USER MANUAL
EASY GO VAC ASPIRATOR
Model No. PM65HG
PM65E
SAVE THESE INSTRUCTIONS
CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician.
PRECISION MEDICAL
300 Held Drive
Northampton, PA 18067
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080
www.precisionmedical.com
CONTENTS
- RECEIVING/INSPECTION
- INTENDED USE
- READ ALL INSTRUCTIONS BEFORE USING
- EXPLANATION OF ABBREVIATIONS
- SAFETY INFORMATION - WARNINGS AND CAUTIONS
- SPECIFICATIONS
- COMPONENT DESCRIPTION
- OPERATING INSTRUCTIONS
- CHARGING BATTERY
- CLEANING/MAINTENANCE
- RETURNS
- DISPOSAL INSTRUCTIONS
- TROUBLESHOOTING GUIDE
- REPLACEMENT PARTS
- LIMITED WARRANTY
RECEIVING/INSPECTION
Remove the Precision Medical Easy Go Vac Aspirator from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Equipment Provider.
INTENDED USE
The Easy Go Vac Aspirator provides a portable, AC/DC powered medical vacuum source. It is intended for use in the homecare/healthcare environment.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Easy Go Vac Aspirator. This is provided for your safety and to prevent damage to the Easy Go Vac Aspirator. If you do not understand this manual, DO NOT USE the Easy Go Vac Aspirator and contact your Equipment Provider.
EXPLANATION OF ABBREVIATIONS
- mmHg: Millimeters of Mercury
- inHg: Inches of Mercury
- kPa: Kilopascal
- lpm: Liters per Minute
SAFETY INFORMATION - WARNINGS AND CAUTIONS
DANGER: Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
CAUTION: Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
- ! CONSULT ACCOMPANYING DOCUMENTS.
- ? Type BF equipment.
- ⚪ Suction "OFF" (Battery Charging mode when plugged into charger.)
- ⚫ Suction "ON"
DANGER
- Equipment not suitable for use in the presence of flammable anesthetic.
- Caution must be exercised when operating the Easy Go Vac Aspirator where oxygen is being administered.
- DO NOT use near any type of flame or flammable/explosive substances.
- DO NOT allow Aspirator to exceed Operating or Storage Temperature specifications.
TO REDUCE THE RISK OF ELECTROCUTION:
- Always unplug the Easy Go Vac Aspirator after using, except when battery is being recharged.
- DO NOT use while bathing.
- DO NOT place or store the Easy Go Vac Aspirator where it can fall or be pulled into a tub or sink.
- DO NOT immerse into water or other liquid.
- DO NOT reach for the Easy Go Vac Aspirator that has fallen into water. Unplug immediately.
- DO NOT attempt to repair the Easy Go Vac Aspirator. Any attempt could result in electrical shock.
WARNING
TO REDUCE RISK OF BURNS, ELECTROCUTION, FIRE, OR INJURY TO PERSONS:
- Close supervision is necessary when this Easy Go Vac Aspirator is used by, on, or near children or invalids.
- Use this Easy Go Vac Aspirator only for its "Intended Use" as described in this manual.
- Keep the cord away from heated surfaces.
- DO NOT use when sleepy or drowsy.
- DO NOT use attachments not recommended by Precision Medical, Inc.
- DO NOT use if the Easy Go Vac Aspirator has a damaged cord or plug.
- DO NOT use if the Easy Go Vac Aspirator has been dropped or damaged.
SPECIFICATIONS
- Dimensions (in Carry Bag): Length: 12" (30 cm), Width: 9.5" (24 cm), Height: 10" (25 cm)
- Weight (with Carry Bag and all Accessories): 9.6 lbs (4.4 kg)
- Shipping Weight: 11 lbs (5 kg)
- Suction Canister: 800cc Max Volume. Vacuum is not to exceed 25 inHg (84.7 kPa). The Canister Lid has a float shut off valve that protects the Aspirator from overfill.
