Agilent Technologies E8257D Option H60/H65 User's and Service Guide Supplement

This manual supplement describes the differences in the E8257D Option H60 and H65 in comparison to the E8257D standard instrument. Refer to the PSG E8257D/67 Signal Generators User's (E8251-90353).

Description

The E8257D Option H60 and H65 provide a double-balanced microwave mixer that is installed internally on the rear panel with connectors for easy access to the mixer ports. These options can be used with another source, such as the E4438C or E8267D, to upconvert an RF frequency to 65 GHz or downconvert an RF frequency to DC (Option H60) or 1 GHz (Option H65).

It is recommended that Option H60 be ordered with Option 567 to utilize the Option H60 frequency range of 65 GHz. It is also recommended that Option H65 be ordered with Option 520, 532, or 540. The Option H65 mixer has an internal LO frequency doubler that enables the use of an E8257D Option 520 to upconvert a 20 GHz RF frequency to 60 GHz. Option 1EA is also recommended for Option H65 due to the mixer requiring more power for optimal performance.

Refer to the microwave mixer data sheets, Figure 1 on page 4 and Figure 2 on page 5 for further information regarding the Option H60 and H65 mixer nominal performance.

Verifying the Shipment

After the instrument has been unpacked, keep the original packaging materials so they can be used if you need to transport the instrument. Inspect the instrument and all accessories for any signs of damage that may have occurred during shipment. If your instrument or any accessories appear to be damaged or missing, refer to "Agilent Support, Services, and Assistance" on page 23.

Content List
ItemAgilent Part NumberQty
RF Cable (2.4 mm)5360-98202
RF Cable (3.5 mm)5062-66741
Adapter1250-21881

Nominal Performance Specification

Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty. Refer to Figure 1 on page 4 and Figure 2 on page 5 for more information. The performance listed below does not include connector and cable losses.

H60 mixer nominal performance:

H65 mixer nominal performance:

Figure 1 E8257D Option H60 Mixer Data Sheet

This data sheet provides specifications for the E8257D Option H60 Double-Balanced Mixer. Key specifications include:

Guaranteed Specifications (From -55°C to +100°C) include Conversion Loss / Noise Figure (SSB) for various RF/IF ranges, Third Harmonic LO Conversion Loss, and Isolation (LO-RF, LO-IF, RF-IF). Input 1dB Compression and Input Two-Tone Third Order Intercept Point are also specified.

Graphs illustrate Conversion Loss vs. GHz, 3rd Harmonic Conv. Loss vs. GHz, Relative IF Response vs. GHz, LO to RF Isolation vs. GHz, and LO to IF Isolation vs. GHz.

Figure 2 E8257D Option H65 Mixer Data Sheet

This data sheet provides specifications for the E8257D Option H65 Double-Balanced Mixer with Integrated LO-Doubler. Key specifications include:

Guaranteed Specifications (From -55°C to +100°C) include Conversion Loss / Noise Figure (SSB) for various RF/IF ranges, Isolation (2LO-RF, 2LO-IF, RF-IF), Input 1dB Compression, and Input Two-Tone Third Order Intercept Point.

A graph illustrates Conversion Loss vs. GHz.

Operational and Verification Procedure

The E8257D Option H60 or H65 can be used to upconvert or downconvert the frequency of your source. Refer to Figure 3.

CAUTION: Inputs are ESD sensitive. It is recommended to keep connector covers on when not in use. Refer to "Electrostatic Discharge Protection" on page 22.

  1. Connect the adapter to the E8257D Option H60/65 RF Output. Refer to Table 2 on page 9.
  2. Connect the RF cable (5063-9820) from the adapter to the LO Port on the E8257D rear panel.

Figure 3 Rear Panel

The rear panel shows connections for ALC INPUT, 1 GHz REF IN, and MIXER PORTS (LO, IF, RF), as well as a 1 GHz OUT port.

Upconverting the Frequency

The upconverted frequency signal can be measured on the RF port on the E8257D Option H60/H65 rear panel.

  1. Preset the instrument with Option H60/H65.
  2. Set the power level to +13 dBm. Refer to Input 1 dB Compression for optimal LO Input Power in Figure 1 on page 4 and Figure 2 on page 5.
  3. Connect your source to the E8257D Option H60/H65 IF port on the rear panel, using a 3.5 mm cable such as the (5062-6674).
  4. Connect the second cable (5360-9820) to the RF Port.
  5. To upconvert your source frequency, set the E8257D Option H60 frequency to the desired RF Output frequency minus your source frequency. For Option H65, set the frequency to the desired RF minus your source frequency and divide by 2. (The IF is subtracted from the upconverted RF for the LO frequency).

