Rice Lake 350-10-4 Dual-Ramp Wheelchair Scale with Seat
Software Version 11525
Introduction
The 350-10-4 digital wheelchair scale with seat is a user-friendly, quality scale, designed for safe weighing of the handicapped and mobility challenged individuals. Movement compensation technology ensures sound, accurate weighments. The scale has roll-on ramp access and a comfortable seat.
The scale has a non-skid rigid platform and two rear heavy-duty wheels for easy maneuverability. The scale is set up to use motion sensing technology to determine the actual weight of a moving patient. The weight can be displayed in either pounds or kilograms, and a tare weight can be entered. Section 3.0 on page 11 of this manual explains scale operation and how to obtain a tare weight.
The 350-10-4 wheelchair scale has a unique folding feature that enables easy transportation and simplifies assembly, avoiding the need for field wiring and recalibration. All that is required is to open the packaging and unfold the scale.
Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals.
Warranty information is available at www.ricelake.com/warranties.
1.1 FCC Compliance
United States
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his own expense.
Canada
This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications.
Le présent appareil numérique n'émet pas de bruits radioélectriques dépassant les limites applicables aux appareils numériques de la Class A prescrites dans le Règlement sur le brouillage radioélectrique édicté par le ministère des Communications du Canada.
1.2 Disposal
Product Disposal
The product must be brought to appropriate separate waste collection centers at the end of its life cycle. Proper separate collection to recycle the product helps prevent possible negative effects on the environment and to health, and promotes the recycling of the materials. Users who dispose of the product illegally shall face administrative sanctions as provided by law.
Battery Disposal
Dispose of batteries at appropriate waste collection centers at the end of their life cycle in accordance with local laws and regulations. Batteries and rechargeable batteries may contain harmful substances that should not be disposed of in household waste. Batteries may contain harmful substances including but not limited to: cadmium (Cd), lithium (Li), mercury (Hg) or lead (Pb). Users who dispose of batteries illegally shall face administrative sanctions as provided by law.
WARNING: Risk of fire and explosion. Do not burn, crush, disassemble or short-circuit lithium batteries.
1.3 Safety
Safety Definitions:
- DANGER: Indicates an imminently hazardous situation that, if not avoided, will result in death or serious injury. Includes hazards that are exposed when guards are removed.
- WARNING: Indicates a potentially hazardous situation that, if not avoided, could result in serious injury or death. Includes hazards that are exposed when guards are removed.
- CAUTION: Indicates a potentially hazardous situation that, if not avoided, could result in minor or moderate injury.
- IMPORTANT: Indicates information about procedures that, if not observed, could result in damage to equipment or corruption to and loss of data.
General Safety
Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
WARNING
Failure to heed could result in serious injury or death.
- Ensure every individual who operates or works with this unit has read and understands all safety information.
- Do not transport the scale while someone is on the scale.
- Do not allow minors (children) or inexperienced persons to operate this scale.
- Do not use in the presence of flammable materials.
- Do not use this product if any of the components are loose or cracked.
- Do not use near water.
- Do not use the scale on slippery surfaces, such as a wet floor.
- Do not use this scale when a person's body or feet are wet, such as after taking a bath.
- Do not place fingers into slots or possible pinch points.
- To avoid cross contamination, the scale should be cleaned regularly.
- Prior to cleaning, make sure the scale is disconnected from the power source.
- People with disabilities, or who are physically frail, should always be assisted by another person when using this scale.
IMPORTANT
- Do not drop the scale or subject it to violent shocks.
- Do not jump on the scale.
- For accurate weighing, the scale must be placed on a flat, stable surface.
- Operating at voltages and frequencies other than specified could damage the equipment.
- Avoid contact with excessive moisture.
- Do not make alterations or modifications to the scale.
Rice Lake Weighing Systems offers optional AC adapters; utilizing an adapter not supplied by Rice Lake Weighing Systems voids all warranties and approvals.
Weight exceeding the maximum capacity may damage the scale.
Assembly
2.1 Unpacking
Place the unopened box in an open area that has ample room for unpacking the scale. Immediately after opening the box, visually inspect the scale and components to ensure all parts are included and undamaged. If parts were damaged in shipment, notify Rice Lake Weighing Systems and the shipper immediately.
