Biosan UVC/T-AR, UVC/T-M-AR, UVT-B-AR, UVT-S-AR PCR Cabinets
User Manual
Manufacturer Information
Manufacturer: Biosan SIA
Rātsupītes iela 7 k-2, Rīga, LV-1067, Latvia
Tel.: +371 674 261 37
Website: https://biosan.lv
Marketing Department: sales@biosan.lv
Technical Support Department: support@biosan.lv
If you have any feedback regarding our products or services, we would be glad to hear it. Please send all feedback to the following address.
Table of Contents
1. Introduction to this User Manual Edition
This manual edition applies to the following models and versions:
Model | Version |
---|---|
UVC/T-AR, PCR cabinet | V.5AD |
UVC/T-M-AR, PCR cabinet | V.7A02 |
UVT-B-AR, PCR cabinet | V.5AD |
UVT-S-AR, PCR cabinet | V.6AA |
Edition 5.-7.02 – August 2025.
2. Safety Measures
2.1. Symbols Used in This Manual
⚠️ Attention! Please read this manual carefully before using the equipment and pay special attention to sections marked with this symbol.
2.2. Symbols Used on the Equipment and Packaging
CE CE marking: The manufacturer confirms compliance with European health, safety, and environmental protection standards, see section 11.1.
WEEE directive marking, see section 11.1.
⚠️ While the UV lamp is on, the front protective screen must be closed. Otherwise, the operator may be exposed to dangerous UV radiation.
⚠️ The total power consumed by devices connected through internal power outlets must not exceed 1000 W.
2.3. General Safety Precautions
- ⚠️ Attention! During operation with open UV light, the front panel must be closed. Otherwise, the operator will be exposed to UV radiation. UV radiation can harm the operator's health.
- ⚠️ Attention! Do not operate the equipment without the UV recirculator cover. UV radiation exposure is harmful and can cause damage to unprotected eyes and skin.
- ⚠️ Attention! The ultraviolet radiation purifier box contains a powerful UV radiation source. Therefore, before operating the equipment, ensure that all personnel working with the UV radiation purifier box are adequately protected. The operator must wear a lab coat with a closed front (fully buttoned), UV-certified safety goggles, and gloves that cover the lab coat or surgical gown cuffs.
- The protection provided by the equipment may be compromised if the equipment is used with accessories (shelves, furniture) not recommended or supplied by the manufacturer, or if the equipment is used for purposes other than those specified in these instructions.
- Protect the equipment from shocks and drops.
- Do not use the equipment if it has visible mechanical damage.
- Store and transport the equipment as described in the Storage and Transportation section.
- Before using cleaning and disinfection agents and methods not recommended by the manufacturer, consult the manufacturer to ensure the proposed method will not damage the equipment.
- Do not modify the equipment's construction.
- When working with hazardous substances, follow the instructions in the safety data sheets of the respective substances used and observe the relevant accident prevention regulations.
2.4. Electrical Safety
- The equipment must only be connected to a power source with a voltage that matches the one indicated on the label, along with the equipment's serial number.
- During operation, the switch and power outlet must be easily accessible.
- If the equipment needs to be moved, disconnect the plug from the power outlet.
- It is prohibited to connect the equipment to a power outlet without grounding, or to use an extension cord without grounding.
- Do not allow liquids to enter the equipment; if this happens, immediately disconnect the equipment from the power and do not attempt to turn it on. Equipment maintenance in this case must only be performed by service personnel.
- It is prohibited to use the equipment in rooms where condensation may occur. The operating conditions of the equipment are described in the Specifications section.
2.5. When Working with the Equipment
- Do not leave the operating equipment unattended.
- Do not operate the equipment in an environment with aggressive or explosive chemical mixtures. Please contact the manufacturer regarding possible equipment operation in specific environments.
- Do not operate the equipment if it is damaged or incorrectly installed.
- Do not use the equipment outside of laboratory premises.
- Do not operate the equipment without installed dust filters.
- Do not operate the box if the open UV lamp remains switched on.
2.6. Biological Safety
- The user is responsible for neutralizing hazardous materials that have spilled onto the equipment's surface or entered the equipment's interior.
