NI PXIe-4302/4303 and RM-4302 User Guide and Specifications

NI PXIe-4302/4303, 32 Ch, 24-bit, Filtered Analog Input Module, RM-4302 Rack-Mount Terminal Block

Introduction

This document explains how to install, configure, and set up the NI PXIe-4302 and NI PXIe-4303 simultaneous filtered modules with the RM-4302 rack-mount terminal block. Driver support for the PXIe-4302/4303 was first available in NI-DAQmx 15.1.1. Driver support for the RM-4302 was first available in NI-DAQmx 15.5. For the list of devices supported by a specific release, refer to the NI-DAQmx Readme, available on the version-specific download page or installation media.

Note: The keying of this terminal block prevents it from connecting to other modules that could be damaged by the voltage present on the terminal block. However, you should only use these terminal blocks with their supported modules.

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment.

This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation.

Furthermore, any modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules.

⚠️ Caution: To ensure the specified EMC performance, operate this product only with shielded cables and accessories. Shielded, twisted pair cable must be used for all RJ-50 ports and the CAL IN port. A shielded cable is not required for the cable connected to the EXT PWR port.

⚠️ Caution: To ensure the specified EMC performance, the length of any cable connected to the EXT PWR and PFI0 ports must be no longer than 3 m (10 ft).

⚠️ Caution: To ensure the specified EMC performance, the length of any cable connected to the Channel (RJ-50) ports must be no longer than 30 m (100 ft).

What You Need to Get Started

To set up and use the RM-4302 rack-mount terminal block with the PXIe-4302/4303 module, you need the following items:

Hardware

Tools

Documentation

You can download needed documents from ni.com/manuals.

Installation

Install the Software

? Note: You must be an administrator to install NI software and devices on your computer.

Ensure that the following software is installed before installing the SC Express hardware:

  1. Your application software, such as LabVIEW, LabWindows™/CVI™ or .NET.
  2. NI-DAQmx—PXIe-4302/4303 modules were first supported in NI-DAQmx 15.1.1. RM-4302 terminal blocks were first supported in NI-DAQmx 15.5.

? Note: For detailed NI software version support, refer to the NI-DAQmx Readme. Back up any applications before upgrading software or modifying the application.

Unpack and Install the Module

Remove the packaging and inspect the module. Contact NI if the module is damaged. Do not install a damaged module.

⚠️ Caution: The module is static sensitive. Always properly ground yourself and the equipment when handling or connecting to the module.

Complete the following steps to install the SC Express module while referring to Figures 1 and 2:

? Note: To maintain forced air cooling in the PXIe system, refer to the Maintain Forced-Air Cooling document.

  1. Make sure the chassis is powered off.
  2. Connect the power cord to your chassis and then plug it into the power supply receptacle before installing the PXIe-4302/4303. The power cord grounds the chassis and protects it from electrical damage while you install the module.

⚠️ Caution: To protect yourself, the chassis, and the PXIe-4302/4303 from electrical hazards, leave the chassis powered off until you complete the PXIe-4302/4303 module installation.

Step 3: Identify Slot

Identify a supported PXI Express slot in the chassis. SC Express devices can be placed only in PXI Express Peripheral slots, PXI Express Hybrid Peripheral slots, and PXI Express System Timing slots. Refer to the chassis documentation for details.

Figure 1. Symbols for PXI Express/PXI Express Hybrid/PXI Slots

Description of Figure 1: The diagram shows symbols for different PXI Express slots. Symbol 1 represents the PXI Express System Controller Slot. Symbol 2 represents the PXI Peripheral Slot. Symbol 3 represents the PXI Express Hybrid Peripheral Slot. Symbol 4 represents the PXI Express System Timing Slot.

Step 4: Discharge Static Electricity

Remove the filler panel, and touch any metal part of the chassis to discharge static electricity.

Step 5: Place Module in Guides

Place the module edges into the module guides at the top and bottom of the slot.

⚠️ Caution: When installing the module, make sure both edges are positioned inside the guides and that the module components do not come into contact with adjacent modules.

Step 6: Slide Module into Slot

Slide the module along the guides until it reaches the rear connector then seat the module by pushing the front panel until it is flush with the front panel of the chassis.

Step 7: Secure Module

Secure the module to the chassis using the front-panel captive screws, shown in Figure 2. Tighten the screws to 0.31 N·m (2.7 lb·in.).