- Patient Tubing: 1/4" ID Non-Conductive PVC tubing, 6 ft. long, with 1/4" - 3/8" ID female connectors, or equivalent
- Electrical: Aspirator Unit: 12 Volts DC, 1.8 Amps; Battery Charger/Power Supply: 100 - 240 Volts AC, 50/60 Hz, 0.80-0.35 Amps
- Internal Battery Type: 12 Volt, Sealed Lead Acid
- Vacuum: 2 - 21 inHg (51 - 533 mmHg)
- Flow: 6.5 ± 0.4 Ipm @ 11.8 inHg (300 mmHg)
- Operating Temperature Range: 50°F to 104°F (10°C to 40°C)
- Storage/ Transport Requirements: Temperature Range: -4°F to 104°F (-20°C to 40°C); Maximum Humidity: 95% Non-condensing
- Equipment Classification: Class I - with respect to protection from electrical shock; Type BF - degree of protection against electrical shock; IPX0 - degree of protection against ingress of liquids
- Mode of Operation: continuous
The Easy Go Vac Aspirator meets the EMC requirements of EN 60601-1-2.
Specifications are subject to change without prior notice.
COMPONENT DESCRIPTION
CAUTION If for any reason any label becomes illegible or lost, contact Precision Medical, Inc..
On/Off Switch
Vacuum Gauge
Yellow Low Battery Light
AC Power Cord
Patient Suction Tubing
External Power Jack
Vacuum Control Knob
Inline Hydrophobic Filter
Auto DC Power Cord (optional)
Patient Tubing Connection
Suction Canister (Built in overfill protection device)
Battery Charger/Power Supply
GROUNDING INSTRUCTIONS
The Battery Charger Power Supply should be grounded. In the event of an electrical short circuit, grounding reduces the risk of electrical shock by providing an escape wire for the electrical current. This Battery Charger Power Supply is equipped with a cord having a grounding wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and grounded.
DANGER Improper use of the grounding plug can result in a risk of electrical shock.
WARNING Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked "Hospital Grade".
OPERATING INSTRUCTIONS
CAUTION Inspect the Easy Go Vac Aspirator for visual damage, DO NOT USE if damaged.
- Place the Easy Go Vac Aspirator in the upright position on a level surface.
- Make sure the On/Off (O/O) Switch is in the “OFF” position.
- Ensure the Inline Hydrophobic Filter and tubing is properly connected to the vacuum port on the Suction Canister lid.
CAUTION DO NOT run the Easy GO Vac Aspirator without the Inline Hydrophobic Filter. This will void the warranty.
- Select appropriate Mode of Operation.
A. AC ELECTRICAL RECEPTICLE OPERATION (100-240 Volts AC)
- Insert the small plug on the cord from the Battery Charger/Power Supply into External Power Jack on the Easy Go Vac Aspirator.
- Connect the female end of the AC Power Cord into the Battery Charger/Power Supply.
- Plug the male end (3 prong plug) into a grounded receptacle.
- Push the On/Off (0/0) Switch to the “ON” position.
- Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
- Vacuum should be set as prescribed by your Physician or Healthcare Professional.
- Connect one end of the Patient Suction Tubing to the port on the canister marked Patient Tubing.
- Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by Healthcare Provider.
- When procedure is completed, press the On/Off (0/0) Switch to the “OFF” position.
- DO NOT overfill the Canister.
- Empty the Suction Canister as follows:
- Remove the canister from its bracket and empty as instructed by Physician or Healthcare Professional.
CAUTION Canister is single patient use only.
- Replace Suction Canister and reconnect all suction tubing.
B. INTERNAL BATTERY OPERATION
- Steps 1 thru 4, see Operating Instructions on page 6.
CAUTION
- Battery should be fully charged before using; to charge battery see Charging Battery.
- When the yellow Low Battery Light is on continuously during normal use, the battery needs to be charged.
- If automatic shut off occurs during battery operation, the Easy Go Vac Aspirator cannot be restarted until the battery has been recharged, or until the Aspirator is connected to either of the external power source options.
- Push the On/Off (0/0) Switch to the "ON" position. If yellow Low Battery Light stays on continuously, battery needs charging. Charge battery or use another Mode of Operation.
