequency measurement range: 9 kHz to 1.5 GHz, 9 kHz to 3.2 GHz
- Display average noise level (DANL) can be as low as -161 dBm (Typical value)
- Phase noise < -98 dBc/Hz (Offset 10 kHz, typical value)
- Full amplitude Precision < 0.7 dB
- Up to 10,001 scanning points
- Minimum resolution bandwidth (RBW) 1 Hz
- Advanced function one key measurement (Option)
- EMI Pre-compliance analysis function (Option)
- Supports analog demodulation analysis (Option)
- Supports tracking generator output function
- 10.1-inch 1280 × 800 HD capacitive touch screen
- Provides USB/LAN interface, supports SCPI protocol
Multi-touch HD Screen for Quick Operation
10.1-inch multi-touch HD capacitive screen with quick menu settings. Supports multiple gesture operations such as dragging, expanding, and zooming on the trace. Convenient human-computer interaction operation solves the problem of cumbersome and difficult operation to the greatest extent.
Excellent Sensitivity to Test Weaker Signals
The weak signal test is easily affected by the noise floor of the spectrum analyzer itself. UTS3000T+ series has a DANL as low as -161 dBm, providing excellent sensitivity to effectively test weak signals.
Removable Dust Mesh
With a detachable dust filter, after the instrument is used for a period of time, the user can remove the dust from the air inlet. To ensure the reliability of the whole machine, it can avoid short-circuit, burn or fire caused by dust.
Scan 10,001 points
UTS3000T+ series provides up to 10,001 sweep points, offering higher frequency resolution and making it easier to capture signals that are difficult to detect.
Excellent Selectivity
It has a stronger capability to resolve signals of adjacent unequal amplitudes.
EMI pre-compliance (Option)
UTS3000T+ series includes optional components that, when used with near-field probes, assist in locating and resolving EMI defects in advance, thereby shortening the development cycle.
Advanced measurement (Option)
The advanced measurement mode provides the test items required by the transmitter test specification: Channel Power, T-power, Occupied BW, TOI, ACP, Spectrum Monitor, CNR, and Harmonics.
Analog demodulation analysis (Option)
Provides AM, FM analog signal for demodulation analysis.
Definitions and Conditions
"Specifications" describes the performance of the parameters covered by the product warranty in detail, unless otherwise noted, these specifications apply to the temperature range of 20°C to 30°C.
"Typical" refers to other product performance information not covered by the product warranty. 80% of the units can exhibit 95% confidence over the temperature range of 20 °C to 30 °C when performance is out of specification. Typical performance does not include measurement uncertainty.
"Nominal Value" means expected performance, or describes product performance that is useful in product applications but not covered by the product warranty.
The analyzer can meet its specifications under the following conditions:
- It is within its calibration cycle and has warmed up for at least 30 minutes.
- If the analyzer has been stored within the allowable storage temperature range but outside the allowable operating temperature range, it must be placed within the allowable operating temperature range for at least two hours before starting.
