LumiraDx Instrument Cleaning and Disinfection Procedure

This Technical Bulletin provides guidance on the cleaning and disinfection of the LumiraDx Instrument, adhering to established site protocols and schedules, as well as local guidelines. It covers general cleaning and disinfection, specific guidance for different sample types (blood-based, swab-based, or both), and lists recommended disinfectant materials.

The difference between cleaning and disinfecting

Cleaning is the physical removal of dirt or other foreign material from the Instrument surface.

Disinfecting is the chemical removal of harmful microorganisms (pathogens) from the Instrument.

General guidance on cleaning the Instrument

General guidance on disinfecting the Instrument

Disinfection guidance – if only testing blood-based samples

Procedure:

  1. Using a LumiraDx recommended disinfecting material, wipe the external surfaces of the Instrument, taking care to avoid the door hinges, Test Strip inlet, power cord, and USB port.
  2. Allow the disinfectant at least 5 minutes contact time with the Instrument before testing the next sample.
  3. Dispose of disinfectant materials in accordance with local biohazardous waste disposal procedures.

Disinfection guidance – if only testing swab-based samples

Procedure:

  1. Using a LumiraDx recommended disinfecting material, wipe the external surfaces of the Instrument, taking care to avoid the door hinges, Test Strip inlet, power cord, and USB port.
  2. Allow the disinfectant at least 1 minute contact time with the Instrument before testing the next sample.
  3. Dispose of disinfectant materials in accordance with local biohazardous waste disposal procedures.

Additional guidance – if testing both blood-based & swab-based samples

If you are testing both blood-based and swab-based samples with your Platform, follow the guidance for disinfection based on the most recent sample-type run.

Recommended disinfectant materials for blood-based samples

The following formulations are compatible for disinfection purposes with the LumiraDx Instrument when testing blood-based samples:

LumiraDx does not recommend the use of products which combine two or more of the above disinfectant formulations.

Please note that due to the potential presence of bloodborne pathogens, alcohol-based wipes alone are not suitable for disinfection.

Recommended disinfectant materials for swab-based samples

The following formulations are compatible for disinfection purposes with the LumiraDx Instrument when testing swab-based samples:

LumiraDx does not recommend the use of products which combine two or more of the above disinfectant formulations.

As an alternative, the use of alcohol-based virucidal/bactericidal wipes is acceptable (ethanol or isopropanol based, ≥70 % v/v) as an effective disinfection procedure between swab-based tests (including COVID-19). For blood-based samples, the effective destruction of any contaminating viruses such as Hepatitis B cannot be guaranteed.

Other notes

Be careful to minimize the risks of cross-contamination when testing patient samples, which can cause false positive results. Insufficient cleaning of the workspace, insufficient disinfection of the Instrument, or inappropriate use of protective equipment (for example, failing to change gloves between patients) can increase the risk of cross-contamination between samples with subsequent false positive results.

Follow local regulations and guidelines for changing gloves and cleaning work area between sample handling and processing.

Product inquiries and technical support

For product inquiries and technical support, please contact LumiraDx Customer Services.

International

E-mail: customerservices@lumiradx.com

Telephone: +44 (0)1172841222

Website: lumiradx.com

US

E-mail: customerservices.US@lumiradx.com

Telephone: 1-888-586-4721

Website: lumiradx.com

Manufacturer Information

Manufactured by:
LumiraDx UK Ltd
Dumyat Business Park Alloa
FK10 2PB, UK
Registration Number: 09206123

Authorized representative in the European Union:
LumiraDx AB
Västra Vägen 5A
16961 Solna, Sweden

Legal and Copyright Information

Copyright © 2021 LumiraDx UK LTD. All rights reserved worldwide.

LumiraDx and Flame logo are trademarks of LumiraDx International LTD. Full details of these and other registrations of LumiraDx can be found at lumiradx.com/IP. All other trademarks are the property of their respective owners.

Content should be used for the use of the LumiraDx products only and in line with instructions provided. You may not, except with our express written permission, distribute or commercially exploit the content. Nor may you transmit it or store it in any other form of electronic retrieval system other than for the purpose of use of the LumiraDx Instrument or LumiraDx Test Strips. Information provided is subject to change without notice.

Product is not available in all countries and regions. Please check with your local LumiraDx sales representative or distributor for availability in specific markets.

The LumiraDx SARS-CoV-2 Ag Test and the LumiraDx SARS-CoV-2 Ab Test have not been cleared or approved by FDA. The LumiraDx SARS-CoV-2 Ag Test has been authorized by FDA under an EUA only for the detection of SARS-CoV-2 nucleocapsid protein. The LumiraDx SARS-CoV-2 Ab Test has been authorized by FDA under an EUA only for detecting the presence of total antibodies to SARS-CoV-2. They have not been authorized for use to detect any other viruses or pathogens. The Tests are authorized in the United States for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostic Tests for detection and/or diagnosis of COVID-19 under Section 564(b)(1) of the Act, 21 U.S.C. § 360bbb-3(b)(1), unless the authorization is terminated or revoked sooner.

PDF preview unavailable. Download the PDF instead.

lumiradx-technical-bulletin-lumiradx-instrument-cleaning-and-disinfection-procedure-ivd-eua-r3 Adobe InDesign 17.0 (Macintosh) Adobe PDF Library 16.0.3

Related Documents

Preview LumiraDx Platform User Manual: Comprehensive Guide
This user manual provides detailed instructions for operating and maintaining the LumiraDx Platform, a point-of-care system for in vitro diagnostic tests. Learn about setup, testing procedures, troubleshooting, and safety information.
Preview LumiraDx Platform Returns: A Comprehensive Guide
This document provides detailed instructions and guidance for returning the LumiraDx Platform, including procedures for data handling, disinfection, packaging, and shipping. It ensures proper return of instruments and associated materials for investigation and quality control purposes.
Preview LumiraDx SARS-CoV-2 Ag Test Specifications and Performance
Detailed specifications and clinical performance data for the LumiraDx SARS-CoV-2 Ag Test, an in vitro diagnostic test for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens.
Preview Optimizing Your LumiraDx Sample Application Process
This technical bulletin provides guidance on optimizing the sample application process for the LumiraDx Test Strip, ensuring accurate and efficient results. It covers sample application, detection, and removal procedures, along with troubleshooting common issues.
Preview LumiraDx SARS-CoV-2 Ag Test Fact Sheet for Healthcare Providers
This fact sheet provides information on the LumiraDx SARS-CoV-2 Ag Test, including its intended use, interpretation of results, and emergency use authorization by the FDA.
Preview LumiraDx SARS-CoV-2 Ag QC Transfer Pipettes Technical Bulletin
Technical bulletin detailing the use of QC transfer pipettes with the LumiraDx SARS-CoV-2 Ag Test, including acceptable pipette types and supplemental instructions for accurate sample dispensing.
Preview LumiraDx SARS-CoV-2 Ag Quality Controls Instructions
Instructions for the LumiraDx SARS-CoV-2 Ag Quality Controls, including intended use, procedure, symbols glossary, customer service information, and limited warranty.
Preview LumiraDx SARS-CoV-2 Ag Test: Quick Reference Instructions
Concise guide for performing the LumiraDx SARS-CoV-2 Ag Test, including sample preparation, running the test on the LumiraDx Platform, and interpreting results. For Emergency Use Authorization (EUA) only.