- The user is responsible for decontaminating the unit before discontinuing its operation and disposal.
3. General Information
The UVC/T-AR, UVC/T-M-AR, UVT-B-AR, and UVT-S-AR PCR cabinets are designed for DNA/RNA decontamination and for creating aseptic conditions in a laboratory workspace. All models are tabletop units, equipped with a metal support frame, glass or acrylic walls, and a work surface that is powder-coated or made of stainless steel. The cabinet is equipped with an inlet for power cables or built-in power outlets inside. Please refer to each model for specific characteristics (see section 7.1).
The PCR cabinets are equipped with an open UV lamp installed at the top inside the cabinet. The open UV lamp's radiation disinfects the work area, inactivating DNA/RNS fragments within 15-30 minutes. A digital timer controls the direct UV radiation duration. A daylight lamp provides illumination for the work surface.
The PCR cabinets are equipped with a circulating bactericidal UV purification recirculator (AR), which ensures continuous decontamination inside the cabinet during operation. This is recommended for operations with DNA/RNA amplicons. The UV recirculator consists of a UV lamp (Fig. 1/1), a fan, and dust filters (Fig. 1/2), all housed in a plastic casing. The operator working with the PCR cabinet is not exposed to UV radiation from the operating UV recirculator; this allows for continuous air treatment without interrupting the work process. Air circulation at a short distance from the UV lamp and reflective panels in the air duct increase the UV light density in the recirculator, ensuring effective deactivation. The UV recirculator can process 100 cabinet volumes per hour, thus ensuring maximum aseptic conditions inside the box.
Figure 1. Recirculator Diagram:
- 1. UV lamp.
- 2. Fan and inlet filter.
- 3. UV indicator.
- 4. Outlet filter.
Microbiological studies by Biosan's scientific department, under the leadership of Doctor of Biology Vasilijs Bankovs, demonstrated a high level of biosafety and effectiveness (maximum contamination level is 1-3 CFU per 100 L of air)1.
1 See http://biosan.lv/eval-en
The PCR cabinet's construction does not provide operator protection, therefore it is not recommended to use the box for work with hazardous infectious and viral materials.
Advantages:
- UV recirculator.
- High-density UV decontamination without ozone.
- Automatic shut-off of the open UV lamp when the front panel is opened.
- Long UV lamp operating life (9000 hours).
- Low noise and energy consumption.
- Compact tabletop model for personal laboratories.
- Shelves for pipettes and reagents.
- Adjustable height tables (LT series), tables with drawers (T-4 / T-4L), laboratory cabinet (LF-1) available upon request.
4. Preparation for Work
4.1. Unpacking
Carefully remove the equipment from the packaging. Keep the original packaging for future transport or storage. Inspect the equipment for any damage sustained during transport. The warranty does not cover damage incurred during transport.
⚠️ Attention! As the equipment is very heavy, unpacking and installation should be performed by two people. Follow the instructions included with the packaging.
4.2. Kit Contents
4.2.1. Standard Kit
- PCR cabinet: 1 pc.
- Power cord (UVC/T-AR only): 1 pc.
- Spare dust filters: 2 pcs.
- Spare fuse for the control unit (see section 8.2): 1 pc.
- Spare fuse for built-in sockets (except UVC/T-AR): 1 pc.
- User Manual, Declaration of Conformity: 1 pc.
4.2.2. Optional Accessories (upon request)
- P-5, pipette stand: 1 compl.
- F-1, flat stand, work surface 400x140 mm: 1 compl.
- PDS-250, DNA/RNS Decontamination solution, 250 ml: 1 pc.
- LT-120, LT-150, LT-180 adjustable height tables (dimensions specified): 1 compl. each
- T-4, T-4L mobile tables with drawers and brakes (dimensions specified): 1 compl. each
- LF-1, laboratory cabinet with 5 drawers: 1 compl.
Image Description: Three tables labeled LT-120, LT-150, LT-180. Two smaller tables labeled T-4, T-4L. One cabinet labeled LF-1.
