1. General Safety and Service Information
Important: Read and follow all instructions in this manual. The device is intended only for the described use.
Safety Precautions:
- No operation in potentially explosive atmospheres! The device is not approved for use in such areas.
- Fire Hazard: Do not use in conjunction with easily flammable gases (e.g., anesthetic gases).
- Portable or mobile RF communication devices may interfere with electromedical equipment.
- Do not connect the device to a PC via data cable while it is in use with a patient.
Commissioning:
Before commissioning, check the calibration sticker on the device's housing. If the date indicated on the sticker has been reached or exceeded, the AlcoQuant 6020 must be recalibrated or sent for servicing and should no longer be used for measurements. Maintenance and calibration may only be performed by appropriately trained and EnviteC-authorized personnel.
Maintenance:
The device must undergo calibration and maintenance every six months by authorized specialist personnel (with a report). Maintenance on the device must be carried out exclusively by authorized service personnel. Only original parts from EnviteC-Wismar GmbH may be used for repairs. The semi-annual calibration is necessary to guarantee the accuracy specified.
Symbol Explanation:
- ⚠️: Indicates that death, severe personal injury, or significant property damage can occur if the corresponding precautions are not taken.
- ℹ️: This is important information about the product or a part of the manual that should be particularly noted.
Disposal:
According to Directive 2002/96/EG (WEEE), electrical and electronic equipment must be returned to the manufacturer for proper disposal after dismantling!
2. Intended Use and Areas of Application
The AlcoQuant 6020 was developed for the rapid and precise determination of breath alcohol concentration and is intended for the following areas:
- Police routine checks
- Correctional facilities
- Occupational safety
3. Device and Function Description
During exhalation by the subject, the AlcoQuant 6020 measures the exhaled volume and takes a sample after 1.2 L (default setting). The AlcoQuant 6020 features a special electrochemical sensor that reacts exclusively to alcohol. The key characteristic of this sensor is its specific response to alcohol and its long-term stability, allowing for a 6-month calibration interval. The influence of other substances that may be present in exhaled air, such as acetone or eucalyptol, is negligible due to the applied electrochemical measurement principle.
The sampling system, developed for highest accuracy, allows for short measurement intervals and thus extremely short waiting times between measurements. The use of the new improved ethanol sensor E 100 and the innovative sampling system guarantees high reproducibility of measurement results. The device incorporates the latest technologies to prevent measurement errors and ensure the highest measurement accuracy, including continuous monitoring of the following functions and components:
- Device temperature
- Sampling
- Signal generator
- LEDs
- Battery
- Alcohol sensor
- Internal memory
- Calibration intervals
The device is equipped with a data memory and a serial data interface (RS232). Data can be transferred to a PC and evaluated using the "Datamanager Alco 60XX" software, including the data cable. Operation is performed via the device's three membrane keys. For hygiene, the membrane keys can be easily cleaned with a damp cloth. All device status and error messages are displayed in plain text on the backlit graphic display.
4. Operation
Device Overview:
Front View:
- 1. ON/OFF Button
- 2. Select Button
- 3. Display
- 4. Red LED - Error Indicator
- 5. Yellow LED - Measurement in progress/Caution
- 6. Mouthpiece connector
- 7. Green LED - Ready for measurement
Rear View:
- 8. OK Button
- 9. RS 232 Data Interface
- 10. Blow-off opening
- 11. Device Label
- 12. Battery compartment opener
- 13. Battery compartment cover
Label Symbols:
- ℹ️: Refer to operating instructions.
- SN: Serial Number
- ?: Date of Manufacture
- Type B application part
- PN: Product Number
- ⚠️: Observe disposal regulations.
Powering On:
Press and hold the ON/OFF button for approximately 1 second. The device automatically performs a self-test of its internal functions and components. For approximately 2 seconds during this self-test phase, the device version, software version (Rev x.x0), and the next calibration date are displayed. After approximately 3 seconds, the device is ready for measurement in active mode (green LED blinks every 1.5 seconds, display shows "Start measurement").
Powering Off:
Press and hold the ON/OFF button for approximately 3 seconds. If no measurement or button input is performed within 2.5 minutes, the AlcoQuant 6020 will automatically shut down.
Important Usage Notes:
- Smoking: Smoking in the exhaled air damages or destroys the measurement system. Therefore, it must be ensured that the test subject has not smoked immediately before the measurement. Waiting time: At least 5 minutes after smoking.
