AEMC MR6292 AC/DC Current Probe
User Manual
Brand: AEMC Instruments
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.
The instrument is guaranteed to meet its published specifications at the time of shipping.
An NIST traceable certificate may be requested at the time of purchase or obtained for a nominal charge by returning the instrument to the repair and calibration facility.
The recommended calibration interval is 12 months from the date of receipt. For recalibration, please use AEMC® Instruments' calibration services. Refer to the repair and calibration section at www.aemc.com.
Serial #: ______________________________ Catalog #: 2129.86 Model #: MR6292
Please fill in the appropriate date as indicated:
Date Received: _________________________________ Date Calibration Due: _______________________
Chauvin Arnoux®, Inc. d.b.a AEMC® Instruments
www.aemc.com
Introduction
Safety Warnings
These safety warnings are provided to ensure the safety of personnel and proper operation of the instrument.
- Read this instruction manual completely and follow all safety information before attempting to use or service this instrument.
- Use caution on any circuit: Potentially high voltages and currents may be present and may pose a shock hazard.
- Read the safety specifications section before using the Model MR6292. Never exceed the maximum working voltage ratings given.
- Safety is the responsibility of the operator!
- Never open the back of the instrument while connected to any circuit or input.
- ALWAYS connect the Model MR6292 to the Micro-Ohmmeter Model 6292 instrument before clamping the probe onto the sample under test.
- Always inspect the probe and lead prior to use. Replace any defective parts immediately.
- NEVER use the Model MR6292 on electrical conductors rated above 600V.
International Electrical Symbols
[Double Insulation Symbol]: This symbol signifies that the instrument is protected by double or reinforced insulation. Use only specified replacement parts when servicing the instrument.
[Warning Symbol]: This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.
[Electric Shock Symbol]: Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.
[Type A Current Sensor Symbol]: This is a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is permitted.
Measurement Categories
- CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary overcurrent protection devices, ripple control units, or meters.
- CAT III: For measurements performed in the building installation at the distribution level such as on hardwired equipment in fixed installation and circuit breakers.
- CAT II: For measurements performed on circuits directly connected to the electrical distribution system. Examples are measurements on household appliances or portable tools.
Receiving Your Shipment
Upon receiving your shipment, ensure the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, providing a detailed description of any damage. Save the damaged packing container to substantiate your claim.
Ordering Information
AC/DC Current Probe Model MR6292 (for use with Micro-Ohmmeter Model 6292 only) - Cat. #2129.86
Includes user manual.
Order Accessories and Replacement Parts Directly Online. Check our Storefront at www.aemc.com/store for availability.
Product Features
Description
The AC/DC Current Probe Model MR6292 is an accessory for the AEMC® Micro-Ohmmeter Model 6292. The Model MR6292 probe is required for using the Model 6292 to take measurements in BSG (both sides grounded) mode.
The probe is designed to meet the latest safety and performance standards. The probe's hook-shaped jaws enable you to “pry” into or “hook” onto cables (it will accept 2 x 500 MCM) or small bus bars.
The Model MR6292 uses Hall effect technology. All electronics are self-contained in the handle. The output of the probe is 10mV/A. No battery is required; the probe operates via voltage provided by the Model 6292. The probe has a proportional mV output for direct reading on the Model 6292.
MR6292 Features
Figure 1:
- Jaws: Accommodates cables up to 1.18" (30mm) diameter or dual cables up to 0.95" (24mm) diameter.
- Conductor: The conductor being measured.
- Protective non-slip guard: For user safety and grip.
- Green light: Indicates power is ON when connected to the Model 6292.
- Lead: 5 ft (1.5m) cable.
- Connector: 5-pin XLR connector.