Product Function and Model Comparison Table
UTS3015T+ | UTS3032T+ | |
---|---|---|
Spectrum analysis | ● | ● |
EMI | ● | ● |
Analog demodulation | ● | ● |
Advanced measurement | ● | ● |
Tracking generator | ● | ● |
Reflection measurement | ● | ● |
Note: ● Standard ○ Option ◍ Not supported
Frequency and Time Specifications
Frequency
UTS3015T+ | UTS3032T+ | |
---|---|---|
Frequency range | 9 kHz to 1.5 GHz | 9 kHz to 3.2 GHz |
Resolution bandwidth | 1 Hz |
10 MHz internal frequency reference
Frequency reference | 10.000000 MHz |
Precision | ± [time since last adjustment x aging rate) + temperature stability +calibration Precision] |
Achievable initial calibration Precision | < 1 ppm |
Temperature stability | < 1 ppm |
Aging rate | ≤ ±1.0 ppm/ year |
Frequency readout Precision (start, stop, center, marker)
Marker resolution | Span / (Sweep point-1) |
Marker frequency uncertainty | ± (Marker frequency x Frequency reference Precision + 1 %x Span + 10 % x RBW+ Marker resolution) |
Marker Mode | Normal, Delta∆, Fixed |
Marker function | Marker Noise, Band Power, Band Density, N dB, Counter |
Counter resolution | 1 Hz |
Uncertainty of frequency counter | ± [Marker frequency x Frequency reference precision + Counter resolution] |
Frequency span (FFT and swept mode)
Sweep range | 0 Hz, 100 Hz to 1.5 GHz | 0 Hz, 100 Hz to 3.2 GHz |
Sweep Precision | Swept ±[0.25%*Span+Span / (Points-1)] | FFT ±[0.10%*Span+Span / (Points-1)] |
Sweep time and triggering
Sweep time | 1 ms to 4,000 s (span ≠ 0) |
1 μs to 4,000 s (span = 0) | |
Sweep Type Rule | Precision, Normal |
Sweep Mode | Swept (1 kHz to 1 MHz), FFT (1 Hz to 30 kHz) |
Sweep Rules | Single, Continuous |
Trigger Type | Free Run, External, Video |
External trigger input | TTL, Rising/Falling |
Resolution bandwidth (RBW)
Range (-3dB bandwidth) | 1 Hz to 1 MHz, 1-3-10 steps | |
Selectivity (-60 dB/-3 dB) | < 4.8: 1 (Nominal) | 60 dB: < 3 dB |
Bandwidth Precision (-3dB) | < 5 % (Nominal) | |
Video bandwidth (VBW) | 1 Hz to 1 MHz,1-3-10 steps | |
Uncertainty of video bandwidth | < 5% |
Amplitude Precision and Range Specifications
Amplitude range | 10 MHz to maximum frequency: (DANL) to +30 dBm | |
Reference level | -100 dBm to+30 dBm, steps 1 dB | |
Preamp | 20 dB, Nominal, 9 kHz to 1.5 GHz (3.2 GHz) | |
Input attenuator range | 0 to 51 dB, 1 dB Step | |
Maximum safe input level | ||
DC volts | 50 V DC | |
Maximum continuous wave RF power | < +33 dBm | 3 minutes, Input attenuation > 20 dB |
Display range | ||
Log scale | 1 dB to 200 dB | |
Linear scale | 0 to Reference level | |
Scale units | dBm, dBmV, dBμV, V, W | |
Sweep (trace) point range | 10,001 | |
Number of traces | 4 | |
Detector | Sample, Peak, Negative, Normal, Average | |
Trace Type | Clear/Write, Average, Max Hold, Min Hold | |
Frequency response | 20°C to 30°C, 30% to 70% relative humidity, Input attenuation 20 dB, be relative to 50 MHz. | |
Preamp Off | 9 kHz to 3.2 GHz | ±0.6 dB; ± 0.3 dB, Typical |
Preamp On | 100 kHz to 3.2 GHz | ±1.0 dB; ± 0.8 dB, Typical |
Error and precision | ||
Resolution bandwidth switching uncertainty | Relative to 10 kHz RBW logarithmic resolution ± 0.2 dB, linear resolution ± 0.01, Nominal | |