Parts contained in the shipping box include:
- Scale
- Manual
- Box containing six AA batteries and four feet
2.1.1 Repacking
Retain the packaging for use in the event that the scale must be returned or moved. The product must be properly packed with sufficient packing materials. Whenever possible, use the original carton and packing materials when shipping the scale back.
IMPORTANT: Damage caused by improper packaging is not covered by the warranty.
2.2 Scale Setup
Move the scale into the area where the weighing process will occur. It's recommended to place the scale on a hard, level surface for the most accurate weighments. Thin carpeting is acceptable, but not recommended.
CAUTION: Rice Lake Weighing Systems recommends using two people for lifting the scale and to use proper lifting techniques to prevent injury.
- Using two people, remove the scale out of the packaging material that it came in by lifting the scale out of the box by the scale base.
IMPORTANT: Do not lift the scale out of the box by its handle as this can cause the hinges to break or affect the scale operation.
- Gently set the scale base down to the floor.
2.2.1 Attach Feet to Scale
The 350-10-4 digital wheelchair scale with seat comes with four feet that must be attached to the scale base for proper operation. The four feet are located in the parts box in the original shipping box. Prior to laying the scale base on the floor, use the following steps to attach the feet to the scale.
- Screw the four round feet clockwise into the scale base as shown in Figure 2-1.
Figure 2-1. Screw Feet Into the Scale Platform
- There must be adequate clearance between the scale base and the floor, so screw each foot out counterclockwise two full turns. This will ensure that there is enough clearance between the scale base and the floor.
- There should be minimal clearance between the scale base and the floor without having the scale base actually touching the floor. Figure 2-2 shows the width of a finger being slid under the scale base and the floor.
IMPORTANT: Not having clearance around the scale base creates inaccurate weighments.
Figure 2-2. Clearance Between Scale Base and Floor
- It is also important to make sure that the scale is completely level. Gently press down on all corners of the scale base to ensure there are no high spots or rocking of the scale base.
- Adjust feet as needed to level the scale base.
IMPORTANT: An unleveled base produces inaccurate weight readings.
2.2.2 Handle Setup
- Loosen the handles and unfold and set upright (handles shown in Figure 2-3).
- Tighten the two hinges on the wheelchair scale by rotating the two handles. This will secure the handrail mounted indicator.
Figure 2-3. Secure the Handrail by Tightening the Durable Hinges
The 350-10-4 wheelchair scale with seat comes fully assembled (shown below).
Figure 2-4. Wheelchair Scale with Seat in an Upright Position
2.2.3 Insert Batteries
The six AA batteries supplied with the scale provide an average of 25 hours of continuous use.
To install the batteries:
- Turn thumbscrew counterclockwise then remove battery cover.
- Insert batteries into the battery chamber as illustrated.
Figure 2-7. Battery Chamber
- Put the cover in place and turn the thumbscrew clockwise to secure.
NOTE: Remove the batteries prior to storing if the product is not going to be used for an extended period of time.
NOTE: If the LO BAT indicator activates, for accurate weighing, replace the batteries or connect the scale to an AC power source as soon as possible.
2.2.4 Power Connection
An optional AC power adapter can be used when a power outlet is available.
IMPORTANT: Only use power adapters supplied by or purchased from Rice Lake Weighing Systems. The use of a power adapter not from Rice Lake Weighing Systems voids the warranty.
Figure 2-8. Power Connection Site
NOTE: The battery annunciator on the display turns off when using an AC power connection.
The brightness of the backlight is reduced to 60% when using battery power.
Operation
This section describes the front panel and includes procedures for operation of the scale.
Figure 3-1. Front Panel Keypad
3.1 Key Descriptions
The display has 10 front panel keys. Key functions are described in the table below.
IMPORTANT: The front panel keys are very sensitive, so only a gentle press is required.