- Alcohol in Mouth: Residual alcohol in the mouth falsifies the measurement. Therefore, it must be ensured that the test subject has not consumed any alcohol immediately before the measurement. Waiting time: At least 15 minutes after the last alcohol intake. Alcoholic mouth sprays, medical juices and drops, as well as belching and vomiting can also lead to temporary mouth alcohol. Rinsing the mouth with water or non-alcoholic beverages does not replace the waiting time!
- Hyperventilation: The test subject should breathe normally and calmly before the measurement, and not inhale or exhale deeply multiple times (hyperventilation), as this can temporarily alter the breath alcohol concentration due to cooling of the breath air, thus falsifying the measurement result.
4.1 Active Measurement (Menu Point 1)
Measurement in the active device mode is used for precise determination of alcohol concentration in exhaled air. This ensures that only deep lung air is supplied to the sensor. This is necessary because there is a direct correlation between the alcohol content of deep lung air and blood alcohol content.
Inserting Original Mouthpiece:
Insert an EnviteC mouthpiece into the designated mouthpiece connector of the AlcoQuant 6020. To do this, open the hygienic packaging of the mouthpiece at the perforation and use it as a hygienic contact shield during insertion (see illustration).
Possible Errors – Measurement Interruption:
If the breath flow is interrupted during the blowing process, a long double tone will be heard, and the error message "Blow error" will appear. The blowing process must be repeated.
- A new mouthpiece must be used for each measurement in active mode.
- The blow-off opening must be kept clear during the blowing process, otherwise the measurement value will be falsified.
- If the subject appears conspicuous due to their behavior or alcohol odor despite a zero measurement, the device must be checked by trained service personnel.
- Patients with reduced lung volume may not reach the required exhalation volume of 1.2 L. In this case, perform a measurement in passive measurement mode (without mouthpiece).
Starting Active Measurement:
After switching on, the device automatically switches to active mode, and "Start measurement" appears on the display. The measurement starts automatically with the exhalation of breath.
Blowing Process:
Blow evenly and with sufficient airflow into the device. An exhalation volume of 1.2 liters must be reached.
Acoustic + Optical Signal:
Correct blowing is indicated by an acoustic signal and the blinking yellow LED.
End of Measurement:
The blowing process is complete when no more sound is heard and the yellow LED turns off. The device analyzes and displays the measurement value. Press a button to switch the device back to measurement readiness.
4.2 Passive Measurement (Menu Point 2)
Select ">Passive measurement<". Measurement in passive mode is performed without a mouthpiece and is used to check ambient air or exhaled air, e.g., from unresponsive patients. It exclusively checks whether alcohol is present in the ambient air. A displayed value other than "0.00" is an indication of alcohol in the ambient air.
Measurement Procedure:
Hold the ready device with the blow-off opening (without mouthpiece) close to the area to be measured, e.g., the patient's mouth and nose, for approximately 5-10 seconds (s). Then, trigger the sample by briefly pressing the "Start" button. The sample should be triggered during the patient's exhalation phase.
Results:
- No Alcohol Detected: If no alcohol is detected, a signal tone sounds, and the green LED lights up.
- Alcohol Detected: If alcohol and a measurement value are displayed, a double tone sounds, and the red LED lights up.
Note: Values measured in passive mode do not correspond to the breath alcohol concentration in deep lung air and therefore do not allow conclusions to be drawn about the blood alcohol concentration. The exact breath alcohol concentration of deep lung air can only be reliably determined in active measurement mode.
4.3 Data Memory
The device has a data memory where measurement results are stored. Each measurement performed is automatically saved with date, time, measurement value, and protocol number. The data memory stores 9,999 data sets.
Storage Warnings:
- Reaching 90% of the storage capacity (9,000 measurements) is indicated by the message ">Caution: Memory 90% full<" on the display. This warning must be confirmed by briefly pressing the "OK" button.
- After the 9,999th measurement, the message ">Caution: Memory full<" appears on the display. This message must also be confirmed by briefly pressing the "OK" button. The oldest measurement value will then be automatically overwritten with the new measurement value!
Measurement values stored in the data memory are retained even when the batteries are changed.
5. Maintenance / Device Care
Calibration:
The device must undergo maintenance every 6 months. This includes the following points:
- Calibration of the device
- Check of the proper device functions
- Calibration verification
Maintenance must be carried out exclusively by EnviteC or by service personnel trained by EnviteC.