4.3. Equipment Installation at the Workplace
- If using a table for the PCR cabinet installation, carefully unpack and assemble the table according to its instructions.
- Install the equipment on a flat horizontal surface no smaller than 720x500 mm (1290x600 mm for UVT-S-AR model), which can support the weight of the PCR cabinet and the devices and materials inside the box, for example, on an LT / T-4 table.
- Plug the power cord into the socket at the rear of the equipment and position the unit so that the power switch and plug are easily accessible.
4.4. Shelf Installation
Follow the instructions provided with the shelf kit.
5. Operation
5.1. Powering On the Equipment
Note: For UVC/T-M-AR and UVT-B-AR, check that the short power cord (Fig. 6/1) is connected to the upper part of the equipment.
- Connect the equipment to a power source equipped with safe grounding.
- Switch the Power button to position I (ON):
- UVC/T-AR: No switch, the equipment will start automatically.
- UVC/T-M-AR and UVT-B-AR: On the right side, rear (Fig. 14/3).
- UVT-S-AR: On the left side, rear (Fig. 15/1).
- Visible Light: To turn the visible light inside the cabinet on or off, use the physical 'Visible light' button on the left side of the display.
- UV Recirculator, UV Light, Display: Press the physical 'UV-recirculator / UV-light' button on the left side of the display. This activates the UV recirculator. The display will also turn on, providing access to UV light status, timers, and other functions.
Note: To visually verify the operation of the enclosed UV recirculator, look for the indicator in the center of its cover (Fig. 1/3), which will be illuminated from the inside.
5.1.5. The interface display turns on with the startup screen (Fig. 2). If the time and date are not set on the equipment, a prompt to set them will appear (Fig. 3). Otherwise, the main display will be shown (see section 5.2).
- Use the '<' and '>' buttons to select the parameter to change.
- Use the '-' and '+' buttons to change the parameter.
- Press the 'OK' button to save the changes and continue.
Figure 2. Power On Screen
Figure 3. Date and Time Setting Screen
5.2. Main Display (Fig. 4)
- Top center: Two lines showing the UV exposure timer (see section 5.3.1 below).
- Bottom center: Eight buttons for controlling the UV exposure timer.
- Top left: 'Settings' button (gear icon, see section 8.1.6) and, if available, 'Alarms' button (exclamation mark icon, see section 8.1.5).
- Top right: Current time, date, and 'Statistics' button (circle icon, see section 8.1.4).
5.3. Timer and UV Radiation for Work Area
⚠️ Attention! The open UV lamp in the cabinet operates only when the protective door is fully closed. Otherwise, the operator may be exposed to dangerous UV radiation levels.
5.3.1. Set the Timer on the Main Screen (Fig. 4). Button functions are as follows:
Button | Short Press | Press and Hold |
---|---|---|
Start/Stop | Starts and stops the UV lamp and timer (see section 5.3.3). | |
- | Decreases the set timer by 1 minute. | Decreases faster. |
+ | Increases the set timer by 1 minute. | Increases faster. |
+15 min | Increases the set timer by 15 minutes. | |
Reset | No action. | Resets the timer to 00:00. |
30 min | No action. | Sets the timer to 30 minutes and starts UV radiation and timer. |
60 min | No action. | Sets the timer to 60 minutes and starts UV radiation and timer. |
Custom | Opens the custom timer setting (see section 5.3.2). | Sets the timer to the custom value and starts UV radiation and timer. |
5.3.2. To set and save a custom UV radiation exposure timer, briefly press the 'Custom' button on the main screen. Set the timer using the '-' and '+' buttons, then press 'Save'. To return without saving, press the button.
Figure 4. Main Display
Figure 5. Custom Timer Screen
5.3.3. To start UV exposure and the timer, press any of the following buttons:
- Pressing the 'Start/Stop' button starts the timer with the set values. If the timer is set to zero, the UV exposure will continue until manually stopped.
- Pressing and holding the '30 min' or '60 min' buttons starts the timer for 30 or 60 minutes, respectively.
- Pressing and holding the 'Custom' button starts the timer with the user-set values. If the user-set timer is set to zero, the UV exposure will continue until manually stopped.