Figure 2. Installing PXIe-4302/4303 Modules

Description of Figure 2: This illustration shows the process of installing PXIe-4302/4303 modules into a PXIe chassis. It highlights the captive screws used for securing the module.

Unpack and Install the Rack-Mount Terminal Block, Chassis, and Cables

Remove the packaging and inspect the terminal block, chassis, and cables. Contact NI if there is any damage. Do not install a damaged product.

⚠️ Caution: The module and terminal block are static sensitive. Always properly ground yourself and the equipment when handling or connecting to the module and terminal block.

⚠️ Caution: Do not supply hazardous voltages (>30 Vrms/42.4 Vpk/60 VDC) to the terminal block.

? Note: The chassis can be powered on or off when connecting the terminal block.

To install the rack-mount terminal block into the rack, complete the following steps while referring to Figure 3:

  1. Place the RM-4302 in the desired position.
  2. Install and tighten the four rack-mounting screws.

If the chassis is not already installed in the rack, complete the following steps:

  1. Attach the PXIe chassis rack-mount hardware to the PXIe chassis. Refer to your rack-mount kit installation guide for more details.
  2. Mount the PXIe chassis in the rack in the desired position. Refer to your rack-mount kit installation guide for more details.

Figure 3. RM-4302 Installation and Front Connections

Description of Figure 3: This diagram illustrates the installation of the RM-4302 rack-mount terminal block and the PXIe chassis into a 19-inch rack. It shows rack-mounting screws, the terminal block itself, and the PFI BNC cable and RJ50 connectors.

Cable Installation

To install the cables, complete the following steps while referring to Figures 3 and 6:

  1. Connect the RJ50 cable connectors to the appropriate RJ50 connector receptacles. Refer to the RJ50 section for more information.
  2. Make any of the following connections as needed for your application:
    • Connect the BNC cable to the PFIO receptacle.
    • Connect the external power source:
      1. Remove the 2-pin EXT PWR connector from the RM-4302.
      2. Install the external power wires. Refer to the EXT PWR section for more information.
      3. Install the 2-pin COMBICON cable into the RM-4302.
    • Connect calibration input:
      1. Remove the 2-pin COMBICON connector from the RM-4302.
      2. Install the calibrator input wires. Refer to the In-Situ Calibration section for more information.
      3. Connect the CAL IN cable shield to AIGND, Pin 2 and 3 of CAL BIAS Connector, as shown in Figure 4.

Figure 4. CAL BIAS Cable Connections

Description of Figure 4: This diagram illustrates the CAL BIAS cable connections. It shows a signal source connected through ferrites and a shield to AIGND, CAL IN+, and CAL IN- terminals on the RM-4302. AIGND is labeled as analog input ground.

d. For CAL IN cable lengths greater than 3 meters, install the included clamp-on ferrite beads on both ends of the CAL IN cable. Refer to Figure 5 to install the clamp-on ferrite beads.

? Note: Clamp-on ferrite beads must be installed on the cable as close to the RM-4302 and to the source as possible. Placing the clamp-on ferrite beads elsewhere on the cable noticeably impairs its effectiveness.

Figure 5. Clamp-On Ferrite Bead Installation

Description of Figure 5: This illustration shows how to install clamp-on ferrite beads onto a cable. It depicts the beads being attached near the connectors.

e. Install the 2-pin COMBICON cable into the RM-4302.

Connect the calibration bias resistors:

  1. Remove the 4-pin COMBICON connector from the RM-4302.
  2. Install the bias resistors. Refer to the CAL IN section for more information.
  3. Install the 4-pin COMBICON cable into the RM-4302.

3. Connect the chassis ground wire to the chassis ground screw.

Figure 6. Cables, Resistor, and Standoff Installation

Description of Figure 6: This diagram shows the complete installation of cables, resistors, and standoffs for the RM-4302 and PXIe module. It labels components such as the RM-4302 terminal block, PXIe chassis, PXIe-4302/4303 module, standoffs, jack screws, SH96-96-2 cable, PXIe controller, calibration input connector, bias resistor connector, bias resistor, EXT PWR connector, chassis ground screw, and chassis ground cable.

