Agilent Technologies E8257D Option H60/H65 User's and Service Guide Supplement
This manual supplement describes the differences in the E8257D Option H60 and H65 in comparison to the E8257D standard instrument. Refer to the PSG E8257D/67 Signal Generators User's (E8251-90353).
Description
The E8257D Option H60 and H65 provide a double-balanced microwave mixer that is installed internally on the rear panel with connectors for easy access to the mixer ports. These options can be used with another source, such as the E4438C or E8267D, to upconvert an RF frequency to 65 GHz or downconvert an RF frequency to DC (Option H60) or 1 GHz (Option H65).
It is recommended that Option H60 be ordered with Option 567 to utilize the Option H60 frequency range of 65 GHz. It is also recommended that Option H65 be ordered with Option 520, 532, or 540. The Option H65 mixer has an internal LO frequency doubler that enables the use of an E8257D Option 520 to upconvert a 20 GHz RF frequency to 60 GHz. Option 1EA is also recommended for Option H65 due to the mixer requiring more power for optimal performance.
Refer to the microwave mixer data sheets, Figure 1 on page 4 and Figure 2 on page 5 for further information regarding the Option H60 and H65 mixer nominal performance.
Verifying the Shipment
After the instrument has been unpacked, keep the original packaging materials so they can be used if you need to transport the instrument. Inspect the instrument and all accessories for any signs of damage that may have occurred during shipment. If your instrument or any accessories appear to be damaged or missing, refer to "Agilent Support, Services, and Assistance" on page 23.
Item | Agilent Part Number | Qty |
---|---|---|
RF Cable (2.4 mm) | 5360-9820 | 2 |
RF Cable (3.5 mm) | 5062-6674 | 1 |
Adapter | 1250-2188 | 1 |
Nominal Performance Specification
Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty. Refer to Figure 1 on page 4 and Figure 2 on page 5 for more information. The performance listed below does not include connector and cable losses.
H60 mixer nominal performance:
- LO port frequency range: 2.7 to 65 GHz
- RF port frequency range: 4.0 to 65 GHz
- IF port frequency range: DC to 3 GHz (up to 4 GHz with 15 dB conversion loss)
- Nominal conversion loss: 12 dB (plus cable and connector losses at 65 GHz RF)
H65 mixer nominal performance:
- IF port frequency range: 1 to 22 GHz
- LO port frequency range: 5 to 30 GHz (LO has built-in freq doubler)
- RF port frequency range: 20 to 65 GHz
- Nominal conversion loss: 13 dB (plus cable and connector losses at 65 GHz RF)
Figure 1 E8257D Option H60 Mixer Data Sheet
This data sheet provides specifications for the E8257D Option H60 Double-Balanced Mixer. Key specifications include:
- RF Range: 4.0 to 65.0 GHz
- LO Range: 2.7 to 65.0 GHz
- IF Range: DC to 3.0 GHz
- 7 dB Typical Conversion Loss
- 25 dB Typical LO to RF Isolation
- 4 to 65 GHz RF Coverage
- 2.40 mm RF/LO connector
Guaranteed Specifications (From -55°C to +100°C) include Conversion Loss / Noise Figure (SSB) for various RF/IF ranges, Third Harmonic LO Conversion Loss, and Isolation (LO-RF, LO-IF, RF-IF). Input 1dB Compression and Input Two-Tone Third Order Intercept Point are also specified.
Graphs illustrate Conversion Loss vs. GHz, 3rd Harmonic Conv. Loss vs. GHz, Relative IF Response vs. GHz, LO to RF Isolation vs. GHz, and LO to IF Isolation vs. GHz.
Figure 2 E8257D Option H65 Mixer Data Sheet
This data sheet provides specifications for the E8257D Option H65 Double-Balanced Mixer with Integrated LO-Doubler. Key specifications include:
- RF Range: 20.0 to 65.0 GHz
- LO Range: 5.0 to 30.0 GHz
- IF Range: 1.0 to 22.0 GHz
- 12 dB Typical Conversion Loss
- Integrated LO Doubler
- 20 to 65 GHz RF Coverage
- 2.40 mm RF and LO connectors
Guaranteed Specifications (From -55°C to +100°C) include Conversion Loss / Noise Figure (SSB) for various RF/IF ranges, Isolation (2LO-RF, 2LO-IF, RF-IF), Input 1dB Compression, and Input Two-Tone Third Order Intercept Point.