NOTE: Filtering may be required to remove unwanted signals. The RF Output frequency will have the difference of the E8257D Option H60/H65, as well as the two original frequencies, along with the sum upconverted product.

Equation 1 Upconverted Frequency (H60)

Upconverted RF = LO(E8257DH60) + IF(yoursource)

Equation 2 Upconverted Frequency (H65)

Upconverted RF = LO(E8257DH65) × 2 + IF(yoursource)

NOTE: The LO Port has a frequency doubler on the LO Input (H65 only). The E8257D Option H65 frequency output will be set to half the desired LO frequency.

Downconverting the Frequency

The downconverted frequency signal can be measured on the IF port on the E8257D Option H60/H65 rear panel.

  1. Preset the instrument with Option H60/H65.
  2. Set the power level to +13 dBm. Refer to Input 1 dB Compression for optimal LO Input Power in Figure 1 on page 4 and Figure 2 on page 5.
  3. Connect your source to the E8257D Option H60/H65 RF port on the rear panel, using a 2.4 mm cable such as the (5063-9820).
  4. Connect the cable (5062-6674) to the IF Port.
  5. To downconvert your source frequency, set the E8257D Option H60 frequency to your source frequency minus the desired IF Output. For Option H65, set the frequency to the source frequency minus the desired IF Output frequency and divide by 2. (The IF is subtracted from the upconverted RF for the LO frequency).

NOTE: Filtering may be required to remove unwanted signals. The IF Output frequency will have the sum of the E8257D Option H60/H65, as well as the two original frequencies, along with the difference down converted product.

Equation 3 Downconverted Frequency (H60)

Downconverted IF = RF(yoursource) - LO(E8257DH60)

Equation 4 Downconverted Frequency (H65)

Downconverted IF = RF(yoursource) - LO(E8257DH65) × 2

NOTE: The LO Port has a frequency doubler on the LO Input (H65 only). The E8257D Option H65 frequency output will be set to half the desired LO frequency.

Verification Procedure

This procedure verifies the mixer is functioning in the E8257D Option H60/H65 for frequency upconversion. There are two methods to verify the mixer's function. The first method tests an upconverted frequency up to 50 GHz using minimal equipment. The second method tests an upconverted frequency up to 65 GHz.

Verifying Mixer Performance up to 50 GHz

This test measures three frequency points: low end, 23 GHz, and 50 GHz. The 50 GHz test requires the DUT to have Option 550 or 567 for Option H60. Units with Option H65 can have Option 540, 532, or 520.

Equipment Required for Measurement up to 50 GHz

Table 2 Recommended Adapters
Instrument OptionsAdapter DescriptionAgilent Part Number
520, 5323.5 mm (f) to (f)5061-5311
540, 550 and 5672.4 mm (f) to 3.5 mm (f)1250-2187
540, 550 and 5672.4 mm (f) to 2.4 mm (f)1250-2188
1ED2.4 mm (f) to Type-N (m)11903D
1ED3.5 mm (f) to Type-N (m)1250-1744

Connect the following equipment for the verification procedure:

  1. Preset all of the test equipment.
  2. Connect the adapter to the RF Output on the E8257D Option H60/H65 front panel. Refer to Table 2 on page 9.
  3. Connect the RF cable (5063-9820) to the adapter and connect the RF cable to the LO Port on the E8257D rear panel.
  4. Connect the E4438C/E8257D or equivalent source to the E8257D Option H60/H65 IF port on the rear panel, using a 3.5 mm cable such as the (5062-6674). Adapters may be required; see Table 2 on page 9. Refer to Figure 3 on page 6 for rear panel connections.
  5. Connect the E4448A or equivalent spectrum analyzer (with a 10 dB attenuator on the input) to the E8257D Option H60/H65 RF port on the rear panel, using a 2.4 mm cable such as the (5063-9820). Adapters may be required.

NOTE: Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.

