HEINE Dermatoscopes

This manual provides instructions for HEINE Dermatoscopes: NC1, DELTA 20 Plus, DELTA 20T, mini3000 LED, and mini3000.

Intended Use

HEINE dermatoscopes are internally powered medical examination lights. They function as reflected-light microscopes for the non-invasive, visual inspection of intact skin by healthcare professionals. The diagnostic use is not limited to any specific patient demographic.

Warnings and Safety Information

Caution! ⚠️ Indicates potential hazardous situations. Ignoring instructions may lead to mild to moderate injuries.

Note! ℹ️ Provides valuable advice for installation, operation, maintenance, or repair, but is not related to hazardous situations.

Product Overview

HEINE DELTA 20 Plus and HEINE DELTA 20T Dermatoscope

Key Components:

HEINE NC1 Dermatoscope

HEINE mini3000 LED/XHL Dermatoscope

Setting Up

To set up the instrument, screw or plug the instrument head into the HEINE battery handle or the HEINE rechargeable handle.

Operation

HEINE DELTA 20 Plus and HEINE DELTA 20T Dermatoscope

Assembly of filter insert and contact plate: Attach the filter insert (9) to the contact plate (1) by rotating it until the bayonet connection locks. Reverse the process to remove the filter.

Attachment of contact plate: Contact plates (1 and 8) attach via a bayonet connection. Turn counterclockwise to remove and reverse to attach. Always ensure the bayonet is securely locked.

HEINE DELTA 20 Plus Dermatoscope:

HEINE DELTA 20T Dermatoscope:

HEINE NC1 Dermatoscope:

HEINE mini3000 LED/XHL Dermatoscope:

Removing the Contact Plate

The contact plate (12) is attached by a bayonet fitting. Rotate the knurled ring counterclockwise and detach. The small contact plate (13) can be used for inaccessible lesions; it inserts into the head. Ensure the light port (16) faces the bulb/LED when replacing.

Hygienic Reprocessing

Follow national standards, laws, and guidelines for hygienic reprocessing. Classification: non-critical. Allow the device to cool before reprocessing. Reprocess immediately if contamination is suspected. HEINE approves only specified agents and procedures. Cleaning and disinfection must be performed by qualified personnel. Observe manufacturer instructions. Do not use spray/immersion disinfection, dripping wet/foaming cloths, ultrasonic, or alcohol-based reprocessing media.

Contact plates must be cleaned/disinfected after each use. Sterilization is recommended only after treating high-risk patients (mini3000: max 4 cycles; DELTA 20 Plus: max 25 cycles). Instrument heads, filter inserts, small contact plates, NC1 contact plates, and DELTA 20T contact plates are not sterilizable.

Procedure:

Changing the Light Source

Allow the device to cool before changing the bulb. The LED cannot be changed for DELTA 20 Plus, DELTA 20T, NC1, and mini3000 LED models. For the mini3000, remove the instrument head from the handle, pull out the bulb, wipe the new bulb with a soft cloth, and insert it fully.

Maintenance and Service

The instruments require no maintenance or service.

General Warnings

General Notes

Warranty is invalidated by the use of non-genuine HEINE products, non-original parts, or unauthorized repairs/modifications. Visit www.heine.com for more information. Remove batteries if the device will not be used for an extended period.

Disposal

The product must be recycled as separate electrical and electronic equipment. Observe relevant country-specific disposal regulations.

Electromagnetic Compatibility

Medical electrical devices are subject to special EMC precautions. Portable and mobile RF communication equipment can affect these devices. This device may cause radio interference; take remedial measures such as reorientation, rearrangement, or shielding. Using non-specified accessories, converters, or cables may increase emissions or reduce immunity. Do not stack or use the device directly next to or with other devices; observe for proper operation in such configurations.

Technical Specifications

Environmental Conditions for Operation: +10°C to +35°C, 30% to 75% relative humidity, 700 hPa to 1060 hPa.

Environmental Conditions for Storage: +5°C to +45°C, 45% to 80% relative humidity, 500 hPa to 1060 hPa.

Environmental Conditions for Transport: -20°C to +50°C, 45% to 80% relative humidity, 500 hPa to 1060 hPa.

Nominal Voltage: 3.0 V - 3.7 V

Nominal Current: 440 - 760 mA

Protection Class: Internal power supply

IP Code: IP20

Device Classification (IEC 62471): Group 2

Applied Part: Type BF (for contact plate)

HEINE mini3000 Dermatoscope: #109 (2.5 V)

Explanation of Symbols

CE Mark: Indicates conformity with the European Medical Device Directive 93/42/EEC.

REF: Catalogue or order number.

Follow Instructions: Symbol indicating mandatory adherence to user instructions.

Fragile: Symbol indicating the product is fragile and requires careful handling.

Keep Dry: Symbol indicating the product should be stored in a dry environment.

Recycling Symbol (Grüner Punkt): Indicates country-specific recyclability.

UDI: Unique Device Identification.

Models: DELTA 20 Plus, DELTA 20T, NC1, mini3000 LED, mini3000, Dermatoscopes

File Info : application/pdf, 28 Pages, 677.76KB

PDF preview unavailable. Download the PDF instead.

HEINE Bedienungsanleitug Dermatoskope

References

Adobe InDesign 15.0 (Macintosh) iLovePDF

Related Documents

Preview Heine Mini3000 Dermatoscope Set: Advanced LED Illumination and Diagnostics
Explore the Heine Mini3000 Dermatoscope Set, featuring advanced LED illumination for accurate diagnosis, a 10x magnification achromatic lens system, and durable, maintenance-free design. Includes kit contents and performance specifications from Advanced Lifecare.
Preview HEINE mini 3000 Series Instrument Instructions
Comprehensive instructions for the HEINE mini 3000 series of medical instruments, including battery handles, ophthalmoscopes, focalux hand lamps, tongue-blade holders, otoscopes, and cliplamps. Covers usage, battery replacement, cleaning, and sterilization.
Preview HEINE Classic+ and SANALON+ Fiber Optics Laryngoscope Blades Instructions
Comprehensive instructions for the safe and effective use, cleaning, disinfection, and sterilization of HEINE Classic+ and SANALON+ fiber optics laryngoscope blades. Includes safety information, warranty details, and applicable standards.