A graph illustrates Conversion Loss vs. GHz.
Operational and Verification Procedure
The E8257D Option H60 or H65 can be used to upconvert or downconvert the frequency of your source. Refer to Figure 3.
CAUTION: Inputs are ESD sensitive. It is recommended to keep connector covers on when not in use. Refer to "Electrostatic Discharge Protection" on page 22.
- Connect the adapter to the E8257D Option H60/65 RF Output. Refer to Table 2 on page 9.
- Connect the RF cable (5063-9820) from the adapter to the LO Port on the E8257D rear panel.
Figure 3 Rear Panel
The rear panel shows connections for ALC INPUT, 1 GHz REF IN, and MIXER PORTS (LO, IF, RF), as well as a 1 GHz OUT port.
Upconverting the Frequency
The upconverted frequency signal can be measured on the RF port on the E8257D Option H60/H65 rear panel.
- Preset the instrument with Option H60/H65.
- Set the power level to +13 dBm. Refer to Input 1 dB Compression for optimal LO Input Power in Figure 1 on page 4 and Figure 2 on page 5.
- Connect your source to the E8257D Option H60/H65 IF port on the rear panel, using a 3.5 mm cable such as the (5062-6674).
- Connect the second cable (5360-9820) to the RF Port.
- To upconvert your source frequency, set the E8257D Option H60 frequency to the desired RF Output frequency minus your source frequency. For Option H65, set the frequency to the desired RF minus your source frequency and divide by 2. (The IF is subtracted from the upconverted RF for the LO frequency).
NOTE: Filtering may be required to remove unwanted signals. The RF Output frequency will have the difference of the E8257D Option H60/H65, as well as the two original frequencies, along with the sum upconverted product.
Equation 1 Upconverted Frequency (H60)
Upconverted RF = LO(E8257DH60) + IF(yoursource)
Equation 2 Upconverted Frequency (H65)
Upconverted RF = LO(E8257DH65) × 2 + IF(yoursource)
NOTE: The LO Port has a frequency doubler on the LO Input (H65 only). The E8257D Option H65 frequency output will be set to half the desired LO frequency.
Downconverting the Frequency
The downconverted frequency signal can be measured on the IF port on the E8257D Option H60/H65 rear panel.
- Preset the instrument with Option H60/H65.
- Set the power level to +13 dBm. Refer to Input 1 dB Compression for optimal LO Input Power in Figure 1 on page 4 and Figure 2 on page 5.
- Connect your source to the E8257D Option H60/H65 RF port on the rear panel, using a 2.4 mm cable such as the (5063-9820).
- Connect the cable (5062-6674) to the IF Port.
- To downconvert your source frequency, set the E8257D Option H60 frequency to your source frequency minus the desired IF Output. For Option H65, set the frequency to the source frequency minus the desired IF Output frequency and divide by 2. (The IF is subtracted from the upconverted RF for the LO frequency).
NOTE: Filtering may be required to remove unwanted signals. The IF Output frequency will have the sum of the E8257D Option H60/H65, as well as the two original frequencies, along with the difference down converted product.
Equation 3 Downconverted Frequency (H60)
Downconverted IF = RF(yoursource) - LO(E8257DH60)
Equation 4 Downconverted Frequency (H65)
Downconverted IF = RF(yoursource) - LO(E8257DH65) × 2
NOTE: The LO Port has a frequency doubler on the LO Input (H65 only). The E8257D Option H65 frequency output will be set to half the desired LO frequency.
Verification Procedure
This procedure verifies the mixer is functioning in the E8257D Option H60/H65 for frequency upconversion. There are two methods to verify the mixer's function. The first method tests an upconverted frequency up to 50 GHz using minimal equipment. The second method tests an upconverted frequency up to 65 GHz.