The 'Actual' timer on the display runs. When the set time expires, the timer automatically switches off the open UV lamp.
Note: The recommended exposure time is two 30-minute sessions, one after the cabinet is switched on, and another before working inside. When using P-5 and F-1 shelves, extend the exposure time to two 45-minute sessions.
5.3.5. The open UV lamp (and timer) can be switched off by pressing the 'Start/Stop' button or the physical 'UV-recirculator / UV-light' button on the left side of the display.
Note: During UV exposure, all screen buttons are locked, except for the 'Start/Stop' and 'Statistics' buttons.
Note: Statistics, settings, and alarm signals are described in the Technical Maintenance section.
5.4. UV Box Ready for Operation. Working in the Box.
Note: The open UV lamp automatically switches off when the front panel is opened (the timer continues to count down the exposure time).
5.4.1. Lift the front screen (Fig. 6/5) upwards to work in the box. The opening heights (Fig. 5/7) are specified in the Specifications section. The screen can be lifted to the following positions:
- UVC/T-AR model: No intermediate positions.
- UVC/T-M-AR, UVT-B-AR, and UVT-S-AR models: Three intermediate positions – one-third, two-thirds, and fully open.
Note: Do not block the recirculator openings (Fig. 6/2)!
5.4.2. To use electrical equipment inside the box, thread the cable through the opening and secure it with the cover (Fig. 6/6), or connect the power cord to the internal socket (Fig. 6/7, for models with a socket).
⚠️ Attention! The power consumption of equipment connected to the internal socket must not exceed 1000 W.
5.5. After work is completed, close the front screen.
5.6. After the operation is finished, press the physical 'UV' and 'Visible light' buttons. Switch off the power switch if present.
Figure 6. DNA/RNA PCR Cabinet, Overview:
- 1. Control panel.
- 2. Recirculator openings.
- 3. Daylight lamp and open UV lamp.
- 4. Upper front panel.
- 5. Movable front protective screen (open).
- 6. Opening.
- 7. Power socket(s) (except UVC/T-AR).
- 8. Opening height.
6. Specifications
Biosan reserves the right to make changes and additions to the product design aimed at improving its usability and work quality without prior notice.
6.2. General Specifications
Model | UVC/T-AR Compact | UVC/T-M-AR Compact | UVT-B-AR Compact | UVT-S-AR Double |
---|---|---|---|---|
Rear panel | Steel | Steel | Steel | Double |
Side panels | PMMA | Glass | Steel with enamel | Glass |
Front panel and screen | Steel with enamel | Steel with enamel | Steel with enamel | Steel |
Work surface | Steel with enamel | Steel with enamel | Steel with enamel | Steel |
Open UV lamp | 1x TUV 25W G13 UV-C | 1x TUV 25W G13 UV-C | 1x TUV 25W G13 UV-C | 2x TUV 30W G13 UV-C |
UV intensity | 18 mW/cm²/s | 18 mW/cm²/s | 18 mW/cm²/s | 18 mW/cm²/s |
Radiation type | Ultraviolet light (λ=253.7 nm), ozone-free | |||
Open UV radiation time setting | 1 min - 24 hrs / continuous (setting step 1 minute) | |||
UV recirculator | 1x TUV 25W G13 UV-C | 1x TUV 25W G13 UV-C | 1x TUV 25W G13 UV-C | 1x TUV 30W G13 UV-C |
Daylight lamp | 1x LED 8W | 1x LED 8W | 1x LED 12W | 1x LED 12W |
Side panel thickness | 4 mm | 4 mm | 2 mm | 4 mm |
Front panel thickness | 8 mm | 8 mm | 4 mm | 5 mm |
Protective screen thickness | 8 mm | 8 mm | 4 mm | 5 mm |
Optical transmission | 92% | 92% | 95% | 95% |
UV protection | > 99.90% | > 99.90% | >96% | >96% |
Noise level | < 45 dB | < 45 dB | < 47 dB | < 47 dB |
Work surface (WxD) | 490 x 645 mm | 490 x 645 mm | 500 x 580 mm | 500 x 1210 mm |
Opening height | 165 mm | 190 mm (max. 210 mm) | 190 mm (max. 210 mm) | 190 mm (max. 220 mm) |
Protective screen positions | Fully closed | 1/3, 2/3, and fully open | 1/3, 2/3, and fully open | 1/3, 2/3, and fully open |
Power socket in box (see section 7.1) | Cable opening | 1 power socket and opening | 1 power socket and opening | 3 power sockets and opening |
Operating voltage | 100-240 V, 50-60 Hz | |||
Power consumption | 64 W | 64 W | 64 W | 113 W |
Overall dimensions (WxDxH) | 720 x 535 x 555 mm | 700 x 580 x 555 mm | 700 x 580 x 555 mm | 1250 x 600 x 590 mm |
Weight, approx. (Netto / Brutto) | 21.4 kg / 30 kg | 27.4 kg / 36 kg | 28.8 kg / 37.5 kg | 51 kg / 62 kg |
Laboratory table | All T & LT models | All T & LT models | T-4L, LT-150, LT-180 | T-4L, LT-150, LT-180 |
⚠️ Attention! The total power consumption of equipment connected to internal sockets must not exceed 1000 W.
