Seaward PV:1525 Handheld PV Tester

The PV:1525 is designed for installers and maintenance technicians to perform electrical safety tests and performance measurements on PV systems up to 1500 VDC and 25 ADC – ideal for residential and commercial installations.

Key Features

Essential Measurement Functions

Technical Specifications

Measurement of Open-Circuit Voltage and Voltage (PV Connections)

ParameterSpecification
Display Range+10.0 VDC ... +1500 VDC, -10.0 VDC ... -440 VDC, 10.0 VAC ... 440 VAC
Measurement Range+10.0 VDC ... +1500 VDC, -10.0 VDC ... -440 VDC, 10.0 VAC ... 440 VAC
Resolution0.1 VDC Maximum
AccuracyFor +DC: ± (0.5% + 2 digits) For –DC and AC: ± (5% + 2 digits)

Measurement of Short-Circuit Current (PV Connections)

ParameterSpecification
Display Range0.00 ADC ... 25.00 ADC
Measurement Range0.50 ADC ... 25.00 ADC
Power Measurement37.5 kW
Resolution0.01 ADC Maximum
Accuracy± (1% + 2 digits)
Maximum Power37.5 kW
Voltage1500 VDC
Current25 ADC

Protective Conductor Continuity / Resistance Measurement

ParameterSpecification
Open-Circuit Test Voltage> 4 VDC, nominal
Test Current at Ω> 200 mA
Display Range0.00 Ω ... 199 Ω
Measurement Range (IEC 61557-4)0.05 Ω ... 199 Ω
Resolution0.01 Ω Maximum
Accuracy0.05 Ω ... 0.09 Ω: ± (2% + 1 digit)
0.10 Ω ... 0.19 Ω: ± (2% + 2 digits)
0.20 Ω ... 1.99 Ω: ± (2% + 3 digits)
2.0 Ω ... 4.9 Ω: ± (2% + 2 digits)
5.0 Ω ... 199 Ω: ± (2% + 5 digits)
Test Cable NullUp to 0.05 Ω
Visual Warning Circuit Protection≥ 30 VAC or VDC at the inputs. Test stops at ≥ 30 VAC or VDC at the inputs.
Repeat Tests per IEC 61557-4Approx. 4000 1-second tests

Insulation Resistance Measurement

ParameterSpecification
Open-Circuit Test Voltage250 V, 500 V, 1000 V, 1500 V (per IEC 61557-2)
Test Voltage Specification-0% +20% (open circuit)
Short-Circuit Test Current< 2 mA
Test Voltage @ 1mA> 1 mA at U × (1000 Ω/V) (per IEC 61557-2)
Display Range0.05 MΩ ... 999 MΩ
Measurement Range (IEC 61557-2)250 V / 500 V String mode: 0.05 MΩ ... 200 MΩ
1 kV / 1.5 kV String mode: 0.05 MΩ ... 999 MΩ
250 V Point-to-Point: 0.05 MΩ ... 300 MΩ
500 V Point-to-Point: 0.05 MΩ ... 500 MΩ
1 kV / 1.5 kV Point-to-Point: 0.05 MΩ ... 999 MΩ
Resistance Resolution0.01 MΩ for 0.05 MΩ ... 1.99 MΩ
0.1 MΩ for 2.0 MΩ ... 19.9 MΩ
1 MΩ for 20 MΩ ... 999 MΩ
Resistance Accuracy±(5% + 1 digit) for 0.05 MΩ ... 0.19 MΩ
±(5% + 3 digits) for 0.20 MΩ ... 1.99 MΩ
±(5% + 2 digits) for 2.0 MΩ ... 5.0 MΩ
±(5% + 5 digits) for > 5.0 MΩ
Insulation Voltage Resolution1 V
Insulation Voltage Accuracy± (2% + 2 digits)
Maximum System Capacitance2 μF
Visual Warning≥ 30 VAC or VDC at the inputs
Repeat Tests per IEC 61557-2Approx. 4000 1-second tests

Operating Current (Wireless via PV:1500 Clamp)

ParameterSpecification
Display Range0.1 ADC ... 400.0 ADC
Measurement Range0.1 ADC ... 400.0 ADC
Resolution0.1 A
Accuracy± (5% + 2 digits)

Operating Power (PV Connections & Wireless via PV:1500 Clamp)

ParameterSpecification
Display Range0.00 kW ... 600 kW
Measurement Range0.50 kW ... 600 kW
Resolution0.01 kW
Accuracy± (6% + 2 digits)