4. Screw the two standoffs into the PXIe-4302/4303 module. Tighten the standoffs to 1.30 N·m (11.5 lb·in.).

5. Connect the male end of the SH96-96-2 cable to the RM-4302 and tighten the two jack screws to 0.90 N·m (8.0 lb·in.).

? Note: For safety purposes and to prevent damage to equipment when high voltages are present, the SH96-96-2 cable is keyed to prevent connection between incompatible terminal blocks, modules, and/or cables.

6. Connect the female end of the SH96-96-2 cable to the PXIe-4302/4303 and tighten the two jack screws to 0.90 N·m (8.0 lb·in.).

? Note: For information about sensors, go to ni.com/sensors.

7. Power on the PXIe chassis if it is powered off.

CAL IN

The CAL IN input connector on the rear of the RM-4302 allows you to connect a voltage reference that can be internally routed to the analog input of each channel (AI+, AI-) on the PXIe-4302/4303. All channels are simultaneously switched to the CAL IN by utilizing the Calibration API within NI-DAQmx. Refer to the In-Situ Calibration section for detailed information.

The calibration source voltage is connected to the CAL IN+ and CAL IN- terminals as shown in Table 1.

Table 1. Calibration Input Connector Pin Assignments

Calibration Input Pin DiagramPinSignal NameSignal Description

Diagram showing two circular connectors with pins 1 and 2.

1CAL IN-Calibration input negative
2CAL IN+Calibration input positive

CAL BIAS

You can use the connector labeled RB1 and RB2 on the rear of the RM-4302 to reference the CAL± input to analog ground (AIGND). For flexibility, access to both CAL+ and CAL- is provided. The CAL BIAS+ pin connects to CAL+ and the CAL BIAS- pin connects to CAL-. You can reference a floating voltage to AIGND by populating the RB1 position. To provide a differential voltage that is referenced to AIGND with zero common mode, populate both RB1 and RB2 with equal value resistors. When choosing resistor values, consider the output impedance of the external reference voltage source. Refer to Table 2 for connections.

Table 2. Bias Resistor Connector

Bias Resistor ConnectorPinSignal NameSignal Description

Diagram showing four numbered terminals (4, 3, 2, 1) in a row.

1CAL BIAS-RB2 calibration bias resistor for CAL-
2AIGNDAnalog input ground
3AIGNDAnalog input ground
4CAL BIAS+RB1 calibration bias resistor for CAL+

In-Situ Calibration

The RM-4302 provides the ability to perform in-situ calibration of input channels referenced to an external traceable voltage source that you can use to connect to the CAL IN connector. In-situ calibration offers a means to determine measurement errors and uncertainties compared to a traceable standard. The calibration data can be used to compensate for changes in the hardware, resulting in a more accurate measurement. In-situ calibration can be performed before and after test acquisition, or on a periodic schedule, to ensure data validity without physically disconnecting any inputs from the RM-4302.

During in-situ calibration, the input channels are disconnected internally from the RJ50 connectors and connected to a calibration bus that is driven by the calibration voltage reference connected to CAL IN±. Any PXIe-4302/4303 circuitry drift since the last full calibration will be available through measurement of these channels. In-situ calibration performs a verification, but not an adjustment. Therefore, the results from in-situ calibration will not rewrite the calibration coefficients to the PXIe-4302/4303. Instead these results can be applied to the test setup, test results, or to the PXIe-4302/4303 programmatically.

To assist you with starting development of an application with in-situ calibration, an example program is available for download from the NI Community. Refer to ni.com/info and enter Info Code exsueg.

EXT PWR

The external power input connector on the rear of the RM-4302 allows you to connect a positive voltage supply for use in powering powered sensors. The external voltage source can be either ground referenced or floating. The EXT PWR- input is connected to the ground of the RM-4302. Do not apply a ground referenced voltage to the EXT PWR- input; doing so can result in excessive current conducted through the ground of the RM-4302. The EXT PWR+ and EXT PWR- are connected to the EXT PWR+ and EXT PWR- of each RJ50 connector on the front of the RM-4302. Refer to Table 4 for signal names and descriptions.

Table 4. External Power Connector Pin Assignments

External Power Connector Pin DiagramPinSignal NameSignal Description

Diagram showing two circular connectors with pins 1 and 2.

1EXT PWR- (GND)Ground
2EXT PWR+External power

After connecting EXT PWR to a powered sensor, a short time delay may need to be introduced in the measurement application to allow the sensor to warm up. Refer to your powered sensor documentation for any power up settling requirements.

