MINI DIGITAL CLAMP-ON MULTIMETER

Operating Instruction

OPERATION MANUAL

1. SAFETY INFORMATION

This Mini-sized digital clamp multimeter has been designed according to IEC61010 oncoming electronic measuring instruments with an over voltage category (CAT II 600V) and Pollution degree 2.

⚠️ WARNING
To avoid possible electric shock or personal injury, follow these Guidelines:

2. INTRODUCTION

This manual provides all safety information, operation instruction, specifications and maintenance for the meter, which is compact, handheld, and battery operated.

This instrument performs AC/DC voltage, DC/AC Current, Resistance, Audible Continuity, Diode, Temperature measurements and Non-Contact ACV Detect (NCV) as well as Capacitance, Hz etc.

It is an auto ranging DMM with AC/DC current clamp meter function.

3. CAUTION ⚠️

To avoid damage to the meter, don't apply input which exceeds the limit shown below:

Function Terminals Input Limits
DCV/ACV VΩ & COM 600V DC or 600V rms AC
~— VΩ & COM 250V DC or rms AC
°C, Hz VΩ & COM 250V DC or rms AC

4. GENERAL CHARACTERISTICS

5. PANEL DESCRIPTION

The multimeter features a clamp jaw, trigger, LCD display, and various input terminals and buttons. Key components are:

6. SPECIFICATIONS

Accuracy is guaranteed for 1 year 23°C±5°C less than 75%RH.

6-1. AC VOLTAGE (Auto ranging)

Range Resolution Accuracy
4V 1mV ± (1.5% + 3d)
40V 10mV
400V 100mV
600V 1V

Input Impedance: more than 10MΩ; Frequency Range: 40 to 400Hz.

6-2. DC VOLTAGE (Auto ranging)

Range Resolution Accuracy
400mV 0.1mV ± (1.5% + 5d)
4V 1mV
40V 10mV ± (1.0% + 3d)
400V 100mV
600V 1V

Input Impedance: more than 10MΩ

6-3. RESISTANCE (Auto Ranging)

Range Resolution Accuracy
400Ω 0.1Ω ± (1.2% + 3d)
4KΩ
40KΩ 10Ω
400KΩ 100Ω
4MΩ 1KΩ ± (1.5% +5d)
40MΩ 10KΩ

6-4. AC CURRENT (Auto ranging)

Range Resolution Accuracy
40A 10mA ± (2.5% + 3d)
400A 100mA

Measuring voltage drop: 200mV; Frequency Range: 40 to 400Hz.

6-5. FREQUENCY (Auto ranging)

Range Resolution Accuracy
10Hz~400kHz 1Hz ± (1.0% +5d)
400kHz~4MHz 1kHz

Input Voltage Limit: Max. 250V (<4M Hz).

6-6. CAPACITANCE (Auto ranging)

Range Resolution Accuracy
1nF~4 µF 0.01n~0.01 µF ± (2.5% + 10d)
4 µ F~400uF 0.1uF ± (4.0% + 20d)
4mF 1uF ± (5.0% + 30d)

6-7. TEMPERATURE (only 3269)

Range Resolution Accuracy
0~750°C 1°C 0~400°C 400~750°C
32~1382°F 1°F 32~752°F 752~1382°F
± (2.0% + 1°C) ± (2.5% + 1°C)

6-8. Non-Contact AC VOLTAGE Detection (NCV ⚡)

One Special Front LED Lighting for indication of over AC100V source without directly contact.

7. CONTROL BUTTON DESCRIPTION

7-1. HOLD AND LCD BACKLIT BUTTON (? H/↔️)

  1. When you press this button briefly, LCD will show the last reading, and "H" symbol will appear till pushed again. Data holding will be cancelled automatically when the function switch is rotated.
  2. When you hold this button down for about 2 seconds. LCD will be backlit and lighting remains about 15 second before automatically turned off.

7-2. Manual range selection (REL ↔️)

7-3. Function select button (SET ▶️)

8. OPERATION INSTRUCTION.

8-1. AC/DC VOLATAGE MEASUREMENT

  1. Connect the BLACK test lead to the COM (?) jack and the RED to the V/Ω jack.
  2. Set the function switch at ACV or DCV position, connect test leads across the source or load under measurement.
  3. Read LCD display. The polarity of RED test lead will be indicated when making a DC measurement.

8-2. RESISTANCE MEASUREMENT

  1. Connect the BLACK test lead to the COM (?) jack and the RED to the V/Ω jack. (NOTE: The polarity of the RED lead is positive "+" and the BLACK is negative "-").
  2. Set the function switch at (OFF/ACV/DCV/Ω/Diode/Continuity/Capacitance/Frequency/Temperature/AC Current/NCV) position, and then Push "SEL" (▶️) button to select Ω Mode, then the symbol "Ω" is shown on LCD.
  3. Connect test leads across the resistance under measurement, and then get the test readings on LCD.

NOTE:

8-3. AC CURRENT MEASUREMENT

  1. Set Function/Range Switch to the AC (~) range(s).
  2. Press the trigger to open the transformer jaws. And clamp one conductor only. It is impossible to make measurements when two or three conductors are clamped at the same time.
  3. Display reading is showing the conductor AC current.