Key Name | Function |
---|---|
On/Off | Powers the scale on or off |
Print LB/KG | Sends data out from the RS-232 port; Allows to toggle between kilograms and pounds providing that it is enabled in Configuration mode; Cannot toggle while in the BMI mode |
Zero >0< | Only functions if the current weight is stable and less than 2% of the capacity of the scale. Anything over 2% requires a recalibration |
Hold Release II | Displays most current weight value on the display and holds that value when the patient is off the scale. A second press releases the weight value. Not active while in BMI mode |
BMI | Pressing the BMI key enables access to the BMI (Body Mass Index) mode (defaults when scale is turned on). The patient gets on the scale, weight stabilizes and press the BMI key. The display then asks for the patient height to calculate out the patient BMI. |
TARE | Used to remove the weight initially of anything on the scale that shouldn't be included in the total weight of the patient on the scale |
CLEAR | When using the BMI function, the display looks for a height entry. Pressing Clear changes this entry back to 190.0 cm (default) or 5 ft, 7.5 in. Once BMI is displayed, pressing the Clear key exits BMI |
ENTER 全 | Used to accept height in BMI mode; accepts the value of the parameter last entered and moves to the next stage. Pressing and holding Enter during startup will display ID. This is the first setup on entering into configuration mode |
Up Arrows ▲ | Adjusts the value of the flashing digit/number. Adjusts height input (0.5 in/0.5 cm) while in BMI mode |
Down Arrows ▼ | Adjusts the value of the flashing digit/number. Adjusts height input (0.5 in/0.5 cm) while in BMI mode |
3.2 Weighing
Use the following steps to weigh.
- Press On/Off to turn on the scale. 0.0 appears on the display along with the ZERO annunciator.
- Place the patient on the scale. The patient's weight is displayed, the LOCK annunciator is on and the indicator beeps to indicate the end of the weighing process.
- Press LB/KG to change the display from lb to kg and vice-versa.
- Press and hold On/Off until OFF displays to turn off the scale.
3.3 Hold/Release Function
Use the following steps to use the Hold/Release function.
- Press On/Off to turn on the scale. 0.0 prompts along with ZERO on the display.
- Press Hold Release once the patient's weight stabilizes. The patient's weight and the HOLD and LOCK annunciators remain on the display when the patient is off the scale.
- Press Hold Release again to return the scale to zero.
NOTE: Pressing Print LB/KG will not work while using Hold/Release function.
NOTE: Pressing Hold Release prior to the patient getting on the scale will also hold the weight display.
3.4 Preset Tare
Use the following steps for the Preset Tare function prior to patient weighing if additional items are being used by the patient.
- Press On/Off to turn on the scale. 0.0 appears on the display along with the ZERO annunciator.
- Place additional item(s) on the scale.
- Press TARE until the display returns to 0.0 and NET annunciator appears on the display.
- Remove additional item(s) from the scale. The weight displays with a negative symbol to the left of it.
NOTE: Not removing the additional item(s) prior to patient weighing will also work.
- Position the patient and additional item(s) on the scale. The display identifies the patient weight. The NET annunciator is still active. The weight of the additional item(s) remains stored in memory for the duration of this weigh-in.
- To cancel the tare weight, remove patient from the scale and press TARE until NET disappears from the display and the display turns back to 0.0 and GROSS appears.
NOTE: Tare weight is also canceled when the scale is turned off.
3.5 Toggle Tare
Use the following steps to use the Toggle Tare function when the additional item to be weighed is known.
- Press TARE when the scale is empty and 0.0 displays. The default values prompt while 0.0 is flashing on the display (default is programmed to be 33.0 lb/15.0 kg).
- Use Up Arrows ▲ and Down Arrows ▼ to adjust the value. Press ENTER 全 to start the tare function. The NET annunciator turns on instead of the GROSS annunciator.
3.6 Using the Body Mass Index (BMI) Function
Use the following steps in determining the BMI.
3.6.1 LB Mode
- Ensure that the scale is at zero.
- Place the patient on the scale to obtain a weight. The LOCK annunciator appears on the display.
- Press BMI. The BMI and FT/IN annunciators appear on the display and a default height value of 5 feet - 7.5 inch (5 - 07.5) is flashing.
- Use Up Arrows ▲ and Down Arrows ▼ to adjust the height value.
- Press ENTER 全.
- The BMI value and BMI annunciator are shown on the display. Press CLEAR to return to the Weighing mode and the BMI function will be turned off.
3.6.2 KG Mode
- Ensure that the scale is at zero.
- Place the patient on the scale to obtain a weight. The LOCK annunciator appears on the display.
- Press BMI. The BMI and CM annunciators appear on the display and a default height value of 170.0 cm (170.0) is flashing.
