Benetech GM1030C Split Type Lux Meter

Instruction Manual

Specific Declarations

Our company shall hold no any responsibility resulting from using output from this product as an direct or indirect evidence. We reserve the right to modify product design and specification without notice.

Standard: Q/GMY 010-2018

Version: 1030-EN-0

CE ♻️

Product Introduction

The Split type illuminometer is a professional instrument for measuring light intensity and brightness. It is used in light intensity measurement engineering, quality control, health care, and light intensity measurement in various surroundings, such as factories, schools, offices, transportation, and homes.

Product Features

This split type illuminometer can measure current value, maximum value, minimum value, and difference value of illumination and temperature. It can also hold illumination and temperature data, calculate illuminance integral and average integral, and record data.

Names and Functions of Parts

  1. Power/Backlight
  2. Back/Difference Value
  3. Mode/Return
  4. Data Holding/Setting
  5. Forward/Integral
  6. Record/Confirm
  7. Illumination Head
  8. Lens Cover
  9. UART Connector (TTL) (Bluetooth version only)

Device Diagram:

The device features a digital display showing values like 69.8 and 32.3, with units indicated as 'Lx' and '°C'. It has several buttons: HOLD, SET, DIF, MODE, Σ, REC, and C. The illumination sensor is a separate unit connected by a cable.

LCD Display

The LCD display shows various indicators:

Note: When the indicator flashes, please change batteries, or the measurement value may not be accurate and Bluetooth communication may be abnormal.

Function Description

The illuminometer has two working modes: Normal mode and Setting mode. Each mode has multiple interfaces.

Normal Mode Interfaces:

  1. Real-time interface: Displays current measured illumination and temperature values.
  2. Max value interface: Records from startup, displays maximum illumination and temperature.
  3. Min value interface: Records from startup, displays minimum illumination and temperature.
  4. Difference value interface: Displays the difference between the latest measured value and a previously saved value.
  5. Holding interface: Freezes the current measured illumination and temperature data.
  6. Integral interface: Calculates the integral of the current illumination value over time (unit: hour).
  7. Integral averaging interface: Calculates the average illumination value over a specified time (unit: hour).
  8. Manual storage interface: Allows manual saving of current illumination values (up to 60 groups). Users can review or delete stored data.
  9. Automatic storage interface: Automatically records illumination values at set intervals (up to 2000 groups). Users can review or delete stored data.

Note: Integral operation and automatic storage can be active simultaneously. The 'Σ' and 'REC' indicators will flash when active.

Setting Mode Interfaces:

  1. Manual storage setting interface: Set manual storage on/off.
  2. Auto storage settings interface: Set automatic storage on/off.
  3. Display unit setting interface: Set illumination (Lux/Fc) and temperature (°C/°F) units.
  4. Timing shutdown setting interface: Set automatic shutdown on/off.
  5. Bluetooth switch setting interface: Set Bluetooth on/off (Bluetooth version only).

Function Operations

1. Turn On/Off

Install batteries. Long press the [POWER] button for 2 seconds to turn on. The device enters the real-time interface after a full-screen display.

2. Backlight On/Off

After startup, short press the [MODE] button in any interface to toggle the backlight. The backlight consumes power and is off by default.

3. Max/Min Value of Illumination and Temperature

In the real-time interface, short press the [MODE] button to switch between Max and Min value interfaces. The screen will display 'MAX' or 'MIN'. Short press [MODE] again to return to the real-time interface.

4. Difference Value of Illumination

In the real-time interface, short press the [DIF] button to switch to the difference value interface. A negative number indicates the real-time value is lower than the saved value, while a positive number indicates it's higher. The screen displays '△'. Short press [DIF] again to return to the real-time interface. The [MODE] button can also be used to return.

5. Data Holding

In the real-time or difference value interface, short press the [HOLD] button to enter the holding interface, displaying 'HOLD'. Short press [HOLD] again to return to the previous interface. The [MODE] button can also be used to return to the real-time interface.

6. Integral Operation

In the real-time interface, press the [REC] button to switch between the integral ('Σ') and integral averaging ('Σ AVG') interfaces. Short press [MODE] to return to the real-time interface from these modes.

Press the [REC] button to start or stop integration. When integration is active, the 'Σ' indicator flashes. The integral value is displayed in Lx·h or Fc·h, and the time span is in hours (h).

Integral Interface Example: Upper screen shows integral illumination (Lx·h), lower screen shows time (h).