Current Fault Protection Reset

The positive temperature coefficient (PTC) resistors used for short circuit fault protection trip by undergoing a large, abrupt change in resistance in response to an overcurrent. The tripped state of the PTC is a latched state that is sustained by a small self-heating current.

The RM-4302 has one PTC per group of 8 channels. The channels in each group are 0..7, 8..15, 16..23, 24..31. A fault on any channel will cause the entire group to trip. Channels not in that group will be unaffected.

To reset the PTC, the fault must be removed and all the current through the PTC must be momentarily interrupted from the faulted group that contains the faulted channel. After removal of the fault, you can interrupt the current by disconnecting EXT PWR or by disconnecting all channels of the faulted group. The following three steps summarize the steps required to reset a tripped PTC:

  1. Remove the short-circuit fault from the affected channel.
  2. Disconnect EXT PWR.
  3. After 10 seconds, reconnect EXT PWR.

Resources

This section describes the resources available on both the front and rear of the RM-4302. Figure 7 shows the locations of these resources.

Figure 7. RM-4302 Front and Rear View

Description of Figure 7: This diagram shows the front and rear panels of the RM-4302. The front panel displays channel connectors (CH0-CH31), STATUS LED, and PFI0. The rear panel shows connectors for EXT PWR, RB1/RB2 (Bias Resistor Connector), CAL IN, and a 96-Pin DIN Cable Connector. Numbered labels indicate specific components like the RM-4302 Product Name, LED Status Indicator, PFI0, RJ50 Connectors, Mounting Flanges, Chassis Ground, External Power COMBICON, Serial Number Label, Calibration Bias Resistor Connector, Calibration Input COMBICON, and 96-Pin DIN Cable Connector.

RJ50

Refer to Table 5 for signal names and descriptions, and the correlation to the pin numbers of the RJ50 (10p10c) modular plug. Refer to Figure 8 for default switch positions.

Table 5. RJ50 Connector Pin Assignments

RJ50* (10p10c) Modular Plug Pin Number DiagramPinSignal NameSignal Description

Diagram of an RJ50 connector with pins labeled 1 through 10.

1NCNo connection.
2AI+Positive input signal—connected to the AI+ input of the PXIe-4302/4303 or disconnected when the channel is connected to CAL_IN+
3AI-Negative input signal—connected to the AI- input of the PXIe-4302/4303 or disconnected when the channel is connected to CAL_IN-
4NCNo connection.
5NCNo connection.
6EXT PWR+External power +.
7EXT PWR-External power -.
8NCNo connection.
9AIGNDAnalog ground.
10NCNo connection.

* NI does not recommend using an RJ45 cable with the RM-4302 as it can physically damage pins 1 and 10 on the RJ50 connector.

Note: Use pins 2 and 3, and 6 and 7 as shielded pairs.

Note: Refer to Figure 8 for default positions

Status LED

The green status LED, when lit, indicates the RM-4302 is connected to a powered PXIe-4302/4303 module. The location of the Status LED is shown in Figure 7.

PFI0

The PFI0 connector provides a female BNC connection to the PFI trigger line of the PXIe-4302/4303. The RM-4302 passes this signal directly through. Refer to the NI PXIe-4302/4303 User Manual for more information about using the PFI trigger line. Refer to Table 6 for signal names and descriptions.

Table 6. BNC Connector Pin Assignments

BNC PFI0 Connector Pin DiagramPinSignal NameSignal Description

Diagram showing a BNC connector with pins labeled 1 and 2.

1PFIOPFI input
2DGNDDigital ground

Switching

The RM-4302 provides software controlled switching that allows you to programmatically control whether the attached PXIe-4302/4303 is connected to the CAL IN source on the back of the RM-4302 or the RJ50 connectors on the front of the RM-4302. Figure 8 illustrates the switching provided.

Figure 8. RM-4302 Switching Configurations

Description of Figure 8: This block diagram illustrates the switching configurations within the RM-4302. It shows how External Power+, External Power-, RJ50 signals (AIGND, AI+, AI-), and CAL IN signals (CAL_IN+, CAL_IN-) are routed, indicating connections to the PXIe-4302.

Confirm SC Express Module Recognition

To confirm module recognition, complete the following steps:

? Note: Software support for the PXIe-4302/4303 is provided by NI-DAQmx. The Read Me First: NI-DAQmx and DAQ Device Installation Guide, which you can download at ni.com/manuals, offers step-by-step NI-DAQmx instructions for installing software and hardware, configuring channels and tasks, and getting started developing an application. For detailed NI software version support, refer to the NI-DAQmx Readme.