- Use Up Arrows ▲ and Down Arrows ▼ to adjust the height value.
- Press ENTER 全.
- The BMI value and BMI annunciator are shown on the display. Press CLEAR to return to the Weighing mode and the BMI function will be turned off.
Troubleshooting
Refer to the following table to check and correct any failure before contacting service personnel.
Symptom | Possible Cause | Corrective Action |
---|---|---|
Scale does not turn on | Dead batteries Faulty electrical outlet Bad power supply |
Replace batteries or connect to AC power Use a different electrical outlet Replace adapter |
Questionable weight or the scale does not zero | External object is interfering with the scale Display did not show 0.0 before weighing |
Remove the interfering object from the scale Help the patient off the scale, zero the scale and begin the weighing process again |
Scale is not placed on a level floor | Scale is out of calibration | Ensure scale is level and begin the weighing process again Check the weight with a certified calibration weight |
Scale base is touching floor during a weighment | Adjust height of feet so fingers can slide between the base of scale and the floor all the way around the platform | |
The load on the scale exceeds the capacity of the scale | Remove the excess weight and use the scale according to manufacturer specifications | |
The display shows a STOP message | ||
The display shows LO Bat message | The battery is low | Replace batteries |
The display shows E and Err messages as detailed below | Identifier - ADC | AD too high |
Overload | AD too low | |
E06 | Scale has been overloaded. Remove load from scale | |
E07 | Battery low, but still usable - one bar left on indicator display | |
E10 | Battery low and unstable - no bars left on indicator display | |
E4L | ||
E4U | ||
E11 | CAL | Calibration Error - recalibrate scale |
Err 1 | Load cell cable may be plugged into wrong connection port | Ensure cable is connected to the load cell connection port. Note: Load cell connection point is located underneath the curved plastic cover of the indicator. Remove four back retaining screws, remove curved back cover to access load cell connection point. |
Err 2 | Low saturation state (low A/D) | The load cell is not connected properly; Check the cables and mechanical connections; if the problem persists, replace the set of load cells |
Err 3 | High saturation state (high A/D) | See Err 2 |
Err 6 | Unstable weight; Cannot calibrate | Check the load cell mechanical surroundings and ensure nothing is contacting the load cell and that the cables are properly welded |
Err 7 | Scale isn't moving | Make sure feet are installed on the scale. Turn the feet all the way in and then back them out three full turns, then level the scale |
SAT | Damaged load cell cable | Replace load cell cable |
Load cell cable may be plugged into wrong connection port | Ensure cable is connected to the load cell connection port. Note: Load cell connection point is located underneath the curved plastic cover of the indicator. Remove four back retaining screws, remove curved back cover to access load cell connection point. |
Configuration
Options and parameter setup are done through the scale configuration.
4.1 Setup Switch
Access to the setup switch is located under the tilt stand cover. Use a Phillips head screwdriver to remove the four screws holding the cover in place (shown below — left photo).
Figure 4-1. Setup Switch Location
4.2 General Navigation
Use the buttons on the front panel to navigate through the menus and parameters.
4.2.1 Change Parameters
- Press BMI to scroll through the menus and/or parameters
- Press ENTER 全 to enter a displayed menu and/or parameter
- Press Up Arrows ▲ or Down Arrows ▼ to scroll through values
- Press ENTER 全 to save the displayed selection and move to the next parameter
4.2.2 Enter Numbers
- Press ENTER 全 to enter parameter value
- Press Up Arrows ▲ or Down Arrows ▼ to increment/decrement numbers
- Press BMI to move to the next digit
- Press ENTER 全 to save value and move to the next parameter
4.2.3 Save and Return to Main Menu
- When a parameters selection/value is correct, press ENTER 全. The next parameter displays
- When all parameters selections/values are correct, SAVE displays
- Press ENTER 全. DONE displays.
- Press ENTER 全 to save settings and return to weigh mode.
4.3 Configuration Mode
Use the following steps to enter into Configuration mode.
- Ensure the scale is turned off.
- Turn the scale on by simultaneously pressing On/Off and ENTER 全. Continue to hold both keys until Id appears. The unit cycles through its startup function and continues to display the software version.