WARNING Operating Easy Go Vac Aspirator after yellow Low Battery Light has been on continuously, will cause automatic shut down.
- Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
- Vacuum should be set as prescribed by your Physician or Healthcare Professional.
- Connect one end of the Patient Suction Tubing to the port on the canister marked Patient Tubing.
- Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by Healthcare Provider.
- When procedure is completed, press the On/Off (0/0) Switch to the “OFF” position.
- DO NOT overfill the Canister.
- Empty the Suction Canister as follows:
- Remove the canister from its bracket and empty as instructed by Physician or Healthcare Professional.
CAUTION Canister is single patient use only.
- Replace Suction Canister and reconnect all suction tubing.
C. AUTO DC OPERATION (Auto Power Port)
- Steps 1 thru 4, see Operating Instructions on page 6.
- Insert small plug on the Auto DC Power Cord into the External Power Jack on the Aspirator.
- Plug the other end into the auto power port receptacle.
CAUTION DO NOT leave the Aspirator connected to the auto power port when not in use. Failure to follow these instructions can cause severe battery failure and will VOID THE WARRANTY, see CHARGING BATTERY instructions.
- Push the On/Off (0/0) Switch to the “ON” position. If yellow Low Battery Light stays on continuously, battery needs charging. Charge battery or use another mode of operation.
WARNING Operating Easy Go Vac Aspirator after yellow Low Battery Light has been on continuously, will cause automatic shut down.
- Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
- Vacuum should be set as prescribed by your Physician or Healthcare Professional.
- Connect one end of the Patient Suction Tubing to the port on the canister marked Patient Tubing.
- Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by Healthcare Provider.
- When procedure is completed, press the On/Off (0/0) Switch to the “OFF” position.
- DO NOT overfill the Canister.
- Empty the Suction Canister as follows:
- Remove the canister from its bracket and empty as instructed by Physician or Healthcare Professional.
CAUTION Canister is single patient use only.
- Replace Suction Canister and reconnect all suction tubing.
CHARGING BATTERY
- Insert small plug on cord from the Battery Charger/Power Supply into External Power Jack on the Easy Go Vac Aspirator.
- Connect the female end of AC Power Cord into the Battery Charger/Power Supply.
- Plug male end (3 prong plug) into a grounded receptacle. Amber light on Battery Charger/Power Supply will come on. When amber light changes to green, battery is fully charged. Average charging time is 2-4 hours. (A new battery may take longer to fully charge the first time.)
Note: No damage will occur if the charger is left connected. - Disconnect AC Power Cord from AC receptacle.
- Disconnect cord from External Power Jack on Aspirator. When fully charged, battery will provide approximately 120 minutes of operation in 8, 15 minute intermittent duration periods.
CLEANING/MAINTENANCE
WARNING
- Turn the Easy Go Vac Aspirator off and disconnect from any external power source before cleaning.
- DO NOT clean the Easy Go Vac Aspirator with Alcohol, Acetone or other solvents.
Inline Hydrophobic Filter:
This filter cannot be cleaned. If it becomes contaminated or clogged, it must be replaced.
To replace Inline Hydrophobic Filter:
- Disconnect filter assembly from Suction Canister and remove Suction Canister from the Aspirator.
- Unscrew filter assembly from the Aspirator.
- Dispose of filter.
- Install new filter (Part # 502438) by screwing filter into the Aspirator, until hand tight. Then reconnect tubing and elbow.
CAUTION DO NOT use sealing material on the filter threads.
Tubing is single patient use only. DO NOT clean.
Easy Go Vac Case:
Use a cloth or sponge dampened with soap and water. Wipe dry with a clean cloth or paper towel.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
DISPOSAL INSTRUCTIONS
Dispose of Bio-waste products as described by your Healthcare Professional or Healthcare Facility protocol.
WARNING
- The Easy Go Vac Aspirator contains a sealed lead acid battery. DO NOT dispose with normal trash. Consult your local waste management authority for proper disposal procedures.
- If Aspirator is used for treatment of infectious disease, consult your Physician or Healthcare Professional for recommended procedures for proper disposal.