Verifying Mixer Performance up to 50 GHz
This test measures three frequency points: low end, 23 GHz, and 50 GHz. The 50 GHz test requires the DUT to have Option 550 or 567 for Option H60. Units with Option H65 can have Option 540, 532, or 520.
Equipment Required for Measurement up to 50 GHz
- E4438C or equivalent 3 GHz signal generator (Option H60)
- E8257D or equivalent 20 GHz signal generator (Option H65)
- E4448A or equivalent spectrum analyzer
- Cable (5063-9820) (x2)
- Cable (5062-6674)
- Adapters, customer supplied (refer to Table 2 on page 9).
Instrument Options | Adapter Description | Agilent Part Number |
---|---|---|
520, 532 | 3.5 mm (f) to (f) | 5061-5311 |
540, 550 and 567 | 2.4 mm (f) to 3.5 mm (f) | 1250-2187 |
540, 550 and 567 | 2.4 mm (f) to 2.4 mm (f) | 1250-2188 |
1ED | 2.4 mm (f) to Type-N (m) | 11903D |
1ED | 3.5 mm (f) to Type-N (m) | 1250-1744 |
Connect the following equipment for the verification procedure:
- Preset all of the test equipment.
- Connect the adapter to the RF Output on the E8257D Option H60/H65 front panel. Refer to Table 2 on page 9.
- Connect the RF cable (5063-9820) to the adapter and connect the RF cable to the LO Port on the E8257D rear panel.
- Connect the E4438C/E8257D or equivalent source to the E8257D Option H60/H65 IF port on the rear panel, using a 3.5 mm cable such as the (5062-6674). Adapters may be required; see Table 2 on page 9. Refer to Figure 3 on page 6 for rear panel connections.
- Connect the E4448A or equivalent spectrum analyzer (with a 10 dB attenuator on the input) to the E8257D Option H60/H65 RF port on the rear panel, using a 2.4 mm cable such as the (5063-9820). Adapters may be required.
NOTE: Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
Option H60 Test
4 GHz Upconverter RF Measurement
- Set the E8257D Option H60 frequency to 2.7 GHz.
- Set the Power Level to +13 dBm > RF On > MOD Off.
- Set your Source frequency to 1.3 GHz and the Power Level to 0 dBm.
- On the E4448A, set the Center Frequency to 4 GHz and Span to 100 MHz.
- Place the marker on the 4 GHz signal, use Trace > Max Hold and measure -30 dBm (± 5 dB). Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.
23 GHz Upconverter RF Measurement
- Set the E8257D Option H60 frequency to 20 GHz.
- Set the Power Level to +13 dBm > RF On > MOD Off.
- Set your Source frequency to 3 GHz and the Power Level to 0 dBm.
- On the E4448A, set the Center Frequency to 23 GHz and Span to 100 MHz.
- Place the marker on the 23 GHz signal, use Trace > Max Hold and measure -30 dBm (± 5 dB). Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.
50 GHz Upconverter RF Measurement
- Set the E8257D Option H60 frequency to 47 GHz.
- Set the Power Level to +13 dBm > RF On > MOD Off.
- Set your Source frequency to 3 GHz and the Power Level to 0 dBm.
- On the E4448A, set the Center Frequency to 50 GHz and Span to 100 MHz.
- Place the marker on the 50 GHz signal, use Trace > Max Hold and measure -35 dBm (± 5 dB). Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.
NOTE: Nominal loss through a 10 dB attenuator and cable is 17 dB at 50 GHz.
Option H65 Test
20 GHz Upconverter RF Measurement
- Set the E8257D Option H65 frequency to 9 GHz.
- Set the Power Level to +13 dBm (or +18 dBm with Option 1EA) > RF On > MOD Off.
- Set your Source frequency to 2 GHz and the Power Level to 0 dBm.
- On the E4448A, set the Center Frequency to 20 GHz and Span to 100 MHz.
- Place the marker on the 20 GHz signal, use Trace > Max Hold and measure -47 dBm (± 5 dB) or -37 dBm (± 5 dB) with Option 1EA. Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.