- Access the setup switch located in the back of the scale to enter the setup parameters for the scale. Use a small paper clip, small screwdriver or other similar object to press the setup switch.
- Once the setup switch is pressed, PROG displays.
- Scale can be configured using a series of menus accessed through the front panel when the scale is in Setup mode.
Figure 4-2. Top Level Menu
- Press BMI to advance to the desired menu.
- Press ENTER 全 and advance in the manual to the related menu selection for further instructions.
4.4 Programming Mode Menu
Various parameters can be set while in Programming mode.
Figure 4-3. Programming Mode Menu Structure
The following table lists the various display messages and sequence when setting up the scale.
Parameter | Description | Choices | Steps |
---|---|---|---|
FULL | Full capacity of the scale | Value (capacity of scale) | The display toggles between a numeric value and FULL; If you do not want to change the value, press the BMI key to move to the next setting; Example: from FULL to LOAD. If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. Press BMI key again to move to the left 6. Use the Up/Down arrow keys to increment/decrement numbers. 7. When done press ENTER key to move to the next parameter (LOAD). |
LOAD | The amount of weight applied during calibration; Can also be changed in the calibration menu | Value (200 lb) | The display toggles between a numeric value and LOAD; If you do not want to change this value, press the BMI key to move to the next setting; Example: from LOAD to ASTART; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to next parameter (ASTART). |
ASTART | Weight process start limit - Maximum (full capacity)/10; Determine when the weight algorithm starts (when the "----" is displayed), below this value the scale will show live weight | Value (2.0) | The display toggles between a numeric value and ASTART; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ASTART to ARW; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to the next parameter (ARW). |
ARW | Auto Reweigh - Restarts the weight algorithm if the weight changed by more than this value. | Value (4.0 lb) | The display toggles between a numeric value and ARW; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ARW to SAL; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to the next parameter (SAL). |
SAL | Semi Auto Live —This value is the interval between weight displays during the algorithm process | Value (0.5) | The display toggles between a numeric value and SAL; If you do not want to change this value, press the BMI key to move to the next setting; Example: from SAL to ROUND; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. When done, press ENTER key to move to the next parameter (ROUND). |
ROUND | Scale Resolution | Values in kg: 1, 2, 5, 10, 20, 50, 100 Values in lb: 1, 2, 5, 10, 20, 50, 100, 200 |
The display toggles between a numeric value and ROUND; The decimal point location is set to the DISP parameter display decimal point location; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ROUND to DISP; If you want to change the value, use the following steps; 1. Press ENTER key to change value. 2. Press the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (DISP). |
DISP | 0.2 0.5 1.0 2.0 5.0 10.0 20.0 0.1 0.0000 0.000 0.00 |
The display toggles between a numeric value and DISP; If you do not want to change this value, press the BMI key to move to the next setting; Example: from DISP to BAUD; If you want to change the value, use the following steps; 1. Press ENTER key to change value. 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (BAUD). | |
BAUD | Baud rate | 96 48 1152 576 384 288 192 144 |
Indicator display illustrates first two digits of baud rate only; The display toggles between a numeric value and baud; If you do not want to change this value, press the BMI key to move to the next setting; Example: from BAUD to ATOL; If you want to change the value, use the following steps 1. Press ENTER key to change value. 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (ATOL). |
ATOL | Algorithm initial tolerance - Maximum value is 255. Values above 255 will not let you proceed and will return to the previous value. | Value (10) | The display toggles between a numeric value and ATOL; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ATOL to ALEN; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press the BMI key to move to the left. 4. When done, press ENTER key to move to the next parameter (ALEN). |
ALEN | Algorithm initial exponent - Maximum value 10. Values above 10, will not let you proceed and will return to the previous value. | Value (8) | The display toggles between a numeric value and ALEN; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ALEN to ATOUT; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. When done, press ENTER key to move to the next parameter (ATOUT). |