Please Recycle
TROUBLESHOOTING
If the Easy Go Vac Aspirator fails to function, consult the Troubleshooting Guide. There are no user serviceable parts inside. DO NOT open case. If the problem cannot be corrected, consult your Equipment Provider.
TROUBLESHOOTING GUIDE
Problem | Probable Cause | Remedy |
---|---|---|
Easy Go Vac Aspirator will not operate on battery | 1. On/Off (0/0) Switch in "OFF" position 2. Battery not charged (Low battery light on) |
1. On/Off (0/0) Switch to "ON" position 2. Recharge battery 3. If battery will not charge, contact your Equipment Provider |
Easy Go Vac Aspirator will not operate when connected to AC Power Source | 1. On/Off (0/0) Switch in "OFF" position 2. AC Power Cord not connected |
1. On/Off (0/0) Switch to "ON" position 2. Check connection at both AC Power Source and Aspirator |
Easy Go Vac Aspirator will not operate when connected to Auto DC Power Cord | 1. On/Off (0/0) Switch in "OFF" position 2. Auto DC Power Cord not connected properly 3. Defective auto power port receptacle 4. Defective Auto DC Power Cord |
1. On/Off (0/0) Switch to "ON" position 2. Check connections at both ends of cord 3. Connect to a known working auto power port 4. Contact your Equipment Provider |
Continued on the next page.
REPLACEMENT PARTS
Description | Part # |
---|---|
Filter | 502438 |
Filter Assembly | 503293 |
800cc Canister, (Case-10 ea.) | 502519-10 |
Patient Suction Tubing | 1955 |
Charger | 503974 |
Carry Bag | 502956 |
AC Power Cord (65) | 503106 |
Auto DC Power Cord | 502954 |
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that EASY GO VAC Aspirator (the Product) and the following component parts thereof will be free of defects in workmanship and/or material for the following periods:
- Easy Go Vac Aspirator: Two (2) years from shipment
- Battery: Six (6) months from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.'s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
Related Documents
![]() |
Precision Medical EasyDialReg Oxygen Regulator User Manual and Specifications User manual and technical specifications for the Precision Medical EasyDialReg Oxygen Regulator, including operating instructions, safety warnings, maintenance, troubleshooting, and warranty information. |
![]() |
Precision Medical Fluximeter User Manual User manual for Precision Medical fluximeters, series 1MFA, 4MFA, 6MFA, and 8MFA. Provides instructions for use, safety precautions, specifications, troubleshooting, and warranty information. |
![]() |
Precision Medical PM63 PowerVac+ Aspirator Parts Identification Detailed parts list and identification for the Precision Medical PM63 PowerVac+ Intermittent Aspirator, including front, rear, side, and internal views. |
![]() |
Precision Medical Air-Oxygen Blender User Manual & Specifications User manual for Precision Medical Air-Oxygen Blenders (PM5200, PM5300). Covers operation, safety, specifications, maintenance, and troubleshooting for medical professionals. |
![]() |
Precision Medical EasyPulse 5 Oxygen Conserving Regulator User Manual User manual for the Precision Medical EasyPulse 5 Oxygen Conserving Regulator (1900 Series, CGA 870 Connection). Provides instructions on installation, operation, maintenance, safety, specifications, and troubleshooting. |
![]() |
Precision Medical Flowmeter Replacement Parts List Comprehensive list of replacement parts for Precision Medical air and oxygen flowmeters, including model numbers for various flow rates and pressures. |
![]() |
Precision Medical EasyAir Compressor PM15 Series User Manual This user manual provides instructions for the Precision Medical EasyAir Compressor, model PM15 Series. It covers receiving, inspection, intended use, safety information, specifications, component identification, grounding, operating instructions, cleaning, maintenance, troubleshooting, and limited warranty. |
![]() |
Precision Medical PM4150 EasyPulse POC Service Manual Comprehensive service manual for the Precision Medical PM4150 EasyPulse POC portable oxygen concentrator. Covers theory of operation, specifications, component descriptions, maintenance, troubleshooting, and detailed replacement procedures for various parts. |