Cleaning:
Clean only with a soft, disposable cloth dampened with a mild cleaning agent, without excessive pressure.
Disinfection:
Perform wipe disinfection with a soft cloth dampened with isopropyl alcohol, without excessive pressure, according to EN 60601-1 (Durability of markings test).
⚠️: Keep the battery compartment closed during cleaning and disinfection! No moisture must enter the device openings. Care must be taken to ensure that the electrical contacts in the device (battery) and at the serial interface remain clean and dry to ensure the proper functioning of the device. After cleaning/disinfection with alcohol, a waiting period of at least 15 minutes must be observed until the next measurement!
5.1 Battery Replacement
A repeating beep and the display message "Replace battery" indicate that the batteries need to be replaced. The device will automatically shut down after 10 seconds.
Batteries can be replaced by gently pressing the battery cover to release the latch and removing the cover. The battery compartment holds a battery pack with 4 x LR6 (AA) Alkaline Mignon cells (Recommendation: Duracell® Procell®). When replacing batteries, all four batteries must generally be replaced. Pay attention to the correct polarity.
Then, reinsert the battery pack into the housing. The battery pack terminals (arrow), as shown in the following image, must be directed towards the device and be at the top. Then, close the battery compartment cover again.
⚠️: Do not insert rechargeable batteries of the same form factor (LR6, AA) into the battery pack! The bar capacity indicator on the display will show incorrect values, and device malfunctions may occur.
6. Disposal of Device, Alcohol Sensor, Batteries
The device and the alcohol sensor must not be disposed of in household waste. Please send both back to EnviteC, stating "Disposal". Batteries must be disposed of as special waste according to the European Battery Directive (EU 91/157/EEC).
EnviteC-Wismar GmbH
Service
Alter Holzhafen 18
D-23966 Wismar
⚠️: Do not throw batteries into fire or open them forcefully. There is a risk of explosion!
7. Error Description and Resolution
Every device error is signaled acoustically by a short double tone and optically either by the yellow LED (error that does not hinder measurement) or by the red LED (error that hinders measurement) and leads to a corresponding error message in the display. The error message must be confirmed by briefly pressing the "OK" button.
Error Display | Possible Cause | Resolution |
---|---|---|
User Errors | ||
Blow Error | This error appears when the breath flow is interrupted or falls below the required level during the blowing process before the minimum exhalation volume of 1.2 L is reached. | After the error output (approx. 2 seconds), the device switches back to the previous measurement mode. A signal tone and the green LED blinking every 1.5 seconds indicate readiness for measurement. The blowing process can be restarted. The subject should blow longer and stronger than before. |
Device Errors Not Hindering Measurement (Yellow LED) | ||
Signal Generator Error | A defective signal generator has been detected. | Measurements are still possible. The error message is only displayed during the self-test (when switching on). Please contact service if the error occurs repeatedly. |
Red LED Error | A defective or dimly lit LED has been detected. | Measurements are still possible. The error message is only displayed during the self-test (when switching on). Please contact service if the error occurs repeatedly. |
Date/Time Error | An error occurred while determining the current date and time. | Measurements are still possible. The incorrect date/time values may be stored with the current measurement. Please contact service if the error occurs repeatedly. |
Device Errors Hindering Measurement (Red LED) | ||
Temperature Error | The temperature measured by the device is outside the specified range of -5 to +50 °C. | Measurements may be inaccurate. Outside the specified temperature range, the measured values may be subject to a greater error tolerance than indicated. Turn off the device and bring it into the specified temperature range (cool or warm). Then turn the device back on. If the error occurs despite the device being obviously within the specified temperature range, please contact service. |
Sampling Error | An error was detected in the sampling system. | Turn the device off and on again. If the error recurs during self-test or measurement, please inform service. A faulty measurement cannot occur, as the measurement value is discarded in case of "Sampling Error". |