ATOUT | Algorithm maximal exponent — Maximum value is 15. Values above 15, will not let you proceed and will return to the previous value. | Value (10) | The display toggles between a numeric value and ATOUT; If you do not want to change this value, press the BMI key to move to the next setting; Example: from ATOUT to TOFF; If you want to change the value, use the following steps; 1. Press ENTER key to change value starting with the right most digit. 2. Use the Up/Down arrow keys to increment/decrement numbers. 3. Press BMI key to move to the left. 4. Use the Up/Down arrow keys to increment/decrement numbers. 5. When done, press ENTER key to move to the next parameter (TOFF). |
TOFF | Auto off timer - Measured in minutes; 0 = always on; Maximum is 9 minutes; When using an external power supply, this parameter is irrelevant | 5 4 3 2 1 0 9 8 7 6 |
The display toggles between a numeric value and TOFF; If you do not want to change this value, press the BMI key to move to the next setting; Example: from TOFF to UNITS; Press the ENTER key to move to the next parameter; (UNITS); If you want to change the value, use the following steps; 1. Press the ENTER key to change values. 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (UNITS). |
UNITS | Units - Selects the unit of measure; It can be either Kg/Lb, Kg only or Lb only | KG/LB KG LB |
The display toggles between unit of measurements and UNIT; If you do not want to change this value, press the BMI key to move to the next setting; Example: from UNITS to OP; If you want to change the value, use the following steps; 1. Press the ENTER key to change values. 2. Press the Up/Down arrow keys to change the value. 3. When done, press the ENTER key to move to the next parameter (OP). |
OP | Binary options: OPO - Live weighing options (0=disable, 1=enable) OP1 - Communication protocol (0=ESC, 1=maintenance) OP2 - BMI menu (0=disable, 1=enable) OP3-RTC power (0=disable, 1=enable) OP4 - Semi-Auto-Live — (0=disable, 1=enable) OP5 - Full calculation — (0=spatial, 1=multiply by (0=disable, 1=enable OP6 - Tare - (0=disable, 1=enable OP7 - Bat type — (0=dry batteries, 1=rechargeable batteries) OP8 - OIML mode - (0=disable, 1=enable | OPO OP1 OP2 OP3 OP4 OP5 OP6 OP7 OP8 |
The display toggles between a binary option and POO-0; If you do not want to change this value, press the BMI key to move to the next settings If you want to change the value, use the following step 1. Press ENTER to change parameters. 2. Use the Up/Down arrow keys to select the display value you want to change. 3. Press BMI key to move the flashing cursor a. Use the Up/Down arrows to change the value. b. Press the BMI key to move the flashing cursor. 4. Press ENTER key to save all of the display parameters. SAVE appears on the display. 5. Press ENTER key again and DONE appears indicating that you are now done entering all of the parameters of the scale. Press the BMI key to cycle back to the top level menu choices, ie: PROG/CALIB/DEF/DONE |
4.5 Default Menu
The default menu is used to return the scale back to its factory settings and is shown below.
Figure 4-4. Default Menu
Use the following steps to return the settings back to their factory default.
- Press ENTER 全. The display shows a default value of NO.
- To change to YES, use Up Arrows ▲ and Down Arrows ▼ to adjust the value.
- Press ENTER 全 and the display shows DONE.
- Press BMI 全 to return to top level level menu.
NOTE: Selecting YES and pressing ENTER will reset to factory defaults settings without changing the calibration and will return you to Weigh mode.
4.6 Scale Calibration
Use the following steps to calibrate the scale.
- Press On/Off and ENTER 全 simultaneously to power on the scale. The unit cycles through its startup function and continues to display the software version. Continue to hold both keys until Id appears.
- Access the setup switch located in the back of the scale to enter the setup parameters for the scale. Use a small paper clip, small screwdriver or other similar object to press the setup switch on the back of the indicator (under the cover).
- Prog displays. Press BMI 全 to toggle along the parameter menu.
- Calib displays and enter the calibration parameters.
- Press ENTER 全 and a numeric value is displayed which represents the amount of weight that is used for calibration. Lb will be flashing. To switch between lb and kg, press LB/KG or On/Off. Once a unit is selected, press ENTER 全 and the right most digit will be flashing.
- To change the calibration load value, press Up Arrows ▲ or Down Arrows ▼ to increment/decrement the flashing digit.
- Use BMI 全 to move the flashing digit to the left or right.
- Once all the digits have been entered, press ENTER 全 and Clear displays.
- Make sure the scale platform is clear of weight and press ENTER 全 again then = displays.
- A request to put the chosen load on the platform is displayed by Put xxx.xx.
- Put the chosen weight on the platform and press ENTER 全 displays then Save displays.