23 GHz Upconverter RF Measurement
- Set the E8257D Option H65 frequency to 10 GHz.
- Set the Power Level to +13 dBm (or +18 dBm with Option 1EA) > RF On > MOD Off.
- Set your Source frequency to 3 GHz and the Power Level to 0 dBm.
- On the E4448A, set the Center Frequency to 23 GHz and Span to 100 MHz.
- Place the marker on the 23 GHz signal, use Trace > Max Hold and measure -45 dBm (± 5 dB) or -35 dBm (± 5 dB) with Option 1EA. Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.
50 GHz Upconverter RF Measurement
- Set the E8257D Option H65 frequency to 20 GHz.
- Set the Power Level to +13 dBm (or +18 dBm with Option 1EA) > RF On > MOD Off.
- Set your Source frequency to 10 GHz and the Power Level to 0 dBm.
- On the E4448A, set the Center Frequency to 50 GHz and Span to 100 MHz.
- Place the marker on the 50 GHz signal, use Trace > Max Hold and measure -47 dBm (± 5 dB) or -40 dBm (± 5 dB) with Option 1EA. Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.
NOTE: Nominal loss through a 10 dB attenuator and cable is 17 dB at 50 GHz.
Verifying Mixer Performance at 65 GHz
This procedure verifies the mixer performance of the E8257D Option H60/H65 at 65 GHz. The upconverted frequency signal can be measured at the RF port on the E8257D Option H60/H65 rear panel.
To test the E8257D Option H60, set your source to 3 GHz and the E8257D Option H60 frequency to 62 GHz. This will allow you to measure the upconverted frequency at 65 GHz, using a spectrum analyzer. (The 11970A mixer must be connected to the E4448A spectrum analyzer when measuring above 50 GHz).
To test the E8257D Option H65, set your source to 15 GHz and set the frequency of the E8257D Option H65 to 25 GHz. This will allow you to measure the upconverted frequency of 65 GHz. (The 11970A mixer must be connected to the E4448A spectrum analyzer when measuring above 50 GHz).
NOTE: The E8257D Option H65 will require Option 532, 540, 550, or 567 for the 25 GHz LO frequency.
Equipment Required 65 GHz Measurement
- E4438C, E8257D, or E8267D or equivalent source (Frequency range 15 GHz for Option H65, Frequency range 3 GHz for Option H60)
- E4448A or E4407B Option AYZ or equivalent spectrum analyzer with external mixer capability (LO and IF ports to meet 11970V harmonic mixer requirements)
- 11970V-009 Mixer Connection Kit
- 11970V Waveguide Harmonic Mixer
- V281B Coaxial to Waveguide Adapter
- Cable (5063-9820) (x2)
- Cable (5062-6674)
- Adapters (customer supplied)
65 GHz Upconverter RF Measurement
Connect the following equipment for the verification procedure: Refer to Figure 3 on page 6.
- Preset all of the test equipment.
- Connect the adapter to the RF Output on the E8257D Option H60/H65 front panel. Refer to Table 2 on page 9.
- Connect the RF cable (5063-9820) to the adapter and connect the RF cable to the LO Port on the E8257D rear panel.
- Connect the E4438C/E8257D or equivalent source to the E8257D Option H60/H65 IF port on the rear panel, using a 3.5 mm cable such as the (5062-6674). Adapters may be required; see Table 2 on page 9. Refer to Figure 3 on page 6 for rear panel connections.
- Connect the V281B adapter to the 11970V harmonic mixer.
- Connect the 11970V harmonic mixer to the E4448A spectrum analyzer, using the 11970V-009 mixer connection kit.
- Connect the 11970V RF Input to the E8257D Option H60/H65 mixer RF port on the rear panel. The V281B adapter must be connected to the 11970V. Refer to "Equipment Required 65 GHz Measurement" on page 13 for waveguide to coaxial adapter.
- Set the E4448A (or equivalent) parameters as follows:
- External Mixer Input on the E4448A or E4407B: Select: [Input/Output] > Input Mixer > Input Mixer = Ext > Ext Mixer Band > 50-75 GHz (V)
- Center Frequency: 65 GHz
- Span: 100 MHz
- Ref Level: 0 dBm
- Place the marker on the 65 GHz signal and verify the approximate power level. (The harmonic mixer has a 35 dB loss).