  1. Launch MAX.
  2. Expand Devices and Interfaces, and expand the chassis in which the module is located to confirm that MAX detects the module and terminal block. The terminal block should appear beneath its associated module. If your module or terminal block is not listed, press <F5> to refresh MAX. If the module is still not recognized, go to ni.com/support/daqmx.

? Tip: Using an NI-DAQmx simulated device, you can test NI-DAQmx applications without installing hardware. Refer to the Create a Simulated Device section for instructions for creating NI-DAQmx simulated devices.

3. Right-click the module name and select Self-Test. When the self-test completes, a verification message appears. If an error occurs, refer to ni.com/support/daqmx.

For information about sensors, go to ni.com/sensors.

Run Test Panels

  1. In MAX, expand Devices and Interfaces select the chassis in which the module is located.
  2. Right-click the device, and select Test Panels.

3. Click Start to test device functions, or Help for operating instructions.

To troubleshoot errors, refer to the NI-DAQmx Help, or go to ni.com/support.

Take an NI-DAQmx Measurement

NI-DAQmx Channels and Tasks

Refer to the NI-DAQmx Help for complete information about channels and tasks.

Use the DAQ Assistant to configure virtual channels and tasks in MAX or in your application.

Configure a Task Using the DAQ Assistant in MAX

Complete the following steps to create a task using the DAQ Assistant in MAX:

  1. In MAX, right-click Data Neighborhood and select Create New to open the DAQ Assistant.
  2. Select NI-DAQmx Task and click Next.
  3. Select Acquire Signals.
  4. Select Analog Input, and the measurement type, such as Voltage.
  5. Select the physical channel(s) to use and click Next.
  6. Name the task and click Finish.
  7. Configure the individual channel settings. Each physical channel you assign to a task receives a virtual channel name. To modify the input range or other settings, select the channel. Click Details for physical channel information. Configure the timing and triggering for your task. Click Run.

Use Your SC Express Device in an Application

For NI software version compatibility, refer to the NI-DAQmx Readme, available from Start»All Programs»National Instruments»NI-DAQmx.

To get started with data acquisition in your application software, refer to the tutorials listed in Table 7.

Table 7. DAQ Assistant Tutorial Locations

ApplicationTutorial Location
LabVIEWGo to ni.com/dataacquisition/videos.
LabWindows/CVIGo to Help»Contents. Next, go to Using LabWindows/CVI»Data Acquisition»Taking an NI-DAQmx Measurement in LabWindows/CVI.
Measurement StudioGo to NI Measurement Studio Help»Getting Started with the Measurement Studio Class Libraries»Measurement Studio Walkthroughs»Walkthrough: Creating a Measurement Studio NI-DAQmx Application.
NI SignalExpress*Go to Help»Taking an NI-DAQmx Measurement in SignalExpress.

* NI SignalExpress, an easy-to-use configuration-based tool for data logging applications, is at Start»All Programs»National Instruments»NI SignalExpress.

Programming Examples

NI-DAQmx includes example programs to help you get started developing an application. LabVIEW and CVI examples are located at Help»Find Examples in your application software. Text-based code examples are located at All Programs»National Instruments»NI-DAQmx» NI-DAQmx Examples. Modify example code and save it in an application, or use examples to develop a new application or add example code to an existing application.

For other examples, go to ni.com/info and enter the Info Code daqmxexp. For additional examples, refer to ni.com/support.

Removal

Terminal Block and Cables Removal

To remove the RM-4302 and PXIe-4302/4303, complete the following steps while referring to Figures 3 and 6:

? Note: The chassis can be powered on or off when removing the terminal block.

  1. Remove the RJ50 connector cables from the terminal block.
  2. Remove any of the following connections as needed for your application:
    • Disconnect the BNC cable from the PFIO receptacle.
    • Disconnect the 4-pin calibration bias resistor connector from the 4-pin bias resistor receptacle.
    • Disconnect the 2-pin EXT PWR connector from the 2-pin EXT PWR receptacle.
    • Disconnect the 2-pin CAL IN connector from the 2-pin CAL IN receptacle.
    • Disconnect ground wire from the ground screw.
  3. Loosen the four jack screws and remove the SH96-96-2 cable from the terminal block and the module.