Specifications
Electrical
- Measurement Range: 1 to 50ADC
- Accuracy: 1 to 50ADC: 1.5% reading ± 0.2A
- Output Signal: 10mV/ADC
- Load Impedance: >100kΩ/100pF
- Insertion Impedance: 0.39mΩ @ 50Hz, 58mΩ @1000Hz
- Working Voltage: 600Vrms
- Common Mode Voltage: 600Vrms
- Influence of Adjacent Conductor: < 10mA/A at 50Hz at 23mm from the probe
- Influence of Conductor in Jaw Opening: 0.5% reading
- Power Source: Powered by Micro-Ohmmeter Model 6292
Mechanical
- Jaw Opening: 1.2" (31mm)
- Maximum Cable Diameter: One 1.18" (30mm) or two 0.95" (24mm) or two bus bars 1.2 x 0.4" (31.5 x 10mm)
- Dimensions: 8.8 x 3.82 x 1.73" (224 x 97 x 44mm)
- Weight: 15 oz (440g)
- Output: 4-conductor shielded cable with 5-pin XLR connector
- Zero Adjustment: Automatic zero by Model 6292
- Case Protection: IP30 per IEC 529
- Drop Test: 1.0m on 38mm of oak on concrete; test according to EN 61010
- Mechanical Shock: 100G, test per IEC 68-2-27
- Vibration: Test per IEC 68-2-6
- Handle: UL 94 V0
- Jaws: UL 94 V0
Environmental
- Operating Temperature Range: 14° to 131°F (-10° to +55°C)
- Storage Temperature Range: -40° to 176°F (-40° to +80°C)
- Temperature Influence: ≤300° ppm/°K or 1%/10°K; ≤0.2A/°K on Zero
- Operating Relative Humidity: 10° to 35°C: 90 ± 5% RH (without condensation); 40° to 55°C: 70 ± 5% RH (without condensation)
- Humidity Influence: 10 to 90% RH @ reference temperature ≤0.1% R
- Altitude: Operating: 0 to 2000m; Non-operating: 0 to 12,000m
Safety
- Electrical: EN 61010-2-32, 600V CAT III, Pollution: 2
- Electromagnetic Compatibility: EN 50081-1 Class B; EN 50082-2 Electrostatic discharge IEC 1000-4-2, Radiated field IEC 1000-4-3, Fast transients IEC 1000-4-4, Magnetic field at 50/60 Hz IEC 1000-4-8
Operation
The Model MR6292 probe is used in conjunction with the Micro-Ohmmeter Model 6292 to perform measurements in BSG (both sides grounded) mode. For additional information about using the Micro-Ohmmeter Model 6292, refer to the user manual on the USB Stick that ships with the instrument.
Making Measurements
- Plug the probe into the Model 6292.
- Turn the power to the Model 6292 ON.
- The green LED on the Model MR6292 should be lit. If the LED does not come ON or goes OFF after a few minutes, check to ensure that the Model 6292 instrument is still powered ON, and that the Model MR6292 is securely plugged into the instrument's socket.
- Set up a test on the Micro-ohmmeter Model 6292, as instructed by the instrument's user manual (located on the USB stick).
- Clamp the probe on the ground conductor to be tested, then start the test. The Model 6292 will display the measured ground conductor current as the "Ig" reading on the instrument's LCD display panel.
Example display output:
R=100.0uΩ It=142A DURATION: 015/060s 10/20/13 03:25 PM Ig=8.0A
In the preceding example, the Ig measurement provided by the MR6292 probe is 8.0A.
Operation Examples
Conductor carrying 25ADC in the direction of the arrow. Model 6292 displays Ig = 25A.
Maintenance
WARNING:
- For maintenance use only specified replacement parts.
- To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
- To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into the case. Turn the current probe OFF and disconnect the unit from all the circuits before opening the case.
Cleaning
- Clean the body of the clamp with a cloth lightly moistened with soapy water.
- Wipe clean with a cloth moistened with clean water and dry.
- Do not use solvent.
Repair and Calibration
To ensure that the Model MR6292 meets factory specifications, it is recommended to schedule it back to the factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
Contact the Service Center for a Customer Service Authorization Number (CSA#). This ensures prompt tracking and processing of the instrument upon arrival. Write the CSA# on the outside of the shipping container. If returning for calibration, specify if a standard calibration or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data) is required.
Ship To:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive
Dover, NH 03820 USA
Phone: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
E-mail: repair@aemc.com
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
For technical problems or assistance with operation or application, please call, fax, or e-mail the technical support team:
Contact: Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
Phone: (800) 945-2362 (Ext. 351)
(603) 749-6434 (Ext. 351)
Fax: (603) 742-2346
E-mail: techsupport@aemc.com
Limited Warranty
The Model MR6292 probe is warranted to the owner for a period of two years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC® Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused, or if the defect is related to service not performed by AEMC® Instruments.
Full warranty coverage and product registration is available on the website at www.aemc.com/warranty.html.
What AEMC® Instruments will do: If a malfunction occurs within the warranty period, return the instrument for repair, provided warranty registration information or proof of purchase is available. AEMC® Instruments will, at its option, repair or replace the faulty material.
REGISTER ONLINE AT: www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair: First, request a Customer Service Authorization Number (CSA#) by phone, fax, or email from the Service Department (address below). Then, return the instrument along with the signed CSA Form. Write the CSA# on the outside of the shipping container. Return the instrument, postage or shipment pre-paid to:
Ship To:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
E-mail: repair@aemc.com
Caution: To protect yourself against in-transit loss, insure your returned material.
NOTE: You must obtain a CSA# before returning any instrument.