Input attenuation switching uncertainty | 20 to 30 °C, fc=50 MHz, Preamp Off, Relative to 20 dB attenuation, Input attenuation 1 to 51 dB | ± 0.5 dB |
Dynamic Range Specifications
1 dB gain compression
20 to 30 °C, fc ≥ 50 MHz, Input attenuation 0 dB, Preamp off
> -5 dBm, Nominal
Displayed average noise level (DANL)
20 to 30 °C, 0dB RF attenuation, RBW=1 Hz, VBW=1 Hz, sample detector, average > 50
UTS3015T+ | UTS3032T+ | |
---|---|---|
Preamp off | ||
9 kHz to 500 kHz | -130 dBm (Nominal) | -105 dBm (Nominal) |
500 kHz to 1 MHz | -143 dBm, -145 dBm (Typical) | -115 dBm, -120 dBm (Typical) |
1 MHz to 10 MHz | -142 dBm, -144 dBm (Typical) | -127 dBm, -130 dBm (Typical) |
10 MHz to 200 MHz | -142 dBm, -143 dBm (Typical) | -142 dBm, -145 dBm (Typical) |
200 MHz to 1.5 GHz | -140 dBm, -142 dBm (Typical) | -143 dBm, -146 dBm (Typical) |
1.5 GHz to 3.2 GHz | --- | -140dBm,-143dBm (Typical) |
Preamp on | ||
9 kHz to 500 kHz | -145 dBm (Nominal) | -125 dBm (Nominal) |
500 kHz to 1 MHz | -155 dBm, -157 dBm (Typical) | -130 dBm, -135 dBm (Typical) |
1 MHz to 10 MHz | -155 dBm, -158 dBm (Typical) | -145 dBm, -147 dBm (Typical) |
10 MHz to 200 MHz | -158 dBm, -160 dBm (Typical) | -158 dBm, -160 dBm (Typical) |
200 MHz to 1.5 GHz | -159 dBm, -161 dBm (Typical) | -161 dBm, -164 dBm (Typical) |
1.5 GHz to 3.2 GHz | -159 dBm, -161 dBm (Typical) | -159 dBm, -161 dBm (Typical) |
Spurious responses
Second harmonic distortion (SHI)
20 to 30 °C, Preamp off, Signal input-30 dBm, 0dB RF attenuation
Absolute amplitude Precision | 20 to 30 °C, fc=50 MHz, RBW=1 kHz, VBW=1 kHz, Peak detectors, Input attenuation20 dB | ± 0.4 dB, Input signal level -20 dBm, Preamp Off |
± 0.5 dB, Input signal level -40 dBm, Preamp On | ||
Total absolute amplitude Precision | 20 to 30 °C, fc > 100 kHz, Input signal level -50 dBm to 0 dBm, RBW = 1 kHz, VBW = 1 kHz, Peak detectors, Input attenuation 20 dB, Preamp Off, 95% confidence | ± (0.4 dB+ Frequency response) |
Input voltage standing wave ratio (VSWR) | 1 MHz to 1.5 GHz | ≤ 1.8 (Nominal) |
1 MHz to 3.2 GHz | ≤ .8 (Nominal) |
Phase noise
20 to 30 °C, fc = 1 GHz, RBW=1 kHz, VBW=10 Hz, Sampling detection, Log avg, avg> 50
Offset | UTS3015T+ | UTS3032T+ |
---|---|---|
10 kHz | -95 dBc/Hz, -98 dBc/Hz (Typical) | -95 dBc/Hz, -98 dBc/Hz (Typical) |
100 kHz | -96 dBc/Hz, -98 dBc/Hz (Typical) | -93 dBc/Hz, -98 dBc/Hz (Typical) |
1 MHz | -115 dBc/Hz, -120 dBc/Hz (Typical) | -115 dBc/Hz, -120 dBc/Hz (Typical) |
Tracking Generator Specifications
Frequency
Frequency range | 100 kHz to 1.5 GHz | 10 MHz to 3.2 GHz |
Counter resolution | 10 Hz | |
Output power level | -40 dBm to 0 dBm | 0.5 dB |
Resolution | be relative to 50 MHz | |
Flatness output | ± 3 dB | |
Maximum safe reverse input level | ||
Average total power | 30 dBm | |
AC coupling | ±50 VDC |
Analog Demodulation Analysis (Option)
Demodulation
Frequency range | 2 MHz to 1.5 GHz | 2 MHz to 3.2 GHz |
Carrier power Precision | ± 2 dB | |
Input power | -30 dB to +20 dBm Automatic attenuation | |
Carrier power display resolution | 0.01 dBm | |
AM measurement | ||
Modulation rate | 20 Hz to 100 kHz | Modulation rate <1 kHz |