? Note: If required for your application, remove the two standoffs from the module.

  1. While supporting the terminal block, remove the four rack-mounting screws.
  2. Remove the terminal block from the rack.
  3. Store the terminal block in an antistatic protective bag.

Module Removal

⚠️ Caution: Do not remove a module with the PXIe chassis powered on. Doing so can damage the module.

To remove the PXIe-4302/4303 module, refer to Figure 9 while completing the following steps:

? Note: To maintain forced air cooling in the PXIe system, refer to the Maintain Forced-Air Cooling document.

  1. Make sure that the chassis is powered off before removing the module.
  2. Loosen the captive screws on the top and bottom of the module.

⚠️ Caution: Do not pull the front panel to remove the module. Doing so may cause module components to come into contact with adjacent modules, causing damage to the modules.

3. Pull the captive screws to unseat the module and slowly slide the module along the guides.

4. Replace the filler panel in the empty slot.

5. Store the module in an antistatic protective bag.

Figure 9. Removing SC Express Modules

Description of Figure 9: This diagram illustrates the process of removing SC Express modules from a PXIe chassis. It shows the captive screws being used to unseat the module and the module being slid out of the slot.

Create a Simulated Device

To run examples without the hardware installed, use an NI-DAQmx simulated device. To create a simulated device in MAX:

  1. Launch MAX.
  2. Right-click Devices and Interfaces»Create New.
  3. From the dialog, select Simulated NI-DAQmx Device or Modular Instrument.
  4. Type 4302 or 4303 in the text box at the top of the window.
  5. Select the device from the list provided.
  6. Click OK.
  7. To simulate an accessory:
    1. Right-click the simulated PXIe-4302(or PXI-4303)»Configure...
    2. Select the Accessory from the Connector 0 drop down.
    3. Click OK.

More Information

After you install NI-DAQmx, the NI-DAQmx documentation is available from Start»All Programs»National Instruments»NI-DAQ. Additional resources are online at ni.com/gettingstarted.

You can access online device documentation by right-clicking your module in MAX and selecting Help»Online Device Documentation. A browser window opens to ni.com/manuals with the results of a search for relevant documents. If you do not have web access, documents for supported modules are included on the NI-DAQmx media.

Troubleshooting

Specifications (RM-4302)

? Note: PXIe-4302/4303 module specifications are located in the NI PXIe-4302/4303 Specifications document.

All specifications are typical for the range of 0 °C to 55 °C unless otherwise specified.

Electrical

Input voltage range ........................................... Refer to the NI PXIe-4302/4303 Specifications document.

Input Characteristics

Number of channels.......................................... 32 analog input channels

AI° differential input

Range ........................................................ Refer to the NI PXIe-4302/4303 Specifications document

Offset voltage¹ .......................................... 10 µV max @ 23 °C ±5 °C
4 µV typ @ 23 °C ±5 °C

Offset voltage stability.............................. 0.5 µV/°C max
0.05 µV/°C

¹ The offset voltage specification of the RM-4302 is an additional offset error that should be added to the offset error of the PXIe-4302/4303 to determine the total system offset error.

AI° to CAL° offset matching ...........................2 µV

CAL IN .............................................................±10 V

RB1, RB2..........................................................application specific

External Power (EXT PWR)

Voltage range ....................................................6 VDC to 28 VDC single ended

Current rating ....................................................3.2 A max
100 mA per EX+

? Note: NI recommends a current limited power supply with 20 A maximum output.

Path resistance...................................................0.7 Ω (EXT PWR to EX+ channel)

Protection ..........................................................Short circuit protected

Protection type ..........................................Positive temperature coefficient (PTC) resistor

Mechanical

RJ50 receptacle

Mean number of insertions before failure.............................................500

Channel configuration switches

Switch type ...............................................Electromechanical relay, latching

Expected switch life

Mechanical........................................5 × 107 cycles

Electrical ...........................................2 × 105 cycles

Connectors

RJ50 ..................................................................32

BNC ..................................................................1

96-pin DIN ........................................................1

COMBICON 2 position ....................................2

COMBICON 4 position ....................................1

Physical Characteristics

Figure 10. RM-4302 Dimensions

Description of Figure 10: This diagram shows the physical dimensions of the RM-4302. It indicates a width of 482.6 mm (19 in.), a height of 43.94 mm (1.730 in.), and a depth of 156.21 mm (6.150 in.). It also shows the front panel layout with status indicators and connectors.