8-4. AUDIBLE CONTINUITY/DIODE TEST

  1. Connect the BLACK test lead to the COM (?) jack and the RED to the V/Ω jack.
  2. Set the function switch at (OFF/ACV/DCV/Ω/Diode/Continuity/Capacitance/Frequency/Temperature/AC Current/NCV) position and push the button "SEL" (▶️) to select continuity (?) or diode (⚡) test mode. Then the symbol "?" or "⚡" is shown on LCD.
  3. In continuity test, if the circuit resistance under test is lower than 50Ω, built-in buzzer will sound.
  4. If diode test mode is selected, connect the RED and BLACK test leads to anode and cathodes of the diode under test separately. The forward voltage drop of diode will be displayed.

8-5. FREQUENCY TEST

  1. Connect the BLACK test lead to the COM (?) jack and the RED test lead to the V/Ω Hz jack.
  2. Put the range selector at "Hz" position, and connect the test leads across the resistance under measurement.
  3. Take the reading from LCD.

Note: When test the high Frequency > 1MHz, the input Voltage must be less than 10V for getting the accurate value.

8-6. CAPACITANCE MEASUREMENT

  1. Connect the BLACK test lead to the COM (?) jack and the RED test lead to the V/Ω jack. (Note: The polarity of the RED test lead is "+")
  2. Put the range selector at (OFF/ACV/DCV/Ω/Diode/Continuity/Capacitance/Frequency/Temperature/AC Current/NCV) position, and connect the test leads across the Capacitor under measurement. Disconnect the capacitor with the outer power source.
  3. Check if the "F" mark on the LCD. If not, push the "SET" (▶️) button one or more times until shown.
  4. Take the reading from LCD. When test the large capacitor, it's normal to take more time for getting the values on LCD.

NOTE: Make sure of discharging all the capacitor's electricity completely before measurements.

8-7. TEMPERATURE MEASUREMENT (only 3269)

  1. Set the function range switch at the °C position.
  2. Make sure the polarity of the thermocouple correct, put the Black pin terminal of the K-type thermocouple sensor into the COM (?) jack and the RED pin terminal into the V/Ω/°C jack.
  3. Set the working end (testing end) on or inside the object under test.
  4. The value of the temperature is shown on the display in degree centigrade (°C).

NOTE:

8-8. Non-Contact AC VOLTAGE Detection (NCV ⚡)

  1. Set the function range switch at the NCV position. Then the NCV indicating LED light is flashing.
  2. Take the clamp jaw close to the AC signal.
  3. Then the NCV indicating LED is constantly flashing and the Beeper sounding at the same time, once ACV > 100V.

NOTE: Don't try to contact the testing place with this tester. No display on LCD at this range.

9. BATTERY REPLACEMENT

If the sign "?" appears on the display, it indicates battery should be replaced. Remove screws and open the back case, replace the exhausted batteries with new ones (AAA 1.5V *2pc or equivalent).

PDF preview unavailable. Download the PDF instead.

Sb76a36a6969d48c79bfadba846d1ff90H

Related Documents

Preview Mini Digital Clamp-On Multimeter Operating Instructions
This document provides operating instructions, safety information, and specifications for the Mini Digital Clamp-On Multimeter. It covers measurements for AC/DC voltage, DC/AC current, resistance, audible continuity, diode, temperature, and Non-Contact ACV detection.
Preview ANENG ST180 Digital Clamp Meter User Manual
User manual providing detailed information, specifications, and operating instructions for the ANENG ST180 Digital Clamp Meter.
Preview ANENG ST180 Mini Digital Clamp-On Multimeter Operating Manual
Operating manual for the ANENG ST180 Mini Digital Clamp-On Multimeter, covering safety information, introduction, specifications, panel description, control buttons, operation instructions for various measurements (AC/DC voltage, resistance, current, frequency, capacitance, temperature, NCV), and battery replacement.
Preview ANENG ST212 Digital Clamp Meter User Manual
Comprehensive user manual for the ANENG ST212 Digital Clamp Meter, covering safety information, specifications, operation guides, and maintenance for AC/DC current, voltage, resistance, capacitance, and temperature measurements.
Preview ANENG Mini Digital Clamp-On Multimeter Operating Instruction Manual
Comprehensive operating instructions and technical specifications for the ANENG Mini Digital Clamp-On Multimeter. This manual covers safety precautions, device overview, panel description, detailed measurement specifications for AC/DC voltage, current, resistance, frequency, capacitance, and temperature, as well as Non-Contact AC Voltage Detection (NCV) and battery replacement procedures.
Preview ANENG ST180 4000 Counts Digital AC Current Clamp Multimeter Specifications and Features
Detailed specifications and features of the ANENG ST180, a 4000 counts digital AC current clamp multimeter. Includes measurements for AC/DC voltage, AC current, resistance, capacitance, frequency, and temperature.
Preview ANENG 683 Touch Meter User Manual: Features, Specs & Operation Guide
Comprehensive user manual for the ANENG 683 digital multimeter. Covers safety, specifications, measurement functions (voltage, current, resistance, etc.), and operational guidance for electrical testing.
Preview ANENG ST181 Mini Digital Clamp-On Multimeter Operation Manual
Operation manual for the ANENG ST181 Mini Digital Clamp-On Multimeter, covering safety, panel description, specifications, and detailed instructions for various electrical measurements including voltage, current, resistance, capacitance, frequency, and temperature.