6.3. Operating Room Requirements
Work area description | Closed laboratories, cold rooms |
Temperature range | +4 °C ... +40 °C |
Humidity parameters | Maximum 80% at 31 °C, linearly decreasing to 50% at 40 °C. Atmosphere without condensation. |
Maximum operating altitude | 2000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |
6.4. Materials Used
- PMMA – polymethyl methacrylate, organic glass, Altuglas® EX.
- Glass – coated with a transparent film (0.1 mm), Euroglas®.
- Steel – polished stainless steel.
- Steel with enamel – steel coated with white powder enamel.
6.5. Cabinet Dimensions
Internal and external dimensions and screen opening heights.
Diagram Description: Three diagrams showing cabinet dimensions. The first shows UVC/T-AR with dimensions 720x535x555 mm and opening height 165 mm. The second shows UVC/T-M-AR and UVT-B-AR with dimensions 700x580x555 mm and opening height 190 mm (max 210 mm). The third shows UVT-S-AR with dimensions 1250x600x590 mm and opening height 190 mm (max 220 mm).
7. Ordering Information
7.1. Available Models and Versions
Model | Catalog Number | Version | Openings | Power Sockets |
---|---|---|---|---|
UVC/T-AR | BS-040102-AAA | V.3AD | 1 | - |
UVC/T-M-AR | BS-040104-A06 | V.6A02 | 1 | 1 x Euro |
UVT-B-AR | BS-040109-AAA | V.4AD | 1 | 1 x Euro |
UVT-S-AR | BS-040107-AAA | V.4AA | 1 | 3 x Euro |
⚠️ Attention! The power consumption of equipment connected to internal sockets must not exceed 1000 W.
7.2. For More Information and to Order Additional Accessories, Contact Biosan.
7.2.1. Additional Accessories
Accessory | Catalog Number |
---|---|
P-5, stand for 5 pipettes | BS-040104-CK |
F-1, stand for reagents, work surface 400x140 mm | BS-040104-DK |
PDS-250, DNA/RNS Decontamination solution, 250 ml | BS-040107-DK |
LT-120, adjustable height table, dimensions 1200x600x(625–1280) mm | BS-040107-EK |
LT-150, adjustable height table, dimensions 1500x600x(625–1280) mm | BS-040107-JK |
LT-180, adjustable height table, dimensions 1800x600x(625–1280) mm | BS-040107-HK |
T-4, mobile table with drawer and wheels with brakes 800x600x745 mm | BS-040101-BK |
T-4L, mobile table with drawer and wheels with brakes 1290x600x770 mm | BS-040107-BK |
LF-1, laboratory cabinet with 5 drawers | BS-050101-BK |
8. Technical Maintenance
8.1. Service
- If the equipment is not working (e.g., the platform is not moving or the equipment does not respond to button presses, etc.) or requires maintenance, disconnect the equipment from the power supply and contact Biosan or your local Biosan representative.
- All technical maintenance and repair work (except for those listed below) must be performed only by qualified and specially trained personnel.