Precision | 1 Hz (Nominal) | < 0.1%Modulation rate (Nominal) Modulation rate ≥1 kHz |
Depth | 5 to 95% | |
Precision | ± 4% (Nominal) | |
FM measurement | ||
Modulation rate | 20 Hz to 100 kHz | Modulation rate < 1 kHz |
Precision | 1 Hz (Nominal) | < 0.1% Modulation rate (Nominal) Modulation rate ≥ 1 kHz |
Frequency offset | 1 kHz to 400 kHz | |
Precision | ± 4% (Nominal) |
EMI (Option)
EMI Resolution bandwidth
Resolution bandwidth (-6dB) | 200 Hz, 9 kHz, 120 kHz, 1 MHz |
Resolution bandwidth Precision | < 5%, (Nominal) |
EMI detector
EMI detector: Peak, Negative Peak, Quasi Peak, EMI Average, Average
EMI Main function
Main function: EMI Standard: CISPR; View: Scan table, Meter, Signal table; Meter control; Avg settings; Limit: AS-NZS, BellCore, DEF-STAN, DO-160, EN, FCC, GB9254, MIL-461, VCCI and Custom; Signal table settings; Scan table settings; Scan Sequence: Scan, Search, Scan-Search-Meas, Scan-Search, Search-Meas, Measure; Sig Detector; Output report
Reflection measurement
Incentive and reflex measurement
Cursor measurement | VSWR, Return loss, Reflection coefficient |
Calibration type | Open |
Excitation power | -20 to 0 dBm |
Advanced measurement kit (Option)
Power Measurement
Channel power | Channel power, Power spectral density |
ACP (Adjacent Channel Power) | Main CH Power, Left channel power, Right channel power |
Occupied bandwidth | Occupied Bandwidth, Transmit Frequency Error |
Time domain power | Zero Span Integrated Power |
CNR (Carrier Noise Ratio) | C/N, Noise Power |
Non-Linear Measurement
TOI, Third-order intercept | Measure the third-order products from two tones |
Harmonic measurement | Max Harmonic number 10 |
Spectrum Monitor Measurement
Spectrogram |
Interface and display
Common interface
RF input | Type-N female, 50 Ω, nominal |
Front panel trace source output | Type-N female, 50 Ω, nominal |
10 MHz Ext Ref In | 10 MHz, > 0 dBm, BNC female, 50 Ω, nominal |
10 MHz out | 10 MHz, -5 dBm to +10 dBm, BNC female, 50 Ω, nominal |
External trigger input | TTL, BNC female |
HDMI display | HDMI 1.4 display interface |
USB-Host | USB-A |
USB-Device | USB-B |
LAN | LAN (VXI11), 10/100/1,000 Base, RJ-45 |
Headphone Jack | 3.5 mm (1/8 inch) miniature stereo audio jack |
Display screen
Display type | 10.1-inch capacitive multi-touch panel |
Display resolution | 1280×800, RGB Vertical pixel |
General Technical Specifications
Specifications
Supply voltage | 100 to 240 VAC (Fluctuations ± 10%) | 100 to 120 VAC (Fluctuations ± 10%) | |
Frequency | 50/60 Hz | 400 Hz | |
Environment | |||
Temperature range | operation: 0°C to +40°C | ||
Non-operating: -20°C to +70°C | |||
Cooling method | Fan forced cooling | ||
Humidity range | Operation: Below + 35 °C ≤ 90% relative humidity; Non-operating: + 35 °C to +40 °C ≤ 60% relative humidity | ||
Altitude | Operation: Below 3,000 m; Non-operating: Below 15,000 m | ||
Pollution degree | 2 | ||
Operating environment | Indoor use | ||
Mechanical specifications | |||
Dimensions | 378mm×218mm×120mm (Width x Height x Length) | ||
Net weight | 4.55 kg | ||
Calibration period | The recommended calibration period is one year |
Regulatory standards
EMC