Memory Error | An error occurred during a read or write access to the internal memory. | No measurement is possible. The measurement value may not be displayed correctly. Data cannot be read correctly from memory, and the measured values cannot be saved in memory. Repeat measurement. Please contact service if the error occurs repeatedly. |
Alcohol Sensor Error | An error was detected in the alcohol sensor. | No measurement is possible. Turn the device off and on again. Please contact service if the error occurs repeatedly. |
Calibration Expired | Calibration has expired. | The device must be recalibrated; please contact service. If the device has already been calibrated, it is possible that the current date settings are incorrect. In this case, please also contact service. |
8. Technical Specifications
Parameter | Specification |
---|---|
Measurement Modes | Active and Passive |
Measurement Range | 0 to 5.5% BAC |
Accuracy |
|
Environmental Conditions |
|
Minimum Exhalation Volume | 1.2 Liter |
Batteries | 4 x LR6 (AA) Alkaline Mignon cells (Recommendation: Duracell® Procell®) |
Measurements per Battery Pack | Approx. 1600 measurements, depending on batteries and environmental conditions |
Device Interface | RS 232 (serial interface). Data can be transferred to a PC. |
Dimensions (L x W x H) | 190 mm x 61 mm x 38 mm |
Weight | 275 g including batteries |
Readiness after Power On | Approx. 6 s |
Readiness after 0% Measurement | Approx. 2 s |
Readiness after 1% Measurement | Approx. 5 s |
Sensor Type | Electrochemical cell E100 |
Cross-sensitivities | Exhaled air substances like acetone etc. are negligible. |
Units | User configurable: ‰, mg/l |
Input | Three-button operation |
Power Saving Mode | Automatic shutdown after 2.5 minutes of inactivity, backlight off after 15 seconds. |
Display | Backlit, full graphic |
Data Memory | Storage of 9,999 measurement values with date and time. |
Mouthpieces | Single-use EnviteC |
Software | PC software for statistical evaluation, data transfer, and logging (optional). |
Calibration | Every 6 months |
Safety Checks | Every 6 months |
Subject to technical changes!
9. Warranty
EnviteC grants a two-year warranty for defects attributable to material or manufacturing errors from the date of purchase. Defects covered by the warranty will be rectified within the scope of our warranty conditions. EnviteC does not grant a warranty if the operator endangers the device's function by not adhering to the operating instructions, improper handling, non-intended use, or unauthorized intervention. In such cases, liability passes to the operator! The warranty expires in case of chemical exposure from leaked batteries or use of batteries beyond their expiry date. Return transport in case of a repair not covered by the warranty claim is at the customer's expense.
Devices requiring repair should be sent with all accessories to the following address:
EnviteC-Wismar GmbH
Service
Alter Holzhafen 18
D-23966 Wismar
Warranty claims can only be asserted in conjunction with the proof of purchase!
10. Ordering Information
AlcoQuant® 6020 – Complete Breath Alcohol Tester Set
Item | Description | Part No. |
---|---|---|
Set | AlcoQuant® 6020, 4 batteries (Alkali, R6/AA, 1.5V), 25 mouthpieces, transport case, and operating manual | 1000300 |
Customer-specific settings (e.g., unit, language) are available upon consultation.
Accessories:
Item | Description | Part No. |
---|---|---|
Transport Case Alco 60XX | Case with insert / 357 mm x 310 mm x 120 mm | 1000644 |
Mouthpieces S-Type | Mouthpieces with backflow preventer, 25 pcs. individually hygienically packed | 31-30-0022 |
Mouthpieces D-Type | Mouthpieces with backflow preventer, 25 pcs. individually hygienically packed | 31-30-0024 |
Software "Datamanager Alco 60XX" | PC software for evaluation and documentation of AlcoQuant® 6020 measurement results | 1000520 |
PC Cable for Data and Service Manager | Data cable for connecting the AlcoQuant® 6020 to RS 232 | 1000519 |
Needle Printer | Mobile needle printer incl. battery with accessories consisting of: printer, charging adapter, and 1 roll of printer paper | 1000616 |
Printer Paper for Printer 1000616 | 5 rolls | 31-30-0026 |
Ribbon for Printer 1000616 | 1 ribbon for printer | 31-30-0079 |
Printer Cable for Printer 1000616 | 25-pin printer cable for connecting the device to the printer | 1000518 |
Car Adapter for Printer 1000616 | 1000768 | |
Thermo Printer | Mobile thermal printer incl. battery with accessories consisting of: printer, charging adapter, and 1 roll of printer paper | 1000977 |
Printer Paper for Printer 1000977 | 1 roll standard | 1001100 |
Printer Paper for Printer 1000977 | 1 roll document-grade | 1001101 |
Printer Cable for Printer 1000977 | 1000563 | |
Car Adapter for Printer 1000977 | 1001103 | |
Yellow Carrying Bag for AlcoQuant® 6020 | 1001105 |