- Operation Integrity Check: If the equipment operates according to the procedure described in the 'Operation' section, no additional checks are required.
- UV Lamp Statistics: Information can be accessed by pressing the 'Statistics' button (circle icon, Fig. 4). In this separate screen (Fig. 8), the operating time is displayed in hours and minutes. Two buttons below lead to screens with QR codes: 'User manual' leads to this manual, 'Consumables' leads to available consumables. Press the '←' button to return to the previous screen.
- Equipment Alarms: Alarms are accessible by pressing the 'Alarm' button (exclamation mark icon, Fig. 4). Alarms indicate when the UV lamp's operating time is nearing its end (see section 8.3) and when dust filters need replacement or cleaning (see section 8.4). Press the 'OK' button to return to the previous screen. Examples of alarms are shown in Fig. 9.
- Settings Menu: The settings menu (Fig. 10) is accessed by pressing the 'Settings' button (gear icon, Fig. 3). Available options:
- Date and Time: To set the correct date and time, see section 5.1.5.
- Service Menu: Accessible by pressing and holding the button for 8 seconds. In this submenu (Fig. 11), there are buttons to reset the UV lamp operating time counters, which also activate by pressing and holding the button for 8 seconds. These actions reset the statistics and corresponding alarms.
8.2. Fuse Replacement
- Control Unit Fuse: Disconnect the equipment from the power. Unplug the power cord from the rear of the equipment (Fig. 13/1, compact cabinets). Open the holder cover by pulling it out (Fig. 12, compact cabinets) or by unscrewing the cover marked FU2 (Fig. 14, UVT-S-AR model). Check the fuse and replace it if necessary with an M 3.15 A (Type M – Medium time-lag) fuse.
- Internal Socket Fuse: Disconnect the equipment from the power. Open the holder cover (Fig. 13/2 or Fig. 14/FU1) at the rear of the equipment by unscrewing it. Check the fuse and replace it if necessary with an M 5.0 A (Type M – Medium time-lag) fuse.
Figure 12. Fuse Holder (Compact)
Figure 13. Fuse Holder (Rear View)
Figure 14. Fuse Holder (UVT-S-AR)
8.3. UV Lamp Replacement
Average lamp operating life is 9000 hours. The UV lamp needs to be replaced after its service life expires or if it stops functioning. Replacement must only be performed by service engineers and specially trained specialists.
Note: The open UV lamp in the cabinet operates only when the protective door is fully closed. Otherwise, the operator may be exposed to dangerous UV radiation levels.
8.4. Dust Filter Replacement
Dust filters (Fig. 6/2) require monthly maintenance. After opening the protective covers of the UV recirculator heads, check if the filters are dirty, replace the filters, or wash them with water if necessary, dry them, and reinsert them into the equipment.
8.5. Maintenance and Cleaning
Disconnect the equipment from the power before cleaning.
⚠️ Attention! Do not wet the control unit and prevent liquid from entering the unit.
- 8.5.1. The touch screen has a soft membrane. Do not use organic or abrasive cleaning agents. Dry excess water with a clean, soft cloth or sponge.
- 8.5.2. Model UVC/T-AR and P-5 and F-1 shelves, full cleaning. The cabinet panels and shelves are made of organic glass (polymethyl methacrylate Altuglas® EX) and are prone to scratching and reduced transparency from improper cleaning. Use a soft sponge or cloth with standard glass cleaner, wipe off excess, and dry.
- 8.5.3. For disinfection, it is recommended to use special solutions for cleaning DNA and RNS (e.g., Biosan PDS-250). Wipe off residual moisture with a moisture-absorbing soft cloth or sponge.
⚠️ Attention! Use mild cleaning agents for cleaning; never use agents containing organic compounds, undiluted alcohol, or cleaning agents containing alcohol (more than 20%) or ammonia on organic glass surfaces. Do not use abrasive cleaning agents or hard material sponges.