Compliance with EMC directives(2014/30/EU), Conform to or better than IEC 61326-1:2021/EN61326-1:2021, IEC 61326-2-1:2021/EN61326-2-1:2021
Conductive disturbance | CISPR 11/EN 55011 | CLASS B group 1, 150kHz-30MHz |
Radiation disturbance | CISPR 11/EN 55011 | CLASS B group 1, 30MHz-1GHz |
(ESD)Electrostatic discharge (ESD) | IEC 61000-4-2/EN 61000-4-2 | ±4.0 kV (Contact), ±8.0 kV (air) |
Radio frequency electromagnetic field immunity | IEC 61000-4-3/EN 61000-4-3 | 3 V/m (80 MHz to 1 GHz) ; 0 V/m (1.4 GHz to 6 GHz) |
(EFT)Electrical fast transient burst (EFT) | IEC 61000-4-4/EN 61000-4-4 | ±1 kV (AC input port) |
Surge | IEC 61000-4-5/EN 61000-4-5 | ±0.5 kV (Live line to zero line) ±1 kV (Fire/zero line to ground) |
Immunity to RF continuous conduction | IEC 61000-4-6/EN 61000-4-6 | 3 V, 0.15-80 MHz |
Voltage dips and short interruptions | IEC 61000-4-11/EN 61000-4-11 | Voltage dip: 0% UT during 0.5 cycle; 0% UT during 1 cycle; 70% UT during 25/30 cycles Short Interruption: 0% UT during 250/300 cycles |
Safety regulations
EN 61010-1:2010+A1:2019
EN IEC61010-2-030:2021+A11:2021
UL 61010-1:2012 Ed.3+ R:19 Jul2019
UL 61010-2-030:2018 Ed.2
CSA C22.2#61010-1:2012 Ed.3+U1;U2;A1
CSA C22.2#61010-2-030:2018 Ed.2
Order Information
Description | Order No. |
---|---|
Spectrum analyzer, 9 kHz to 1.5 GHz | UTS3015T+ |
Spectrum analyzer, 9 kHz to 3.2 GHz | UTS3032T+ |
Standard accessories | Power cord ×1 USB cable x1 UT-D14 |
Recommended options & accessories
Options | Advanced measurement kit UTS3000-AMK EMI measurement option UTS3000-EMI Analog demodulation analysis option UTS3000-AMA |
UT-CK01 accessories kit | SMAJ-NJ-0.7M DC-6G cable x1 UT-W02-6GHz NJ-NJ-0.7M DC-6G cable x1 UT-W01-6GHz Adapter SMA-N-KJ-T DC-6GHz x2 UT-C01-6GHz Adapter N-BNC-JK DC-4GHz x2 UT-C02-4GHz Antenna 2400 MHz-2500 MHz x2 UTS-T01 Antenna 824-960 MHz/1710-1990 MHz x2 UTS-T02 50Ω-SMA-SMB cable x1 UT-W03 |
UTS-EMI01 | Adapter SMA-N-KJ-T DC-6 GHz x1 UT-C01 Near field probe, frequency range 30 MHz-3 GHz, Detection range 10 cm x1 NFP-3G-P1 Near field probe, frequency range 30 MHz-3 GHz, Detection range 3 cm x1 NFP-3G-P2 Near-field probes kit Near field probe, frequency range 30 MHz-2 GHz, resolving power 5 mm x1 NFP-2G-P3 Near field probe, frequency range 30 MHz-3 GHz, resolving power 2 mm x1 NFP-3G-P4 |
Reflection Measurement | VSWR bridge UT-RB60 |
Options ordering and installation
- Purchase options: Based on your requirements, please purchase the specified function options from Uni-t Sales Personnel and provide the serial number of the instrument that needs the option installed.
- Receive certificate: You will receive the license certificate based on the address provided in the order.
- Register and obtain license: Visit the Uni-t official website license activation session for registration. Use the license key and instrument serial number provided in the certificate to obtain the option license code and license file.
- Install the option: Download the option license file to the root directory of a USB storage device, and connect the USB storage device to the instrument. Once the USB storage device is recognized, the Option Install menu will be activated. Press this menu key to begin installing the option.