General Technical Data

ParameterSpecification
Mechanical ConstructionWeight: Approx. 1 kg / 2.2 lbs
Dimensions: 265 mm × 115 mm × 78 mm (10.4" × 4.6" × 3.1")
Display Type3.5-inch color display with backlight
Display Resolution480 × 320 pixels
Display Field of View70° left to right, 60° top to bottom
Power Source11.55 V 2930 mAH rechargeable Lithium-ion battery
Auto Power OffUser configurable
Internal Memory1000 memory slots
Environmental Conditions (Operation)Dry, without condensation for indoor and outdoor use
Environmental Conditions (Storage)-25 °C ... +65 °C / -13 °F ... +149 °F; dry, without condensation; battery removed
+5 °C ... +40 °C / +41 °F ... +104 °F
Barometric AltitudeMax. 2000 m / 6562 ft.
Electrical SafetyOvervoltage Category: IEC 61010-01 CAT III / 1500 V
Degree of Pollution: 2 (per IEC 61010-1)
Protection System: Housing: IP40 per IEC 60529 (Protection against solid foreign objects: ≥ 1.0 mm Ø; Protection against water ingress: not protected)
Protection ClassII
Electromagnetic Compatibility (EMC)Emissions: IEC 61326-1, Class A
Immunity: IEC 61326-1

Data Interfaces

Bluetooth® LE

ParameterSpecification
Frequency Band2402 MHz ... 2480 MHz
Transmit Power Range20 dBm
Modulation TypeGFSK
Channel Spacing2 MHz
Spectrum Efficiency (Art. 3.2)STSI EN 300 328 V2.2.2
Antenna Type and GainPCB Antenna, 3.26 dBi
UsageTransfer of test data; connection to peripherals, software updates

RF Long Range

ParameterSpecification
Frequency Band433.375 MHz
Transmit Power Range4.89 dBm ERP
Modulation TypeCSS
Channel Spacing125 kHz
Spectrum Efficiency (Art. 3.2)ETSI EN 300 220-2 V3.2.1
Antenna Type and GainFPC Antenna, 2.8 dBi peak gain
UsageConnection to SS:200LR Wireless Irradiance Meter

Delivery Scope

PV:1525 Complete Set

** Full product. Total delivery scope: see below

SS:200LR Wireless Irradiance Meter (Wireless Irradiance Meter)

PV:1500 Clamp (Wireless Current Clamp)

Ordering Information

Devices

DevicePart Number
PV:1525 Complete Set (Global Version)*601A910
SS:200LR Wireless Irradiance Meter (Global Version)*396A942
PV:1500 Clamp601A1200

*Not for USA.

Optional Accessories & Spare Parts

PDF preview unavailable. Download the PDF instead.

Seaward pv1525-data-sheet de Adobe PDF Library 17.0

Related Documents

Preview Seaward PV:1525 Handheld PV Test Instrument - Specifications and Features
Detailed specifications, features, and scope of delivery for the Seaward PV:1525 handheld PV test instrument, designed for electrical safety and performance testing of solar PV systems.
Preview ERT1557 Operating Instructions - Seaward
Comprehensive operating instructions for the Seaward ERT1557 Earth Resistance and Soil Resistivity Tester, covering setup, usage, memory functions, data download, specifications, maintenance, and safety.
Preview How to Set Up the Black Bluetooth Scanner: Basic Setup and Pairing with Apollo Series Instruments
A guide to troubleshooting connection problems with the 1D black Seaward Bluetooth scanner and Apollo Series portable appliance testers. Learn basic setup and pairing procedures.
Preview Seaward AGL-5 Airfield Ground Lighting Tester: Operating Instructions
Operating instructions and specifications for the Seaward AGL-5 Airfield Ground Lighting Tester, a device for testing airfield ground lighting systems. Includes warranty, maintenance, and calibration information.
Preview How to Perform a Checkbox Sequence on Seaward Apollo 400+ Test Instrument
A step-by-step guide for users of the Seaward Apollo 400+ to perform a checkbox sequence, including firmware updates and connection procedures.
Preview How to Setup Text and Comment Lines on Seaward Apollo and PATGuard 3
A troubleshooting guide for setting up text and comment lines on Seaward Apollo 400+, 500+, 600+ and PATGuard 3 software. Learn how to configure these fields for asset descriptions and notes.
Preview How to Transfer Data from PrimeTest 250+ to Datalogger Software
Step-by-step instructions on how to transfer data from a Seaward PrimeTest 250+ electrical safety tester to the Datalogger software using a USB connection.
Preview Setting Up Large Notes Fields in PATGuard 3
A troubleshooting guide for setting up large notes fields in Seaward's PATGuard 3 software, including firmware and software update instructions.