- Option H60: -52 dBm (± 5 dB) nominal
- Option H65: -60 dBm (± 5 dB) LO Input +13 dBm (nominal), -52 dBm (± 5 dB) LO Input +18 dBm (nominal)
NOTE: Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
Replaceable Parts
NOTE: Special options are built to order; long lead times may be encountered when ordering replacement parts.
Description | Agilent Part Number | Qty |
---|---|---|
Adapter (coax SMA female to SMA female) | 1250-1753 | 1 |
Adapter (coax SMA male to SMA male) | 1250-1788 | 1 |
Adapter (coax 2.4 mm female to 2.4 mm female) | 1250-2188 | 1 |
Cable Assembly (36 in, 2.4 mm) | 5063-9820 | 1 |
Hole Plug (Dome-HD for 0.312-D-hole stl) | 6960-0023 | 2 |
Hole Plug (.75-D-hole steel) | 6960-0003 | 1 |
H60/H65 Rear Panel (analog, UNM and IEM) | E8251-00368 | 1 |
RF Cable (2.4 mm male to 2.4 mm female) | E8251-20393 | 2 |
H60/H65 Rear Panel Label (LO IF RF, 0.7 in spacing silver) | E8251-80055 | 1 |
User's and Service Guide (Option H60/H65) | E8251-90099 | 1 |
U-Wave Mixer Dbl Bal (Option H60) or U-Wave Mixer Dbl Bal (Option H65) | 0955-1848 or 0955-1871 | 1 |
Washer (1/4 in) | 2190-0421 | 3 |
Mixer Assembly
Refer to Figure 4 for replacement part identification.
Figure 4 Cable and Mixer Configuration
This diagram illustrates the configuration of cables and mixer components. It shows RF Cables (E8251-20393) connected to the mixer's LO and RF ports. An IF port is also visible. The assembly includes an SMA Barrel (1250-1788) and an SMA Connector (1250-1753), secured by washers (2190-0421 and 3050-0161).
Safety and Regulatory Information
Introduction
Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Before Applying Power
Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply.
WARNING: To prevent electrical shock, disconnect the Agilent Technologies E8257D Option H60/H65 from mains electrical supply before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Connector Care and Cleaning
If alcohol is used to clean the connectors, the power cord to the instrument must be removed. All cleaning should take place in a well ventilated area. Allow adequate time for the fumes to disperse and moist alcohol to evaporate prior to energizing the instrument.
WARNING: Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Declaration of Conformity
A copy of the Declaration of Conformity is available upon request, or a copy is available on the Agilent Technologies web site at http://regulations.corporate.agilent.com/DoC/search.htm
Statement of Compliance
This instrument has been designed and tested in accordance with CAN/CSA 22.2 No. 61010-1-04, UL Std No. 61010-1 (Second Edition), and IEC 61010-1 (Second Edition).
General Safety Considerations
Cautions
Cautions applicable to this instrument.
- CAUTION: The Mains wiring and connectors shall be compatible with the connector used in the premise electrical system. Failure to ensure adequate earth grounding by not using the correct components may cause product damage and serious injury.
- CAUTION: Always use the three prong AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause product damage and the risk of electrical shock.
- CAUTION: This product is designed for use in Installation Category II and Pollution Degree 2.
- CAUTION: Verify that the premise electrical voltage supply is within the range specified on the instrument.
- CAUTION: Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.
Servicing
Warnings applicable to this instrument.
- WARNING: Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to manufacturer's instructions.
- WARNING: This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall be inserted only into a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the product, is likely to make the product dangerous. Intentional interruption is prohibited.
- WARNING: For continued protection against fire hazard, replace line fuse only with the same type and rating. The use of other fuses or material is prohibited.
- WARNING: These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
- WARNING: The opening of covers or removal of parts is likely to expose the user to dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
- WARNING: This product is designed for use in Installation Category II and Pollution Degree 2.
- WARNING: No operator serviceable parts inside. Refer servicing to qualified personnel.