Solvent | Effect on Glass |
---|---|
Biosan PDS-250 | No effect |
DNA-Exitus Plus™ | No effect |
RNase-Exitus Plus™ | No effect |
H2O2 6% | No effect |
Ethyl alcohol ≤20% | No effect |
Ethyl alcohol >20% | Intensifying effect. Do not use! |
Note: The formation of microcracks is a normal process for organic glass panels exposed to ultraviolet radiation. Microcracks appear over time; this process can only be slowed down by minimizing UV exposure time. Microcracks may appear during the warranty period; this is considered normal wear and tear, not covered by the warranty. Organic glass panels can be replaced.
8.5.4. Models UVC/T-M-AR, UVT-B-AR, and UVT-S-AR, external cleaning. The glass panel is covered on the outside with a 4 mil thick film for UV radiation protection. Use a soft sponge or cloth with standard cleaning agent, wipe off excess, and dry. Do not use ethanol or other agents containing organic solvents.
8.5.5. Models UVC/T-M-AR, UVT-B-AR, and UVT-S-AR, internal cleaning and decontamination. Recommended disinfection solutions include: 75% ethanol solution, sodium hypochlorite solution, special solutions for DNA and RNS cleaning (e.g., Biosan PDS-250). Wipe off residual moisture with a moisture-absorbing soft cloth or sponge.
9. Storage and Transportation
- Store and transport the equipment in a horizontal position (see label on packaging) at temperatures between -20°C and +60°C and a maximum relative humidity of 80%.
- If the equipment was transported or stored in a warehouse, allow it to stand for approximately 2-3 hours at room temperature before connecting it to power.
- No additional operations are required for long-term storage.
10. Warranty
- 10.1. The manufacturer guarantees the equipment's compliance with the specified specifications, provided the user adheres to the requirements for equipment operation, storage, and transportation.
- 10.2. The warranty period for operation is 24 months from the date the equipment is delivered to the consumer. For extended warranty options, see section 10.5.
- 10.3. The warranty applies only to equipment transported in its original packaging.
- 10.4. If equipment defects are found, the user must draw up and confirm a claim act, which must be sent to the manufacturer or distributor. The claim form can be found on our website, in the Technical Support section, via the link below.
- 10.5. Extended Warranty.
- For UVC/T-M-AR and UVT-S-AR, which are Premium class models, an additional year of warranty is available free of charge after registration, within 6 months of purchase. Online registration is available in the Warranty Registration section, via the link below.
- For UVC/T-AR and UVT-B-AR, which are Basic Plus class models, an additional warranty is a paid service. Contact our service department via the form on our website, in the Technical Support section, via the link below.
- 10.6. A description of the equipment classes is available on our website, in the Product Class Comparison section, via the link below.
Technical Support: biosan.lv/lv/support
Warranty Registration: biosan.lv/register-lv
Product Class Comparison: biosan.lv/classes-lv
10.7. The following information will be required in case of warranty or post-warranty repair needs. Fill out and save this form:
Model | Serial Number | Date of Sale |
---|---|---|
UVC/T-AR, UVT-B-AR, UVC/T-M-AR, UVT-S-AR, PCR cabinet |
10.8. Manufacturing Date: The manufacturing date is indicated in the serial number on the equipment label. The serial number consists of 14 digits formed by XXXXXXYYMMZZZZ, where XXXXXX is the model code, YY and MM are the manufacturing year and month, ZZZZ is the unit number.
11. EU Declaration of Conformity
11.1. The UVC/T-AR, UVC/T-M-AR, UVT-B-AR, UVT-S-AR PCR cabinets comply with the following relevant Union legal acts:
- LVD 2014/35/EU: LVS EN 61010-1:2011 + A1:2019 Safety requirements for electrical measuring, control, and laboratory use apparatus. General requirements.
- EMC 2014/30/EU: LVS EN 61326-1:2021 Electrical measuring, control, regulation, and laboratory analysis apparatus. Electromagnetic compatibility (EMC) requirements. General requirements.
- RoHS3 2015/863/EU: Directive on the restriction of the use of certain hazardous substances in electrical and electronic equipment.
- WEEE 2012/19/EU: Directive on waste electrical and electronic equipment.
11.2. The Declaration of Conformity is available for download on the respective model's page on our website, using the link below.