Option H60 Test

4 GHz Upconverter RF Measurement

  1. Set the E8257D Option H60 frequency to 2.7 GHz.
  2. Set the Power Level to +13 dBm > RF On > MOD Off.
  3. Set your Source frequency to 1.3 GHz and the Power Level to 0 dBm.
  4. On the E4448A, set the Center Frequency to 4 GHz and Span to 100 MHz.
  5. Place the marker on the 4 GHz signal, use Trace > Max Hold and measure -30 dBm (± 5 dB). Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.

23 GHz Upconverter RF Measurement

  1. Set the E8257D Option H60 frequency to 20 GHz.
  2. Set the Power Level to +13 dBm > RF On > MOD Off.
  3. Set your Source frequency to 3 GHz and the Power Level to 0 dBm.
  4. On the E4448A, set the Center Frequency to 23 GHz and Span to 100 MHz.
  5. Place the marker on the 23 GHz signal, use Trace > Max Hold and measure -30 dBm (± 5 dB). Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.

50 GHz Upconverter RF Measurement

  1. Set the E8257D Option H60 frequency to 47 GHz.
  2. Set the Power Level to +13 dBm > RF On > MOD Off.
  3. Set your Source frequency to 3 GHz and the Power Level to 0 dBm.
  4. On the E4448A, set the Center Frequency to 50 GHz and Span to 100 MHz.
  5. Place the marker on the 50 GHz signal, use Trace > Max Hold and measure -35 dBm (± 5 dB). Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.

NOTE: Nominal loss through a 10 dB attenuator and cable is 17 dB at 50 GHz.

Option H65 Test

20 GHz Upconverter RF Measurement

  1. Set the E8257D Option H65 frequency to 9 GHz.
  2. Set the Power Level to +13 dBm (or +18 dBm with Option 1EA) > RF On > MOD Off.
  3. Set your Source frequency to 2 GHz and the Power Level to 0 dBm.
  4. On the E4448A, set the Center Frequency to 20 GHz and Span to 100 MHz.
  5. Place the marker on the 20 GHz signal, use Trace > Max Hold and measure -47 dBm (± 5 dB) or -37 dBm (± 5 dB) with Option 1EA. Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.

23 GHz Upconverter RF Measurement

  1. Set the E8257D Option H65 frequency to 10 GHz.
  2. Set the Power Level to +13 dBm (or +18 dBm with Option 1EA) > RF On > MOD Off.
  3. Set your Source frequency to 3 GHz and the Power Level to 0 dBm.
  4. On the E4448A, set the Center Frequency to 23 GHz and Span to 100 MHz.
  5. Place the marker on the 23 GHz signal, use Trace > Max Hold and measure -45 dBm (± 5 dB) or -35 dBm (± 5 dB) with Option 1EA. Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.

50 GHz Upconverter RF Measurement

  1. Set the E8257D Option H65 frequency to 20 GHz.
  2. Set the Power Level to +13 dBm (or +18 dBm with Option 1EA) > RF On > MOD Off.
  3. Set your Source frequency to 10 GHz and the Power Level to 0 dBm.
  4. On the E4448A, set the Center Frequency to 50 GHz and Span to 100 MHz.
  5. Place the marker on the 50 GHz signal, use Trace > Max Hold and measure -47 dBm (± 5 dB) or -40 dBm (± 5 dB) with Option 1EA. Cable loss and amplitude accuracy of the spectrum analyzer will affect the measured value.

NOTE: Nominal loss through a 10 dB attenuator and cable is 17 dB at 50 GHz.

Verifying Mixer Performance at 65 GHz

This procedure verifies the mixer performance of the E8257D Option H60/H65 at 65 GHz. The upconverted frequency signal can be measured at the RF port on the E8257D Option H60/H65 rear panel.

To test the E8257D Option H60, set your source to 3 GHz and the E8257D Option H60 frequency to 62 GHz. This will allow you to measure the upconverted frequency at 65 GHz, using a spectrum analyzer. (The 11970A mixer must be connected to the E4448A spectrum analyzer when measuring above 50 GHz).

To test the E8257D Option H65, set your source to 15 GHz and set the frequency of the E8257D Option H65 to 25 GHz. This will allow you to measure the upconverted frequency of 65 GHz. (The 11970A mixer must be connected to the E4448A spectrum analyzer when measuring above 50 GHz).

NOTE: The E8257D Option H65 will require Option 532, 540, 550, or 567 for the 25 GHz LO frequency.

Equipment Required 65 GHz Measurement

65 GHz Upconverter RF Measurement

Connect the following equipment for the verification procedure: Refer to Figure 3 on page 6.