- WARNING: If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
Regulatory Information
This section contains information that is required by various government regulatory agencies.
Instrument Markings
- Instruction Documentation Symbol: Indicates the user should refer to the documentation.
- AC Input Symbol: Indicates the instrument requires alternating current (ac) input.
- WEEE Symbol: Indicates separate collection for electrical and electronic equipment, mandated under EU law.
- Power Line Switch ON Symbol: Indicates the power line switch is ON.
- Power Line Switch STANDBY Symbol: Indicates the power line switch is in the STANDBY position.
- Power Line Switch OFF Symbol: Indicates the power line switch is in the OFF position.
- Chassis Ground Symbol: Identifies a terminal internally connected to the product frame or chassis.
- CE Mark: Registered trademark of the European Community.
- CSA Mark: Registered trademark of CSA International. Complies with Canada: CSA 22.2 No. 61010-1-04.
- ISM1-A: Symbol for an Industrial Scientific and Medical Group 1 Class A product.
- ICES/NMB-001: Marking for compliance with the Canadian Interference-Causing Equipment Standard.
- DC Symbol: Indicates Direct Current.
- IP 20: Instrument designed to meet IP 20 requirements for egress and operational environment.
- N10149 (C-Tick Mark): Required mark signifying compliance with an EMC requirement.
- China RoHS Packaging Symbol: Indicates compliance with China RoHS regulations for paper/fiberboard packaging.
Battery Collection
Do not throw batteries away but collect as small chemical waste, or in accordance with your country's requirements. You may return the battery to Agilent Technologies for disposal. Refer to "Contacting Agilent" on page 23 for assistance.
Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland).
LpA<70 dB | |
---|---|
Operator Position | am Arbeitsplatz |
Normal Operation | normaler Betrieb |
per ISO 7779 | nach DIN 45635 t. 19 |
EMC Information
Complies with European EMC Directive 2004/108/EC:
- IEC/EN 61326-1
- CISPR Pub 11 Group 1, class A
- AS/NZS CISPR 11
- This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB du Canada.
Electrostatic Discharge Protection
Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the network analyzer. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. To prevent damage to the instrument:
- Always have a grounded, conductive table mat (9300-0797) in front of your test equipment.
- Always wear a grounded wrist strap (9300-1367) with grounding cord (9300-0980), connected to a grounded conductive table mat, having a 1 MΩ resistor in series with it, when handling components and assemblies or when making connections.
- Always wear a heel strap (9300-1126) when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
- Always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
- Always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
- Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the outer conductor.
- While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
- Connect the other end of the cable to the test port and remove the short from the cable.
Figure 3 ESD Protection Setup
This diagram illustrates the ESD protection setup, showing a grounded wrist strap connected to an anti-static mat, which is placed in front of the test equipment. A note indicates that a 3-wire line power cord with proper connection to ground must be used for ESD protection.
Agilent Support, Services, and Assistance
Service and Support Options
The analyzer's standard warranty is a one-year return to Agilent Technologies service warranty.
NOTE: There are many other repair and calibration options available from the Agilent Technologies support organization. These options cover a range of service agreements with varying response times. Contact Agilent for additional information on available service agreements for this product.
Contacting Agilent
Assistance with test and measurements needs and information or finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist
You can also purchase accessories or documentation items on the Internet at: http://www.agilent.com/find
If you do not have access to the Internet, contact your field engineer.
NOTE: In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine the warranty status of your unit.
Shipping Your Analyzer to Agilent for Service or Repair
IMPORTANT: Agilent Technologies reserves the right to reformat or replace the internal hard disk drive in your analyzer as part of its repair. This will erase all user information stored on the hard disk. It is imperative, therefore, that you make a backup copy of your critical test data located on the analyzer's hard disk before shipping it to Agilent for repair.
If you wish to send your instrument to Agilent Technologies for service or repair:
- Include a complete description of the service requested or of the failure and a description of any failed test and any error message.
- Ship the analyzer using the original or comparable antistatic packaging materials.
- Contact Agilent for instructions on where to ship your analyzer.