Average Integral Interface Example: Upper screen shows average illuminance (Lx or Fc), lower screen shows time (h).

7. Manual Storage Operations

In the real-time interface (normal mode), long press the [MODE] button for 1 second to enter setting mode. Short press [MODE] to access the manual storage setting interface. 'Hand CLO' indicates manual storage is off, and 'Hand OPE' indicates it is on.

Short press [SET] to turn manual storage on, and [REC] to turn it off.

After confirming the setting, long press [MODE] for 1 second to exit and return to the real-time interface. Pressing [REC] will manually save the data.

Manual Storage Query: In the real-time interface, long press [REC] for 1 second. If in auto-storage interface, short press [MODE] to switch to manual storage. Short press [REC] or [DIF] to navigate through stored records. Long press [REC] for 1 second to return to the real-time interface.

Manually Delete Stored Records: In the real-time interface, long press [REC] for 1 second. If in manual storage interface, short press [MODE] to switch to automatic storage. Then long press [HOLD] for 1 second to delete all manually stored data. Long press [MODE] for 1 second to return to the real-time interface.

8. Manual Storage Query

In the real-time interface, long press [REC] for 1 second. If it is the automatic storage interface, short press [MODE] to switch to the manual storage interface. Short press [REC] or [DIF] to navigate through records. The upper part shows the illumination value, and the lower part shows the record number. Long press [REC] for 1 second to return to the real-time interface.

9. Manually Delete Stored Records

In the real-time interface, press [REC] for 1 second. If it is the auto-save interface, short press [MODE] to switch to the manual storage interface. Then long press [HOLD] for 1 second to delete all manually stored data. Long press [MODE] for 1 second to return to the real-time interface.

10. Automatic Storage

In the real-time interface (normal mode), long press [HOLD] for 1 second to enter setting mode. Press [HOLD] again or [MODE] to exit. In setting mode, switch to the automatic storage setting interface. 'Auto CLO' means off, 'Auto OPE' means on.

Short press [DIF] to turn automatic storage on, and [REC] to turn it off.

Note: Automatic and manual storage cannot be active simultaneously. Manual storage is the factory default.

After confirming automatic storage, long press [SET] for 1 second to exit setting mode. In the real-time interface, press [REC] for 1 second. If in manual memory interface, press [MODE] to switch to automatic storage interface.

Automatic Storage Timer: Short press [REC] to start the automatic storage timer. The 'REC' indicator flashes. Long press [REC] for 1 second to enter time interval setting. Use [HOLD] to adjust the decimal point position and [DIF]/[SET] to adjust the value. The default interval is 0.001h (3.6 seconds).

Automatic Storage Records Query: In the real-time interface, long press [REC] for 1 second. If it is the manual storage interface, short press [MODE] to switch to the automatic storage interface. The 'MAX' indicator shows the latest stored record. Short press [REC] or [DIF] to navigate through records. Long press [REC] for 1 second to return to the real-time interface.

Delete Automatic Stored Records: In the real-time interface, long press [REC] for 1 second. If it is the manual storage interface, short press [MODE] to switch to the automatic storage interface. Then long press [HOLD] for 1 second to delete all automatically stored data. Long press [MODE] for 1 second to return to the real-time interface.

Performance and Indicators

Refer to JJG245-2005 “Illuminometer verification procedures”.

Illumination probe Silicon Diode
Illumination range 0~200,000Lux (0~20,000Fc), divided into four gears
Gear Gear range Minimum resolution Accuracy
X1 gear 0.0~199.9Lux 0.1 Lux ±3%rdg+5dgts
X10 gear 20.0*10~199.9*10Lux 1 Lux ±3%rdg+10dgts
X100 gear 20.0*100~199.9*100Lux 10 Lux ±4%rdg+10dgts
X1000 gear 20.0*1000~199.9*1000Lux 100 Lux ±4%rdg+10dgts
Repeatability ±2%
Refresh rate 2 times/second
Temperature probe NTC Thermistor
Temperature range -20~50°C (-4~122°F)
Temperature accuracy ±1.0 °C
Power Three AAA batteries

App Operation Instruction (Bluetooth version only)

App Main Interface:

  1. After Bluetooth connection, the app synchronizes instrument states (unit, auto-storage time interval).
  2. The upper part displays real-time illumination and temperature values, as well as maximum and minimum values. The blue circle in the middle allows viewing saved max/min values. The lower part shows the illumination trend curve.
  3. Click 'Storage' to enter the storage interface. Click 'Start' to record data, and 'Clear' to clear saved data.
  4. Click the Bluetooth icon to enter the Bluetooth interface. Click the upper right button for 'settings', 'language switch', and 'folder' options.

