EXTECH INSTRUMENTS

UV-AB Light Meter Model UV505

User Manual

Additional translations of the User Manual are available at www.extech.com

Introduction

Thank you for selecting the Extech UV-AB Light Meter Model UV505. This device measures UV-AB ultraviolet light in the 290~390 nm wavelength range. This device is supplied fully tested and calibrated and, with appropriate use, will provide many years of reliable service. Please visit our website (www.extech.com) to check for the latest version of this User Manual, Product Updates, and Customer Support.

Features

General Safety

Read the entire User Manual and Quick Start Guide before using the device. Use the meter only as specified and do not attempt to repair or open the meter case. Do not allow children to handle the meter or the protective cap. Dispose of batteries and the meter responsibly and in accordance with all applicable laws and regulations.

UV Safety Information

Exercise extreme caution when using this meter to test an ultraviolet light source. Exposure to UV radiation can be dangerous depending on the exposure time, radiation intensity, wavelength, and individual sensitivity to ultraviolet rays.

Meter Description

The UV-AB Light Meter Model UV505 features a compact design with an LCD display and several control buttons.

Physical Components:

Note: The battery compartment and tripod mount are located on the rear of the meter.

LCD Display Elements:

  1. Measurement Reading
  2. Data Hold Indicator
  3. Zero Mode Indicator
  4. Low Battery Indicator
  5. Measurement Unit (μW/cm² or mW/cm²)

Button Functions:

Operation

Powering the Meter

Press the power button (?) to turn the meter on or off. If the LCD does not turn on, check the batteries located in the rear battery compartment. The meter automatically shuts off after 15 minutes of inactivity.

Zeroing the Sensor

With the sensor protective cap in place, press the Zero button (?️) to zero the display before taking measurements. A "ZERO" indicator will appear on the display while the meter is zeroing. Press the Zero button (?️) again to exit zero mode; the "ZERO" indicator will turn off.

Taking Measurements

Remove the sensor cap and point the light sensor towards the UV light source. View the measurement on the meter's display.

Selecting Range and Measurement Unit

Press the Unit button (?) briefly to alternate the range between (mW/cm² and μW/cm²); the lower display area will show the selected range.

Data Hold

Press the Hold button (⏸️) briefly to freeze or unfreeze a reading on the display.

Backlight

The LCD is equipped with a backlight for easier viewing, especially in dimly lit areas. Press the Backlight button (?) briefly to turn on the backlight. The backlight automatically turns off after 10 seconds.

Measurement Considerations

Maintenance

Battery Replacement

  1. Turn off the meter.
  2. Press and hold the lock button above the battery compartment (rear of the meter) and slide the battery compartment door off in a downward motion (⬇️).
  3. Replace the two 1.5 V AAA batteries, observing correct polarity. Reassemble the meter before use.

Battery Safety: Dispose of batteries responsibly; never throw batteries into a fire, as they may explode or leak. If the meter is not used for more than 60 days, remove the batteries and store them separately.

Never dispose of used or rechargeable batteries with household waste. As consumers, users are legally obligated to return used batteries to appropriate collection points, the store where the batteries were purchased, or other locations where batteries are sold.

Disposal: Do not dispose of this instrument with household waste. The user is obligated to return end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment.

Cleaning and Storage

Periodically clean the outer casing with a damp cloth and mild detergent; do not use abrasive products or solvents. Use a dry, lint-free dusting cloth to clean debris and dust from the sensor. Avoid any materials that could potentially scratch the sensor surface.

Specifications

Display 3-3/4 digit LCD with backlight (4000 counts)
Size: 35 x 30 mm (1.4 x 1.2")
Sampling Rate Approx. 3 seconds
Measurements Ultraviolet (UV-AB) Light
UV Sensor Spectrum 290 to 390 nm
Peak Sensitivity Wavelength 365 nm
Input Overload Display "OL"
Tripod Mount On rear panel
Power 2 x 1.5 V 'AAA' batteries
Battery Life Approx. 50 hours
Auto Power Off After approx. 15 minutes of inactivity
Operating Temperature/Humidity 5 to 40 °C (41 to 104 °F) / 80% RH max
Operating Altitude 2000 meters (7000 feet) maximum
Storage Temperature/Humidity -10 to 60 °C (14 to 140 °F) / 70% RH max
Dimensions/Weight 133 x 48 x 23 mm (5.2 x 1.9 x 0.9") / 90 g (3.2 oz)
Safety For indoor use only. EMC: EN61326 (1997), A1 (1998), A2 (2001); Pollution Degree 2
Electrical Specifications (23±5 °C) Ultraviolet (UV-AB) Unit Range Resolution Accuracy
mW/cm² 0 to 39.99 0.01 ± 4% of full scale
μW/cm² 0 to 3999 1 ± 4% of full scale

Customer Support

Support Lines: U.S.A. (877) 239-8324 | International: +1 (603) 324-7800

Technical Support: Option 3; Repairs/Returns: Option 4

Technical Support Email: support@extech.com

Service Email: repair@extech.com

Website: www.extech.com

PDF preview unavailable. Download the PDF instead.

UV505 UM-pt Acrobat Distiller 9.0.0 (Windows)

Related Documents

Preview Extech UV505 Pocket UV-AB Light Meter Datasheet
Datasheet for the Extech UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and applications for measuring UV-AB light in various environments.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. Provides instructions on features, operation, safety, maintenance, and technical specifications for measuring UV-AB light.
Preview Manuale Utente Misuratore di Radiazione UV-AB Extech UV505
Manuale utente completo per il Misuratore di Radiazione UV-AB Extech UV505. Include istruzioni operative, specifiche tecniche, informazioni di sicurezza e manutenzione.
Preview EXTECH UV505 Pocket UV-AB Light Meter Datasheet | Extech Instruments
Detailed datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, featuring specifications, features, and applications for measuring UV-AB light. Includes product overview, technical details, and ordering information.
Preview Extech UV505 Pocket UV-AB Light Meter Datasheet
Product datasheet for the Extech UV505 pocket-sized UV-AB light meter, detailing its features, specifications, and applications in various fields.
Preview Extech UV505 Pocket UV-AB Light Meter Datasheet
Product datasheet for the Extech UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and ordering information for measuring UV-AB light. Includes technical specs and contact details.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. This document provides detailed information on the device's features, operation, specifications, safety precautions, and maintenance for measuring UVA and UVB light.
Preview EXTECH UV505 Pocket UV-AB Light Meter Datasheet
Product datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and ordering information for UV-AB light measurement applications.