- Press ENTER 全 again and the display indicates Done.
- Press BMI 全 three times to exit back out to the top level Done parameter.
- Press ENTER 全 to return to Weigh mode.
- To exit calibration without changing zero or span existing calibration, press CLEAR, then press On/Off.
4.7 Test Menu
To access the TEST menu, use the following steps.
- Press On/Off and ENTER 全 simultaneously to power on the scale until ID flashes.
- Press ENTER 全 again.
- Continue to press BMI 全 to scroll through the various menu items.
- Once complete, press BMI 全 again then Done displays.
- Press ENTER 全 to start the weighing process.
Figure 4-5. Test Menu
Parameter | Choice | Description |
---|---|---|
RID | Value | Displays internal ID number |
VER | Value | Displays the current software version |
BAT | Value | Displays the current battery level |
VALUE | Value | Displays the actual value |
A2D | Value | Displays the actual raw counts of the scale |
Communications
The unit comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing Systems. That connection is shown in USB Connection section.
The RS-232 parameters are:
- 9600 baud (selectable in the programming mode)
- 8 data bits
- 1 stop bit
- no parity
- no handshaking
There are three methods of communication:
- Push-button keypad print
- Standard remote protocol
- Escape protocol
5.1 Push-button Keypad Print
With a stable, in-range weight, press and hold Print for at least three seconds, or until the scale emits two quick beeps.
NOTE: If the scale does not beep after five seconds, release as the weight was either in motion or out of range.
If displaying weight and not BMI, the scale will send out the following 21 character string:
xxxxxxxxx<SP>uu<SP>mmmmm<SP><CR><LF>
Token | Description |
---|---|
XXXXXXXXX | Weight with decimal point and “-” sign |
<SP> | Space |
uu | Unit - lb or kg |
mmmmm | Mode - gross or net |
<CR> | Carriage return |
<LF> | Line feed (moves cursor down to the next line) |
Example:
80.1 KG<PATIENT><SP><WEIGHT><SP>-60.1<SP>KG<SP><CR><LF>
In BMI mode (displaying the BMI value), the scale will send out the following data:
PATIENT WEIGHT 60.1 KG
PATIENT HEIGHT170.0 CM
PATIENT BMI 20.8
Example in KG:
<PATIENT><SP><WEIGHT><SP>-60.1<SP>KG<SP><CR><LF>
<PATIENT><SP><HEIGHT><SP>-170.0<SP>CM<SP><CR><LF>
<PATIENT><SP><B><SP><M><SP><I><SP><SP><SP>20.8<SP><SP><SP><SP><CR><LF>
Example in LB:
<PATIENT><SP><WEIGHT><SP>132.4<SP>LB<SP><CR><LF>
<PATIENT><SP><HEIGHT><SP>-5-07.5<SP>FT<SP><CR><LF>
<PATIENT><SP><B><SP><M><SP><I><SP><SP><SP>20.4<SP><SP><SP><SP><CR><LF>
In case of under weight or over weight, the word Under or Over will be sent correspondingly.
5.2 Communication Protocols
The scale has two communication protocols, escape and maintenance protocol.
5.2.1 Escape Protocol
An escape protocol is where the escape (0X1B or ASCII 27) is used to indicate that there is a command following. On the PC side there must be a listener created by the vendor that will interpret this protocol. This listener must also take care of all the issues regarding data integrity to make sure that the data that was sent and received is valid.
Two examples include:
- Scale initiated communication
- PC initiated communication
The escape protocol commands table shows (below) what can be sent across communications lines.
PC Initiated | ESC Value | PC Initiated | ESC Value | |
---|---|---|---|---|
Request current values/settings | R | Send single reading | R | |
Diagnostics | A | Send diagnostic response | ||
Send scale control messages | C |
Table 5-2. Escape Protocol Commands
ESC characters that will be used is shown below.