Weight............................................................... 1579 g (55.7 oz)

⚠️ Caution: Clean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to service.

Environmental

Maximum altitude............................................. 2,000 m (800 mbar), at 25 °C ambient temperature

Pollution Degree ............................................... 2

Indoor use only

Operating Environment

Ambient temperature range .............................. 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.)

Relative humidity range.................................... 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

Storage Environment

Ambient temperature range ..............................-40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.)

Relative humidity range....................................5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

Shock and Vibration

Operating shock ................................................30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.)

Random vibration

Operating ..................................................5 Hz to 500 Hz, 0.3 grms

Non-operating ...........................................5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Non-operating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

Safety

Measurement Category .....................................I¹

⚠️ Caution: Do not connect the RM-4302 to signals or use for measurements within Measurement Categories II, III, or IV.

⚠️ Caution: The protection provided by the RM-4302 can be impaired if it is used in a manner not described in this document.

Safety Standards

This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:

? Note: For UL and other safety certifications, refer to the product label or the Online Product Certification section.

¹ Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:

? Note: In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.

? Note: Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

? Note: For the standards applied to assess the EMC of this product, refer to the Online Product Certification section.

CE Compliance

This product meets the essential requirements of applicable European Directives as follows:

Online Product Certification

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

Waste Electrical and Electronic Equipment (WEEE)

EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit ni.com/environment/weee.

Worldwide Support and Services

The National Instruments website is your complete resource for technical support. At ni.com/support you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.

Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services.

Visit ni.com/register to register your National Instruments product. Product registration facilitates technical support and ensures that you receive important information updates from NI.

National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information, support phone numbers, email addresses, and current events.

Legal and Trademarks

Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2016 National Instruments. All rights reserved.

PDF preview unavailable. Download the PDF instead.

377021a Acrobat Distiller 15.0 (Windows); modified using iText 4.2.0 by 1T3XT FrameMaker 8.0

Related Documents

Preview NI PXIe-4304/4305 and RM-4304 User Guide and Specifications
Comprehensive user guide and technical specifications for National Instruments NI PXIe-4304/4305 Analog Input Module and RM-4304 Rack-Mount Terminal Block. Covers installation, configuration, system setup, and detailed product specifications.
Preview National Instruments PXIe-4304/4305 and RM-4304 User Guide and Specifications
Comprehensive user guide for National Instruments PXIe-4304/4305 filtered analog input modules and the RM-4304 rack-mount terminal block. Covers installation, configuration, software setup, hardware connections, calibration, troubleshooting, and detailed specifications.
Preview PXIe-4302/4303 and TB-4302C Calibration Procedure
This document provides the calibration procedure for the National Instruments PXIe-4302/4303 module and the TB-4302C terminal block, including verification and adjustment steps, required test equipment, and test conditions.
Preview NI PXIe-4304/4305 and RM-4304 User Guide and Specifications
Comprehensive user guide and technical specifications for National Instruments PXIe-4304, PXIe-4305 Analog Input Modules and RM-4304 Rack-Mount Terminal Block. Covers installation, configuration, calibration, removal, troubleshooting, and specifications.
Preview National Instruments PXIe-4300 and TB-4300/B/C User Guide and Specifications
Comprehensive user guide and technical specifications for the National Instruments PXIe-4300 module and TB-4300/B/C terminal blocks, covering installation, signal connection, operation, safety, and troubleshooting.
Preview National Instruments NI PXIe-4138/4139 Getting Started Guide
A guide to installing, configuring, and testing the National Instruments PXIe-4138 and PXIe-4139 Precision System PXI Source Measure Units. Covers system requirements, unpacking, software and hardware installation, pinouts, configuration in MAX, self-calibration, programming, and troubleshooting.
Preview National Instruments Oscilloscope Probe Compensator User Guide
A user guide for the National Instruments Oscilloscope Probe Compensator (SMB (f)-to-Probe Compensation Tab Cable), detailing its compatibility, physical specifications, and step-by-step instructions for compensating passive oscilloscope probes for accurate measurements.
Preview NI PXIe-6555/6556 Specifications: Digital Waveform Generator/Analyzer
This document provides detailed specifications for the NI PXIe-6555 and NI PXIe-6556 Digital Waveform Generator/Analyzers, covering channel specifications, timing, memory, scripting, triggers, and environmental considerations.