  1. Preset all of the test equipment.
  2. Connect the adapter to the RF Output on the E8257D Option H60/H65 front panel. Refer to Table 2 on page 9.
  3. Connect the RF cable (5063-9820) to the adapter and connect the RF cable to the LO Port on the E8257D rear panel.
  4. Connect the E4438C/E8257D or equivalent source to the E8257D Option H60/H65 IF port on the rear panel, using a 3.5 mm cable such as the (5062-6674). Adapters may be required; see Table 2 on page 9. Refer to Figure 3 on page 6 for rear panel connections.
  5. Connect the V281B adapter to the 11970V harmonic mixer.
  6. Connect the 11970V harmonic mixer to the E4448A spectrum analyzer, using the 11970V-009 mixer connection kit.
  7. Connect the 11970V RF Input to the E8257D Option H60/H65 mixer RF port on the rear panel. The V281B adapter must be connected to the 11970V. Refer to "Equipment Required 65 GHz Measurement" on page 13 for waveguide to coaxial adapter.
  8. Set the E4448A (or equivalent) parameters as follows:
    • External Mixer Input on the E4448A or E4407B: Select: [Input/Output] > Input Mixer > Input Mixer = Ext > Ext Mixer Band > 50-75 GHz (V)
    • Center Frequency: 65 GHz
    • Span: 100 MHz
    • Ref Level: 0 dBm
  9. Place the marker on the 65 GHz signal and verify the approximate power level. (The harmonic mixer has a 35 dB loss).
    • Option H60: -52 dBm (± 5 dB) nominal
    • Option H65: -60 dBm (± 5 dB) LO Input +13 dBm (nominal), -52 dBm (± 5 dB) LO Input +18 dBm (nominal)

NOTE: Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.

Replaceable Parts

NOTE: Special options are built to order; long lead times may be encountered when ordering replacement parts.

Replaceable Parts
DescriptionAgilent Part NumberQty
Adapter (coax SMA female to SMA female)1250-17531
Adapter (coax SMA male to SMA male)1250-17881
Adapter (coax 2.4 mm female to 2.4 mm female)1250-21881
Cable Assembly (36 in, 2.4 mm)5063-98201
Hole Plug (Dome-HD for 0.312-D-hole stl)6960-00232
Hole Plug (.75-D-hole steel)6960-00031
H60/H65 Rear Panel (analog, UNM and IEM)E8251-003681
RF Cable (2.4 mm male to 2.4 mm female)E8251-203932
H60/H65 Rear Panel Label (LO IF RF, 0.7 in spacing silver)E8251-800551
User's and Service Guide (Option H60/H65)E8251-900991
U-Wave Mixer Dbl Bal (Option H60) or U-Wave Mixer Dbl Bal (Option H65)0955-1848 or 0955-18711
Washer (1/4 in)2190-04213

Mixer Assembly

Refer to Figure 4 for replacement part identification.

Figure 4 Cable and Mixer Configuration

This diagram illustrates the configuration of cables and mixer components. It shows RF Cables (E8251-20393) connected to the mixer's LO and RF ports. An IF port is also visible. The assembly includes an SMA Barrel (1250-1788) and an SMA Connector (1250-1753), secured by washers (2190-0421 and 3050-0161).

Safety and Regulatory Information

Introduction

Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.

Before Applying Power

Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply.

WARNING: To prevent electrical shock, disconnect the Agilent Technologies E8257D Option H60/H65 from mains electrical supply before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.

Connector Care and Cleaning

If alcohol is used to clean the connectors, the power cord to the instrument must be removed. All cleaning should take place in a well ventilated area. Allow adequate time for the fumes to disperse and moist alcohol to evaporate prior to energizing the instrument.

WARNING: Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.

Declaration of Conformity

A copy of the Declaration of Conformity is available upon request, or a copy is available on the Agilent Technologies web site at http://regulations.corporate.agilent.com/DoC/search.htm

Statement of Compliance

This instrument has been designed and tested in accordance with CAN/CSA 22.2 No. 61010-1-04, UL Std No. 61010-1 (Second Edition), and IEC 61010-1 (Second Edition).

General Safety Considerations

Cautions

Cautions applicable to this instrument.

Servicing

Warnings applicable to this instrument.

Regulatory Information

This section contains information that is required by various government regulatory agencies.