Name | ESC Character | ESC Value with Parameters | Description |
---|---|---|---|
Reading | R | R | Tells PC the scale is sending a reading; immediately following this is the value that is sent Example: <ESC><R><ESC><W0200.0<ESC>Nm<ESC>E |
Weight | W | Wnnn.n | The patient weight (Example: W02000 means 200.0). If scale is overloaded or under loaded, 999.99 is returned |
Height | H | Hnnn.n | Patient height |
BMI | B | Bnn.n | Patient BMI |
Units | N | Nc | Indicates the units the values have been taken (m=metric, c=constitutional). |
End of Packet (EOP) | E | E | Indicates the end of the command has been reached. |
Diagnostics (request) | A | Accc | A request for a diagnostic test on certain parts of the scale (like battery life, load cells). |
Diagnostics (response) | Z | Zccc | The response of the diagnostics done on the scale; values include error codes to indicate an issue, or all zeros (Z000) to indicate the scale is performing properly |
Control (set a value) | C | Cccc=c | Sets the value of the scale's global settings Example: <ESC><CUOM=m><ESC>E sets the unit of measurement |
Table 5-3. ESC Characters
Name of Control | Unit of Measure (metric or constitutional) | Identifier | Unit |
---|---|---|---|
UOM | c (m or c) |
Table 5-4. Scale Global Values and Identifiers
5.2.2 Maintenance Protocol
Maintenance protocol commands are listed below.
Command | Definition |
---|---|
R | Reboot |
V | Firmware ID + development version |
W | Current weight |
A | Current AD |
Z | Zero the scale |
F | Show flash values (used for the first flash process) |
L | USB On/Off (not available on USB communication |
Table 5-5. Maintenance Protocol Commands
USB Connection
The scale has the capability of connecting to a Windows® computer (PC) using a USB cable (not included) and a terminal emulation program. A terminal emulation program allows the transfer of data between the scale and PC using a serial port.
Figure 5-3. Connection Ports
NOTE: Apple® and Macintosh® computers are unable to transfer the necessary data to the scale. Only use a PC for data transfer.
Connecting software and downloads should always be addressed by the IT department for safety reasons and can vary depending on what type of computer platform is being used.
NOTE: Consult the IT department if driver protections are preventing the use of the USB driver. Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver.
- Connect the scale's indicator to a PC using a USB-Type B to USB-Type A cable (not included).
- Turn the indicator on.
NOTE: In most cases, the PC should find the driver and automatically configure the driver when the scale is plugged into a USB port.
- Open a terminal emulation program, such as Advanced Serial Port Terminal, pUtty or Hercules (used in this example).
- Connect to the serial port assigned by the PC (COM5 in example). This can be found in Device Manager. Once selected, press Open.
Figure 5-4. Connecting to a Serial Port
- With weight on the scale, press and hold the Print button on the indicator for three seconds. The patient's weight is sent to the PC.
Figure 5-5. Patient Weight Displayed
Maintenance
The following section provides instructions for maintaining and cleaning the unit.
IMPORTANT: Do not immerse the scale in cleaning or other liquid solutions.
Do not use Isopropyl alcohol or other solutions to clean the indicator display surface.
6.1 Basic Maintenance
Before the first use of the scale and after periods of non-use, check the scale for proper operation and function. If the scale does not operate correctly, contact a qualified service personnel.
Go through the following steps for basic maintenance.
- Check the overall appearance of the entire scale for any obvious signs of damage
- Inspect the condition of the AC power adapter cord for cracking, fraying or for broken or bent prongs
6.2 Cleaning
Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Before beginning the cleaning process, disconnect the scale from the AC power source.
- Clean all external surfaces with a clean, damp cloth or tissue. Mild soap and water solution may be used. Dry with a clean soft cloth
- Do not immerse the scale into cleaning or other liquid solutions
- Do not use Isopropyl alcohol or other solutions to clean the display surface
Specifications
Power
120 VAC - 9 VDC - 60Hz / 230 VAC - 9 VDC - 50Hz
Battery Type
6 AA size Alkaline batteries
Battery Use
25 hours continuous use
Automatic power-off can be configured
Data Communications
RS-232 with RJ-45 jack
USB Connection
Selectable baud rate, default - 9600
8 bits
No parity
1 stop bit
No handshaking
Environmental
Operating Temperature: 50°F to 95°F (10°C to 35°C)
Storage Temperature: 32°F to 122°F (0°C to 50°C)
Humidity: 85% relative humidity
Capacity and Graduation
1000 lb x 0.2 lb (453 kg x 0.1 kg)
Dimensions
Platform Dimensions: 48.5 in W x 29.5 in L x 2.3 in H