Extreme Value Interface: Click minimum/maximum value buttons to view respective values.

Saving Interface: Click 'Storage' to save data.

Bluetooth Connection Interface: Click 'Search' to find nearby Bluetooth devices (search takes about 5 seconds).

Setting Interface:

  1. Set illumination and temperature units.
  2. Configure Automatic (2000 groups) and Manual (60 groups) storage modes. These modes cannot be active simultaneously. Storing, reading, and clearing require a mode to be started. Click 'Read' to access stored data, and 'Clear' to clear it.
  3. Only one mode among integral, average integral, and difference value modes can be active at a time. Switch to the corresponding mode to view data.

File List

Short press to read files, long press for multi-select and delete.

Data Display Interface: The upper part shows the illumination trend curve, average, maximum, and minimum values. The lower part shows remark information. The upper right corner button allows exporting data to Excel (up to 65535 items).

Host and PC Connection (Bluetooth version only)

The computer terminal App (LuxMeter) facilitates UART serial port communication for convenient operation and parameter setting, including:

The app allows reading and deleting stored data, and synchronizing real-time data for curve generation and analysis.

Note: Refer to the app's help options for specific operations.

Software Download: Download the LuxMeter software and instructions from http://www.benetechco.net. Navigate to 'Support → Download center' and download 'LuxMeter UART v2.1.zip'. Extract the package, install the UART-USB cable driver, and then install LuxMeter v2.1.exe.

Installation: Follow the installation wizard prompts. It is recommended not to install in the system partition to protect data during OS reinstallation.

UART Interface: Designed for professionals. Requires a specialized UART-USB line. The Prolific PL2303 driver must be installed (usually included with WIN10). Before connecting, set the COM port in the software via the device manager. The default baud rate is 9600bps.

For protocol development, contact the sales department. The protocol purchase price is RMB 500.

Models: GM1030C Split Type Lux Meter, GM1030C, Split Type Lux Meter, Type Lux Meter, Lux Meter, Meter

PDF preview unavailable. Download the PDF instead.

benetech gm1030c manual GPL Ghostscript 9.52

Related Documents

Preview BENETECH GM1030 Split Type Lux Meter: Instruction Manual
User manual for the BENETECH GM1030 Split Type Lux Meter. This guide details product features, operation, data logging capabilities, and mobile app integration for accurate light intensity and brightness measurements in diverse environments.
Preview BENETECH Lux Meters: GM1010, GM1020, GM1030, GM1030C - Features & Specifications
Explore the BENETECH range of Lux Meters, including models GM1010, GM1020, GM1030, and GM1030C. This document details their features, specifications, and applications for measuring illuminance in various environments.
Preview BENETECH GT8913 Digital Anemometer Instruction Manual
Instruction manual for the BENETECH GT8913 Digital Anemometer, covering setup, operation, specifications, troubleshooting, and maintenance. Provides detailed guidance on using the device for measuring wind speed, temperature, humidity, and air volume.
Preview BENETECH GM1371 Humidity and Temperature Meter Instruction Manual
Instruction manual for the BENETECH GM1371 Humidity and Temperature Meter, detailing its features, operation, technical specifications, and important announcements.
Preview Benetech GM1352 Sound Level Meter Instruction Manual
Instruction manual for the Benetech GM1352 Sound Level Meter, covering introduction, component identification, LCD display, operating instructions, considerations, and technical parameters.
Preview Benetech GM8902 Digital Air Flow Anemometer: Specifications & Features
Comprehensive details on the Benetech GM8902 Digital Air Flow Anemometer, covering its technical specifications, measurement capabilities, package contents, and software features for industrial and home applications.
Preview BENETECH GT8911 Hot Wire Anemometer Instruction Manual
Comprehensive instruction manual for the BENETECH GT8911 Hot Wire Anemometer, detailing its features, specifications, operation, PC connectivity, troubleshooting, maintenance, and warranty information.
Preview BENETECH GM86 Micro Power Monitor Instruction Manual
Instruction manual for the BENETECH GM86 Micro Power Monitor, detailing its features, safety instructions, operating procedures, application cases, maintenance, and technical specifications for measuring voltage, current, power, and energy consumption.