Instrument Markings

Battery Collection

Do not throw batteries away but collect as small chemical waste, or in accordance with your country's requirements. You may return the battery to Agilent Technologies for disposal. Refer to "Contacting Agilent" on page 23 for assistance.

Compliance with German Noise Requirements

This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland).

Acoustic Noise Emission/Geraeuschemission
LpA<70 dB
Operator Positionam Arbeitsplatz
Normal Operationnormaler Betrieb
per ISO 7779nach DIN 45635 t. 19

EMC Information

Complies with European EMC Directive 2004/108/EC:

Electrostatic Discharge Protection

Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the network analyzer. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. To prevent damage to the instrument:

Figure 3 ESD Protection Setup

This diagram illustrates the ESD protection setup, showing a grounded wrist strap connected to an anti-static mat, which is placed in front of the test equipment. A note indicates that a 3-wire line power cord with proper connection to ground must be used for ESD protection.

Agilent Support, Services, and Assistance

Service and Support Options

The analyzer's standard warranty is a one-year return to Agilent Technologies service warranty.

NOTE: There are many other repair and calibration options available from the Agilent Technologies support organization. These options cover a range of service agreements with varying response times. Contact Agilent for additional information on available service agreements for this product.

Contacting Agilent

Assistance with test and measurements needs and information or finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist

You can also purchase accessories or documentation items on the Internet at: http://www.agilent.com/find

If you do not have access to the Internet, contact your field engineer.

NOTE: In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine the warranty status of your unit.

Shipping Your Analyzer to Agilent for Service or Repair

IMPORTANT: Agilent Technologies reserves the right to reformat or replace the internal hard disk drive in your analyzer as part of its repair. This will erase all user information stored on the hard disk. It is imperative, therefore, that you make a backup copy of your critical test data located on the analyzer's hard disk before shipping it to Agilent for repair.

If you wish to send your instrument to Agilent Technologies for service or repair:

Models: H60, H65, E8257D PSG Analog Signal Generator, E8257D, PSG Analog Signal Generator, Analog Signal Generator, Signal Generator, Generator

PDF preview unavailable. Download the PDF instead.

agilent-e8257d manual-new ?ext= Acrobat Distiller 8.1.0 (Windows)

Related Documents

Preview Agilent 4072A/4073A Parametric Tester Installation Guide
This installation guide provides detailed instructions for setting up and verifying the Agilent 4072A and 4073A Advanced Parametric Testers, covering system cabinet installation, component mounting, software setup, and safety procedures.
Preview 8662A Synthesized Signal Generator Operating and Service Manual
This document is the operating and service manual for the 8662A Synthesized Signal Generator, covering specifications, safety considerations, and errata updates. It details the instrument's functionality from 10 kHz to 1280 MHz and provides essential information for its operation and maintenance.
Preview Agilent B1500A Semiconductor Device Analyzer Programming Guide
Comprehensive programming guide for the Agilent B1500A Semiconductor Device Analyzer, detailing remote control via GPIB, measurement modes, programming examples, and command reference for semiconductor testing.
Preview Agilent G5541A NGS Bravo Option A Installation Checklist
Installation checklist for the Agilent G5541A NGS Bravo Option A, ensuring correct installation and functionality of the instrument and associated systems.
Preview Agilent 34980A Multifunction Switch/Measure Unit User's Guide
Comprehensive user guide for the Agilent 34980A Multifunction Switch/Measure Unit, detailing its features, operation, and connectivity for data acquisition and measurement control systems. Explore its capabilities for signal routing, switching, and precise measurements.
Preview Agilent ICP-MS Troubleshooting and Maintenance Guidelines
Comprehensive guide from Agilent Technologies on troubleshooting and maintaining Inductively Coupled Plasma - Mass Spectrometry (ICP-MS) systems, covering common issues, causes, cleaning procedures, and best practices for optimal performance.
Preview Agilent 55B AA Spectrometer User's Guide
Comprehensive user guide for the Agilent 55B AA Atomic Absorption Spectrometer, covering installation, operation, maintenance, safety, and troubleshooting for laboratory use.
Preview Agilent PNA Series RF Network Analyzers Service Guide: E8356A, E8357A, E8358A
Comprehensive service guide for Agilent Technologies PNA Series RF Network Analyzers (E8356A, E8357A, E8358A), detailing maintenance, testing, troubleshooting, and adjustment procedures for optimal performance.