Limited Warranty and Liability
UNI-T guarantees that the Instrument product is free from any defect in material and workmanship within three years from the purchase date. This warranty does not apply to damages caused by accident, negligence, misuse, modification, contamination, or improper handling. If you need a warranty service within the warranty period, please contact your seller directly. UNI-T will not be responsible for any special, indirect, incidental, or subsequent damage or loss caused by using this device. For the probes and accessories, the warranty period is one year. Visit instrument.uni-trend.com for full warranty information.
Learn more at: www.uni-trend.com
Register your product to confirm your ownership. You will also get product notifications, update alerts, exclusive offers and all the latest information you need to know.
UNI-T is the licensed trademark of UNI-TREND TECHNOLOGY (CHINA) CO., Ltd. UNI-T products are protected under patent laws in China and internationally, covering both granted and pending patents. Licensed software products are the properties of UNI-Trend and its subsidiaries or suppliers, all rights reserved. This manual contains information that replaces all earlier published versions. The product information in this document subject to update without notice. For more information on UNI-T Test & Measure Instrument products, applications, or service, please contact UNI-T instrument for support, the support center is available on www.uni-trend.com ->instruments.uni-trend.com https://instruments.uni-trend.com/ContactForm/
Headquarter
UNI-TREND TECHNOLOGY (CHINA) CO., Ltd.
Address: No.6, Industrial North 1st Road, Songshan Lake Park, Dongguan City, Guangdong Province, China
Tel: (86-769) 8572 3888
Europe
UNI-TREND TECHNOLOGY EU GmbH
Address: Steinerne Furt 62, 86167 Augsburg, Germany
Tel: +49 (0)821 8879980
North America
UNI-TREND TECHNOLOGY US INC.
Address: 2692 Gravel Drive, Building 5, Fort Worth, Texas 76118
Tel: +1-888-668-8648
Copyright © 2025 by UNI-Trend Technology (China) Co., Ltd. All rights reserved.
Related Documents
![]() |
UTS3000T+ Series Spectrum Analyzer Data Sheet Comprehensive data sheet for the UNI-T UTS3000T+ Series Spectrum Analyzer, detailing product features, specifications, and ordering information for models UTS3036T+ and UTS3084T+. |
![]() |
UTS1000B/T Series Spectrum Analyzer Datasheet Datasheet for the UNI-T UTS1000B/T Series Spectrum Analyzer, detailing its features, specifications, and options. Covers frequency and time specifications, amplitude precision, dynamic range, phase noise, and regulatory standards. |
![]() |
UTS3000B Series Spectrum Analyzer User's Manual User's manual for the UNI-T UTS3000B Series Spectrum Analyzer, detailing its functions, features, front and rear panel operations, and safety precautions. |
![]() |
MSO3000HD Series High-Resolution Oscilloscope Datasheet Datasheet for the UNI-T MSO3000HD Series High-Resolution Oscilloscope, detailing its features, specifications, and applications including digital oscilloscope, logic analyzer, and power analyzer functions. |
![]() |
UNI-T MSO3000HD Series High-Resolution Oscilloscope Datasheet Explore the UNI-T MSO3000HD Series High-Resolution Oscilloscope Datasheet. Discover its advanced features, including a 500 MHz bandwidth, 2.5 GSa/s sampling rate, 12-bit resolution, and a 10.1-inch capacitive touch screen. Learn about its multiple instrument functions, such as digital oscilloscope, logic analyzer, and spectrum analyzer, designed for efficient and versatile operation. |
![]() |
UNI-T MSO1000HD Series High-Resolution Oscilloscopes Datasheet Comprehensive datasheet for the UNI-T MSO1000HD Series High-Resolution Oscilloscopes, detailing features, specifications, performance characteristics, application scopes, and accessories. |
![]() |
UNI-T MSO1000HD Series High-Resolution Oscilloscopes Data Sheet Data sheet for the UNI-T MSO1000HD Series High-Resolution Oscilloscopes, detailing features, specifications, performance characteristics, and accessories. |
![]() |
UNI-T MSO1000HD Series High-Resolution Oscilloscopes Data Sheet Comprehensive data sheet for the UNI-T MSO1000HD Series High-Resolution Oscilloscopes, detailing features, specifications, performance characteristics, probes, accessories, and order information. |